OPERATION and MAINTENANCE
INSTRUCTION MANUAL
ADU-03 Single Handpiece Manual Control System
The Aseptico ADU-03 is one of the finest pneumatic handpiece control systems
available to the professional. Offering manual activation and easily mounted
on a wall or post, the ADU-03 provides an economic control system for a dental
operatory,
hygienist’s room, or satellite office.
Congratulations!
This system is engineered to provide many years of reliable service. Please
read the instructions provided in this manual to receive the best and longest
service from your Aseptico equipment.
Separate manuals may be provided to cover the operation and maintenance of
handpieces or other accessories for your unit. Handpieces are purchased
separately
PACKAGE CONTENTS:
- ADU-03 Pneumatic Control Unit
- TA-90D 3-Way Air/Water Syringe with Tip
- AA-42 Foot Control and Tubing Options:
- NWS-7 Self-Contained Clean Water System with AA-61 Water Connector for scalers
SETTING UP THE UNIT:
- Use the WALL MOUNTING HOLES to secure cabinet back to a flat vertical surface (see Fig. 1).
- Attach the front of the unit to the CABINET MOUNTING TABS using the 8 screws provided (see Fig. 1).
- Connect the braided airline to a clean regulated 70-100 PSI compressed air supply.
- Connect the blue water line to 35psi max city water supply -OR- if using the Optional NWS-7 Water System, fill the white plastic water-supply bottle with clean water and attach the bottle to the white fitting on the underside of the unit (see Fig. 2).
OPERATION FUNCTIONS:
Water On/Off Toggle
Controls the on/off flow of water coolant to the handpiece.
Water Flow Control
Adjusts the water coolant flow to the handpiece.
Pressure Gauge
Gives a visual indication of the drive air pressure to the handpiece.
Drive Air Pressure Control
Adjusts the drive air pressure to the handpiece.
Foot Control
Controls the drive air pressure to the handpiece. Apply light foot pressure to
any part of the foot control cover.
3-Way Air /Water Syringe
Delivers air, water, or a combination mist. Supplied with autoclavable tip.
NWS-7 Water System (Optional)
White 1-liter bottle dispenses water through the 3-Way Air/ Water Syringe and
handpiece.
AA-61 Water Connector
Provides water via standard 1/4” quick disconnect to independent Cavitron or
scaler units requiring a water supply.
UNIT OPERATION:
HANDPIECE
Connect either a high-speed or low-speed handpiece to the handpiece connector.
Depress foot control for operation.
3-WAY AIR/WATER SYRINGE:
Depress the right button for air operation, and the left button for water
operation. Depressing both buttons will create a mist. The syringe features
quick-change autoclavable tips: To remove a tip, press on the locking collar
surrounding the tip socket and pull the used tip straight out of the socket
(see Fig 3). To insert a new tip, press locking collar and push tip into
socket as far as it will go. Release collar and gently tug on tip before using
to ensure that tip is securely locked into socket.
MAINTENANCE:
Syringe Tip Sterilization:
- Remove contaminated syringe tip.
- Remove all visible signs of contamination before autoclaving.
- Autoclave tip at 132° C (270° F) for ten minutes.
- Sterilize between each patient use.
NOTE: Since only the tips can be autoclaved, it is recommended that the
Air/Water Syringe be bagged with a disposable, singleuse plastic sleeve
between each patient use.
Water Lines:
Disinfect the water lines weekly. Prepare a 1:10 bleach solution (1 part
household bleach to 10 parts water). Remove water reservoir and discard
residual water. Replace empty water supply tank and air purge all waterlines.
Fill water supply tank with bleach solution. Run bleach solution through all
lines. Allow bleach solution to stand in lines for 10 minutes. Remove water
supply tank and discard bleach. Flush water supply tank and all lines
thoroughly with clean water. Air purge and leave lines dry until next clinical
use.
WARRANTY:
Aseptico Inc. warrants its new products against defects in material and
workmanship under normal and proper use, care, and maintenance for a period of
two (2) years from date of original invoice. This two (2) year warranty does
NOT apply nor is it extended to products that are not manufactured by
Aseptico. These products may be covered by a separate limited warranty
provided by the particular manufacturer, and all claims and questions
regarding the same are to be directed to the particular manufacturer.
Expendable components, such as batteries, fuses, light bulbs, and tubing sets
installed on Aseptico products are specifically excluded and have no warranty.
Consumable goods are warranted for the stated expiration date of such goods.
Repair or replacement of any product(s) or part(s) under this warranty does
not extend the term of this warranty, and such product(s) or part(s) shall
remain covered by the unexpired portion of the warranty period, or for ninety
(90) days from the date of return to Aseptico, whichever is later. This
limited warranty applies only to the initial or first installation of the
product or part.
During the specific warranty periods set forth above, Aseptico will, at its
option, repair or replace the product(s) or particular part(s) that are found
to be defective in either material or workmanship in part or whole. Aseptico
shall be the sole arbiter of such action. In the event of alleged defect under
warranty, the purchaser is to notify Aseptico’s Customer Service department
promptly. Customer Service will provide Return Material Authorization (RMA)
instructions, usually directing that the product be returned for service,
shipping prepaid by the buyer or end user, to Aseptico or its designated and
authorized warranty service center. This warranty shall not apply to products
(1) that have been subjected to neglect, abuse, misuse, improper installation,
inadequate maintenance, or damage due to improper use of cleaning materials or
chemicals, or noncompliance with Aseptico’s storage, installation, operation,
maintenance or environmental requirements; (2) that have undergone any
modification
or repair not previously authorized by Aseptico in writing, or service, repair
or modification by or from any facility other than an authorized Aseptico
service center or technician, or that use non-authorized software or spare or
replacement parts; or (3) that fail due to reasonable and normal use or wear
and tear, or materials made, furnished or specified by the buyer or end user.
Aseptico does not assume under this warranty any risks or liabilities arising
from the clinical use of its products, whether or not such use involves
coincidental utilization of products manufactured by others. Under no
circumstances will Aseptico be liable or responsible for special,
compensatory, incidental, consequential or punitive damages, lost profits,
lost sales, or loss of use or loss of business opportunity by or through the
use of the product. Aseptico’s sole and maximum liability with respect to
the product, other than its obligations set forth above, shall be the total
purchase price paid for the product
P.O. Box 1548 • Woodinville, WA 98072
8333 216th Street SE • Woodinville, WA 98072
(425) 487-3157 • (800) 426-5913
www.aseptico.com
•info@aseptico.com
4. P/N: 420284
Rev. H ECO 15498
04/2022
PRINTED IN THE
U.S.A.
Documents / Resources
|
Aseptico ADU-03 Single Handpiece Manual Control
System
[pdf] Instruction Manual
ADU-03 Single Handpiece Manual Control System, ADU-03, Single Handpiece Manual
Control System, Handpiece Manual Control System, Control System
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>