Aseptico AMC15 Mobile Dental Cart User Manual

October 30, 2023
Aseptico

Aseptico logo AMC‑15
Mobile Dental Cart

Aseptico AMC15 Mobile Dental Cart

OPERATION &
MAINTENANCE MANUAL

SAFETY PRECAUTIONS

To prevent injury to people and damage to property, please heed relevant warnings and remarks. They are marked as follows:

  • WARNING: Serious injury or death may result if ignored.
  • CAUTION: Damage to property or the environment may result if ignored.
  • NOTE: Important additional information and hints.

Aseptico accepts no liability for direct or consequential injury or damage resulting from improper use, arising in particular through the non-observance of the operating instructions, or improper preparation and maintenance.

  • WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
  • WARNING: Connecting electrical equipment to this device creates a Medical Electrical (ME) system. If connecting equipment that is not part of the ME system, have a trained service person evaluate the ME system to ensure safety leakage currents are maintained per IEC 60601-1.
  • WARNING: Clean, disinfect, and sterilize new or repaired handpieces and instruments before first use and between each patient use. Only use sterilized handpieces and instruments during treatment. Non-sterilized handpieces and instruments may cause bacterial or viral infections. Always sterilize handpieces and instruments after operation.
  • CAUTION: Always examine unit components for damage before commencing treatment. Damaged components must not be used and must be replaced.
  • CAUTION: For device transport, always keep the pull handle in hand during transport and ensure stable platform before releasing.
  • WARNING: Use for intended purposes only. Failure to observe the operating instructions may result in the patient or user suffering serious injury or the product being damaged, possibly beyond repair. Before using this product, make sure that you have studied and understood the operating instructions.
  • WARNING: For use by qualified and trained personnel only.
  • WARNING: Do not install where there is a risk of an explosion. The Mobile Dental Cart is not intended for operation in the presence of flammable anesthetics or gases.
  • WARNING: Always operate a high-speed handpiece with water coolant. Operating a high-speed handpiece without water coolant can cause thermal injury to the patient.
  • CAUTION: Observe local regulations concerning disposal of amalgam waste. s WARNING: Never operate unit or work on patients when the top lid is open.
  • WARNING: Do not use this device for dental implant procedures.
  •   CAUTION: Do not run saline solutions through the water system—saline will corrode the plumbing.
  • WARNING: Do not position the unit so that the power cord is inaccessible to unplug during normal use.
  • WARNING: No modification of this equipment is allowed.
  • CAUTION: Dispose of all medical waste products, residues, equipment and etc. in accordance with local Regulations.
  • CAUTION: All repairs are to be performed by Authorized Repair Personnel only. Aseptico offers servicing of equipment that is out of warranty and this is highly recommended to ensure optimal performance and product life. Circuit diagrams, parts list and requested relevant servicing information will be provided upon request.
  • CAUTION: Do not use CaviCide™ cleaner on the unit.
  • CAUTION: Do not plug 230V power supply into the 115V version of the unit.
  • NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
  • WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  • WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  • WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the unit, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

Congratulations!

Congratulations! Your new Aseptico AMC-15 Mobile Dental Cart is the finest affordable system available. The system provides an array of standard and optional dental equipment features and functions designed to serve a wide range of dental applications—all conveniently packaged into a single self- contained transportable cart.
This system is engineered to provide many years of reliable service. Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment. Separate manuals may be provided to cover the operation and maintenance of other accessories for your unit.
NOTE: Shaded paragraphs throughout this manual indicate optional equipment that can be purchased separately.

STANDARD FEATURES

  • Automatic handpiece controls for 3 pneumatic handpieces.
  • Connections for two high-speed pneumatic handpieces (air and water provided), connection for one low-speed pneumatic hand- piece (air provided).
  • 3-way air/water syringe with a self contained water system.
  • High Volume (HVE) and low volume vacuum lines with adjustable suction valves and an amalgam solids trap.
  • Independent 1/2 HP compressor and vacuum pumps with venturi boosters
  • Aluminum tank for rust-free air supply.
  • The AMC-15 is available in either a 115V -or- 230V power supply configuration.
  • Amalgam Separator Filter and Filter Bypass

OPTIONAL EQUIPMENT (Purchased Separately)

  • (PN AA-20A) Auxiliary Arm with Accessory Tray
  • (PN 730631-01) Bottle 1000ml Zirc Kit Modified (Clear 1-liter Water Bottles w/ Level Indicators)

INDICATIONS FOR USE

The AMC-15 Mobile Dental Cart is a mobile self-contained dental system that is used for endodontic and general dentistry applications.
RX: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A DENTIST.
Expected Service Life: The expected service life is seven years from the date of first use.

PACKAGE CONTENTS

  • Self Contained Mobile Cart
  • (PN TA-90D) Syringe 3-Way Air/Water Quick Change Tip
  • (PN 730471) Bottle 1 Liter (32 Oz) Blow Molded HDPE (Two 1-Liter Water Bottles)
  • (PN AA-43W) Foot Control, Wet/Dry Disc, White Pedal, Gray Tubing
  • (PN 840049) Linecord Remote Us 10ft Hosp Grey 15a/125v -or- (PN 840007) Linecord Remote Euro Black 2.5m

HANDPIECES ARE PURCHASED SEPARATELY

FRONT VIEWAseptico AMC15 Mobile Dental Cart - AMC‑15

SIDE VIEW

Aseptico AMC15 Mobile Dental Cart - SIDE VIEW

TOP VIEW

Aseptico AMC15 Mobile Dental Cart - REAR VIEW

REAR VIEW

Aseptico AMC15 Mobile Dental Cart - TOP VIEW

SETTING UP THE UNIT

  1.  Unpack the AMC-15 from its cardboard crate following the PACKING INSTRUCTIONS on page 16, in the reverse order.

  2. The AMC-15 is configured for 3 Midwest-style (4-hole) non-fiber optic air handpieces: two high-speed (with water) A , and one low-speed (no water) B . Connect compatible handpieces to the connectors in the HP-1, HP-2 and LOW SPEED instru- ment holders. Attach saliva and HVE tips to the valves in the low and high vacuum instrument holders C .
    Aseptico AMC15 Mobile Dental Cart - configured

  3.  Before installing the amalgam separator, remove the two caps from the inlet and outlet ports of the separator. Ensure that the two O-rings that seal the separator against the manifold remain in place on the ends of the port fittings and are not lodged inside the caps. Save the caps for reuse later when sending the amalgam separator to be recycled. Open the waste compartment door on the back side of the cart and place the amalgam separator A into the amalgam manifold B . You will need to loosen the retaining screw C located under the top lid to get the bracket down and around the lip of the amalgam separator. Tighten the retaining screw to secure the amalgam separator.
    Aseptico AMC15 Mobile Dental Cart - amalgam separator

  4. Inside the waste compartment, make sure that the cable from the waste tank sensor is plugged into the connector labeled WASTE SENSOR A on the back wall of the waste compartment. Ensure that all tubes and the bungee tie-down cord are securely attached to the waste tank.
    NOTE: The AMC-15 will not operate without the waste sensor wire plugged into the waste sensor connector. The yellow waste indicator LED on the front panel will illuminate if the waste sensor  cable is disconnected.

  5. Fill the water bottles A with clean water, then screw the bottles back into their fittings. The optional clear water bottles B (PN 730631-01) include floating level indicators C , color coded to identify different fluids such as sterile water or dis- tilled water, as desired.
    Aseptico AMC15 Mobile Dental Cart - fluids CAUTION: Do not run saline solutions through the water system—saline will corrode the plumbing.

  6. Place the foot control A on the floor.

  7. Connect the power cord to the power inlet A on the rear of the cart, then plug the other end of the power cord into a grounded electrical outlet of the proper voltage corresponding to the unit version (115V or 230V). Turn the main power switch B to the  N  (I) position.
    Aseptico AMC15 Mobile Dental Cart - power WARNING: Do not connect the power cord to the AMC-15 in the presence of anesthetics or flammable gases.
    CAUTION: Do not connect 230V power to a 115V unit to avoid circuit breaker damage not covered under warranty.

