MPM-45 Series 45W AC-DC High Reliable
PCB-Mount Green Medical Power Module
Instruction Manual
MPM-45 Series 45W AC-DC High Reliable PCB-Mount Green Medical Power
Module
https://www.meanwell.com/Upload/PDF/PCB_EN.pdf ANSI AAMI ES60601 -1 BS
EN/EN60601 -1 IEC60601-1 TPTC004
Features
- 3.43″x2.05″ compact size
- PCB chassis or screw terminal mounting version
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- No load power consumption<0.1W
- Extremely low leakage current
- Wide operating temp. range -30 – +80 C
- EMI Class B without additional components
- Isolation Class II
- Protections: Short circuit / Overload / Over voltage
- No minimum load required
- Operating altitude up to 4000 meters( Note.7)
- 50W peak(10 sec.)
- 3 years warranty
Applications
- Portable medical device
- Mobile clinical workstation
- Medical computer monitor
- Medical examination instrument
■ GTIN CODE
MW Search: https:
www.meanwell.com/serviceGTIN.aspx
Description
MPM-45 is a 45W high density and small size (87x52x29.5mm) AC/DC PCB-mount
type medical grade power supply. It features the operation for 80-264VAC, a
low no load power consumption less than 0.1W, a high efficiency up to 92.5%,
Class II (no FG) double insulation, outstanding dissipation,2-5G anti-
vibration by model, high EMC performance, 4KVAC isolation, etc. The design
observes IEC/BS EN/EN60601-1 and ANSI/AAMI ES60601-1version three with 2 x
MOPP level and ultra-low leakage current (<100pA). It is very suitable for BF
(patient contact) type medical device or relevant equipment.
■ Model Encoding
SPECIFICATION
MODEL | MPM-45-5 | MPM-45-12 | MPM-45-5 | MPM-45-24 | MPM-45-8 |
---|---|---|---|---|---|
OUTPUT | DC VOLTAGE | 5V | 12V | 15V | 24V |
CURRENT | Peak(10 sec.) | 8.8A | 4.13A | 3.3A | 2.1A |
Convection | 8A | 3.75A | 3A | 1.88A | 0.94A |
RATED POWER | Peak(10 sec.) Note.2 | 44W | 49.5W | 49.5W | 50.4W |
Convection | 40W | 45W | 45W | 45.1W | 45.1W |
RIPPLE & NOISE (max.) Note.3 | 80mVp-p | 120mVp-p | 120mVp-p | 200mVp-p | 240mVp-p |
VOLTAGE TOLERANCE Note.4 | ±2.0% | ±2.0% | ±2.0% | ±2.0% | ±2.0% |
LINE REGULATION | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ±0.5% |
LOAD REGULATION | ±1.0% | ±1.0% | ±0.5% | ±0.5% | ±0.5% |
SETUP, RISE TIME | 1000ms, 30ms/230VAC 1000ms, 30ms/115VAC at full load | ||||
HOLD UP TIME (Typ.) | 50ms/230VAC 12ms/115VAC at full load | ||||
INPUT | VOLTAGE RANGE Note.5 | 80 ~ 264VAC 113 ~ 370VDC | |||
FREQUENCY RANGE | 47 ~ 63Hz | ||||
EFFICIENCY (Typ.) | 88% | 91.5% | 92.5% | 92.5% | 92% |
AC CURRENT (Typ.) | 1.2A/115VAC 0.6A/230VAC | ||||
INRUSH CURRENT (Typ.) | COLD START 30A/115VAC 60A/230VAC | ||||
LEAKAGE CURRENT (max.) Note.6 | Touch current <100μA/264VAC | ||||
PROTECTION | OVERLOAD | 115% ~ 135% rated output power |
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 5.3 ~ 7.2V| 12.6 ~ 16.2V| 15.8 ~ 20.3V| 25.2 ~ 32.4V| 50.4 ~
64.8V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| Protection type : Shut down o/p voltage, re-power on to
recover
ENVIRONMENT| WORKING TEMP.| -30 ~ +80℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP.| -40 ~ +85℃
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 60℃)
SOLDERING TEMPERATURE| 260℃±5℃/10sec.max.
VIBRATION| Blank:10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
ST:10 ~ 500Hz, 5G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.7| 4000 meters / OVCⅡ
SAFETY & EMC
(Note 8)| SAFETY STANDARDS| IEC60601-1, BS EN/EN60601-1, EAC TP TC 004,
UL ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved; Design
refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted| BS EN/EN55011 (CISPR11)| Class B
Radiated| BS EN/EN55011 (CISPR11)| Class B
Harmonic Current| BS EN/EN61000-3-2| Class A
Voltage Flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m(80MHz~2.7GHz) Table
9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| >95% dip 0.5 periods, 30% dip
25 periods, >95% interruptions 250 periods
OTHERS| MTBF| 4590.4K hrs min.
Telcordia SR-332 (Bellcore) ; 563.4K hrs min.
MIL-HDBK-217F (25℃)
DIMENSION| PCB mounting style : 875229.5mm (LWH)
Screw terminal style : 1095233.5mm (LWH)
PACKING| PCB mounting style : 0.185Kg;60pcs/12.1Kg/0.94CUFT
Screw terminal style :0.206Kg;50pcs/11.3Kg/0.56CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC input,
rated load and 25℃ of ambient temperature.
2. 33% Duty cycle maximum within every 30 seconds. Average output power
should not exceed the rated power.
3. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted
pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
4. Tolerance : includes set up tolerance, line regulation and load
regulation.
5. Derating may be needed under low input voltages. Please check the derating
curve for more details.
6. Touch current was measured from primary input to DC output.
7. The ambient temperature derating of 3.5℃/1000m with fanless models and of
5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
8. The power supply is considered a component which will be installed into a
final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer
to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
※ Product Liability Disclaimer:For detailed information, please refer to
https://www.meanwell.com/serviceDisclaimer.aspx
-
Block Diagram
-
Derating Curve
-
Output Derating VS Input Voltage
-
Mechanical Specification
‧ PCB mounting style (MPM-45)
‧ Screw terminal style (MPM-45-xxST)
-
Installation Manual