Micsig DP750-100 High Voltage Differential Probe User Manual

June 9, 2024
Micsig

Micsig DP750-100 High Voltage Differential Probe

Micsig-DP750-100-High-Voltage-Differential-Probe

Overview

Designed with unconventional appearance, the DP750 adopts the solution by building in the test line and probe which with shielding function, integrated the interface end in a compact box, no need to twist the cables, leads to a lower noise floor and has less interference.
Max 100MHz bandwidth, Max input voltage 750V, switchable 50X / 500X attenuation ratio, high accuracy, the noise floor is low and is not affected when switching ratio, support one-press Zero calibration, and overvoltage alarm. Support USB power supply, suitable for any brand of oscilloscope. Support direct power supply with Micsig UPI multi-function interface oscilloscope. Automatically matches the scope attenuation coefficient without manual adjustment.
The DP750 can be widely used in the research and development, commissioning or maintenance of switching power supplies, frequency converters, electronic ballasts, frequency conversion home appliances and other electrical power devices.

Specifications

Model name DP750 -100
Bandwidth 100MHz
Rise Time ≤ 3.5ns
Attenuation Ratio 50X; 500X
Gain Accuracy ±2%
Max Differential Test Voltage(DC+AC PK) 50X: 75V; 500X: 750V
Max Common Mode Input Voltage CAT II 600V
Input Referred Noise ≤ 12mVrms(50X); ≤ 120mVrms(500X)
Common Mode Rejection Ratio (CMRR) > 80dB(DC); > 80dB(100KHz); > 60dB(1MHz)
Input Resistance Differential input impedance: 8MΩ
ground: 4 MΩ 2.5pF
Output Voltage ≤ 1.5V
Power Rate 1W
Power Supply Micsig UPI Interface; USB (Adaptor)
Overrange Alarm Indicator flashes

Dimensions

| Power Supply Box: 803722 mm
Signal In/Output Box: 492512.5 mm
Input Cable Length| 150mm
Output Cable Length| 1000mm
Working Temperature| 0℃ -40℃
Working Humidity| 10%-85%

Panel Description

Differential probe power supply box

Micsig-DP750-100-High-Voltage-Differential-Probe-1

Differential probe signal box

Micsig-DP750-100-High-Voltage-Differential-Probe-2

Power AdapterMicsig-DP7Micsig-DP750-100-High-Voltage-Differential-
Probe-350-100-High-Voltage-Differential-Probe-3

Precautions

  1. Before starting the measurement, the zero point of the probe needs to be calibrated:
    First short-circuit the two input terminals, power on, press and hold the “Zero” button for 3s, the two indicators will start flashing at the same time, and the Zero point calibration is successful after 3s.

  2. When the differential probe output is connected to the oscilloscope, the bandwidth of the oscilloscope is required to be no less than 100MHz, and the input impedance of the channel is no less than 1MΩ.

  3. It is recommended to use it after warming up for 10 minutes to get more accurate data.

  4. If there is a functional failure of the product, please contact local dealer or Micsig directly.

Warranty

  1. The main body of this differential probe is guaranteed for 1 year. During the product warranty period, under normal use, the company will be responsible for free maintenance without dismantling and repairing the fault caused by the quality of the product itself.
  2. The warranty is invalid in the following cases, but maintenance services are provided, labor costs are waived, and only spare parts are charged:
    • Consumers damage any accessories due to improper use, maintenance and storage.
    • Damage caused by force majeure factors, such as natural disasters, etc.
  3. In the following cases, the company will refuse to provide maintenance services or provide maintenance services for a fee:
    • Unable to provide product packaging or anti-counterfeiting labels on product packaging.
    • The content of the anti-counterfeiting label has been altered, or is blurred and unrecognizable.
    • It has been dismantled by any person not authorized by Micsig (eg: changing wires, dismantling internal components, etc.).
    • No sales voucher or the content of the sales voucher does not match the product.

Safety

  1. Non-professionals should not open the product shell
  2. Do not use with the product shell open
  3. When measuring, do not touch any bare metal
  4. When overload occurs, please disconnect the power supply and input immediately
  5. Do not use in flammable and explosive environments

Shenzhen Micsig Technology Co., Ltd.
Phone: +86 755-88600880
Email: sales@micsig.com
Web: www.micsig.com
Add: 1F, Huafeng International Robot Industrial Park, Hangcheng Rd, Bao’an District, Shenzhen, Guangdong, China

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals