Aseptico ADU-03ST Manual Dental Control System Instruction Manual

October 30, 2023
Aseptico

Aseptico - logo ADU-03ST Manual Dental Control System
Instruction ManualAseptico ADU-03ST Manual Dental Control
System

SPECIFICATIONS:

Case Size: 11 in. x 11 in. x 40 in.
Case Weight: 17 lbs.
Power Source: 80 – 100 PSI Compressed Air
To prevent injury to people and damage to property, please heed relevant warnings and remarks. They are marked as follows:

  • WARNING: Serious injury or death may result if ignored.
  • CAUTION: Damage to property or the environment may result if ignored.
  • NOTE: Important additional information and hints.

The Aseptico ADU-03ST is one of the finest high volume evacuator systems available to the professional. Offering manual activation and a remote stand, the ADU-03ST provides an economic HVE control system for a dental operatory, hygienist’s room, or satellite office.
Congratulations!
This system is engineered to provide many years of reliable service. Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment.
Separate manuals may be provided to cover the operation and maintenance of hand-pieces or other accessories for your unit. Handpieces are purchased separately.

PACKAGE CONTENTS:

  • TA-90D 3-Way Air/Water Syringe with Tips

  • NWS-8 Self Contained Clean Water System

  • 25-Foot Extension Line

  • AA-61 Water Connector for scalers
    Options:

  •  AA-25ST Saliva Ejector System

  • AA-21ST High Volume Evacuator

SETTING UP THE UNIT:Aseptico ADU-03ST Manual Dental Control System -
SETTING UP

  1. The ADU-03ST comes factory-mounted on a tripod stand.
  2.  Remove the system from the carrying case (Fig. 1), fully extend tripod legs, and stand the system up (Fig. 2).
  3. Unit height is adjusted by loosening the adjustment handle on tripod, raise or lower system, then tighten adjustment handle.

Aseptico ADU-03ST Manual Dental Control System - ADU-03ST
Setup

REMOTE SETUP:

  1. Remove components from the soft carrying case as required. Note item locations to ease repackaging.

  2.  Attach one end of the 25 ft. grey extension line to the grey air supply line from the unit. Connect the other end to a clean regulated 80-100 PSI compressed air supply.

  3.  Locate the white plastic water supply bottle in the soft carrying case. Fill the bottle with clean or sterile water and attach it to the white fitting on the underside of the ADU-03ST.

  4.  Optional AA-25ST Saliva Ejector System – The Saliva Ejector
    System consists of a plastic waste container and a single lowvelocity hose. Mount the waste container to the quick connect on the mounting block located on the back left-hand side of the chassis (see Fig. 3). Push the waste container straight up into the Quick Connector, until the snap ring locks it into place. Attach saliva ejector hose to the quick connect located on top of waste container and then place ejector valve into the holder located on left side of unit. To remove waste container assembly, push in on snap-ring tab and pull the assembly out of Connector.

  5. Optional AA-21ST HVE System – The High Volume Evacuator consists of a plastic waste container and a single large-diameter low velocity hose. Mount the waste container and hose assembly to the quick connect on the mounting block located on the back right-hand side of the chassis (see Fig. 3), then place HVE valve in the holder on the right side of unit. To attach the assembly, push the waste container and hose assembly straight into the Quick Connector, until the snap ring locks it into place. To remove assembly, push in on snap-ring tab and pull the assembly out of Connector.

  6.  Slide the accessory tray between the ADU-03ST chassis back and the mounting block.

OPERATION FUNCTIONS

Aseptico ADU-03ST Manual Dental Control System - OPERATION
FUNCTIONS Pressure Gauge:
Gives a visual indication of the system pressure.
3-Way Air /Water Syringe:
Delivers air, water, or a combination mist. Supplied with autoclavable tip.
NWS-8 Water System:
White 1-liter bottle dispenses water through the 3-Way Air/ Water Syringe and high-speed handpiece.
AA-25ST Saliva Ejector On/Off Toggle (Optional):
Turns vacuum On/Off to the saliva ejector valve.
AA-21ST High Volume Evacuator (Optional):
Provides vacuum flow to the High Volume Evacuator when the HVE valve button is depressed.
AA-61 Water Out Connector:
Provides water via standard 1/4” quick disconnect to independent Cavitron or scaler units requiring a water supply. Remove the black rubber cap, then connect water line to this connector. FOR WATER OUT CONNECTIONS ONLY!
UNIT OPERATION:
3-Way Air/Water Syringe:

Depress the right button for air operation, and the left button for water operation. Depressing both buttons will create a mist. The syringe features quick-change autoclavable tips: To remove a tip, press on the locking collar surrounding the tip socket and pull the used tip straight out of the socket (see Fig. 5). To insert a new tip, press locking collar and push tip into socket as far as it will go. Release collar and gently tug on tip before using to ensure that tip is securely locked into socket. Optional Saliva Ejector:
Remove the saliva ejector valve from its holder and insert a standard saliva ejector tip. Turn the low vacuum toggle switch ON. Waste from the vacuum system will collect in the attached plastic bottle. When the waste container is 3/4 full, it must be emptied to prevent a back-up in the system.
Optional HVE System:
Remove the vacuum hose from its holder and insert a standard oral evacuator tip. Depress the button switch on the side of the HVE valve for vacuum. Waste from the vacuum system will collect in the attached plastic bottle. When the waste container is 3/4 full, it should be emptied.

