Aseptico AEU-14CF Emergency Field Dental Unit User Manual

October 30, 2023
Aseptico

Aseptico AEU-14CF Emergency Field Dental Unit

Aseptico-AEU-14CF-Emergency-Field-Dental-Unit-PRODUCT

The AEU-14CF Emergency Field Dental Unit is a portable dental unit designed for use in emergency field situations. The unit is designed to run off of multiple power systems, making it easy to use in a variety of settings.

Product Usage Instructions

To verify that the AEU-14CF unit is functioning properly, follow the set-up procedure in the Operation & Maintenance manual. The  system is designed to run off of multiple power systems, so it isbest to isolate systems during diagnostics. Use the 120295 Rev G, Sheet 8 steps E-L of the schematic drawing set for a step-by-step functional check. If a problem is identified, refer to the troubleshooting guide for possible causes.

Cleaning and Lubrication

  • When servicing the AEU-14CF Emergency Field Dental Unit, thoroughly clean and inspect any disassembled components before reassembly. Use a hot detergent solution to clean all non-electrical parts. Flush all non-electrical parts with clear, hot water. Avoid using abrasive cleaners as they can damage surface finishes. Wipe all parts with a soft lint-free cloth.
  • Clean electrical parts with appropriate electrical parts cleaner or air. Use a silicone-based lubricating grease, such as Dow Corning No. 103, to lubricate O-rings and seals in the system. Before performing any reassembly of parts that contain O-rings or seals, apply a light coat of silicone grease to make installation easier and prevent damage to the O-rings or seals.

Routine Maintenance

Regular maintenance is important to keep the AEU-14CF unit functioning properly. Follow the recommended cleaning and lubrication procedures outlined above. In addition, be sure to perform routine maintenance checks on a regular basis to ensure optimal performance.

ESD Precautions

When working on the AEU-14CF unit, use electrostatic discharge (ESD) precautions to prevent damage to electronic sub-assemblies or their components. Designate an ESD static controlled area and mark it off with appropriate signs. Only allow trained personnel in this area to prevent ESD damage from occurring.

GENERAL SERVICE INFORMATION

This service and parts manual offers information, parts lists and an expanded troubleshooting guide not available in the AEU-14CF Operation and Maintenance Instruction Manual. It will help you diagnose any issue that may arise while operating your unit, as well as better understand the AEU-14CF Expedition Emergency Field Dental Unit as a whole, thereby reducing service time. For a more in-depth breakdown of the AEU-14CF’s interior components, refer to the Schematic Drawing Set. This drawing set includes part drawings and schematic diagrams which show components in their actual places in the unit relative to one another. A plumbing schematic is provided with water lines indicated in blue. Plumbing parts are called out by Aseptico part number in the assembly drawings. Electrical schematics are provided for the chassis and battery pack with wires color coded. Electrical parts are called out by Aseptico part number in the assembly drawings.

INSPECTION & OPERATION VERIFICATION:
To verify that the AEU-14CF unit is functioning properly, first follow the set-up procedure in the Operation & Maintenance manual. The system is designed to run off of multiple power systems, so it is best to isolate systems during diagnostics. 120295 Rev G, Sheet 8 steps E-L of the schematic drawing set are a step by step functional check. If a problem is identified, refer to the troubleshooting guide for possible causes.

CLEANING AND LUBRICATION

When servicing the AEU-14CF Emergency Field Dental Unit, the parts of any component disassembled should be thoroughly cleaned and inspected before reassembly. A hot detergent solution is an effective cleaner on all non- electrical parts. Flush all non-electrical parts with clear, hot water. Abrasive cleaners have the potential to damage surface finishes and should be avoided. Any wiping should be done with a soft lint-free cloth.
Electrical parts should be cleaned with appropriate electrical parts cleaner or air.
Use a silicone base lubricating grease, such as Dow Corning No. 103 to lubricate O-rings and seals in the system. Before performing any reassembly of parts that contain O-rings or seals, apply a light coat of silicone grease. This will make installation easier and prevent the O-rings or seals from being damaged.

ROUTINE MAINTENANCE:

WATER LINES – Disinfect the water lines weekly. Prepare a 1:10 bleach solution (1 part household bleach to 10 parts water). Remove water reservoir and discard residual water. Replace empty water supply tank and air purge all waterlines. Fill water supply tank with bleach solution. Run bleach solution through all lines. Allow bleach solution to stand in lines for 10 minutes. Remove water supply tank and discard bleach. Flush water supply tank and all lines thoroughly with clean water. Air purge and leave lines dry until next clinical use.

ESD PRECAUTIONS

The following electrostatic controls must be used when working on this unit:

  • ESD Training and Standards:
    Employees handling electronic sub-assemblies and ESD-sensitive components are expected to be trained. Training should be based on IPC-A-610 or equivalent ESD standard ANSI/ESD-S-20-20 – Protection of Electrical and Electronic Parts, Assemblies and Equipment.

  • ESD Static Controlled Area:
    Areas that are designated for handling and working on electronic sub- assemblies or their components should be marked off with signs indicating the area where ESD controls are to be enforced. These areas are to be kept clear from persons that are not trained to prevent ESD damage from occurring.

  • ESD Environment:
    The work area is to be free of all static generating materials, such as plastic containers, water bottles, plastic bags, plastic objects, such as plastic pens, heat guns
    (unless made for the ESD environment).

  • ESD Jackets:
    Clothing should be non-static generating (cotton).
    Static generating clothing (e.g. wool, acrylic, nylon) must be covered with an ESD jacket that is buttoned closed.

  • Optional gloves:
    Nitrile gloves may be used to cover the hands when working, but are not required.

  • Seating:
    ESD Chairs should be used in place of static generating chairs (e.g. modern office seating use static generating materials).

  • Storage and packaging:
    All circuit boards and components are to be stored on or in static dissipative or static shielding material, throughout shipping and storage.

  • ESD Wrist Strap and Mat Routine Checks:
    The wrist strap should be checked daily using an ESD wrist strap testing station. See chart below.
    ESD mats should be checked at least quarterly.Aseptico-AEU-14CF-Emergency-
Field-Dental-Unit-FIG- \(2\)

BASIC DISASSEMBLY

Aseptico-AEU-14CF-Emergency-Field-Dental-Unit-FIG-
\(3\)

  1. Make sure the unit is turned off and unplugged from all AC and DC power sources.
  2. Make sure all accessories have been removed from the unit.
  3. Stand the unit upright, into its operating position. Remove the water bottle. Go to the rear of the case and remove the two upper screws with a #2 Phillips head screwdriver. Go to the top of the unit and remove the two screws while supporting the chassis with your hand. Remove the chassis from the case (Fig. 1).
  4. Reassemble the unit in the reverse order. Check the O-rings on the screws to ensure they are not damaged. If the O-rings are damaged, replace with new screws or RTV screws in place.

For details on subcomponent removal see the system schematic drawing set for:
330855 UPPER CHAMBER SUBASSEMBLY, AEU-14CF
330365 BATTERY PACK ASSY AEU-14
These documents show how the unit is assembled and can be followed in reverse order to dis-assemble the unit.

PARTS LIST

System Contents (Drawing No. 120295)Aseptico-AEU-14CF-Emergency-Field-
Dental-Unit-FIG- \(3\)

Item Part No QTY
1 Syringe, 3-Way Air Water TA-90D 1
2 Upper Chamber Subassembly 330855 1
3 PANPHL, #6-32 X 3/8, Stnls, Self Sealing 510561 4
4 Tubing, 2H Poly Strt Syringe Grey AA-85G 1
5 Battery Pack Assembly 330365 2
6 Linecord Remote US 10Ft Gry 840001 1
7 Cable Battery Assy AEU-14 875043 1
8 Battery & Cable Pouch 461286 1
9 Cable Assy, Battery Pack AEU-14 875209 1
10 Footswitch, On/Off 8-Pin AE-7P 1
11 Motor/Cable Assy Pouch 461285 1
12 Motor/Cable Assy w/Light & Water AE-200-30 1
13 Solar Panel 330863 1
14 Label, Mylar Air 420307 1
Label, Mylar Water 420299 1
16 Bottle, 16oz. Pet 730427 2
17 Manual, AEU-14CF Operation 420382 1
18 Manual, AEU-14CF Service 420417 1
19 Manual, Schematic Set 420591 1
20 Instruction Sheet, Yellow Paper 420717 1
21 Re-Packing Guide 420460 1
22 Case, AEU-14 Expedition 410157-01 1

