MEAN WELL MFM-05 Series 5W High Reliable Green Medical On Board Type Owner’s Manual
- June 10, 2024
- MEAN WELL
Table of Contents
5W High Reliable Green Medical On Board Type
Owner’s Manual
Features
- 65″x0.88″ compact size
- Medical safety approved (2 x MOPP) according to
- ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- No load power consumption<0.075W
- Extremely low leakage current
- Wide operating temp. range -40 — +85’C
- EMI class B for class II configuration
- Protections:
- Short circuit / Overload / Over voltage / Over temperature
- No minimum load required
- Typical lifetime > 52K hours
- 3 years warranty
Applications
- Portable medical device
- Mobile clinical workstation
- Medical computer monitor
- Medical examination instrument
GTIN CODE
MW Search https://www.meanwell.comiserviceGTIN.aspx
Description
MFM-05 is a 5W high density and small size (4222.320.5mm) AC/DC on board type medical grade power supply series. It features the operation for 80-264VAC, a low no load power consumption less than 0.075W, a high efficiency up to 82%, Class II (no FG) double insulation, outstanding dissipation, 5G anti-vibration, high EMC performance, 4KVAC isolation, etc. The design observes IEC/BS EN/EN60601-1 and ANSI/AAMI ES60601-1 version three with 2xMOPP level and ultra-low leakage current (<80 it A). It is very suitable for BF (patient contact) type medical device or relevant equipment.
Model Encoding
SPECIFICATION
MODEL | MFM-05-3.3 | MFM-05-5 | MFM-05.12 | MFM-05-15 | MFM-05-24 |
---|---|---|---|---|---|
OUTPUT | DC VOLTAGE | 3.3V | 5V | 12V | 15V |
RATED CURRENT | 1.25A | 1A | 0.42A | 0.33A | 0.23A |
CURRENT RANGE Note.2 | 0 – 1.25A | 0 – 1A | 0 – 0.42A | 0 – 0.33A | 0 – 0.23A |
PEAK CURRENT | 1.38A | 1.1A | 0.46A | 0.36A | 0.25A |
RATED POWER | 4.1W | 5W | 5W | 5W | 5.5W |
PEAK LOAD(105eC.) Note.3 | 4.6W | 5.5W | 5.5W | 5AW | 6W |
RIPPLE & NOISE (MaX.) Note.4 | 100mVp-p | 100mVp-p | 150mVp-p | 150mVp-p | 180mVp-p |
VOLTAGE TOLERANCE Note.5 | ±2.5% | ±2.5% | ±2.5% | ±2.5% | 13.% |
LINE REGULATION | ±0.3% | ±0.3% | ±0.3% | ±0.3% | ±0.3% |
LOAD REGULATION | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ±0.5% |
SETUP, RISE TIME | 1000ms, 30ms/230VAC 1000ms, 30ms/115VAC at full load | ||||
HOLD UP TIME (Typ.) | 40ms/230VAC 12ms/115VAC at full bad | ||||
INPUT | VOLTAGE RANGE Note.6 | 80- 264VAC | |||
FREQUENCY RANGE | 47 – 440Hz | ||||
EFFICIENCY (Typ.) | 74% | I 78% | I 80% | I 81% | I 82% |
AC CURRENT (Typ.) | 0.2A/115VAC 0.1Al230VAC | ||||
INRUSH CURRENT (Ty p.) | COLD START 25A/115VAC 45A/230VAC | ||||
LEAKAGE CURRENT (max.) Note.7 | Touch current <80fiA/264VAC | ||||
PROTECTION | OVERLOAD | 110% – 180% rated output power |
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 3.8 – 5V| I 5.75 – 6.8V| I 13.8 – 16.2V| 1 17.3 -20.3V| 27.6 –
32.4V
Protection type : Shutoff o/p voltage, clamping by zener diode
OVER TEMPERATURE| Protection type : Shutdown o/p voltage, recovers
automatically after temperature goes down
EWBONMENT| WORKING TEMP.| -40 – +85°C (Refer to Merating Curve’)
WORKING HUMIDITY| 20-90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 – +100°C , 10 -95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/°C (0 – 60°C)
SOLDERING TEMPERATURE| Wave soldering: 265°C,5s (max.); Manual soldering:
390°C ,3s (max.)
VIBRATION| 10 -500Hz, 5G 10m initcycle, period for 60min. each along X, Y, Z
axes
OPERATING ALTITUDE Note.8| 5000 meters
SAFETY &
EMC (Note 9)| SAFETY STANDARDS| I EC60601-1, BS EN/EN60601-1, EAC TP TC 004,
ULANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3m Edition approved ; Design
refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-0/P:4KVAC
ISOLATION RESISTANCE| I/P-0/P:100M Ohms/ 500VDC / 25°C/ 70T RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted| BS EN/EN55011 (CISPR11)| Class B
Radiated| BS EN/EN55011 (CISPR11)| Class B
Harmonic Current| BS EN/EN61000-3-2| Class A
Voltage Fficker| BS EN/EN61000-3-3|
EMC IMMUNITY| BS EN/EN55035. BS EN/EN60601-1-2
Parameter| Standard| Test Level I Note
ESD| BS EN/EN61000.4-2| Level 4, 15KV air ; Level4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m(80MHz-2.7GHz )
Table 9, 9-28V/m( 385MHz-5.78GHz )
EFT bursts| BS EN/EN61000.4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods, 100% interruptions 250 periods
OTHERS| MT BF| 9337.3K hrs min. Tetcordia SR-332 (Be core) :1799.5K hrs min.
MIL-HDBK-217F (25°C)
DIMENSION| 42’22.3’20.5mm (L’W’H) or 1.65″0.88’0.80′ inch
PACKING| 0.018Kg; 270pcs/5.8Kg/0.94CUFT
NOTE| 1. All parameters NOT specie ly mentioned are measured at 230VAC input,
rated load and 25°C of ambient temperature.
2. No minimum load required.
a 33% Duty cycle maximum within every 30 seconds. Average output power should
not exceed the rated power
4. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted
pair-wire terminated with a 0.1“f & 47;4 parallel capacitor.
5. Tolerance : includes set up tolerance, line regulation and load
regulation.
6. Deiating may be needed under low input voltages. Please check the deiating
curve for more details.
7. Touch current was measured from primary input to DC output.
8. The ambient temperature derating of 3.5’C/1000m with fanless models and of
5°C /1000m with fan models for operating altitude higher than 2000m(6500ft).
9. The power supply is considered a component which will be installed into a
final equipment. The final equipment must be reconfirmed that it still meets
EMC drectives. For guidance on how to perform these EMC tests, pleas refer to
‘EMI testing of component power supplies.” (as available on
http://www.meanweil.com)
Product Liability Disclaimer : For detailed information, please refer to