  8. ACCESSORY TRAY — To install the optional accessory tray, insert the post A into the holder on the top lid of the cart. Use the included 3/16” hex wrench to adjust the height of the pivot block B . Place the end of the arm C onto the pivot block, then place the tray on the other end of the arm.

Aseptico AMC15 Mobile Dental Cart - accessory

WARNING: Maximum weight capacity of the accessory tray is: 10 lb (4.5 kg)

SYSTEM OPERATION

  1. FOOT CONTROL — Press the foot pedal A to activate the working handpiece. Handpiece speed is controlled by how far down you press the foot pedal. To enable water coolant to be activated with the foot pedal, toggle the water coolant switch B towards the  dot C on the foot control, as shown. Toggle the switch away from the dot to disable water coolant.

  2. INSTRUMENT HOLDERS — Each instrument holder (except the syringe holder) provides an on/off toggle switch A and an automatic instrument shut-off switch B . The toggle switch must be ON (towards the dot C , as shown) and the instrument must be out of the instrument holder to operate the instrument. (NOTE: Turn toggle switches OFF to ensure that any instruments that may not be fully seated in their holders will not turn ON accidentally when the foot pedal is activated.)

  3. DRIVE AIR — Pneumatic handpiece air pressure should be adjusted to the pressure recommended by the manufacturer. To determine the current pressure setting, lift the handpiece from the HP-1 instrument holder and press the foot pedal while looking at the handpiece pressure gauge A on the front panel. If adjustment is necessary, lift the top lid on the AMC-15 cart and lock the hinge in the open position. Rotate the HP  1 adjustment screw B counterclockwise to increase pressure or clockwise to decrease pressure. Repeat this procedure for the HP-2 handpiece using the HP 2 adjustment screw C and for the LOW SPEED handpiece using the LOW SPEED HP adjustment screw D. The 2 remaining screws ( E and F ) are not used.

  4. WATER BOTTLES — The AMC-15 incorporates a self-contained pressurized water system which includes: two 1-liter water bottles, a bottle select switch A and a bottle pressure switch B . Use the bottle select switch to direct flow from either water bottle.  The bottle pressure switch allows both bottles to be refilled without draining the air tank.
    (NOTE: Pressure is released from both bottles; you cannot remove one bottle while using the other).
    Aseptico AMC15 Mobile Dental Cart - incorporates To refill the water bottles:
    a. Toggle the bottle pressure switch B OFF to release pressure from both bottles.
    b. Unscrew the bottles from the unit and fill with water. Never use saline as it will corrode the plumbing.
    c. Screw bottles back into their fittings on the unit.
    d. Toggle bottle pressure B ON to pressurize both bottles.
    e. Select the bottle to be used with the bottle select switch A .

  5. 3‑WAY AIR/WATER SYRINGE — Press the left button A to dispense water, press the right button B to dispense air, and press both buttons simultaneously for an air/ water mist. To adjust the air and water flow to the syringe, lift the top lid and lock the
    hinge in the open position. Locate the adjustment block in the right front corner, then use a small flat-blade screwdriver to adjust the left screw C for air flow and the right screw D for water flow. Counterclockwise increases flow and clockwise decreases flow.

  6. WATER COOLANT — The HP 1 A and HP 2 B valves control water coolant flow to the HP-1 and HP-2 handpieces respectively. Switch ON the water coolant toggle on the foot control and switch ON the air coolant toggle on the side control panel. Lift the
    HP-1 handpiece from its holder and run at a midrange speed. Rotate the HP 1 water coolant flow knob clockwise until it seats softly (closed). Begin turning counterclockwise until a fine mist is visible. If a mist is not generated, see AIR COOLANT adjustment below. Repeat water coolant adjustment steps for the HP-2 handpiece using the HP 2 valve.
    Aseptico AMC15 Mobile Dental Cart - WATER COOLANT

  7. AIR COOLANT — Air coolant (chip air) works in combination with the water coolant setting to create a fine mist for the working handpiece. Make sure the water coolant switch  on the foot control is switched ON and the air coolant A is switched ON. Operate a handpiece at a midrange speed. Rotate the air coolant flow knob B control clockwise until it seats softly. Begin turning counterclockwise until a fine mist is visible. Adjust air coolant in combination with water coolant for optimum mist.

  8. FLUSH TOGGLE — Use this toggle switch to flush each handpiece and motor with water. Flush each working instrument for about 5 seconds after each patient, and about 20 seconds at the beginning of each day. To flush an instrument, remove it from its holder and direct the spray nozzle into a basin. Hold the flush toggle A in the ON position for the desired number of seconds, then release. (NOTE: If water bottles empty while flushing, air will be flushed through the lines).

  9. HIGH AND LOW VOLUME VACUUMS — The AMC-15 is equipped with a 5-liter waste tank, a high volume evacuator (HVE) A hose, and a low volume saliva ejector B hose. Both high and low volume hoses operate simultaneously from the same vacuum source. Use the valves on the HVE hose and low volume hose to adjust the relative vacuum pressure between the two hoses, as needed.
    The best vacuum performance will be achieved by closing the saliva ejector valve when operating the HVE, or closing the HVE valve when operating the saliva ejector.
    Aseptico AMC15 Mobile Dental Cart - VOLUME VACUUMS

  10. WASTE INDICATOR — The yellow LED A located on the front panel will light up (or flash) to indicate when the waste tank is full, or the waste tank sensor cable is discon nected. When the waste tank is full, the unit may cycle on and off, or shut down completely, indicating the waste tank must be emptied. See WASTE REMOVAL (page 11) to empty the waste tank. See SETTING UP THE UNIT (page 8) for the location of the waste tank sensor cable.
    Aseptico AMC15 Mobile Dental Cart - WASTE INDICATOR

  11. AUXILIARY WATER CONNECTION — An auxiliary water connection A is provided on the front panel of the AMC-15 cart for connecting optional equipment. The water connection provides 35 psi (2.41 bar) of pressure.
    Aseptico AMC15 Mobile Dental Cart - AUXILIARY WATER

MAINTENANCE AND STERILIZATION

  1. WASTE REMOVAL — Empty the waste tank daily. Waste removal steps:
    a. Open the waste compartment door on the rear of the unit.
    b. Remove the end of waste hose from tethered hose clamp A , then direct the waste hose to a suitable drain.
    Aseptico AMC15 Mobile Dental Cart -WASTE REMOVALc. Switch ON the waste purge pump B to empty the waste tank. It may take a minute before waste water begins to flow from the end of the hose.
    d. Return the end of the waste hose into the clamp A , then pinch clamp closed to prevent leakage.
    e. Check that peristaltic pump latch C is closed fully and waste hose clamp A is secured on waste tube to prevent vacuum leakage. Refer to label in waste compartment.
    The type of amalgam separator used with the AMC-15 requires no settlement time before emptying the waste tank.

  2. WASTE SYSTEM CLEANING — After emptying the waste tank each day, the waste system should be cleaned.
    The waste system consists of:
    • Waste tank
    • Waste discharge hose
    • HVE (high volume ejector) valve and vacuum line
    • Saliva ejector (low volume) valve and vacuum line
    Cleaning steps:
    a. Prepare 1 liter of a 1:10 bleach solution (1 part household bleach to 9 parts water).
    b. Remove the disposable ends from both the HVE and saliva ejector valves.
    c. Vacuum half of the bleach solution through the HVE valve, then the other half through the saliva ejector valve by putting the tip of each valve in the bleach solution.
    d. Flush each valve and line by vacuuming 1/2 liter of clean water through each valve.
    e. Empty the waste tank following the instructions in step 1 above.
    f. Clean the outside of the ejector valves and hoses with a disinfectant solution.

  3. AMALGAM SEPARATOR — The amalgam separator A is located in the waste compartment. The separator filters out particulate matter while the waste water is being pumped out of the unit. A significant drop in waste discharge pressure will occur when the amalgam separator becomes full of solids. Replacement amalgam separators (PN 730595) are available through Aseptico. Refer to the label on the amalgam separator for the address of an approved amalgam recycler. Replace the separator by unscrewing the amalgam separator retaining knob B located under the top lid, and removing the amalgam separator from the waste compartment. NOTE: You can operate the AMC-15 with the bypass filter C installed in place of the amalgam separator, where permitted.
    Aseptico AMC15 Mobile Dental Cart - AMALGAM SEPARATOR1

  4. AMALGAM SOLIDS TRAP — Replace the disposable amalgam solids trap routinely. The performance of the vacuum system will decrease significantly as the trap fills up with solids. The trap sits inside the central vacuum canister located under the top lid. Keep spare traps in stock. The AMC-15 uses standard 1 7/8” traps (both open and closed type traps are compatible):
    Aseptico AMC15 Mobile Dental Cart - AMALGAM SOLIDS
TRAP NOTE: Check with state laws regarding the recycling of amalgam solids waste.

  5. AIR TANK PURGE — Empty the air tank daily to remove condensation from air tank. Place a container underneath the air tank drain hose A to catch water. Rotate the air tank pressure knob B counter-clockwise to empty the air tank, then rotate knob clockwise to close valve when done.

  6. PURGING WATER LINES — Purge all water from the system with air before cleaning the water lines (step 7 below). Also purge the water lines with air if the cart will be unused for more than a week or exposed to freezing temperatures. Purging steps:
    a. Empty both water bottles and reinstall them into their fittings.
    b. Rotate open (counter-clockwise) the HP 1 and HP 2 water coolant knobs (open fully for fastest purging).
    c. Toggle ON all instrument holder switches.
    d. Switch ON the main power.
    e. Toggle ON the bottle pressure switch.
    f. Purge all water lines simultaneously or one at a time: Hold handpieces over a basin while holding the flush toggle switch A in the ON position until all water is purged from the lines with air.
    g. Finally, hold the water syringe over the basin and hold down the water button until all water is purged from the line with air.

  7. CLEANING WATER LINES — Clean the water lines weekly.
    Cleaning steps:
    a. Prepare a 1:10 bleach solution (1 part household bleach to 9 parts water).
    b. Purge all water lines with air (see step 6 above).
    c. Fill water bottles with bleach solution.
    d. Run bleach solution through all lines using the flush toggle switch.
    e. Allow bleach solution to stand in the lines for 10 minutes but no longer: Immediately remove water bottles and discard the bleach, then flush water bottles and all lines thoroughly with clean water.
    f. Purge all water lines with air (see step 6 above) and leave dry until the next clinical use.
    CAUTION: Do not run saline solutions through the water system—saline will corrode the plumbing.

  8. 3‑WAY AIR/WATER SYRINGE — The syringe tip is autoclav able. To remove a tip from the syringe, press down on the locking collar A surrounding the tip socket and pull the tip B straight out of the socket. Clean with a moist cloth, then autoclave at 270° F (132° C) for 10 minutes. To insert a tip, press down on the locking collar and push the tip into the socket as far as it will go. Release the locking collar and gently tug on tip before using to ensure that the tip is securely locked into the socket.
    (NOTE:   Since only the tip can be autoclaved, it is recommended that the air/water syringe be bagged with a disposable, single-use plastic sleeve between each patient use.)

  9. GENERAL CLEANING — The external surfaces of the AMC-15 cart should be cleaned using a soft cloth moistened with a mild detergent solution. Any external surfaces of the cart that are contacted during use should be wiped down with a soft cloth moistened with a disinfectant at the beginning of each day and between each patient use.
    CAUTION: Do not use CaviCide™ cleaner on the unit to avoid damage to the surface.

  10. HANDPIECES — When sterilizing handpieces, follow the instructions provided by the handpiece manufacturer.

  11. SYSTEM PRESSURE — System pressure is set at the factory and should not need adjustment. The system pressure gauge A is located under the top lid and should read approximately 80 psi (5.5 bar). Pressure is controlled by the system pressure regulator B . Contact Aseptico customer service in the unlikely event that adjustment in necessary.

Aseptico AMC15 Mobile Dental Cart - SYSTEM PRESSURE

TROUBLESHOOTING GUIDE

Unit will not start:

  • Check system power connection.
  • Check circuit breakers.
  • Check if waste tank sensor is connected.
  • Check if waste tank is full.

Unit starts but trips circuit breakers:

  • Check source circuit to see if it is a minimum of 15A.
  • NOTE: The use of extension cords is not recommended.

No water pressure:

  • Check water bottles for water level and verify that both caps are tight.
  • Check that water bottle pressure toggle is in the ON position.

Insufficient vacuum:

  • Check that peristaltic pump latch is closed fully and waste tube clamp is secured on waste tube. Refer to label in waste compartment.
  • Check vacuum hose assemblies for blockage.
  • Check amalgam solids trap under the top lid for blockage.
  • Check that the waste tank lid is properly seated and tightly secured.

Insufficient handpiece operation:

  • Check the pressure gauge on the front of the cabinet and ensure that sufficient air is being delivered to the handpiece.
  • Check that handpiece tubing is untangled and not crimped.
  • Check handpiece connection for missing gasket.
  • Check the system pressure gauge under the top lid. System pressure is factory set and should be approximately 80 psi (5.5 bar). Call aseptico customer service if system pressure needs to be adjusted.

No water to handpiece:

  • Check that the toggle on the foot control is in the ON position.
  • Check that bottle select toggle is selecting a bottle filled with water (or that both bottles are full).
  • Check that the bottle pressure toggle is in the ON position.
  • Check that the water coolant valve to the selected handpiece is open.

No coolant air to handpiece:

  • Check that the air coolant toggle is in the ON position.
  • Check that the air coolant flow control valve is open.

Waste pump is not working:

  • Check the circuit breaker next to the waste pump switch.
  • Check to see if the amalgam separator is clogged by replacing it with the bypass filter.
  • Test if the check valve to the amalgam separator is clogged by removing the tube from the check valve closest to the separator and seeing if a cup of water can be extracted.

Unit is turning On and Off and loses suction:

  • Check if waste tank sensor is firmly connected.
  • Check if waste tank is full. (Yellow waste indicator light on the front panel will be on or flashing.)

Amalgam filter or bypass filter leaks:

  • Check that the O-rings are properly installed onto the ends of fittings.

MESSAGES

SYSTEM MESSAGES
None
ERROR MESSAGES
None
FAULT MESSAGES
None

REPLACEMENT PARTS LIST

The following items can be purchased from Aseptico for replacement (or as spares) of standard and optional equipment.

PART NO. ITEM QTY
730595 FILTER REPLACEMENT AMALGAM SEPARATOR 1
730631-01 BOTTLE 1000ML ZIRC KIT MODIFIED (CLEAR 1-LITER WATER BOTTLES W/
LEVEL INDICATORS) 2
AA-20A AUXILIARY ARM WITH ACCESSORY TRAY 1
AA-29 SOLIDS COLLECTOR DISPOSABLE SCREEN 1 7/8” SINGLE (AMALGAM SOLIDS TRAP
(OPEN TYPE)) 1
AA-29-100 SOLIDS COLLECTOR DISPOSABLE SCREEN 1 7/8” SINGLE PKG 100
(AMALGAM SOLIDS TRAP (OPEN TYPE)) 100
AA-290VS SOLIDS COLLECTOR DISPOSABLE COVERED 1 7/8” SINGLE
(AMALGAM SOLIDS TRAP (CLOSED TYPE)) 1
AA-290VS-100 SOLIDS COLLECTOR DISPOSABLE COVERED 1 7/8” SINGLE PKG 100
(AMALGAM SOLIDS TRAP (CLOSED TYPE)) 100
TA-1 SYRINGE TIP AUTOCLAVABLE 3-WAY 1

SYMBOL DEFINITIONS

| Caution – Consult Accompanying Documents
---|---
| Consult Instructions For Use
| Do Not Use If Damaged
| Use By Date
| Part Number
| Follow instructions for use
| Type B Applied Part
| Warning—Potential danger to patient or user (consult accompanying documents)
| Dangerous Voltage
| Alternating Current
| This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately.
| Air Coolant Control
| Water Coolant Control
| Air Coolant Toggle -or- Flush Toggle
| Motor Direction
****| Protective earth (ground)
| Protect Against Dripping Water
| Serial Number
| Footswitch
| On/Off Switch – Auxiliary
| On/Off Switch – Mains
| Bottle Pressure Release Toggle
| Bottle Select Toggle
| Manufacturer
| Atmospheric Pressure Limitation
| Temperature Limitation
| Humidity Limitation
| Sterilizable up to the temperature indicated
| Waste Tank full -or- Waste Level Sensor Cable disconnected

SPECIFICATIONS

Cart Size………. 23 5” W x 30 0” L x 36 5” H (56 7 cm x 76 2 cm x 92 7 cm)
Cart Weight (with all optional equipment)…………… 139 lb (63 kg)
Shipping Crate Size…………………… 30 0” W x 37 0” L x 50 0” H (76 2 cm x 93 9 cm x 127 cm)
Shipping Crate Weight……………………….72 lb (32 7 kg)
Power Source…………………………….115V version: 115V @ 60Hz
230V version: 230V @ 50Hz
Power Rating……………………… 115V version: 11 1A @ 60Hz, 115VAC
230V version: 6 5A @50Hz, 230VAC
Circuit Breakers
……………..115V version: Input: 15A, Pressure: 8A, Vacuum: 8A, Waste Pump: 1A & 3/4A
……………230V version: Input: 15A, Pressure: 5A, Vacuum: 5A , Waste Pump: 1A & 3/4A
Operating Pressure……………………..80 psi (5 5 bar)
High Volume Vacuum …………………6 0 SCFM @ 0 in Hg (169 9 liters/min @ 0 cm Hg)
………….3 6 SCFM @ 4 in Hg (101 9 liters/min @ 10 2 cm Hg)
Low Volume Vacuum ………………………..13 SCFM @ 1 5 in Hg (36 8 liters/min @ 4 in Hg)
Pressure Pump………………………………….1 8 SCFM @ 100 psi (51 0 liters/min @ 6 9 bar)
Air Storage Capacity……………………..2 08 gal (7 9 liters) nominal
Air Storage Pressure………………………….100 psi (6 9 bar)
Water Bottle Capacity…………………. 67 8 fl oz (2 0 liters)
Water Flow ………………………..5 07 fl oz /min (0 15 liter/min)
Waste Tank Capacity……………1 08 gal (4 08 liters)
Noise Level…………………………….Compressor: 58 dBA @ 3’4” (1 meter)
……………Vacuum: 62 dBA @ 3’4” (1 meter)
……………..Compressor & Vacuum Combined: 63 dBA @ 3’4” (1 meter)
AMC-15 Cart Duty Cycle Continuous
Environmental Conditions
Operating Temperature: …………….50° to 104° F (10° to 40° C)
Transport/Storage Temperature:………………………. -40° to 160° F (-40° to 71° C)
Relative Humidity: ……………..10 to 95% non-condensing
Altitude: ………………………0 to 10,000 feet (0 to 3048 meters)
IMPORTANT When running the AMC-15 unit at 50Hz, expect approximately 17% less vacuum and pressure volume due to slower turning of the compressor.
NOTE With regard to setting the handpieces pressure, ‘kg/cm2’ and ‘bar’ are equivalent. This device was tested to the parameters for Electromagnetic Emission and Immunity as stated in IEC 60601-1-2:2007 and is within those parameter limits. These requirements provide reasonable protection against harmful electromagnetic interference in a typical medical installation. However, high levels of radio-frequency (RF) emissions from electrical devices, such as cellular phones, may disrupt the performance of this device. To mitigate disruptive electromagnetic interference, position this device away from RF transmitters and other sources of electromagnetic energy.

The AMC-15 has the following applied parts that come into physical contact with the Patient during normal use. All are type B applied parts as defined in IEC 60601-1 safety standard.

  1. Pneumatic Handpieces with integrated air and water cooling
  2. 3-way air/water syringe
  3. High Volume Suction (HVS)
  4. Low Volume Suction

Classifications:

  • Class I
  • IPX0 – Cart; IPX1 – Footswitch
  • Type B Applied Parts
  • Ordinary Protection
  • Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
  • Note: For reliable grounding, connect to receptacle labeled “Hospital Grade.”