Chassis Front View (Drawing No. 330855)Aseptico-AEU-14CF-Emergency-Field-
Dental-Unit-FIG- \(4\)

Item Part No QTY
1 HOLDER HPCE ASSY, AEU-14 330860 1
2 Chassis Frame 461176-08 1
3 Overlay, AEU-14 420370 1
4 Display Board Insulator 461284 1
5 CABLE ASSY AC MAINS, AEU-14CF 875202 1
6 Power Inlet w/Fuse Holder 2Amp 840081 1
7 Fuse, 5x20mm Slo-Blo 1.25A 830126 2
8 M/S Stnls FlaPhl 4-40×3/8 510112 2
9 Knob Gry Plastic 1/4D Style II 850012 1
10 Valve, Needle Control w/o Knob 730066 1
11 CABLE ASSY BATTERY VEHICLE INPUT, AEU-14CF 875200 1
12 Battery Pack Input 875199 1
13 Wire Set AEU-14 Foot Pedal Con 875042 1
14 CABLE ASSY BATTERY SWITCH, AEU-14CF 875201 1
Motor Harness 875060 1
16 Label, Mylar Air 420307 1
17 Label, Mylar Water 420299 1
18 Bottle, 16oz. Pet 730427 2

Chassis Back (Drawing No. 330855)Aseptico-AEU-14CF-Emergency-Field-Dental-
Unit-FIG- \(5\)

Item Part No QTY
1 Water Solenoid Assembly 875206 1
2 Pressure Sensor Assembly 875207 1
3 Shield Left Chassis Bracket 461280 1
4 Fuse Block 830064-02 1
5 Fuse 5A 830110 1
6 Fuse 12A 830084 1
7 Power Supply Cable 875203 1
8 PANPHIL #4-40 x 1/4 STNLS 510144 4
 Aluminum Standoffs 511003 4
10 Power Supply 840151 1
11 Cover, Pump Front AEU-14CF 461177 1
12 PANPHIL #6-32 X 1/4 STNLS 510977 6
13 1/2 inch Grommet 870331 1
14 Cover, Pump Back 461181 1

Main Internals (Drawing No. 330855)Aseptico-AEU-14CF-Emergency-Field-Dental-
Unit-FIG- \(6\)

Item Part No QTY
1 Chassis Base 461176-08 1
2 COMPRESSOR SUBASSEMBLY, AEU-14CF 330858 1
3 CABLE ASSY WATER SOLENOID TO PCB, AEU-14CF 875206 1
4 CABLE ASSY PRESSURE SENSOR TO PCB, AEU-14CF 875207 1
5 Shield Left Chassis Bracket 461280 1
6 Fuse Block 830064-02 1
7 Main Board 330364 1
8 Foot Pedal 875042 1
9 Wire Assy, Motor Conn 875060 1
10 Display Board Clamp 461283 1
11 PANPHL, #4-PLASTITE X 1/4, STNLS 510819 2
12 Brkt AEU-14CF Heat Sink Diode 461359 1
13 Thermal Pad 461649 1
14 AEU-14 Heatsink 461335 1
15 Thermo Heatsink Compound 490007 1

Chassis Front (Drawing No. 330855)Aseptico-AEU-14CF-Emergency-Field-Dental-
Unit-FIG- \(7\)

Item Part No QTY
1 PANPHILS #6-32, 1/4 510977 2
2 Gasket For NWS-8 Bottle Lid 730473 1
3 #6 Plastite 3/8 510792 2
4 Cap, Air Bottle AEU-14 461458 1
5 Brkt AEU-14CF Cap Cmpl 461179 1
6 Handpiece Holder Assembly 330860 1
7 #6 x 3/8 Plastite Screws 510792 3
8 FAN CABLE ASSEMBLY 875204 1
9 FAN MOUNT GASKET, 40MM 540015 1
10 NUT HEX NYLOC, #4-40, STNLS 510394 3
11 PANPHILS, #4-40, 1-1/8 STNLS 510999 3
12 Power Inlet w/Fuse Holder 2Amp 2 Pole 840081 1
13 Fuse 5x20mm Slo-Blo 1.25A 830126 2
14 M/S Stnls FlaPhl 4-40 x 3/8 510112 2
15 CABLE ASSY BATTERY VEHICLE INPUT, 875200 1
16 #4-40 Nyloc Nuts 510394 4
17 Battery Pack Input 875199 1
18 #4-40 Nyloc Nuts 510394 4
19 Foot Pedal Cable Assembly 875042 1

Plumbing Components (Drawing No. 330855)Aseptico-AEU-14CF-Emergency-Field-
Dental-Unit-FIG- \(8\)

Item Part No QTY
1 T Fitting 1/16 Barbs 730152 1
2 Check Valve 730428 1
3 Gasket, Nylon #10 730074 1
4 Ftn Barb 10-32 x 1/8 Plated 730073 1
5 Sleeve Clamp 1/4 730095 1
6 Ftn Barb 10-32 x 1/16 Plated 730062 1
7 Sleeve Clamp Ftn 1/8ID Clear 730015 1
8 Grey Tubing AA-95G 1
9 FTN STRAIGHT, 1/16 BARBS PANEL MOUNT, ALU 730798 1
10 1/8 Uni-clamp 730096 1
11 TUBING, POLY 1/8 OD Cear AA-94C 1
12 TUBING, POLY 1/8 OD BLU AA-94B 1

Chassis Internals (Drawing No. 330855)Aseptico-AEU-14CF-Emergency-Field-
Dental-Unit-FIG- \(9\)

Item Part No QTY
1 Ribbon Cable 870271 1
2 NUT HEX NYLOC, #4-40, STNLS 510394 1
3 AEU-14 Heatsink 461335 3
4 Thermal Pad 461649 3
5 Thermo Heatsink Compound 490007 3
6 Clamp, AEU-14 Display 461283 1
7 Display Board 330391 1

Case & Battery Pack (Drawing No. 120295)Aseptico-AEU-14CF-Emergency-Field-
Dental-Unit-FIG- \(10\)

Item Part No QTY
1 Case AEU-14 Expedition 410157-01 1
2 Chassis Label 420432-37 1
3 UDI/BT1 3×1 Prodyct UD SN Label 420457-00 1
4 Battery Pack Assy AEU-14 330365 2
5 PANPHL, #6-32 X 1/4, Stnls, Self Sealing 510605 2
6 (Not Shown) Foot, Case AEU-14 461290 2
7 PANPHL, #6-32 X 3/8, Stnls, Self Sealing 510561 2
8 Battery Cage 330859 1
9 PANPHILS #6-32, 3/8 Stnls, Self Sealing 510561 4
10 PANPHILS #6-32, 3/4 Stnls, Self Sealing 510589 2

Battery Pack Assembly (Drawing No. 330365)Aseptico-AEU-14CF-Emergency-Field-
Dental-Unit-FIG- \(11\)

Item Part No QTY
1 Cable Assy Battery Pack Solar Panel 875212 2
2 Standoff AEU-14CF Battery Case 461333 2
3 Tubing Silicone .281IDx.375OD 70 Duro 730489 2
4 M/S Stnls FlaPhl 4-40×3/8 Flat Undercut 510112 16
5 Spring Clip Catch Back Plate 462787 1
6 Spring Clip Catch AEU-14 462786 1
7 Chassis AEU-14 Battery Top 461194 1
8 Cable Assy Battery Pack Output 875211 2
9 CHASSIS AEU-14 BATTERY BASE CMPL 461195-08 1
10 FOAM INSERT FLAME RETARDANT BATT BOX 462788 1
11 Cable Assy Battery Pack Leads AEU-14 875210 1
12 PCB Assy AEU-14 Temperature 330392 1
13 Pad, AEU-14CF Battery Heatsink 461334 1
14 FLATPHL, #4-40 x 1/4, STNLS, 100 Degree 510270 4
15 Washer, Flat Stnls #4 510192 4

TROUBLESHOOTING

Chassis does not light when turned on:| •    Check chassis to power connection.

•    Check that voltage is proper voltage.

•    Check fuse. If blown, replace with 1.25A/250V slo-blo fuse for 120V operation, & 0.6A/250V slo- blo fuse for 230V operation.

*    Check internal wiring harness connections.

*    Check the two internal DC current fuses. If either fuse is open, it is likely that the AC/DC power supply or main board has a short that caused excessive current draw.

*    Check ribbon cable connection between main PCB and display PCB.

---|---
Chassis lights when turned on, but handpiece does not turn:| •    Check motor plug connection.

•    Check foot switch connection.

•    Depress foot switch.

•    Increase RPM.

•    Check that an instrument is properly seated in the handpiece and the collet is closed.

Chassis does not light when 12V/24V power is applied (* requires further dis- assembly, refer to schematic drawing set):| •    Check connections at DC source, chassis, and cable.

•    Check voltage of DC source.

*    Check internal wiring harness connections.

*    Check the internal 12A fuse on the red DC power line. If the fuse is open it is likely that the main board has a short that caused excessive current draw.

Chassis does not light when battery pack is applied:| •    Check connections at the battery pack, chassis, and cable.

•    Check voltage of the battery pack.

*    Check internal wiring harness connections.

*    Check the internal 5A fuse on the orange DC power line. If the fuse is open it is open it is likely that the main board has a short that caused excessive current draw.

Motor slowing down or sluggish:| •    Check for dirty, under-lubricated handpiece.

•    Check if handpiece lubricant is running into motor.

•    After lubricating, set handpiece with head down to let excess lubricant drain out.

Unit fails to build pressure:| •    Check that bottles are tight.

•    Check for air leaks.

Battery fault light on:| •    Check if battery is too cold (<0°C). Disconnect battery and allow it to warm to room temperature.

•    Check if battery is too hot (>55°C). Disconnect battery and allow it to cool to room temperature

•    Check if battery voltage is too low (<23V). Disconnect battery and confirm voltage is low (pins A to D). An under-voltage battery may be able to be brought above 23V by trickle charging the battery with the console or the solar panel.

•    Check if battery voltage is too high (>34V). Disconnect battery and confirm voltage is high (pins A to D). An over-voltage battery should be replaced.

•    Charging timeout at 4 hours. Reset the charger. If two consecutive charging attempts fail, replace the battery pack.

Unit fault light on:| •    Unit operational error observed. Remove all power from unit for 30 minutes before attempting to use the system again. If unit fault light is on during two consecutive attempts to operate the sys- tem, return system for repair.

*    Check if there is overheating or faulty temperature monitoring of voltage booster transistors Q1 and Q2 by temperature sensor RT2 on the main PCB. Repair or replace PCB.

*    Check if there is overheating or faulty temperature monitoring of voltage booster diodes D2, D5, D6, and D14 by temperature sensor RT1 on the main PCB. Repair or replace PCB.

*    Check for faulty regulation of boost voltage VPP to 36V on the main PCB. Repair or replace PCB.

*    Check for faulty regulation of compressor drive voltage VPMP to 24V on the main PCB. Repair or replace PCB.

*    Check for faulty regulation of battery charge current CHG_I_MON on the main PCB. Repair or replace PCB.

SPECIFICATIONS

  • Size: . . . . . . . . . . . . . . . . . . . . . . . . . .17.5” x 11.6” x 7.0” (44.5 cm x 29.5 cm x 17.8 cm)

  • Volume: . . . . . . . . . . . . . . . . . . . . . . .0.82 ft3 (0.23 m3)

  • Weight: . . . . . . . . . . . . . . . . . . . . . . .27 lbs (12.25 kg)

  • Power Source: . . . . . . . . . . . . . . . . . .AC Dual Voltage 120V/230V @ 50Hz/60Hz 12V/24V vehicle battery 28.0V battery pack (x 2) 36V solar panel to charge battery packs

  • Water Reservoir Capacity: . . . . . . . .16 fl. oz. (0.473 liter)

  • Air Reserve Capacity, Standard: . . .28.8 in3 (473 cc)

  • Noise Levels . . . . . . . . . . . . . . . . . . .66 dBa @ 1 meter (3’4”)

  • Fuses: . . . . . . . . . . . . . . . . . . . . . . . .120V: 1.25A, 250V slo-blo fuse (Installed at factory) 230V: 0.6A, 250V slo-blo fuse

  • Fuse Size: . . . . . . . . . . . . . . . . . . . . .5 x 20mm

  • Duty Cycle for AE-200-30 Motor: . .Intermittent use: 1 minute On / 3 Minutes Off

  • Mains Current Ratings: . . . . . . . . . .120V: 1.25A, 230V: 0.6A
    NOTE : The appliance inlet is the mains disconnect means.

  • Environmental Conditions:

    • Operating Temperature: 5° to 45° C (41° to 113° F)
    • Transport/Storage Temperature: -20° to 55° C (-4° to 131° F)
    • Relative Humidity: 5 to 95% non-condensing
    • Altitude: 0 to 2591 meters (0 to 8,500 feet)

NOTE : Air pressure instruments are set to 2591 meters (8,500 ft) elevation.
WARNING : Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air, or with oxygen or nitrous oxide.
CAUTION : If the integrity of the protective earth ground is in doubt, only use battery pack or vehicle battery power for operation.
MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1:2003

Type of Protection Against Electric Shock:

  • Class I when system is powered by mains power source.
  • Class II when system is powered by Vehicle Battery power source.
  • Internally Powered Equipment for Battery Pack power source.
    • Degree of Protection Against Electric Shock: Type B Applied Part
    • Degree of Protection Against Ingress of Water: Ordinary Equipment
    • Mode of Operation: Continuous

This device has been tested and found to comply with the emissions requirements of IEC 60601-1-2:2014. These requirements provide reasonable protection against harmful electromagnetic interference in a typical medical installation. However, high levels of radio-frequency (RF) emissions from electrical devices, such as cellular phones, may disrupt the performance of this device. To mitigate disruptive electromagnetic interference, position this device away from RF transmitters and other sources of electromagnetic energy.

SYMBOL DEFINITIONS

  • Type B Equipment
  • Attention , consult accompanying documents
  • Dangerous Voltage
  • Do Not Throw Into Trash
  • Alternating current
  • Protective earth (ground)
  • Protect Against Dripping Water

WARRANTY

  • Aseptico Inc. warrants its new products against defects in material and workmanship under normal and proper use, care, and maintenance for a period of two (2) years from date of original invoice. This two (2) year warranty does NOT apply nor is it extended to products that are not manufactured by Aseptico. These products may be covered by a separate limited warranty provided by the particular manufacturer, and all claims and questions regarding the same are to be directed to the particular manufacturer.
  • Expendable components, such as batteries, fuses, light bulbs, and tubing sets installed on Aseptico products are specifically excluded and have no warranty. Consumable goods are warranted for the stated expiration date of such goods.
  • Repair or replacement of any product(s) or part(s) under this warranty does not extend the term of this warranty, and such product(s) or part(s) shall remain covered by the unexpired portion of the warranty period, or for ninety (90) days from the date of return to Aseptico, whichever is later. This limited warranty applies only to the initial or first installation of the product or part.
  • During the specific warranty periods set forth above, Aseptico will, at its option, repair or replace the product(s) or particular part(s) that are found to be defective in either material or workmanship in part or whole. Aseptico shall be the sole arbiter of such action. In the event of an alleged defect under warranty, the purchaser is to notify Aseptico’s Customer Service department promptly. Customer Service will provide Return Material Authorization (RMA) instructions, usually directing that the product be returned for service, shipping prepaid by the buyer or end user, to Aseptico or its designated and authorized warranty service center.
  • This warranty shall not apply to products (1) that have been subjected to neglect, abuse, misuse, improper installation, inadequate maintenance, or damage due to improper use of cleaning materials or chemicals, or non-compliance with Aseptico’s storage, installation, operation, maintenance or environmental requirements; (2) that have undergone any modification or repair not previously authorized by Aseptico in writing, or service, repair or modification by or from any facility other than an authorized Aseptico service center or technician, or that use non-authorized software or spare or replacement parts; or (3) that fail due to reasonable and normal use or wear and tear, or materials made, furnished or specified by the buyer or end user.
  • Aseptico does not assume under this warranty any risks or liabilities arising from the clinical use of its products, whether or not such use involves coincidental utilization of products manufactured by others. Under no circumstances will Aseptico be liable or responsible for special, compensatory, incidental, consequential or punitive damages, lost profits, lost sales, or loss of use or loss of business opportunity by or through the use of the product. Aseptico’s sole and maximum liability with respect to the product, other than its obligations set forth above, shall be the total purchase price paid for the product.

For Further Service And/Or Technical Assistance Contact:
8333 216th St.
Woodinville, WA 98072-1548
(800) 426-5913
(425) 487-3157
Fax : 425-487-2608
www.aseptico.com
info@aseptico.com
PRINTED IN THE USA

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals