eyc THS13-14 Indoor Duct Temperature and Humidity Transmitter Instruction Manual
- June 4, 2024
- eyc
Table of Contents
C2107-THS1314
Temp. & Humidity-THS13/14
Temperature & Humidity Transmitter for Wall type/ Duct type
www.eyc-tech.com
Features
- Compact design, good look and easy to install
- An economical choice, high C/P ratio
- High-Tech sensor component with high accuracy and long-term stability
Applications
Monitoring for HVAC process / Air conditioning / Environmental ventilation
control /
Environmental monitoring for building / Factory / Clean room / Lab. /
Monitoring for storeroom / Crisper / Agriculture / Food industry
Specification
Input
Input type | Capacitive humidity sensor & Pt1000 |
---|
Measuring range
Temperature | 0 … 50°C, Suggestion:15 … 35°C |
---|---|
Humidity | 0 … 100 %RH(Non-condensing) |
Suggestion:20 … 80%RH
Accuracy (at 25°C)
Temperature | 0.3°C+0.002°C x tactual |
---|---|
Humidity | ±3%RH(30…80%RH) |
Output
Output | 4 … 20 mA |
---|---|
Signal connection | 2-wire |
Warm-up time | <2 min., stable time 20 mins. |
Response time | t63(15 … 45°C / 33 … 75 % RH)≦10 sec. |
t90(at +25°C)<20 sec.
Sampling time| about 3 sec.
Temperature influence| 0.2%RH / °C
Load resistance| 2-wire current :<(V-10) / 0.02<500 Ω
Output calibration(ZERO &SPAN)| adjustment range
10% of zero-point and full-range
Environment
Medium | Air |
---|---|
Operating Temp. | 0 … +50°C |
Storage Temp. | -25 … +60°C |
Electrical
Power supply | 2-wire: DC 11 … 35 V |
---|---|
Current consumption | 2-wire: DC:41 mA |
Electrical connection | 2-wire:3P connection |
Installation
Installation | Wall type / Duct type |
---|
Protection
IP Rating | IP54(Sensor: IP20) |
---|---|
Electric protection | ■ Polarity protection |
■ Over-voltage
■ Short circuit
Certification
Certification| CE
Material
Housing| PC fire-proof class(PC-110)(UL94V-2)
Weight| THS13:132 g / THS14:157 g
Diagram
Dimension
Ordering Guide
Additional Option Test Report
For more detailed information please contact us.
ILAC / TAF
YUDEN-TECH CO.,LTD. Calibration Laboratory – ( ILAC / TAF ) Test report.
(TAF accreditation:3032, complying with ISO / IEC 17025 ) TAF has mutual
recognition arrangement with ILAC MRA
Project | Measurand level or range |
---|---|
Hygrometer | Temperature:0°C … 69.5°C |
Humidity:10% … 95%
RTD / Thermocouple| Resistance thermometer:-40°C … +300°C
Thermocouple thermometer:-20°C … +300°C
References