ICI Titan HD Series High Resolution Titan HD Infrared Camera User Guide

June 9, 2024
ICI

ICI Titan HD Series High Resolution Titan HD Infrared Camera

THE MANUFACTURER RESERVES THE RIGHT TO ALTER THE SPECIFICATIONS OF THE PRODUCT WITHOUT PRIOR NOTIFICATION. THE MANUFACTURER ALLOWS HIMSELF THE RIGHT TO MODIFY WITHOUT ANY PRELIMINARY OPINION THE TECHNICAL SPECIFICATIONS OF THE PRODUCT. THIS MANUAL MAY CONTAIN TECHNICAL INACCURACIES OR TYPOGRAPHICAL ERRORS.

NOTICE ALL STATEMENTS, INFORMATION, AND RECOMMENDATIONS IN THIS MANUAL ARE BELIEVED TO BE ACCURATE BUT ARE PRESENTED WITHOUT WARRANTY OF ANY KIND. NOTWITHSTANDING ANY OTHER WARRANTY HEREIN, ALL DOCUMENT FILES AND SOFTWARE ARE PROVIDED “AS IS” WITH ALL FAULTS. ICI DISCLAIMS ALL WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING, WITHOUT LIMITATION, THOSE OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT OR ARISING FROM A COURSE OF DEALING, USAGE, OR TRADE PRACTICE. THE SOFTWARE LICENSE AND LIMITED WARRANTY FOR THE ACCOMPANYING PRODUCT ARE SET FORTH IN THE INFORMATION PACKET THAT SHIPPED WITH THE PRODUCT AND ARE INCORPORATED HEREIN BY THIS REFERENCE. IF YOU ARE UNABLE TO LOCATE THE SOFTWARE LICENSE OR LIMITED WARRANTY, CONTACT YOUR ICI REPRESENTATIVE FOR A COPY. IN NO EVENT SHALL ICI BE LIABLE FOR ANY INDIRECT, SPECIAL, CONSEQUENTIAL, OR INCIDENTAL DAMAGES, INCLUDING, WITHOUT LIMITATION, LOST PROFITS OR LOSS OR DAMAGE TO DATA ARISING OUT OF THE USE OR INABILITY TO USE THIS MANUAL, EVEN IF ICI HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.

Package Includes

ICI-Titan-HD-Series-High-Resolution-Titan-HD-Infrared-Camera-
fig-1

ENSURE ALL SYSTEM EQUIPMENT AND COMPONENT ITEMS ARE PRESENT BEFORE BEGINNING INSTALLATION

Set Up

  • The camera can be used as a handheld device.
  • Mount the fully charged device to a tripod using the ¼”-20 mount located on the body.ICI-Titan-HD-Series-High-Resolution-Titan-HD-Infrared-Camera-fig-2

MAKE SURE TRIPODS DO NOT BLOCK THE DIRECT PATH OF PERSON(S) TO BE IMAGED TO ENSURE THE EQUIPMENT WILL NOT BE MOVED OR KNOCKED DOWN. USING A DIVIDING BARRIER WILL HELP KEEP TRIPODS SEPARATE FROM THE PATH.

TURN ON THE CAMERA BY PRESSING AND HOLDING THE POWER BUTTON FOR AT LEAST TWO SECONDS.

Power on device

PRESS AND HOLD THE POWER BUTTON FOR MORE THAN 5 SECONDS TO SHUT DOWN THE CAMERA. THE THERMAL IMAGER NEEDS SUFFICIENT WARM-UP TIME FOR THE MOST ACCURATE TEMPERATURE MEASUREMENTS AND BEST IMAGE QUALITY. THIS TIME CAN OFTEN VARY BY ENVIRONMENTAL CONDITIONS. IT IS BEST TO WAIT A MINIMUM OF 10 MINUTES FOR THE DEVICE TO COMPLETELY WARM-UP.

Battery Installation

MAKE SURE TO HOLD THE CAMERA FIRMLY IN BOTH HANDS WHEN REMOVING OR REPLACING THE BATTERY.

  • Slide the battery bay cover latch.
  • Open the battery bay cover.
  • Ensure metal contacts are aligned; then, push the battery into the battery bay until battery is secure under battery buckle.
  • To replace a battery held in the battery bay push and hold the battery buckle; then, pinch the battery handle with the index finger and thumb; pull to remove battery.

DO NOT FORCE THE BATTERY INTO THE BATTERY BAY. IF USER FEELS SIGNIFICANT RESISTANCE REMOVE THE BATTERY AND ENSURE METAL CONTACTS ARE ALIGNED BEFORE TRYING TO PLACE BATTERY INTO THE BATTERY BAY.

MAKE SURE TO HOLD THE CAMERA FIRMLY IN BOTH HANDS WHEN REMOVING OR REPLACING THE BATTERY.

  • Close the battery bay.
  • Slide the battery bay latch to secure the battery bay cover.

ENSURE BATTERY COVER IS PROPERLY CLOSED. IMPROPER CLOSURE OF BATTERY COVER MAY EXPOSE THE BATTERY AND INTERNAL ELECTRONICS TO ELEMENTAL DAMAGE.

Memory Card

  • Align the contact pins of a memory card to the front of the device and insert the card into the proper slot (SD or Micro-SD) on the side of the device until a clicking sound is heard.ICI-Titan-HD-Series-High-Resolution-Titan-HD-Infrared-Camera-fig-10
  • To eject the memory card, press down until a click is heard; then, release. The memory card will be ejected from the device.

DO NOT FORCE MICRO-SD CARD INTO THE SLOT. THE SIDE OF THE CARD WITH CONTACT PINS MUST BE INSERTED FIRST. IF THERE IS RESISTANCE FLIP THE CARD AROUND AND TRY AGAIN. TO EJECT THE MEMORY CARD PRESS UNTIL A CLICK IS HEARD AND THEN RELEASE. MEMORY CARD WILL BE EJECTED.

Charging Instructions: Option 1

  • While a battery is inside the device, plug the Type-C end of either the Type-C Cable or USB Cable into the Type-C port of the device.
  • Plug the other end of the cable into the proper slot (Type-C or USB) of the power adapter.ICI-Titan-HD-Series-High-Resolution-Titan-HD-Infrared-Camera-fig-11
  • Plug power adapter into a 110/120V electrical outlet.

Charging Instructions: Option 2

REMOVE THE BATTERIES FROM THE DEVICE BEFORE CONTINUING WITH THIS METHOD OF CHARGING.

  • Insert batteries into the battery charging base.
  • Plug the Type-C end of either the Type-C Cable or USB Cable into the back of the battery charging base.
  • Plug the other end of the cable into the proper slot (Type-C or USB) of the power adapter.
  • Plug power adapter into a 110/120V electrical outlet.

WHEN THE BATTERY IS NOT PLACED IN THE BATTERY CHARGING BASE THE INDICATOR LIGHT IS BLUE. THE LIGHT FLASHES BLUE WHEN THE BATTERY IS PLACED AND CHARGING. WHEN THE BATTERY IS FULLY CHARGED THE LIGHT IS STEADY BLUE. CHARGE THE BATTERY FOR A MINIMUM OF 4 HOURS BEFORE USING THE DEVICE. ICI DOES NOT RECOMMEND CHARGING THE DEVICE WHILE IT IS ON A TRIPOD. LAY DEVICE FACE DOWN ON A FLAT AND STABLE DESK OR TABLE WHILE CHARGING. ENSURE THE IMAGER IS NEAR ROOM TEMPERATURE BEFORE CHARGING THE DEVICE. DO NOT CHARGE IN EXTREMELY HOT OR COLD AREAS. THE BATTERY CAPACITY MAY BE DECREASED IF CHARGED IN AN EXTREME TEMPERATURE ENVIRONMENT. IF YOU DO NOT USE THE IMAGER FOR A LONG PERIOD OF TIME, PUT THE DEVICE IN A COOL AND DRY ENVIRONMENT. THE BATTERY WILL DISCHARGE OVER TIME; THEREFORE, THE IMAGER SHOULD BE CHARGED OCCASIONALLY TO PREVENT DAMAGE TO THE BATTERY OR THE DEVICE.

Lens Cover

MAKE SURE TO HOLD THE CAMERA FIRMLY IN BOTH HANDS WHEN REMOVING OR REPLACING THE LENS COVER.

  • Firmly grip around the outer lens ring.
  • Pull the lens cover from the lens. Replace lens cover by firmly pressing lens cover onto the lens.

REPLACE THE LENS COVER WHEN THE DEVICE IS NOT IN USE.

Lens Installation

MAKE SURE TO HOLD THE CAMERA FIRMLY IN BOTH HANDS WHEN CHANGING THE LENS.

  • Firmly grip around the outer lens ring and rotate the lens 90° counter-clockwise.
  • Pull the lens from the lens mount.
  • Align the index marks on the new lens and lens mount.
  • Push the lens into the lens mount; then, rotate the lens 90° clockwise.

DO NOT FORCE LENS INTO THE SLOT AS THREADS MAY STRIP. IF USER FEELS SIGNIFICANT RESISTANCE REMOVE THE LENS AND ENSURE INDEX MARKS AND THREADS ARE PROPERLY ALIGNED; THEN, PUSH THE LENS INTO THE LENS MOUNT AND TRY AGAIN.PLACE THE LENS IN ITS LENS CONTAINER WHEN NOT IN USE TO AVOID DAMAGE TO THE LENS. AVOID DROPPING OR SMASHING THE LENS TO AVOID LENS DAMAGE.

About ICI / Contact

Infrared Cameras, Inc. 2105 W. Cardinal Dr. Beaumont, TX 77705 Phone: 409-861-0788 | Toll Free: 866-861-0788 | International: 409-861-0788 General Inquiry: [email protected] Website: www.infraredcameras.com You may reach a representative by phone or email Mon – Fri 8:00AM – 5:00PM CST. ICI manufactures complete systems and software. We can provide complete engineering, software, and OEM solutions. Our Fortune 500 clients rely on us for infrared equipment and thermography training (which we offer through the Infrared Training Institute). In addition to providing custom germanium, silica, and sapphire optics, we also build windows for enclosures, as well as custom pan and tilt units. We can even provide customizable explosion-proof systems. Our knowledge and experience stems from years of using infrared imaging and temperature measurement instruments to provide solutions to: managers, engineers, scientists, inspectors and operators in space, power companies, medical, pulp and paper, food industry, research and development, and various process industries. You can see our products and services used in industrial, commercial, and government applications worldwide. Originally named Texas Infrared (still DBA), Infrared Cameras, Inc. has been in business since March, 1995. Thank you for your dedicated and continued support.

ICI cameras fall under US Federal Law and Export Control. 2105 W. Cardinal Dr. Beaumont, TX 77705 Phone: 409-861-0788 Toll Free: 866-861-0788 infraredcameras.com [email protected] We Are IR ™ ©Copyright 2022, Infrared Cameras, Inc. – All rights reserved. The contents of this document may not be reproduced in whole or in parts without the written consent of the copyright owner. Printed in the United States of America. 2105 W. Cardinal Dr. Beaumont, TX 77705 866-861-0788 www.infraredcameras.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

ICI User Manuals

Related Manuals