CLEANING

Syringe Tip Sterilization:

  1. Remove contaminated syringe tip.
  2.  Remove all visible signs of contamination before autoclaving.
  3.  Autoclave tip at 132° C (270° F) for ten minutes.
  4.  Sterilize between each patient use.

NOTE: Since only the tips can be autoclaved, it is recommended that the Air/Water Syringe be bagged with a disposable, singleuse plastic sleeve between each patient use.
Saliva Ejector:
The valve assembly on the saliva ejector snaps apart at the swivel. After each use, the vacuum system should be cleaned by running clean water or vacuum system cleanser through the evacuator hose.
HVE System:
After each use, the high volume evacuator system should be cleaned by running clean water or a vacuum system cleanser through the evacuator hose.
Vacuum System Waste Bottles:
Waste from the vacuum systems will collect in their respective plastic waste bottles. When the waste containers are 3/4 full, they should be emptied. After each use, the vacuum systems should be cleaned by running clean water or a vacuum system cleanser through the evacuator hoses. CAUTION: Use only NON- foaming  cleansers in the vaccum lines. Thoroughly clean the waste bottles.
Water Lines:
Disinfect the water lines weekly. Prepare a 1:10 bleach solution (1 part household bleach to 10 parts water). Remove water reservoir and discard residual water. Replace empty water supply tank and air purge all waterlines. Fill water supply tank with bleach  solution. Run bleach solution through all lines. Allow bleach solution to stand in lines for 10 minutes. Remove water supply tank and discard bleach. Flush water supply tank and all lines thoroughly with clean water. Air purge and leave lines dry until next clinical use.

WARRANTY:

Aseptico Inc. warrants its new products against defects in material and workmanship under normal and proper use, care, and maintenance for a period of two (2) years from date of original invoice. This two (2) year warranty does NOT apply nor is it extended to products that are not manufactured by Aseptico. These products may be covered by a separate limited warranty provided by the particular manufacturer, and all claims and questions regarding the same are to be directed to the particular manufacturer.
Expendable components, such as batteries, fuses, light bulbs, and tubing ets installed on Aseptico products are specifically excluded and have no warranty. Consumable goods are warranted for the stated expiration date of such goods.
Repair or replacement of any product(s) or part(s) under this warranty does not extend the term of this warranty, and such product(s) or part(s) shall remain covered by the unexpired portion of the warranty period, or for ninety (90) days from the date of return to Aseptico, whichever is later. This limited warranty applies only to the initial or first installation of the product or part.
During the specific warranty periods set forth above, Aseptico will, at its option, repair or replace the product(s) or particular part(s) that are found o be defective in either material or workmanship in part or whole. Aseptic shall be the sole arbiter of such action. In the event of alleged defect under warranty, the purchaser is to notify Aseptico’s Customer Service department promptly. Customer Service will provide Return Material Authorization (RMA) instructions, usually directing that the product be returned for service, shipping prepaid by the buyer or end user, to Aseptico or its designated and authorized warranty service center. This warranty shall not apply to products (1) that have been subjected to neglect, abuse, misuse, improper installation, inadequate maintenance, or damage due to improper use of cleaning materials or chemicals, or noncompliance with Aseptico’s storage, installation,  operation, maintenance or environmental requirements;  (2) that have undergone  any modification or repair not previously  uthorized by Aseptico in writing, or service, repair or modification by or from any facility other than an authorized Aseptico service center or technician, or that use non-authorized software or spare or replacement parts; or (3) that fail due to  reasonable and normal use or wear and tear, or materials made, furnished or specified by the buyer or end user.
Aseptico does not assume under this warranty any risks or liabilities arising from the clinical use of its products, whether or not such use involves coincidental utilization of products manufactured by others. Under no circumstances will Aseptico be liable or responsible for special, compensatory, incidental, consequential or punitive damages, lost profits, lost sales, or loss of use or loss of business opportunity by or through the use of the product. Septic’s sole and maximum liability with respect to the product, other than its obligations set forth above, shall be the total purchase price paid for the product.

Aseptico - logoP.O. Box 1548 • Woodinville, WA 98072
8333 216th Street SE • Woodinville, WA 98072
(425) 487-3157 • (800) 426-5913
www.aseptico.com info@aseptico.com
P/N: 420284-01
Rev. E
ECO 15498
04/2022
PRINTED IN THE
U.S.A

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals