Micsig RCP Series Rogowski AC Current Probe User Guide
- May 15, 2024
- Micsig
Table of Contents
Micsig RCP Series Rogowski AC Current Probe
Specifications
- Model: RCP300-XS
- Bandwidth: 10Hz-30MHz
- Measurement range: 200mApk-300Apk
- Output sensitivity: 20mV/A50X
- Accuracy (typical): 1%
- Output noise: 18mVpp
- Output interface: BNC
- Peak di/dt: 20kA/s
- Droop Effect of conductor position: 9%/ms
- Offset voltage: Not specified
- Peak coil isolation voltage: Not specified
- Output impedance: Not specified
- Measurable conductor diameter: Not specified
- Power supply: Not specified
- Integrator size: Not specified
- Wire length (integrator to Rogowski coil): Not specified
- Coil inner diameter: Not specified
- Coil circumference: Not specified
- Coil cross-section diameter: Not specified
- Interface: Not specified
- Environment: Not specified
- Working temperature: Not specified
- Storage temperature: Not specified
- Working humidity: Not specified
- Storage humidity: Not specified
Product Usage Instructions
Operation Steps:
- Connect to the oscilloscope: Set the oscilloscope input impedance to 1M, connect the probe BNC end to the oscilloscope (ensure proper grounding).
- Power the probe: Use a standard adapter to power the probe. The indicator light turns green after power on.
- Connect the Device Under Test (DUT): Ensure the coil plug is inserted correctly and the wire or pin under test passes through the appropriate position of the coil.
- Power up the DUT.
- After test: Disconnect the circuit first, then unplug the coil.
- Disconnect probe power.
Safety Precautions:
- Use within safe voltage range.
- The equipment connected to the probe must be reliably grounded.
- Inspect the outer skin of the Rogowski coil before use. Stop using it if damaged.
- Ensure the circuit under test is turned off before connecting the probe.
- Use the adapter that comes standard with the probe.
FAQ
Warranty Information:
Micsig Warranty:
- Micsig warrants the main body of this current probe for 1 year.Free maintenance is provided for failures caused by product quality under normal use during this period.
Circumstances Excluded from Warranty:
- No packaging or anti-counterfeiting label.
- Altered or blurred anti-counterfeit label.
- Unauthorized disassembly.
- No valid sales voucher.
Contact Information:
If you have any questions or need further assistance, please contact Shenzhen
Micsig Technology Co., Ltd. at Tel: +86-755-88600880 or Email:
sales@micsig.com.
Overview
The RCP series Rogowski current probe measures AC currents up to 600Apk, max. bandwidth up to 30 MHz, delivers 1% typical accuracy, able to measure high- frequency, large current signals easily and accurately.
A 1.6mm thin, flexible, clip-around Rogowski coil allow user to conduct measurements without damaging the conductor and have no interference to the DUT.
Specifications
Model | RCP300-XS | RCP600-XS |
---|---|---|
Bandwidth | 10Hz-30MHz | 10Hz-30MHz |
Measurement range | 200mApk-300Apk | 200mApk-600Apk |
Output sensitivity | 20mV/A(50X) | 10mV/A(100X) |
Accuracy (typical) | 1% | 1% |
Output noise | < 18mVpp | < 12mVpp |
Output interface | BNC | |
Peak di/dt | 20kA/μs | 40kA/μs |
Droop | 9%/ms | 6%/ms |
Effect of conductor position | Within ±1% (deviation from center part) | |
Offset voltage | <±1mV | |
Peak coil isolation voltag | AC 2kVrms (1 min) (50Hz/60Hz) ( Rogowski coil part |
only )
Output impedance| High impedance
Measurable conductor
diameter
| ≤ 20mm
Power supply| DC 12V
Integrator size| 704017mm
Wire length (integrator to Rogowski coil)| 1.5m(customizable)
Coil inner diameter| 25mm(customizable)
Coil circumference| 80mm(customizable)
Coil cross-section diameter| Appx. 1.6mm
Interface| 1MΩ BNC
Environment
Working temperature| Base unit : 0℃ – 55℃ Coil : -20℃ – 125℃
Storage temperature| -30℃ -70℃
Working humidity| ≤ 85%RH
Storage humidity| ≤ 90%RH
Appearance
The RCP series current probe are composed of two parts: Integrator and Rogowski coil.
- Output: Standard BNC, compatible with all BNC oscilloscopes.
- Power supply: DC 12V, adapter
- Power indicator: Turn Green after powered on.
- Cable length: 1.5m,from integrator to coil, customizable.
- Rogowski coil inner diameter: 25mm, measures wires within 20mm in diameter.
- Rogowski coil cross-section diameter: 1.6mm
- Rogowski coil circumference: 80mm, customizable.
- Current direction: When the current flows in the marked direction, the output ispositive, otherwise it is negative.
Precautions
- to ensure accuracy, the wire being measured should be positionedas much as possible between X and Y in the right diagram, where X is the center of coil and Y is the midpoint of the coil circumference.
- to ensure accuracy, the wire should avoid the coil junction asmuch as possible during measurement (shadow area).
- try to stay away from strong magnetic field interference sources asmuch as possible to avoid measurement errors.
- the coil can be placed around the wire being measured to measurethe interference signals in the surrounding area, to determinewhether there is strong interference nearby.
Operation steps
- Connect to the oscilloscope: Set the oscilloscope input impedance to 1MΩ, connect the probe BNC end to oscilloscope (make sure the oscilloscope is properly grounded);
- Power the probe: Use standard adapter to power the probe. Indicator light turns green after power on;
- Connect the DUT: make sure that the coil plug is inserted in place and the wire or pin under test passes through the appropriate position of the coil.
- Power up the DUT.
- After test, disconnect the circuit first, then unplug the coil.
- Disconnect probe power.
Warranty
-
Micsig warrants the main body of this current probe for 1 year.
During the warranty period, Micsig will be responsible for free maintenance for any failure caused by the quality of the product under normal use. -
Under the following circumstances, Micsig will refuse to provide maintenance services or charge for a fee:
-
No packaging or anti-counterfeiting label.
-
Anti-counterfeit label has been altered or blurred beyond recognition.
-
Unauthorized disassembly, such as: changing wires, dismantling internal components, etc.
-
No sales voucher or the content of sales voucher does not match the product.
Safety Precautions
- Please use within safe voltage range.
- The equipment connected to the probe must be reliably grounded.
- The outer skin of the Rogowski coil should be inspected before use. If it is damaged,stop using it.
- Before connecting the probe to the circuit under test, make sure the circuit undertest is turned off.
- Please use the adapter that comes standard with the probe.