LUMEX Series 588W Bariatric Clinical Care Recliner Instruction Manual
- June 13, 2024
- LUMEX
Table of Contents
LUMEX SERIES 588W BARIATRIC
CLINICAL CARE RECLINER
OPERATING MANUAL CLASS 1 EQUIPMENT – ORDINARY – TYPE
B
SAVE THIS MANUAL FOR FUTURE USE.
INTRODUCTION
This manual contains assembly, operation, and maintenance instructions for
your Model 588W Bariatric Clinical Care Recliner. Read the entire manual
carefully before using your recliner, and refer to it during use if you have
questions.
The Lumex 588W Bariatric Clinical Care Recliner features a soft seat and back
package with wide, soft armrests for improved patient comfort and support
during long-term sitting.
The easy-to-activate gas cylinder mechanism provides quick Trendelenburg
positioning and improved lift.
Info: A Replacement Parts List is available at
www.grahamfield.com.
Intended Use
The Lumex 588W Bariatric Clinical Care Recliner is intended for use by
bariatric patients of all age groups with varying medical and physical
conditions. The recliner is ideal for use by clinics, hospitals, and
rehabilitation centers.
The Lumex 588W Bariatric Clinical Care Recliner supports patient weights up to
700 lb. (318 kg) when evenly distributed.
IMPORTANT SAFETY PRECAUTIONS—PLEASE READ BEFORE USE
The safety statements presented in this chapter refer to the basic safety information that the 588W user shall pay attention to and abide by. There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specific to the operations. Please note the following special statements, used throughout this manual, and their significance:
WARNING: Indicates a potential hazard situation or unsafe practice that,
if not avoided, could result in death or serious personal injury.
CAUTION: Indicates a potential hazard or unsafe practice that, if not
avoided, could result in minor or moderate personal injury.
NOTICE: Indicates a potential hazard or unsafe practice that, if not
avoided, could result in product or property damage.
Info: Provides application recommendations or other useful information to
ensure that you get the most from your product.
WARNINGS
WARNING: Periodically inspect all parts and replace as needed.
WARNING: Do not use near an open flame, as the chair could be a
“combustion source”, resulting in fire, property damage, and severe physical
injury.
WARNING: Service and repair of the Lumex 588W Bariatric Clinical Care
Recliner must be performed by an authorized Lumex distributor. GF Health
Products, Inc. (“Graham-Field”) is not responsible for any consequence
resulting from any unauthorized service or repair.
WARNING: The electronics and actuator contain no serviceable components.
DO NOT attempt to open the electronics; this may void the warranty and may
result in damage to the unit. If service is required, consult your Graham-
Field authorized distributor.
WARNING: Do not open the batteries. If service is required, consult your
Graham-Field authorized distributor. Replace the batteries when damaged,
faulty, or every two years.
WARNING: This product has a maximum weight capacity of 700 lb. (318 kg),
EVENLY DISTRIBUTED.
WARNING: Caster wheels must be locked during occupant transfer.
WARNING: The Lumex 588W Bariatric Clinical Care Recliner is equipped with
a built-in battery backup system: nevertheless, the unit should remain plugged
into wall receptacle during normal use.
WARNING: Use of the battery backup is recommended only for transport,
power failure, emergency, and when there is no reliable AC power or properly
grounded electrical wall outlet.
WARNING: When using the battery backup, continuous operation must not
exceed two hours with 1.5 min. of operation to 15 min. of rest.
WARNING: Unauthorized modification of your recliner or the use of non-
Lumex replacement parts may change the structure of the recliner and could
create a hazardous condition, which may result in serious injury and will void
the warranty.
WARNING: GF Health Products, Inc. specifically disclaims responsibility
for any bodily injury or property damage which may occur during any use which
does not comply with federal, state or local laws or ordinances.
WARNING: Keep all moving parts free of obstructions (i.e. blankets,
heating blankets/pads, tubing, wiring, furniture, walls, and other types of
products).
WARNING: Check the range of movement on all motors to ensure they do not
bind in the Full Up or Full Down positions.
WARNING: Ensure that the power cord DOES NOT get tangled around the bed,
bed rails or legs during normal operation of the bed.
WARNING: DO NOT open assemblies such as the Actuators, Hand Control
Pendant, or Control Box. If unauthorized personnel perform work on these
components, the manufacturer’s warranty becomes void.
WARNING: The Trendelenburg Quick Release trigger when activated will
quickly lower the patients head below their feet. Only use in emergency
situations. Make sure that there are no obstructions that can cause the
trigger to be released unexpectedly. When activating the trigger help guide
the back down.
WARNING: DO NOT activate Trendelenburg Quick Release unless for an
emergency or maintenance check. Doing so may result in serious injury.
PRODUCT DESCRIPTION
|
---|---
Hardware Included
Description| Qty
Bolt with 1/2″ hex head| 4
Bolt with 7/16″ hex head, split washer| 2
Screw with M5 socket head| 4
M5 hex key| 1
Tools Needed
Socket wrench for 1/2 & 7/16 hex head bolts| 1
ASSEMBLY
Unpacking
- Check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field authorized distributor.
2. Remove all packaging materials from recliner base.
Setup
- Plug the recliner’s power cord into a properly grounded electrical wall outlet. Using the hand control pendant, raise the back fully into the upright position (Fig. 4).
- Align the tubes on the backrest with back support tubes on the mainframe assembly (Fig. 5). Place the backrest onto the support tubes until the holes on the back are aligned with the threads of the inner nuts (Fig. 6).
- Attach and fully tighten all 1/2″ and 7/16″ bolts to the back support bars with a wrench (Fig. 7, Fig. 8).
- Align the holes on the push bar with the holes on the back bar. Attach socket head screws to the push bar and fully tighten with supplied hex key (Fig 9). Repeat until push bar is tightly secured to the backrest (Fig. 10).
Footboard Installation
- Align attachment arms of footboard assembly inside the hinges on the crossbar located behind the legrest upholstery. Ensure that the holes on each hinge are aligned with the holes on the attachment arms of the footboard.
- Insert one bridge pin through the holes on one hinge and attachment arm (Fig. 11).
- Insert the cotter clip into the hole on the bridge pin so that the attachment is secure (Fig. 12).
- Repeat steps 2 and 3 for the other side.
OPERATION
WARNING: Disconnection of this device: In case of emergency, always
disconnect the device by removing the power cord plug from the power outlet.
WARNING: This device is equipped with a built-in battery backup system:
nevertheless, the unit should remain plugged into wall receptacle during
normal use.
WARNING: Use of the battery backup is recommended only for transport,
power failure, emergency, and when there is no reliable AC power or properly
grounded electrical wall outlet.
WARNING: When using the battery backup, continuous operation must not
exceed two hours with 1.5 min. of operation to 15 min. of rest.
Seat Positioning: Trendelenburg Position (Shock Position)
With the exception of Trendelenburg Position, all seat positioning must be
controlled by the hand pendant.
TRENDELENBURG POSITION:
To attain Trendelenburg position from any position, grasp the push handle with
one hand while actuating the emergency quick-release trigger with the other
(see fig. 2 detail below), then guide the back down fully and release the
trigger.
To return to seated position: Press the “up” button on the hand control
until the backrest is in the upright position.
Seat
Positioning: Hand Pendant All seat positioning, with the exception of the
Shock Position, must be controlled by the hand pendant.
The hand pendant has four buttons that control the positioning of the chair.
The two top buttons control the backrest and the two bottom buttons control
the legrest (Fig. 13).
Caster Operation
WARNING: Caster wheels must be locked except during transport.
Info: This recliner is equipped with four swivel casters for improved
maneuverability.
The rear casters are also equipped with locking features.
To lock rear caster: To prevent the rear casters from swiveling and the
wheel from turning, step on the rear pedal of the caster until it latches.
To unlock rear caster: Step on middle button to unlatch.
Table Operation
To raise table: Raise the table all the way up and slide in toward the
armrest until tabs engage slots in mount bracket. Press down lightly on table
to ensure lock tabs are engaged.
To lower table: Lift table up until tabs disengage slots and pull table
away from armrest and lower table to selfstoring position.
PRODUCT SPECIFICATIONS
Upholstery| Flame and stain resistant vinyl upholstery meets CA-TB117-2013
Specifications
---|---
Power| 120Vac, 60 Hz, Max. 2.7A
Class| Class 1 Equipment; Type B
Total Weight| 264 lb. (120 kg)
Overall Dimensions| 54.06″ H x 57.95″ L x 36.89″ W (1373 mm H x 1472 mm L x
937 mm W)
Maximum Weight Capacity for Motorized Recliner| Recliner: 700 lb. (318 kg),
EVENLY DISTRIBUTED
Upholstered Back: 440 lb. (200 kg), EVENLY DISTRIBUTED
Upholstered Seat: 660 lb. (300 kg), EVENLY DISTRIBUTED
Upholstered Armrest: 110 lb. (50 kg), EVENLY DISTRIBUTED
Upholstered Legrest: 110 lb. (50 kg), EVENLY DISTRIBUTED
Side Table: 22 lb. (10 kg), EVENLY DISTRIBUTED
All weight capacities and maximum loads are to be evenly distributed over
component / product as described above
Applied Parts| Upholstered Back, Upholstered Headrest, Upholstered Seat,
Upholstered Armrest, Side Table, Upholstered Legrest, and Footrest
Operator Accessible Parts| Applied Parts, Power Supply Cord, Pendant Control,
Padded Push Handle, and Switch Wheels Assembly
Operation Environment (Normal Use)| An ambient temperature range of 50°F to
104°F (10°C to 40°C)
A relative humidity range of 30% to 75%
An atmospheric pressure range of 10.15 psi to 15.37 psi (70 kPa to 106 kPa)
Transport and storage| This equipment shall be capable, while packed for
transport or storage, of being exposed for a period not exceeding 15 weeks to
environmental conditions not outside the following ranges:| an ambient
temperature range of -4°F to +140°F (-20°C to +60°C)
a relative humidity range of 0% to 85%, including condensation
an atmospheric pressure range of 7.25 psi to 15.37 psi (50 kPa to 106 kPa)
Disposal| Follow all applicable federal, state, and local laws and regulations
to dispose of the unit
Battery| 24V 1.3A BA16 battery: LINAK BA1616121100
Fuse| 250 V 6.0A (Use listed fuse ONLY)
| MEDICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY IN ACCORDANCE WITH UL 60601-17
MAINTENANCE
WARNING: Inspect all components periodically and replace as needed.
Electronics, Actuator, and Batteries
WARNING: The electronics and actuator contain no serviceable components.
If service is required, the unit must be returned to the factory for repair.
DO NOT attempt to open the electronics or actuator or obtain local service;
this may void the warranty and may result in damage to the unit. Consult your
Graham-Field authorized distributor for further information.
WARNING: Do not open the batteries. If service is required, consult your
Graham-Field authorized distributor. Replace the batteries when damaged,
faulty, or every two years.
EVERY THREE MONTHS
- Check and clean casters.
- Tighten bolts in each leg that secure caster into frame.
- Check caster brakes.
- Check recliner mechanism and actuator movement for operation in all positions, including Trendelenburg Position, with assistant seated in recliner.
- Check the movement of the recline mechanism for any squeaking noise. If you hear squeaking, apply lithium grease to the white rollers on the mechanism.
EVERY DAY
- Check the battery back-up to ensure that it is charged
- There is LED indicator on the battery
LED Indication of operation
- Solid yellow – Charging (battery not ready)
- No LED light – Fully charged (battery ready)
- Flashing yellow – Error during charging
The battery back-up has a Buzzer functionality
The buzzer will make a warning when a button on the hand control is pressed
and the battery capacity is low. If the battery capacity is low ensure that
plug is plugged into the wall and also is plugged into the power plug on the
left side of the recliner (see figure 16)
Cleaning for Upholstered Parts (Non-Live Parts)
Recliner upholstery requires little maintenance. Everyday dirt can usually be
removed using a clean, soft cloth or sponge with mild soap and water. Spills
and accidents require immediate attention for best results.
Always begin cleaning by using steps 1 or 2, then progress to the stronger
cleaners if needed. Most soil can be easily removed with soap and water or the
concentrated cleaners suggested in step 2. If use of a disinfectant is
desired, the Centers for Disease Control and Prevention recommends using a
solution consisting of 1/4 cup of bleach to 1 gallon of water, then rinsing
the solution thoroughly from the surface with clear water. Not rinsing
thoroughly and/or using stronger mixtures will likely result in discoloration
of the upholstery.
Staining Agent | Step | Cleaning Procedure |
---|---|---|
Eye Shadow / Mascara | 1 | Remove excess spill with clean, damp cloth. Clean |
with a 1:1 solution of mild dish detergent (such as Ivory® liquid) and water.
Rinse with clear water and dry with a clean cloth.
Grease / Suntan Lotion / Chocolate
Crayon / Black Felt Tip Pen
Blood / Urine / Feces| 2| Use undiluted application of concentrated
antibacterial cleaner (such as Formula 409® or Fantastik® Spray Cleaner), then
wipe with a clean cloth.
Red Lipstick
Oil Base Paint / Tar / Asphalt| 3| Use a 1:1 solution of ammonia and water or
a 1:4 solution of bleach and water, rinse with clear water, and dry with a
clean cloth.
Other Tough Stains| 4| Use undiluted application of naptha (lighter fluid).
Saturate a soft, clean cotton cloth with this solution. Rub the stain in a
circle 10 times. Rinse thoroughly with clear water. Use another soft, clean
cotton cloth to pat surface dry. Check results. If stain remains, proceed to
step 5.
5| Use 1:1 solution of isopropyl alcohol and water. Saturate a soft, clean
cotton cloth with this solution. Rub the stain in a circle 10 times. Rinse
thoroughly with clear water. Use another soft, clean cotton cloth to pat
surface dry. Check results.
If stain remains, use undiluted isopropyl alcohol. Saturate a soft, clean
cotton cloth with this solution. Rub the stain in a circle 10 times. Rinse
thoroughly with clear water. Use another soft, clean cotton cloth to pat
surface dry. Check results.
If stain remains, use 1:1 solution of acetone and water. Saturate a soft,
clean cotton cloth with this solution. Rub the stain in a circle 10 times.
Rinse thoroughly with clear water. Use another soft, clean cotton cloth to pat
surface dry.
Service
WARNING: Consult your Graham-Field authorized distributor for further
information.
LIMITED WARRANTY
SCOPE OF WARRANTY
GF Health Products, Inc. (“GF”) warrants to the original purchaser only that
it will replace or repair components, at GF’s sole discretion, that are
defective in material or workmanship under normal use and service. All
warranties are conditioned upon the proper use of the products strictly in
accordance with good commercial practice and applicable GF instructions and
manuals, including proper use and maintenance. To the extent that a third
party warrants a component, GF conveys all of its rights under that warranty
to the original purchaser, to the extent permitted. This limited warranty
shall only apply to defects that are reported in accordance with the
provisions set forth in this warranty document, within the applicable warranty
period and which, upon examination by GF or its authorized representative,
prove to be a warranty item. (See Obtaining Warranty Service below) This
limited warranty is not transferable.
The warranted components and time period are set forth below:
Frame: ……………………………………………… ten years
Casters:………………………………………………….. one year
Electronics, motors, electrical components: …………………….. one year
Upholstery: …………………………………… one year
All other durable components not listed above:………… three years
- Labor is not included in the warranty.
