Aseptico AEU-525CF Portable Dental System User Manual
- October 30, 2023
- Aseptico
Table of Contents
- AEU-525CF Portable Dental System
- PACKAGE CONTENTS:
- SAFETY PRECAUTIONS:
- SETTING UP THE UNIT(*):
- OPERATION:
- STERILIZATION AND MAINTENANCE:
- AIR FILTER SETUP INSTRUCTIONS:
- TRANSPORT III REPACKING INSTRUCTIONS:
- TROUBLESHOOTING:
- REPLACEMENT PARTS LIST
- SYMBOL DEFINITIONS:
- SPECIFICATIONS:
- WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
AEU-525CF AEU-525CFH
Portable Dental System
Transport III
OPERATION & MAINTENANCE MANUAL
AEU-525CF Portable Dental System
Classifications in accordance with ANSI/AAMI ES60601-1:
- Class I
- Types B and BF Applied Parts
- Ordinary Protection
- Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Note: For reliable grounding, connect to receptacle marked “Hospital
Grade”.
Note: The emissions characteristics of this equipment make it suitable
for use in industrial areas and hospitals (CISPR 11class A). If it is used in
a residential environment (for which CISPR 11 class B is normally required)
this equipment might not offer adequate protection to radio-frequency
communication services.
The user might need to take mitigation measures, such as relocating or re-
orienting the equipment.
INDICATIONS FOR USE:
The AEU-525CF/AEU-525CFH Transport III is a portable selfcontained dental
system that is used for endodontic and general dentistry applications.
EXPECTED SERVICE LIFE:
7 years,
ESSENTIAL PERFORMANCE:
This device does not have ESSENTIAL PERFORMANCE as defined in IEC 60601-1.
RX: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A DENTIST .
P.O. Box 1548 • Woodinville, WA 98072
8333 216 th Street S.E. • Woodinville, WA 98072
(425) 487-3157 • (800) 426-5913
www.aseptico.com • i
To prevent injury to people and damage to property, please heed relevant
warnings and remarks. They are marked as follows:
WARNING: Serious injury or death may result if ignored.
CAUTION: Damage to property or the environment may result if ignored.
NOTE: Important additional information and hints.
This device has been tested and found to comply with the emissions
requirements of IEC 60601-1-2:2014. These requirements provide reasonable
protection against harmful electromagnetic interference in a typical medical
installation.
However, high levels of radio-frequency (RF) emissions from electrical
devices, such as cellular phones, may disrupt the performance of this device.
To mitigate disruptive electromagnetic interference, position this device away
from RF transmitters and other sources of electromagnetic energy.
MEDICAL GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY, IN ACCORDANCE WITH: ANSI/AAMI ES606011:2005/(R)2012
and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical
equipment—Part 1: General requirements for basic safety and essential
performance.
CAN/CSA C22.2 NO. 60601108 (2013) Medical Electrical Equipment Part 1:
General Requirements for Basic Safety and Essential Performance (Adopted IEC
606011:2005, third edition, 200512), Includes Corrigendum 1:2011.
Any malfunction or deterioration in the characteristics and/or performance of this device, as well as any inadequacy in its labelling or the instructions for use which might lead to or might have led to the death of a patient or user or to a serious deterioration in his state of health (e.g. Serious Adverse Event); this information shall be reported immediately to Aseptico Inc. and/or the Authorized Representative (within the European Union)
Your new Aseptico AEU-525CF/AEU-525CFH Transport III Dental System is the finest portable electric dental system available. The System features a high- torque brushless micro motor with digital controls for the handpiece ratios, speed, torque, motor direction, and LED illumination. The dual-voltage system includes an ultrasonic scaler, 3-way air/water syringe, HVE and saliva ejector vacuum systems, a self contained water system, self contained waste system with a near-full level warning and vacuum pump safety shutoff, oilless air compressor, and vacuum pump. The system comes in a case with built-in wheels and a handle for maximum portability.
Congratulations!
This Transport III System is engineered to provide many years of reliable
service. Please read the instructions provided in this manual to receive the
best and longest service from your Aseptico equipment. Separate manuals may be
provided to cover the operation and maintenance of other accessories for your
unit.
PACKAGE CONTENTS:
- Portable Case with Electronic Control, Compressor, and Vacuum Pump
- Electric Motor Control Console includes an Autoclavable 40K Brushless Micro Motor with ISO E-type coupling and LED light source
- Ultrasonic Scaler and 3 Instrument Tips
- TA-90D 3-Way Air/Water Syringe
- HVE and Saliva Ejector Vacuum Hoses with Valves
- Water Supply and Air Reserve Bottles
- AE-7PM On/Off Foot Control
- Power Cords (One each U.S. and Euro)
- Operation & Maintenance Manual
- Air Filter Assembly (PN: 330634)
AEU-525CF Version:
• Handpieces purchased separately
AEU-525CFH Version:
- AHP-72MK-FO Highspeed 1:5 Increaser Contra-Angle Handpiece with Fiber Optic (Qty 3)
- AHP-63MK Lowspeed 1:1 Contra-Angle Handpiece (Qty 3)
- AHP-64 Lowspeed 1:1 Straight Handpiece (Qty 3)
SAFETY PRECAUTIONS:
Aseptico accepts no liability for direct or consequential injury or damage
resulting from improper use, arising in particular through the non-observance
of the operating instructions, or improper preparation and maintenance.
WARNING: Clean, disinfect, and sterilize new or repaired handpieces and
instruments before first use and between each patient use. Only use sterilized
handpieces and instruments during treatment. Non-sterilized handpieces and
instruments may cause bacterial or viral infections. Always sterilize
handpieces and instruments after operation.
CAUTION: Always examine unit components for damage before commencing
treatment. Damaged components must not be used and must be replaced.
WARNING: Use for intended purposes only. Failure to observe the operating
instructions may result in the patient or user suffering serious injury or the
product being damaged, possibly beyond repair. Before using this product, make
sure that you have studied and understood the operating instructions.
WARNING: Do not install where there is a risk of an explosion. The 525CF
Transport III System is not intended for operation in the presence of
flammable anesthetics or gases.
WARNING: CONTRAINDICATIONS Ultrasonic oscillations emitted by the
Ultrasonic Scaler may prevent the proper function of cardiac pacemakers.
Therefore, Aseptico recommends that patients with a cardiac pacemaker should
not be treated with the Ultrasonic Scaler component of the AEU-525CF/AEU-
525CFH Transport III System.
WARNING: The Ultrasonic Scaler instrument tips oscillate at high
frequency and can fracture during operation. To help prevent the tips from
fracturing and possibly injuring the patient, always follow the scaler
manufacturer’s operating instructions and recommended ultrasonic power
settings.
CAUTION: The lens for the electric motor LED is soft and can be damaged.
If the lens needs to be cleaned, use a lintfree swab and isopropyl alcohol –
do not use other solvents as they might adversely react with the LED assembly.
CAUTION: The AEU-525CF/AEU-525CFH Transport III electric motor is not
recommended for use with endodontic files that have torque limit requirements
less than 100 g-cm.
WARNING: Do not use this device for dental implant procedures.
CAUTION: Do not run saline solutions through the water system — saline
will rust the water filters.
WARNING: For use by qualified and trained personnel only.
CAUTION: Operating the AEU-525CF/AEU-525CFH without using the High or Low
Volume Evacuation can cause the compressor motor to cycle on/off, possibly
affecting other electrically powered equipment sharing the same circuit.
CAUTION: Dispose of electronic waste and waste containing amalgam in
accordance with local regulations.
WARNING: Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the unit, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.
WARNING: Do not stare into the handmotor LED! The light from the hand
motor LED may cause eye injury.
WARNING: Do not stare into the scaler leds! The light from the scaler
LEDs may cause eye injury.
WARNING: Under certain conditions of use, the tip of the dental drill,
handpiece, or hand motor may exceed 43 °C (109 °C) and can cause thermal
injury to the patient. To avoid the risk of overheating, observe the following
practices:
- Always operate a high-speed handpiece with water coolant.
- Do not operate the handmotor continuously for long periods of time. Use the appropriate duty cycle stated for the hand motor and handpiece selected.
- Only touch the bur or file to the tooth of the patient; avoid contacting other parts of the patient with the device.
- Ensure that the ratio setting is set to match the gear ratio of the handpiece being used.
- Thoroughly clean and lubricate the E-type handpieces after each use and before sterilization.
- Follow the instructions provided with the handpiece from the manufacturer for complete operating and maintenance instructions.
WARNING: Under certain conditions of use, the tip of the optional ultrasonic scaler may exceed 43 °C (109 °C) and can cause thermal injury to the patient. To avoid the risk of overheating, observe the following practices:
- Do not use the scaler dry. If used dry, the Instrument Tip will heat immediately.
- Verify that the scaler coolant flow available at the tip is no less than 20ml/min before use on a patient.
- Follow the scaler manufacturer’s operating instructions and recommended ultrasonic power settings for each tip.
- Never use the scaler on metal or porcelain restorations.
SETTING UP THE UNIT(*):
-
Unpack the Transport III System, small luggage case, air filter assembly from the shipping case and the bag of hoses from the cutout in the shipping case lid (Fig. 1).
-
Lay Transport III case flat on floor with case lid facing upward (Fig. 2). Open lid.
-
Remove accessories and place them aside for assembly later (Fig. 3).
-
Remove System components from case and place aside temporarily (Fig. 3):
a. Loosen water-bottle holding strap on inside of lid and remove water bottle.
b. Unhook the waste-tank bungee cord and remove waste tank.
c. Loosen air/electric-module holding strap and remove air/electric module assembly. -
Detach fastener tool from inside case lid (Fig 4a). Use the tool to loosen the four fasteners on the outside of the rear metal door (Fig. 4b). Turn fastener counterclockwise 1/4 turn, to loosen. Stand unit upright. Use the fastener tool to loosen the two fasteners on the small metal door located on the top of the case (Fig. 4c).
Open case lid and reattach fastener tool to its Velcro ® mounting pad, then close and lock case lid. NOTE: This fastener tool can also be used to loosen the air/electric module hold-down screw shown in Figure 13b, page 5. -
On front (handle) side of unit, pull handle locking tab outward while raising the handle (Fig. 5a). Lift handle all the way up until the locking tab engages the bottom slot (Fig 5b).
(*) Also refer to the Aseptico Transport III video at www.aseptico.com for basic setup instructions. -
Open the small metal door on the top of the caseand locate the high and low vacuum hose connectors in the recess under the door. Pull both hoses all the way out of the recess so they extend to the top of the handle as shown in Figure 6.
-
(NOTE: The user may choose to interchange Step 8 with Step 9, to facilitate easier setup of the scaler and syringe.) On the back side of the unit, position the air/ electric module assembly over the top of the handle as shown in Figure 7a, then lower the assembly so its mounting bracket rests on the opening in the handle (Fig. 7b). Rotate the bottom of the assembly forward so that its locking knob aligns withand engages the vertical slot in the handle(Fig. 7b). Rotate the locking knob 90° to secure the air/electric assembly to the handle (Fig. 7c).
-
(NOTE: The user might have chosen to interchange Step 9 with Step 8, for easier setup of the scaler and syringe.) Place scaler and air/water syringe into their respective instrument holders on the air/electric module (refer to the location-ID label on top of the module). (See Fig. 8)
-
Install electric motor connector into receptacle located on bottom of air/electric module (Fig. 9). Align the round dimple on the motor cord connector with the mark on the receptacle and push cord straight into receptacle. To remove motor/cord, pull motor cord connector straight out of receptacle. Place motor into its instrument holder on the module and attach appropriate E-Type handpiece.
-
Install foot control connector into receptacle located on bottom of air/electric module (Fig. 10). Once the connector pins are properly engaged, tighten the outer sleeve to lock into place. Place foot control on floor.
-
Fill the water-supply bottle with clean water or other suitable irrigation fluid and attach to the water bottle cap on the left side of the air/electric module (Fig. 11a).
Optional Water-Supply Adapter – Attach the optional water-supply adapter (PN
- to the end of the water bottle filter tube (Fig. 11b). Attach the opposite end of the Adapter to the water supply. Toggle the water bottle purge switch to the purge position.
-
On the back side of the air/electric module, rotate the rear housing downward 180° (Fig. 12). (Note the connector block in the recess directly below the rear housing.)
-
Disengage the locking tab on the handle.
Slowly and carefully lower the air/electric module down until the rear housing engages the connector block in the recess below (Fig 13a). (NOTE: Ensure that the three O-rings on the housing are lubricated with an appropriate lubricant, preferably containing PTFE. Dry O-rings will damage
the seals, resulting in poor performance.) Lock the module into place by handtightening the hold-down thumb-screw on the housing, until snug (Fig 13b). -
Ensure that the lid on the waste container is securely latched onto the container. Hang the container onto the bracket located on the back side of case lid (Fig. 14a). Plug the level-sensor cord connector into the socket located on the motor housing (Fig. 14b). The connector plug and the socket are keyed and fit only in one position. Gently press the connector against the socket and rotate until the connector aligns with and enters the socket. NOTE: If the sensor is accidentally unplugged, a warning beep will sound and the compressor will automatically shut off.
-
Attach the HVE and saliva ejector hose connectors (white) to their respective ports on the top of the waste tank (Fig. 15a, page 6). Note that the notches in the hose connectors align to the white tabs on the tank. Rotate the white locking latch over the top of the two hose connectors to secure them into position (Fig. 15b, page 6). Place the HVE and saliva ejector valves into their respective instrument holders on
the air/electric module (refer to the location-ID label on top of the module). Attach the desired evacuator/saliva ejector tips into the valves. -
Attach the HVE and saliva ejector hose connectors (black) from Step # 6, page 3, to their respective ports on the top of the waste tank (Fig. 16). Rotate the black locking latch overtop the two hose connectors to secure them in place.
-
Toggle the water-bottle purge switch located on the back side of the air/ electricmodule to the “PRESSURE” position as shown (Fig. 17).
-
Verify that the voltage-selector switch located on the motor housing is set to the proper 115V or 230V power source (Fig. 18). Attach the power cord to the power inlet receptacle and plug into a grounded electrical outlet.
NOTE: Ensure that the power cord is correct for the voltage source in the country of use.
Turn both rocker switches on the motor housing to the ‘ON’ (I) position. The lefthand switch turns the unit power On/Off and also acts as a circuit breaker for the
whole unit. The right-hand switch acts as a circuit breaker for the compressor. Electrical overloads will trip the two switches. If either switch is found in a neutral position (midway between ‘ON’ and ‘OFF’), check for overload causes and correct the problem. Then, place the switch in the ‘ON’ position. Green LED’s adjacent to each switch illuminate when the switches are On. NOTE: When the unit is in use, it’s compressor motor will occasionally turn On and Off, to maintain proper pressure. -
Ensure that the system operating pressure is maintained at 45-55 PSI (3.10-3.79 bar). The pressure gauge is located on the side of the air/electric module (Fig. 19).
-
Turn all instrument holder and manifold control panel toggle switches to the ‘ON’ position: toggle the instrument holders toward the red dot (see Fig. 20); toggle the manifold panel switches to the UP position (see Fig. 19). NOTE: If the vacuum comes on, reseat the vacuum handpieces in their holders.
IMPORTANT: AIR FILTER SETUP
Refer to Air Filter Setup Instructions, Page 22, for information on installing
the air filter assembly onto the 525CF Unit.
OPERATION FUNCTIONS:
-
ON/OFF POWER SWITCH & CIRCUIT BREAKER (Fig. 24) – Controls power On/Off to the Transport III. When the unit is turned On, the green LED next to the power switch will illuminate.
-
COMPRESSOR CIRCUIT BREAKER SWITCH (Fig. 24) When the switch is On (circuit closed/continuity complete), the compressor will start and the green LED next to the switch will illuminate. When the switch trips open (continuity interrupted) due to a compressor overload, the switch and LED turn Off. Simply turn the switch back On to reset breaker. NOTE: If switch is found in a neutral position (midway between On and Off), find and correct the fault then turn the switch to the On position. IMPORTANT: The switch should be left in the On position when not in use.
-
ON/OFF FOOT CONTROL (Fig. 25) – The Foot Control provides On/Off operation to the motor, water, and air coolant to the handpiece, and the handpiece light, when the electric motor is removed from its holder and its water & air toggle switches are toggled ‘ON’. The Foot Control also provides On/Off operation to the ultrasonic scaler and water coolant to the scaler when the scaler is removed from its holder and the water toggle switch is toggled ‘ON’. For instrument operation, apply foot pressure to any part of the disk. The instrument holder toggle must be ‘On’ and the instrument removed from its holder before operation can begin. NOTE: The HVE and saliva ejector are not controlled by the Foot Control – they will automatically start when removed from their holders and their toggle switches are On.
-
INSTRUMENT CONTROLS (Figs 26 & 27) The Transport III includes controls for the motor, scaler, and vacuum valve holders. These holders include an On/Off toggle switch (Fig. 26) which when turned ‘Off’ (toggle to the left), keeps the instrument off, even when removed from its holder. When the toggle switch is turned ‘On’ (toggle toward the red dot), the instrument will automatically activate when removed from its holder.
-
ELECTRIC MOTOR AND CONTROL
PANEL (Fig. 28) – There are two different control modes that can be used to operate the Transport III electric motor: Manual Mode using the console buttons, or Preset Mode using the preprogrammed Presets:
A. MANUAL OPERATION – When the electric motor is not running, the user can adjust its ratio, speed, torque, motor direction, and motor LED settings using the control panel keypad (The Preset indicator in upper left of LCD display will blink on and off, indicating the change): -
Depress the Standby button to turn electric motor control panel on or off. If the console was turned off using the Standby button, the Display will darken and all the buttons, except the Standby button, will become inoperative. Press the Standby button again to turn on the control panel and return the motor to its last used settings.
-
Press the RATIO button repeatedly to select the ratio that matches the handpiece being used. Available ratios are: 1:5, 1:2, 1:1, 5:1, and 8:1.
NOTE: The user can reprogram the unit to display only a preferred set of ratio(s) from the five available options. Refer to page 12 for complete instructions on Customizing Ratios. -
Press the SPEED Up/Down buttons to select the desired operating speed for the handpiece being used. See Figure 29 for handpiece speed ranges.
-
Press the TORQUE p/Down buttons to select the desired operating torque percentage.
-
Press MOTOR DIRECTION button repeatedly to cycle through and select the Forward (FWD) or Reverse (REV) settings. An audible beep indicates reverse direction when the motor is running.
-
To switch the electric motor to Endodontic operating mode, first select the 8:1 handpiece ratio, then press the MOTOR DIRECTION button repeatedly until the word “ENDO” appears in the Motor Direction window. Next, select the desired Torque level using the UP/DOWN Torque buttons. The unit is then ready to operate in ENDO mode – the rotation of the handpiece will automatically alternate between forward and reverse when the selected torque is reached (in order to free the instrument).
-
Press and release the ILLUMINATION button to turn the handpiece LED On/Off. To activate the handpiece and LED, remove the handpiece from its holder, turn the Illumination button ‘On’, and press the foot pedal to operate the motor. When the foot pedal is released and the motor has stopped, the light will turn off after approximately 20 seconds. Press and hold the Illumination button for three seconds to enter the LED intensity adjustment mode. When in this mode, press the Torque Up/Down buttons to select the desired illumination in 10% increments, from 10% to 100%. The handpiece LED will automatically turn on and change intensity as the adjustments are made. The display icon will also blink as the adjustments are made. To exit the adjustment mode, press and release the Illumination button.
B. PRESET OPERATION – The Transport III electric motor module is shipped with five preprogrammed factory Presets for each ratio (see Fig. 31, page 13). This allows the user to instantly access a wide range of settings. After the handpiece ratio has been selected, press the Preset Button repeatedly to cycle through five different Presets (P1 – P5).
When the desired Preset is displayed, its operating parameters are automatically activated. Each Preset can store the following parameters:
– Bur/Drill Speed
– Motor Direction
– Torque Setting
– LED Light ‘On’ or ‘Off’
– Endo Mode Activation (8:1 handpiece ratio only)
NOTE: The electric motor is programmed at the factory to initially start up with Ratio 1:5 and Preset No.1 active. Subsequently, when the System power is turned on, the last settings used will be activated.
CUSTOMIZING RATIOS:
The user can enable or disable individual handpiece ratio options, so that
only preferred or commonly used ratios are displayed during the ratio
selection process. Follow these steps to reprogram the unit with customized
ratio options:
-
Press and hold the Ratio button to enter the Ratio Customization Mode.
-
Press the Ratio button repeatedly to switch between the available ratio options. Each Ratio displays in the Ratio window as the user cycles through the five options.
-
When the desired ratio is displayed, press the Preset button to toggle the ratio to either enabled or disabled status. The word “On” will appear in the Preset display window when
enabled, and two dashes “ ” will appear when disabled (see Figs. 30a & 30b). -
Repeat Step 3 above for the remaining four ratio options until only the desired ratios are enabled.
-
Press and hold the Ratio button to exit the Ratio Customization Mode. Only the enabled ratios will be displayed when selecting Ratio options; any ratio that has been disabled will not be displayed (unless you re-enable it).
NOTE: The system must always have at least one ratio option enabled. If the user attempts to disable the last operational ratio, the system will emit a beep sound and ignore the user’s command to disable this remaining option.
NOTE: If the factory defaults are restored, all customized Ratio option settings will be overwritten.
Figure 31 – Factory Default Presets
RATIO| PRESET NO.| SPEED (RPM)| DIRECTION| TORQUE|
ENDO MODE
(8:1 Ratio Only) ,| HANDPC
LED
---|---|---|---|---|---|---
1:05| 1| 200,000| FWD| 100%| —| On
2| 150,000| FWD| 100%| —| On
3| 150,000| FWD| 50 % Torque Limit| —| On
4| 100,000| FWD| 100%| —| On
5| 100,000| FWD| 50 % Torque Limit| —| On
1:02| 1| 80,000| FWD| 100%| —| Off
2| 60,000| FWD| 100%| —| Off
3| 60,000| FWD| 50 % Torque Limit| —| Off
4| 40,000| FWD| 100%| —| Off
5| 40,000| FWD| 50 % Torque Limit| —| Off
1:01| 1| 40,000| FWD| 100%| —| On
2| 40,000| FWD| 50 % Torque Limit| —| On
3| 20,000| FWD| 100%| —| On
4| 10,000| FWD| 100%| —| On
5| 8,000| FWD| 100%| —| On
5:01| 1| 2,000| FWD| 100%| —| On
2| 1,500| FWD| 100%| —| On
3| 1,200| FWD| 100%| —| On
4| 800| FWD| 100%| —| On
5| 400| FWD| 100%| —| On
8:1 (ENDO
MODE)| 1| 300| FWD| 100%| YES| Off
2| 300| FWD| 70%| YES| Off
3| 300| FWD| 50%| YES| Off
4| 300| FWD| 30%| YES| Off
5| 300| FWD| 10%| YES| Off
CUSTOMIZING PRESETS:
Each Preset can be customized by the user with its own unique set of
operating parameters:
1. First select the Ratio, then select the Preset number of the Preset you
wish to customize.
2. Adjust the Preset’s speed, torque, handpiece illumination, and motor
direction as desired. NOTE: The ENDO Mode feature can only be enabled when an
8:1 ratio is selected. The Preset indicator will blink on and off, indicating
a change is in process.
3. Press and hold the Preset button to save its new settings (a beep will
sound).
4. Press the foot pedal to activate the motor and begin operation.
FACTORY DEFAULTS: The electric motor module will retain the factory
default settings (Fig. 31) in memory until changed by the user. To recall the
original factory defaults, press and hold the PRESET and RATIO buttons
simultaneously for approximately 3 seconds. IMPORTANT: When factory
defaults are recalled, all usercustomized Ratio and Preset settings will be
overwritten.
SLEEP MODE:
After 20 minutes of motor inactivity, the electric motor module
automatically enables an energy-saving Sleep Mode which turns off the display.
When in this mode, three square symbols will blink consecutively across the
darkened LCD. To turn the unit on again and light up the display, the user can
either press the standby button on the control panel or remove the handpiece
from its holder and then press on the foot pedal. The last used settings will
be restored.
-
SCALER (Fig. 32) – The Transport III includes a piezoelectric scaler system that provides adjustable ultrasonic levels and a water coolant system. The scaler includes 3 instrument tips: The #37 instrument is specially developed for subgingival scaling, furcations, supragingival fine scaling and spot removal. The #38 tip is used for lingual and buccal subgingival scaling and furcations. The #39 tip is used for universal lingual and buccal supragingival scaling. The scaler is controlled by the scaler ultrasonic setting adjustment knob which is located on the left side of the air/electric module, on the manifold control panel (see Fig. 32). The ultrasonic intensity can be adjusted by turning the control knob clockwise to maximum or counterclockwise to minimum. When scaling, follow the tip manufacturer’s recommended ultrasonic settings for each tip. The coolant level is controlled by the scaler water On/Off toggle switch and the coolant flow adjustment knob, located just below the ultrasonic adjustment knob (see Fig. 32). Turn knob counter-clockwise to increase coolant flow; or clockwise to decrease coolant flow.
Verify that scaler coolant flow of no less than 20ml/min is available at the tip.
SEE ACCOMPANYING SCALER DOCUMENTS
WARNING:
**** Ensure proper coolant flow before using Scaler on a patient. Heat emitted by the Scaler may damage teeth if coolant water is not atomized at the Scaler tip. -
3-WAY AIR/WATER SYRINGE (Fig. 33a)
• Pressing the left button dispenses water.
• Pressing the right button dispenses air.
• Pressing both buttons simultaneously dispenses an air/water mist. This mist can be adjusted with the two syringe flow control knobs located on the manifold control panel (see Fig. 33b). The top knob adjusts the water flow and the bottom knob adusts the air flow. Turn the knobs clockwise to decrease the flow or counterclockwise to increase the flow. -
HANDPIECE CONTROLS: Water – The handpiece water toggle valve allows water to flow to the working handpiece when the foot pedal is depressed. The toggle valve is located on the manifold control panel (see Fig. 34). Place the toggle in the upward position to turn handpiece water On or downward to turn it Off. A water control valve is also provided to adjust the volume of water flow o the handpiece. This valve is located next to toggle valve (Fig. 34).
Air – The handpiece air toggle valve allows air to flow to the working handpiece when the foot pedal is depressed. The toggle valve is located on the manifold control panel (see Fig. 34). Place the toggle in the upward position to turn handpiece air On or downward to turn it Off. An air flow control valve is also provided to adjust the volume of air to the handpiece. This valve is located next to the toggle valve (Fig. 34). -
WATER SUPPLY BOTTLE (Fig. 35) – The Transport III incorporates a self-contained pressurized water system. The system consists of one 1-liter bottle with a pressure release toggle valve that allows pressure to be released from the bottle while maintaining system pressure. To refill the water supply bottle:
a.) Toggle the bottle pressure relief valve to “PURGE” (see Fig. 35) to remove pressure from the water bottle.
b.) Unscrew the bottle from its cap.
c.) Fill bottle with water or other nonsaline solution.
d.) Screw bottle back into its cap.
e.) Toggle the pressure relief valve back to “PRESSURE” to repressurize the bottle.
WARNING: Before removing the Water Supply Bottle, toggle the water pressure relief valve to the “PURGE” position, allowing pressure to release. -
AIR BOTTLE (Fig. 36) – The 750ml air bottle is located inside the upper right corner of the case and stores pressurized air for the 3-way air/water syringe. The bottle includes a filter equipped with a drain plug that allows condensation to be removed. To drain condensation, place a container or towel or below the filter and use pliers to carefully loosen the black knob on the bottom of the filter.
IMPORTANT: To open the drain, turn the black knob clockwise (follow arrow on “DRAIN → “ label); to close drain, turn the knob counter-clockwise. Do not overtighten.
WARNING:
Before removing the Air Bottle, vent all reserve air pressure in the bottle. Press the right (Air) button on the 3-Way Air/ Water Syringe, or turn either the HVE or LVE toggle switches ‘On’ with unit power off, to depressurize the air. -
HVE AND SALIVA EJECTOR VACUUMS
The unit is equipped with a high volume evacuator (HVE) and low volume saliva ejector system (Fig. 37). Both high and low volume systems can operate simultaneously. When the vacuum holder toggle switches are toggled Off (to left side), the vacuum system will stay off, even when the vacuum valves are removed from their holders. When the vacuum holder toggle switches are toggled On (toward the red dot), the vacuum will start automatically when the vacuum valves are removed from their holders. -
WASTE CONTAINER — When the HVE or saliva ejector instruments are in use, the vacuum pump will run and waste from the vacuum system will collect in the 1.8 liter plastic waste container located on the back side of the unit (Fig. 38). The container provides two separate waste compartments: one for HVE and the other for the saliva ejector. (The HVE compartment contains a vacuum-intake screen for collecting large debris.) Each compartment includes its own 85%-full sensor. When either one of these sensors reaches the 85%-full level, an audible alert will start to
IMPORTANT:
To prevent the HVE and Low Volume (Saliva Ejector) compressor system from running excessively, ensure that the vacuum heads are firmly cradled in their holders after each use and that their control levers are toggled ‘Off’’.
beep and a red LED on the tank lid will flash (see Fig. 38). If both compartments continue to fill, the compressor/cacuum pump will automatically shut down before either compartment reaches 100% full. The alert will sound until the container is emptied – To empty Waste Container:
1. Disconnect vacuum hoses and level sensor wire connector.
2. Detach waste container assembly from unit and carry container and contents to proper waste disposal location.
3. Unsnap lid and empty both container compartments (side-pour out), then rinse out container.
4. Replace lid, reattach container assembly to unit, and reconnect hoses and level sensor. -
AUXILIARY WATER OUTPUT (Fig. 21)– Water or other solutions placed in the water bottle can be withdrawn through the auxiliary water output using an Aseptico fitting (PN AA-60) attached to tubing. The output has an auto shutoff to prevent leakage.
