**MEAN WELL RPD-60A 60W Dual Output Medical Grade Owner’s Manual


**

Features

  • 4″×2″ compact size
  • Medical safety approved (2 x MOPP) accroding to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • Cooling by free air convection
  • EMI class B for class configuration Ⅰ
  • Extremely low leakage current
  • Protections: Short circuit / Overload / Over voltage
  • years warranty

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

GTIN CODE

MW Search: hitps:/www.meanwell.com/serviceGTIN.aspx

Description RPD-60 is a 60W highly reliable green PCB type medical power

supply with a high power density on the 4” by 2” footprint. It accepts 90~264VAC input and offers dual output voltages.

RPD-60is able to be used for Class I (with FG) system design. The extremely low leakage current is less than 1501A. In addition, it conforms to international medical regulations (2*MOPP)and EMC BS EN/EN55011.

Model Encoding

Model Encoding

SPECIFICATION

MODEL RPD-60A RPD-60B
OUTPUT OUTPUT NUMBER CH1
DC VOLTAGE 5V 12V
RATED CURRENT 5A 2A
CURRENT RANGE 0.5 ~ 5.5A 0.1 ~ 2.2A
RATED POWER 49W 53.5W
PEAK LOAD(10sec.) Note.2 53.9W 58.85W
RIPPLE & NOISE (max.) Note.3 80mVp-p 80mVp-p
VOLTAGE TOLERANCE Note.4 +3,-2% ±6.0%
LINE REGULATION ±0.5% ±1.0%
LOAD REGULATION ±1.5% ±2.0%
SETUP, RISE TIME 300ms, 15ms/230VAC        300ms, 15ms/115VAC at full

load
HOLD UP TIME (Typ.)| 70ms/230VAC        14ms/115VAC at full load
INPUT| VOLTAGE RANGE| 90 ~ 264VAC      127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 78%| 82%
AC CURRENT (Typ.)| 1.1A/115VAC        0.7 A/230VAC
INRUSH CURRENT (Typ.)| COLD START 60A/230VAC      30A/115VAC
LEAKAGE CURRENT **Note.5| Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC
PROTECTION| OVERLOAD| 115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
ENVIRONMENT| WORKING TEMP.| -20 ~ +65℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 45℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.6| 3000 meters
SAFETY & **EMC (Note 9)| SAFETY STANDARDS| UL62368-1,TUV BS EN/EN62368-1,IEC62368-1,IEC60601-1, EAC TP TC 004, UL ANSI/AAMI ES60601-1,CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved,TUV BS EN/EN60601-1 approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC  O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted  susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,100% interruptions 250 periods
OTHERS| MTBF| 5031.3K hrs min.    Telcordia SR-332 (Bellcore) ; 679.5K hrs min.    MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 101.650.829mm or 4″ 2″ 1.14″ inch
PACKING| 0.15Kg; 96pcs/15.4Kg/0.89CUFT
NOTE|

  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
  3. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
  4. Tolerance : includes set up tolerance, line regulation and load regulation.
  5. Touch current was measured from primary input to DC output.
  6. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  7. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
  8. Heat Sink HS1,HS2 can not be shorted.
  9. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

| Arrow| .com.| File Name: RPD-60-SPEC ____2022-09-20

Block Diagram

Block Diagram

Derating Curve

AMBIENT TEMPERATURE (C)
Derating Curve

Output Derating VS Input Voltage

INPUT VOLTAGE (VAC) 60Hz
Output Derating

Mechanical Specification

Dimensions

AC Input Connector (CN1) : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N JST VHRor equivalent JST SVH-21T-P1.1or equivalent
2 No Pin
3 AC/L

DC Output Connector (CN2) : JST B6P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1,2 V1 JST VHRor equivalent JST SVH-21T-P1.1or equivalent
3,4 COM
5 V2
6 NC

: Grounding Required

  1. HS1,HS2 cannot be shorted.
  2. M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1,M2 and chassis grounding.

Installation Manual

Please refer to: http://www.meanwell.com/manual.html