Medical electrical equipment with respect to electric shock, fire and mechanical hazards only in accordance with ANSI/AAMI ES60601-1:2005/(R)2012) (Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance) CSA CAN/CSA-C22.2 NO. 60601-1:14 (Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance). Particular requirements for basic safety and essential performance of dental equipment: IEC 80601-2-60(First Edition): 2012

Disposing of your old product:
Contact your local disposal authority and comply with the correct disposal method and follow national guidelines for unit disposal.

WARRANTY

Aseptico Inc. warrants its new products against defects in material and workmanship under normal and proper use, care, and maintenance for a period of two (2) years from date of original invoice. This two (2) year warranty does NOT apply nor is it extended to
products that are not manufactured by Aseptico. These products may be covered by a  separate limited warranty provided bythe  particular manufacturer, and all claims and questions regarding the same are to be directed to the particular manufacturer. Expendable components, such as batteries, fuses, light bulbs, and tubing sets installed on Aseptico products are specifically excluded and have no warranty. Consumable goods are warranted for the stated expiration date of such goods.
Repair or replacement of any product(s) or part(s) under this warranty does not extend the term of this warranty, and such product(s) or part(s) shall remain covered by the unexpired portion of the warranty period, or for ninety (90) days from the date of return to Aseptico, whichever is later. This limited warranty applies only to the initial or first installation of the product or part.
During the specific warranty periods set forth above, Aseptico will, at its option, repair or replace the product(s) or particular part(s) that are found to be defective in either material or workmanship in part or whole. Aseptico shall be the sole arbiter of such action. In the event of alleged defect under warranty, the purchaser is to notify Aseptico’s Customer Service department promptly. Customer Service will provide Return Material Authorization (RMA) instructions, usually directing that the product be returned for service, shipping prepaid by the buyer or end user, to Aseptico or its designated and authorized warranty service center.
This warranty shall not apply to products (1) that have been subjected to neglect, abuse, misuse, improper installation, inadequate maintenance, or damage due to improper use of cleaning materials or chemicals, or non- compliance with Aseptico’s storage, installation, operation, maintenance or environmental requirements; (2) that have undergone any modification or repair not previously authorized by Aseptico in writing, or service, repair or modification by or from any facility other than an authorized Aseptico service center or technician, or that use non-authorized software or spare or replacement parts; or (3) that fail due to reasonable and normal use or wear and tear, or materials made, furnished or specified by the buyer or end user.
Aseptico does not assume under this warranty any risks or liabilities arising from the clinical use of its products, whether or not such use involves coincidental utilization of products manufactured by others. Under no circumstances will Aseptico be liable or responsible for special, compensatory, incidental, consequential or punitive damages, lost profits, lost sales, or loss of use or loss of business opportunity by or through the use of the product. Aseptico’s sole and maximum liability with respect to the product, other than its obligations set forth above, shall be the total purchase price paid for the product.

PACKING INSTRUCTIONS

  1. Remove all liquid wastes, water from water bottles, and the bypass filter and amalgam separator.

  2. Remove all accessories from the unit and purge air pressure from air tank.

  3. Place shipping crate base on level floor. Lift the AMC-15 cart onto the pedestal A in the crate base.
    CAUTION: Two people are required to lift the cart.

  4. Strap the ends of both handles A to the four anchor eye-bolts B on the base.

  5. Place the boxed foot control pedal A under the cart.

  6. Wrap the cart with pallet stretch wrap.
    Aseptico AMC15 Mobile Dental Cart -cart 2

  7. Insert the cardboard end panels and side panels into the crate base, making sure the “End Up” labels A arepointing up. Use the adhesive strips on the inside of the end panel flaps to attach to the side panels.

  8. Place the packing material A on top of cart, then place the accessories box B on top of a piece of bubble wrap.
    Aseptico AMC15 Mobile Dental Cart -material29. Place the lid on top of the crate. Position the four strapping blocks A onto the edges of the lid, then connect the straps over the blocks and tighten.

Aseptico logo

P.O. Box 1548 • Woodinville, WA 98072-1548
(800) 426-5913 • (425) 487-3157 • Fax: 425-487-2608
www.aseptico.com[email protected]
P/N: 421208
Rev. E ECO 15628
12/2022
PRINTED IN THE USA

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Aseptico User Manuals

Related Manuals