OPERATION:
-
After the Transport III unit has been set up and you have made yourself familiar with the operation functions, you are ready for operation as follows:
-
Turn the power and circuit breaker switches on the rear of the case to the ‘On’ (I) position. The green LEDs next to the switches and the electric motor control panel display will light up, and the unit will pressurize. NOTE: If electric motor control panel doesn’t illuminate, press the Standby Button to turn the panel On.
-
Toggle the instrument holder toggle switch of the instrument to be used to the ‘On’ position (pointed toward the red dot).
-
Check the pressure gauge on the side of the air/electric module and ensure that the system operating pressure is within the acceptable 45 – 55 PSI range.
-
Using the Electric Motor:
a. Attach the proper sterile E-type handpiece to the motor and set the RATIO Selector to match the gear ratio of the handpiece being used (i.e. 1:5 for a 1:5 increaser handpiece, or 8:1 for a 8:1 reduction handpiece).
b. Select the desired preprogrammed PRESET on the display panel, OR, manually adjust the speed, torque, motor
direction, electric motor LED, or ENDO settings using the panel keypad: SPEED – Adjust the up/down SPEED arrows on the display to the desired RPM. NOTE: When selecting the SPEED, ensure that the Ratio Selector matches the handpiece ratio.The RPM display will indicate bur speed when the Ratio Selector is set to the proper gear ratio.
TORQUE – Adjust the up/down TORQUE arrows on the display to the desired torque.
MOTOR DIRECTION – Select either forward (FWD) or reverse (REV) direction on the display.
ILLUMINATION – If using a handpiece with LED illumination, turn the LED button on the display On or Off.
ENDO MODE – To select endodontic operating mode (8:1 handpiece ratio must first be selected) press the Motor Direction button repeatedly until “ENDO” appears in the Motor Direction window. -
Handpiece Water/Air:
Toggle the handpiece water and air toggle valves, located on the side of the air/electric module, to the On position (pointing upward). Adjust the handpiece water and air flow control knobs to the desired settings. (Turn clockwise to decrease flow, counterclockwise to increase.)
NOTE: Some compressed air is used to cool the motor and handpiece. This air vents around the LED and axially through the E-coupling to the handpiece.
IMPORTANT: Turn Off the handpiece water and air when using handpieces that do not support internal irrigation. -
Using the Ultrasonic Scaler: See accompanying documents.
a. Toggle the Scaler Water Toggle Switch, located on the side of the air/electric module, to the On position (pointing upward).
b. Screw on a sterile Instrument Tip to the Scaler handpiece and tighten it firmly. Turn carefully backwards approximately 1mm. This will put the tip in a neutral position so it can move freely.
IMPORTANT: Rinse the water line before attaching tip by pressing the foot switch and allowing the water to flow for 30 seconds.
c. Turn the Scaler Water Flow Control knob to the desired setting (clockwise to decrease flow, counterclockwise to increase). The water should create a mist around the Scaler tip.
d. Adjust the Scaler Ultrasonic Setting knob to the desired setting.
WARNING: Ensure proper coolant flow before using Scaler on a patient. Heat emitted by the Scaler may damage teeth if coolant water is not atomized at the Scaler tip. -
Using the 3-Way Air/Water Syringe:
a. Use the syringe as necessary for irrigation or drying. Press left button for water, right button for air, or both buttons simultaneously for a mist. IMPORTANT: A plastic dental barrier sheath made for the 3-way syringe must be used on syringes that have not been autoclaved.
b. Adjust the water and air flow by using a screwdriver to turn the two small adjustment screws located on the side of the air/electric module. Upper screw controls water, lower screw controls air. Turn screws clockwise to decrease flows, counterclock-wise to increase flows. -
Using the HVE and Saliva Ejector Vacuums:
Adjust the vacuum pressure on the HVE (high vacuum) and saliva ejector (low vacuum) by moving the control levers on their respective valve heads. -
Remove the instrument to be used from its holder.
-
Press the foot control to activate motor or scaler and begin operation.
-
Empty the waste container when the 85%full alert sounds or when the red LED located on the container lid flashes.
NOTE: The unit’s vacuum system will automatically shut down if the waste tank is not promptly emptied. The audible alert will also continue to sound until the container is emptied.
STERILIZATION AND MAINTENANCE:
Because of its simple design, the Aseptico AEU-525CF/AEU-525CFH Transport III
System requires very little maintenance.
Any maintenance that is needed can be performed in minutes.
PURGING THE SYSTEM:
If the unit will not be used for an extended period of time, or if the unit
might be subjected to freezing conditions, the user should purge the system of
all water.
Simply empty the contents of the water bottle and install it back into its
cap, then operate the air/water syringe, scaler, and handpiece with water
coolant ‘ON’ until only air comes through the water lines.
Pack the unit and store as normal.
HANDPIECES:
Thorough cleaning and lubrication of Etype handpieces after each use and
before sterilization is very important to ensure proper operation and service
life of the handpiece. Follow the instructions provided with the handpiece for
complete maintenance instructions.
IMPORTANT! Protect motor from excess oil draining from handpieces. After
lubricating and before autoclaving, stand handpiece on its base, on a paper
towel, and allow excess oil to drain (see Fig. 39).
ELECTRIC MOTOR & CORD ASSEMBLY:
The entire Electric Motor and Cord Assembly is fully autoclavable (Fig. 40).
Steam autoclave motor/cord assembly at 132° C (270° F) for ten minutes.
Loosely coil the motor cord when autoclaving. Avoid sharply bending the cord
when autoclaving. Alternatively, wipe down the motor cord with disinfecting
solution, and/or sleeve the cord between each patient.
ELECTRIC MOTOR O-RINGS:
Replace electric motor O-rings when worn or damaged (see Fig. 41). Gently peel
old O-rings out of grooves and replace with
WARNING
- Do not attempt to disassemble the motor or motor connector.
- Do not oil or lubricate the motor.
- Do not attach a handpiece to the motor while the motor is running.
- Do not bend motor cord sharply.
Failure to comply with any of the above instructions may void your warranty.
CAUTION: The Electric Motor is sensitive to shock. Do not drop or impact motor against a hard surface.
CAUTION FOR ALL STERILIZATION:
- Do not exceed 135° C or 275° F
- Do not submerge in any solutions
- Do not use ultrasonic cleaners
new O-rings (PN 520069). Occasionally apply non-toxic lubricant (preferably containing PTFE) to O-rings to maintain flexibility.
LARGE O-RINGS (On Air/Electric Module and High/Low Vacuum Lines):
Keep all O-rings lubricated with PTFE lubricant. Periodically inspect rings
for damage or excessive wear — damaged or worn O-rings will create vacuum
leaks, degrading system performance.
MOTOR LED LENS CLEANING:
The lens of the LED light on the motor (see Fig. 41) is soft and can be
damaged. It should not be exposed to dust and debris. Excessive dust and
debris may cause a drastic decrease in optical output. In the event that the
light requires cleaning, first try a gentle swabbing, using a lint-free swab.
If needed, use a lint-free swab and isopropyl alcohol to gently remove dirt
from the lens.
Do not use other solvents as they may adversely react with the LED assembly.
GENERAL CLEANING:
The external surfaces of the chassis should be cleaned using a soft cloth
moistened with a mild detergent solution. Any external surfaces of the unit
that are contacted during use should be wiped down with a soft cloth moistened
with a disinfectant at the beginning of each day and between each patient use.
WATER LINES:
Disinfect the water lines weekly. Prepare a 1:10 bleach solution (1 part
household bleach to 9 parts water). Remove water bottle and discard residual
water. Replace empty water bottle and air purge all waterlines. Fill water
bottle with bleach solution. Run bleach solution through all lines. Allow
bleach solution to stand in lines for 10 minutes. Remove water bottle and
discard bleach. Flush water bottle and all lines thoroughly with clean water.
Air purge and leave lines dry until next clinical use.
VACUUM SYSTEM:
CAUTION:
Do not run saline solutions through the water system — saline will rust the
water filters.
The HVE and low volume saliva ejector valves are fully autoclavable. Remove
the valves from their hoses before autoclaving.
The vacuum hoses should not be autoclaved. Clean hoses with a disinfectant
solution. CAUTION: Use only NON-foaming cleansers in the vaccum lines.
AIR BOTTLE AND FILTER:
Routinely check the air bottle filter once a day for condensation (see Fig.
36). To drain condensation, place a towel or container below the filter and
use pliers to carefully loosen the
black knob on the bottom of the filter.
IMPORTANT: To open the drain, turn the black knob clockwise (follow arrow
on “DRAIN → “ label); to close drain, turn the knob counterclockwise. Do not
overtighten.
NOTE: For highest reliability of operation, it is recommended to replace
the air bottle (PN 730657) after 1 year of operation.
WASTE SYSTEM CLEANING:
Empty and clean the waste system whenever the level alert occurs. Also empty
and clean it routinely once a day or before the unit is to be shipped or
stored. Follow these steps:
- Empty all waste from the waste container, including any solids trapped in the HVE strainer in the waste container lid.
- Prepare approximately 2/3 liter of 10% bleach/water solution in a separate container. Submerge the end of the high-vacuum ejector (HVE) into this bleach solution and pull no more than 1/3 liter of the solution through the line into the waste container. Repeat this process for the saliva ejector (low vacuum) line. IMPORTANT: The solution will enter the waste container at a very high rate — Care must be taken not to overfill the container’s waste compartments.
- Discard bleach solution. All components of the waste container, including the lid assembly and waste container strainer, can be safely rinsed with 10% bleach solution. (NOTE: Take care to ensure that water is kept off the level-sensor electrical connector on the case.) Rinse and dry tanks and the lid. If unit is to be shipped or stored, hang vacuum lines vertically to allow any residual water to drain before packing.
3-WAY AIR/WATER SYRINGE:
Depress the right button for air operation, and the left button for water
operation. Depressing both buttons will create a mist. The syringe features
quick-change autoclavable tips: To remove a tip, press on the locking collar
surrounding the tip socket and pull the used tip straight out of the socket
(Fig. 42). To insert a new tip, press locking collar and push tip into socket
as far as it will go. Release ring and gently tug on tip before using to
ensure that tip is securely locked into socket.
Syringe Tip Sterilization:
- Remove contaminated syringe tip.
- Remove all visible signs of contamination before autoclaving.
- Autoclave tip at 132° C (270° F) for ten minutes.
- Sterilize between each patient use.
NOTE: Since only the tips can be autoclaved, it is recommended that the air/water syringe be bagged with a disposable, single-use plastic sleeve between each patient use.
ULTRASONIC SCALER:
The scaler handpiece cover and scaler tips are fully autoclavable. Disinfect
and clean the cover and tips before autoclaving. Autoclave at a maximum
temperature of 135º C (275º F) for 10 minutes or 120º C (248º F) for 20
minutes. Wipe off the scaler handpiece and its silicone hose with a soft
cloth. Use a 45% isopropal and detergent solution. DO NOT IMMERSE the
handpiece in any fluid or spray any fluid directly on the handpiece.
MOTOR/CORD RECEPTACLE O-RINGS The O-rings (PN 520081) for the three water/air
ports in the motor/cord receptacle should be replaced if damaged or worn. Use
the provided O-ring installer pin and sleeve to replace the O-rings:
-
Remove old O-ring from water or air port fitting.
-
Slide new O-ring over pointed end of installer pin, onto the pin’s shank (see Figure 43).
-
Insert pointed end of installer pin into open end of installer sleeve until O-ring stops against end of tool.
-
Position concave end of installer pin against end of water/air port fitting (see Figure 44).
-
Push installer sleeve inward, until new O-ring seats into groove on fitting (see Figure 45).
AIR FILTER ASSEMBLY:
The foam filter on the Air Filter Assembly should be cleaned regularly if the
AEU- 525CF/AEU-525CFH is used in a dusty environment. Gently pull the foam
sheet off the white mounting studs on the Assembly frame and clean with soap
and water or compressed air. If washed, allow filter to dry before operating
unit.
AIR FILTER SETUP INSTRUCTIONS:
Whenever the unit is operated in environments where dust or other airborne particulates are present, it is recommended that the Air Filter Assembly, P/N 330634, be attached to the AEU-525CF/AEU525CFH to prevent it’s cooling-air fan and filters from clogging, and possibly causing it to overheat and shut down. Follow the steps below to install the Air Filter:
-
Lay the AEU-525CF/AEU-525CFH case flat on the floor with case lid facing upward. Open lid. Detach fastener tool from inside of case lid and use this tool to loosen the four fasteners on the outside of the rear metal door (see Fig. 46). Turn fastener counterclockwise 1/4-turn, to loosen.
-
Open the AEU-525CF/AEU-525CFH rear metal door to expose the fan and motor housing (see Fig. 47).
-
Place Air Filter Assembly, P/N 330634, into the open metal door frame as shown in Figure 48. Use the fastener tool to tighten the four (4) fasteners, securing the Filter Assembly to the unit’s lid. Turn the fasteners clockwise 1/4-turn to tighten. Note the clear plastic access holes provided for the power switches and waste-tank sensor connector.
The 525CF Air Filter Assembly is now ready to use.
TRANSPORT III REPACKING INSTRUCTIONS:
-
Remove all water and liquid wastes from water bottle and/or waste container. Clean the bottle and container per the instructions found on page 20.
-
Remove all accessories (handpieces, evacuator tips, syringe tips, etc.) from the unit and pack into luggage case provided.
-
Disassemble the system’s components in the reverse order of the instructions provided in the Setup Section, pages 3-9.
-
Strap water bottle into place inside the case lid as shown in Fig. 49.
-
Secure waste tank in place with the tiedown provided. Note “This Side Out” label.
-
Strap air/electric module into case as shown in Fig. 49 with strap provided.
-
Coil motor and cord, put into plastic bag and place next to waste tank as shown.
Place syringe assembly in bottom of case, just to the right of the air/electric module. -
Coil high and low suction hoses and place in bag in cutout in lid of shipping case (Fig. 50).
-
Coil foot pedal cable around pedal, put into plastic bag and place on top of motor housing, in the recess shown in Fig. 49.
-
Coil the 115V and 220V power cords and place in luggage case.
-
Place scaler accessories in a plastic bag in the pocket between the air/electric module and the top of the case (Fig. 49).
-
Carefully close case lid, taking care not to pinch any internal components.
-
Carefully lower the large shipping case down onto its back side as shown in Fig. 50. (NOTE: This task might require two or more people.)
-
Place the luggage case and Air Filter Assembly into the smaller rectangular cutouts in the large shipping case (Fig. 50).
-
Carefully lower the dental unit into the large cutout in the shipping case (Fig. 50). Close the seven clasps tightly on the three sides. Raise entire case to a vertical position, resting on its built-in wheels. Raise handle (Fig. 51).
CAUTION:
BEFORE REPACKING THE TRANSPORT III INTO ITS CASE, ALWAYS REMOVE ALL WATER AND
LIQUID WASTES FROM WATER BOTTLE AND WASTE CONTAINER
ALWAYS LIFT DENTAL UNIT AND SHIPPING CASE USING THE HANDLES PROVIDED
WHEN TRANSPORTING UNIT, DO NOT PLACE HEAVY OBJECTS ON TOP OF SHIPPING CASE.
TROUBLESHOOTING:
Problem: | Correction: |
---|---|
Unit will not start: | • Check system power connection. |
• Check if both circuit breaker switches are On.
• Check if waste container sensor is connected.
• Check if waste container is full.
• Check if fan is blocked or inoperative.
Unit starts but trips circuit breaker:| • Check source circuit to see if it is
a minimum of 15A.
• Check voltage selector switch for proper voltage.
NOTE: Operating the unit off an extension cord is not recommended.
No water pressure:| • Check water supply bottle water level. Verify that cap
is tight and not cross-threaded.
• Check that water supply pressure toggle is in the ‘PRESSURE’ position.
Insufficient vacuum:| • Check HVE and saliva ejector vacuum hose assemblies
for blockage; empty the HVE strainer.
• Check that the waste container lid is properly seated and tightly secured.
• Check that holder toggle switch is On.
Insufficient handpiece operation:| • Check the pressure gauge on the side of
the air/electric module and ensure that system pressure is sufficient.
• Check that handpiece tubing is untangled and not crimped.
• Check handpiece connection for missing or broken gasket.
No water to handpiece:| • Check that handpiece water toggle on side of
air/electric module is ‘On’.
• Check that the water flow control valve to the handpiece is open
(counterclockwise).
No coolant air to handpiece:| • Check that handpiece air toggle on side of
air/electric module is ‘On’.
• Check that the air flow control valve to the handpiece is open
(counterclockwise).
Electric motor control panel does notlight up when on:| • Press Standby Button
on control panel.
Electric motor control panel lights upwhen turned on, but handpiece doesnot
turn:| • Check motor plug connection.
• Depress foot switch.
• Turn holder toggle switch toward red dot.
• Increase Torque setting
• Check that a file or bur is properly seated in the handpiece.
Electric motor slowing down or sluggish:| • Increase torque setting.
• Check for dirty, under-lubricated handpiece.
• Check if handpiece lubricant is draining into motor. After lubricating and
before
autoclaving, stand handpiece on its base to let excess lubricant drain out.
Electric motor handpiece light does not turn on:| • Confirm that handpiece is
a fiberoptic illumination type.
• Press illumination button on control panel to turn light On and/or increase
light intensity
Vacuum doesn’t turn off when hoses are in their holders:| • Ensure that HVE
and saliva ejector vacuum valves are placed firmly in their holders.
• Check vacuum On/Off toggles on holders.
Pressure fails to stabilize:| • Check that air and water bottles are tight.
Check line and fittings for air leaks.
Pressure fails to turn off at 55 PSI:| • Check for broken cable to pressure
switch.
Unit fails to build pressure:| • Check that bottles are tight. Check wires for
breaks to pressure regulation switch.
Red LED on the waste container is lit:| • Empty full waste container.
• Check waste level sensor connection. Verify that connector contacts are dry.
• Check that floats in waste container move freely.
Compressor is on, but gauge shows no increase in pressure:| • Check unloader
valve to see if the valve is switched. Check the line from the compressor to
the valve for a rupture. Check the electrical connections to the valve.
HVE vacuum switch is On (toward red dot), but vacuum doesn’t activate when
hose is removed:| • Inspect the toggle lever on the vacuum holder to see if
it is stuck. Loosen the set- screw under the switch and adjust the switch for
proper operation. Check instrument control switches on the air/electric module
assembly.
HVE vacuum pressure seems lower:| • Check to see if the waste container lid is
properly seated and tightly secured.
Saliva ejector switch is On (toward red dot), but vacuum doesn’t activatewhen
hose is removed:| • Inspect the toggle lever on the vacuum holder to see if it
is stuck. Loosen the setscrew under the switch and adjust the switch for
proper operation. Check the toggling connections to the air pilot valve and
pump. Check to see if the air pilot valve is functioning by removing the line
to the pump and seeing if air is present.
REPLACEMENT PARTS LIST
ITEM | PART NO | QTY |
---|---|---|
WASTE TANK ASSEMBLY AEU-525CF/AEU-525CFH | 330603 | 1 |
INLINE STRAINER WASTE CONTAINER (HVE STRAINER) | 462124 | 1 |
O-RING .125ID X .070CS VT70 (AIR-JUNCTION) | 520025 | 3 |
O-RING .320ID X .O28 CS VITON 70 (E-HEAD) | 520069 | 3 |
O-RING .042ID X .142 X .050 W VITON (MTR RECEPTCL) | 520081 | 3 |
O-RING .208 ID X .070 CS VITON AS568-009 (AIR-JUNCTION) | 520121 | 3 |
O-RING .234ID X .070 CS SILICONE 2-010 (LVE FITTING) | 520100 | 2 |
O-RING .549ID X .103 CS SILICONE 2-113 (HVE FITTING) | 520101 | 2 |
FILTER INLINE ONLY 35 MICRON (FOR UNLOADER) | 730001 | 1 |
FILTER 10-32 THRD STNLS 100 MICRON (WATER BOTTLE) | 730326 | 1 |
BOTTLE 1 LITER (32 OZ) BLOW MOLDED HDPE (WATER) | 730471 | 1 |
GASKET (FOR AIR & WATER BOTTLE LID) | 730473 | 3 |
FILTER AIR 1/4”NPT X 1.19 GAST (INLET FILTER) | 730494 | 1 |
BOTTLE 750 ML (25 OZ) BLOW MOLDED HDPE (AIR) | 730657 | 1 |
LINECORD REMOTE US 10 FT HOSP GREY 15A/125A | 840049 | 1 |
LINECORD REMOTE EURO BLACK 2.5 M | 840007 | 1 |
O-RING FOR AA-35/36 PKG OF 4 AUTOCLAVABLE | AA-35LR | 1 |
O-RING FOR AA-37 SET OF 3/PKG AUTOCLAVABLE | AA-37LR | 1 |
SCALER TIP THIN SUBGINGIVAL | ASC-10-PE37 | 1 |
SCALER TIP SLIM UNIVERSAL | ASC-10-PE38 | 1 |
SCALER TIP POWER UNIVERSAL | ASC-10-PE39 | 1 |
SYRINGE TIP AUTOCLAVABLE | TA-1 | 1 |
SYRINGE 3-WAY AIR/WATER QUICK CHANGE TIP | TA-90D | 1 |
MOTOR/CABLE ASSY AE-240SC-40 W/SHORT CABLE | AE-24SC-40 | 1 |
FOAM FILTER INNER AEU-525CF/AEU-525CFH | 730684 | 1 |
FOAM FILTER OUTER AEU-525CF/AEU-525CFH | 730685 | 1 |
FOAM FILTER TUBE AEU-525CF/AEU-525CFH | 730687 | 1 |
SYMBOL DEFINITIONS:
| Type B Equipment
---|---
| Type BF Equipment
| Attention, consult accompanying documents
| Consult Instructions For Use
| Manufacturer
| Dangerous Voltage
| Alternating current
| Air
| Water
| Counterclockwise to Increase/Open Clockwise to Decrease/Close
| Electric Handpiece
| Syringe
| Scaler
| Scaler Tip Ultrasonic Control
| Model Number
| Saliva Ejector
| High Volume Evacuator (HVE)
| Pressure Gauge
| Motor Direction
| Light Controls
| Footswitch
| On/Off Switch – Mains
| Atmospheric Pressure Limitation
| Temperature Limitation
| Humidity Limitation
| Protective earth (ground)
IPX1| Protect Against Dripping Water
SN| Serial Number
| Do Not Lift by Top Lid Or Latches
| Pressurize / Purge Water Bottle Pressure
DO NOT THROW INTO TRASH:
Dispose of electronic waste and waste containing amalgam in accordance with
local regulations.
SPECIFICATIONS:
Transport III Case Size:| 17.93” W x 22.06” L x 10.43” H (45.5 cm x 56 cm x
26.5 cm)
---|---
Shipping Case Size:| 30.5” W x 25.75” L x 24.75” H (77.47 cm x 65.40 cm x
62.86 cm)
Weight (fully loaded):| 57 lbs (25.8 kg)
Shipping Case Weight: (fully loaded)| 108 lbs (49 kg)
Power Source:| AC Manual-Switching
Power Rating:| 115 VAC, 50/60 Hz, 4.8/5.6 A or 220 VAC, 60 Hz, 2.8 A or 230
VAC, 50/60 Hz, 2.5/2.8 A
Operating Pressure:| 50-60psi (3.45-4.14 bar)
High Volume Vacuum:| 5.6 SCFM @ 0.8″ Hg (159 liters/min @ 2 cm Hg)
3.5 SCFM @ 4″ Hg (99 liters/min @ 10 cm Hg)
Low Volume Vacuum:| 1.2 (+/-.1) SCFM @ 1.5″ Hg (34 (+/-3) liters/min @ 4 cm
Hg)
Simultaneous Vacuum:| High 2.4 SCFM @ 4″ Hg (68 liters/min @ 10 cm Hg)
Low 1.2 (+/-.1) SCFM @ 1.5″ Hg (34 (+/-3) liters/min @ 4 cm Hg)
Vacuum/Compressor Pump:| 1.1 SCFM @ 50 PSI (31.0 liters/min @ 3.45 bar)
Oilless Compressor
Water Reservoir Capacity:| 33.9 fl. oz. (1.0 liters)
Air Storage Capacity:| 25.4 fl. oz. (750 ml) nominal
Water Flow:| 5.07 fl. oz./min (0.15 liter/min)
Waste Container Capacity:| 0.48 gallon total/0.24 gal. per side (1.8 liters
total/0.90 L per side)
(NOTE: Capacities reflect liquid volumes up to shutoff levels.)
Noise Level:| 70 dBA @ 3’4” (1 meter)
Case Duty Cycle:| Continuous
Compressor Duty Cycle:| Continuous when operating at 50/60 Hz
Electric Motor Duty Cycle:| 17% (1 minute ON / 5 minutes OFF)
Environmental Conditions:| Operating Temperature: 0° to 35° C (32° to 95° F)
Transport/Storage Temperature: -20° to 65° C (-4° to 149° F)
Relative Humidity: 5% to 95% non-condensing
Altitude: 0 to 3048 meters (0 to 10,000 feet)
IMPORTANT
When running the AEU525CF/AEU-525CFH unit at 50Hz, expect approximately 17%
less vacuum and pressure due to slower turning of the compressor.
WARRANTY
Aseptico Inc. warrants its new products against defects in material and
workmanship under normal and proper use, care, and maintenance for a period of
two (2) years from date of original invoice. This two (2) year warranty does
NOT apply nor is it extended to products that are not manufactured by
Aseptico. These products may be covered by a separate limited warranty
provided by the particular manufacturer, and all claims and questions
regarding the same are to be directed to the particular manufacturer.
Expendable components, such as batteries, fuses, light bulbs, and tubing sets
installed on Aseptico products are specifically excluded and have no warranty.
Consumable goods are warranted for the stated expiration date of such goods.
Repair or replacement of any product(s) or part(s) under this warranty does
not extend the term of this warranty, and such product(s) or part(s) shall
remain covered by the unexpired portion of the warranty period, or for ninety
(90) days from the date of return to Aseptico, whichever is later. This
limited warranty applies only to the initial or first installation of the
product or part.
During the specific warranty periods set forth above, Aseptico will, at its
option, repair or replace the product(s) or particular part(s) that are found
to be defective in either material or workmanship in part or whole. Aseptico
shall be the sole arbiter of such action. In the event of alleged defect under
warranty, the purchaser is to notify Aseptico’s Customer Service department
promptly.
Customer Service will provide Return Material Authorization (RMA)
instructions, usually directing that the product be returned for service,
shipping prepaid by the buyer or end user, to Aseptico or its designated and
authorized warranty service center.
This warranty shall not apply to products (1) that have been subjected to
neglect, abuse, misuse, improper installation, inadequate maintenance, or
damage due to improper use of cleaning materials or chemicals, or non-
compliance with Aseptico’s storage, installation, operation, maintenance or
environmental requirements; (2) that have undergone any modification or repair
not previously authorized by Aseptico in writing, or service, repair or
modification by or from any facility other than an authorized Aseptico service
center or technician, or that use non-authorized software or spare or
replacement parts; or (3) that fail due to reasonable and normal use or wear
and tear, or materials made, furnished or specified by the buyer or end user.
Aseptico does not assume under this warranty any risks or liabilities arising
from the clinical use of its products, whether or not such use involves
coincidental utilization of products manufactured by others. Under no
circumstances will Aseptico be liable or responsible for special,
compensatory, incidental, consequential or punitive damages, lost profits,
lost sales, or loss of use or loss of business opportunity by or through the
use of the product. Aseptico’s sole and maximum liability with respect to the
product, other than its obligations set forth above, shall be the total
purchase price paid for the product.
P.O. Box 1548 • Woodinville, WA 98072
8333 216th Street SE • Woodinville, WA 98072
(425) 487-3157 • (800) 426-5913
www.aseptico.com •
info@aseptico.com
P/N: 420957
Rev. Q
ECO 15646 02/2023
PRINTED IN THE U.S.A.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>