† Upholstery is only warranted on material supplied by GF.
‡ The warranty period is as designated above. If a part is replaced under warranty, the original warranty period will not be affected. All other replacement parts will be subject to the warranty period specified.
The applicable warranty period shall commence from date of shipment to the original customer, unless there is an expiration date on the component in which case the warranty shall expire on the earlier of warranty period or the expiration date.
OBTAINING WARRANTY SERVICE
Customers located in the United States who wish to report a warranty issue,
must contact GF directly by calling 1.770.368.4700 or by e-mailing a request
to cs@grahamfield.com. Customers located outside
the United States must contact the Distributor from whom they purchased the
products. In both cases, further directions will be provided once the initial
contact is made. This limited warranty shall only apply to defects that are
reported within the applicable warranty period. Failure to abide by the
specific directions will result in denial of the warranty claim.
The warranty does not cover and GF shall not be liable for the following:
- Defects, damage, or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage, tampering or failure to seek and obtain repair or replacement in a timely manner;
- Products which are not installed, used, or properly cleaned and maintained as required in the official manual for the applicable product; Products considered to be of a non-durable nature including, but not limited to: filters, fuses, gaskets, lubricants, and charts;
- Accessories or parts not provided by GF;
- Matching of color, grain or texture except to commercially acceptable standards;
- Changes in color caused by natural or artificial light:
- Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such products which are not expressly authorized in writing, in advance, by GF;
- Any labor or shipping charges incurred in the replacement part installation or repair;
- Costs and expenses of regular maintenance and cleaning; and
- Representations and warranties made by any person or entity other than GF.
ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL DAMAGES DISCLAIMER
THIS WARRANTY IS GF’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED. GF MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IF
ANY MODEL OR SAMPLE WAS SHOWN TO THE CUSTOMER, SUCH MODEL OR SAMPLE WAS USED
MERELY TO ILLUSTRATE THE GENERAL TYPE AND QUALITY OF THE PRODUCT AND NOT TO
REPRESENT THAT THE PRODUCT WOULD NECESSARILY CONFORM TO THE MODEL OR SAMPLE IN
ALL RESPECTS. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF THE
DEFECTIVE PARTS. GF SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT,
SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES, INCLUDING,
BUT NOT LIMITED TO: DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE,
DOWNTIME, COVER, OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND
BENEFITS. CERTAIN STATES MAY CONFER ADDITIONAL RIGHTS REGARDING WARRANTIES AND
IN THOSE STATES GF’S LIABILITY AND THE LIABILITY OF GF’S SUPPLIERS, SHALL BE
LIMITED TO THE FULLEST EXTENT PERMITTED BY LAW.
The warranties contained herein contain all the representations and warranties
with respect to the subject matter of this document, and supersede all prior
negotiations, agreements and understandings with respect thereto. The
recipient of this document hereby acknowledges and represents that it has not
relied on any representation, assertion, guarantee, warranty, collateral
contract or other assurance, except those set out in this document.
For additional information on this product or this warranty, please contact a
GF Customer Service Representative.
NOTES:
- Additional terms and conditions may apply.
- Freight claims must be notated on the appropriate shipping documents and must be made with immediacy. International, federal and state regulations govern specific requirements for freight claims. Failure to abide by those regulations may result in a denial of the freight claim. GF will assist you in filing the freight claim.
- Claims for any short shipment must be made within three (3) days of the invoice date.
Manufactured For:
LX_588W-INS-LAB-RevF23
GF Health Products, Inc.
One Graham-Field Way, Atlanta, GA 30340
© 2007, GF Health Products, Inc. All Rights Reserved.
Lumex is a registered trademark of GF Health Products, Inc.
Made in Taiwanwww.grahamfield.com
+1 770.368.4700
Information contained herein is subject to change without notice.
The most current and complete product information can be found on our website.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>