MEAN WELL GSM36E Series 36W AC-DC Reliable Green Medical Adaptor User Guide

September 26, 2024
MEAN WELL

MEAN WELL GSM36E Series 36W AC-DC Reliable Green Medical Adaptor

Specifications

Model DC Voltage Rated Current Rated Power (max.) Efficiency (Typ.)
GSM36E05-P1J 5V 4.5A 22.5W 80%
GSM36E07-P1J 7.5V 4.32A 32.4W 83%

Product Usage Instructions

Safety Precautions

  • Before using the AC-DC Reliable Green Medical Adaptor, read the user manual carefully and adhere to the safety instructions provided.

Powering On the Adaptor

  • Plug the adaptor into a compatible power source within the specified voltage range (80 ~ 264VAC) and frequency range (47 ~ 63Hz).

Connecting to Device

  • Use the appropriate DC plug as per the device’s requirements.
  • Ensure a secure connection between the adaptor and the device.

Operating Conditions

  • Ensure the working temperature is between -25°C to +60°C and the working humidity environment is suitable for operation.

Overload Protection

  • The adaptor is equipped with overload protection. In case of an overload, the protection mechanism will activate to prevent damage.

Frequently Asked Questions

Q: What should I do if the adaptor gets overheated during use?

A: If the adaptor becomes excessively hot, disconnect it from the power source immediately and allow it to cool down before using it again.

Q: Can I use this adaptor with devices that require different voltage outputs?

A: No, it is essential to match the DC voltage output of the adaptor with the requirements of your device to avoid damage.

Features

  • 2 pole EURO plug, Class Il power unit
  • Medical safety approved (2 x MOPP) according to EN60601-1/-1-11
  • Extremely low leakage current
  • No load power consumption<0.1W
  • Energy efficiency Level VI and meet CoC version 5 (except 5~9V for Level V)
  • -25~+60°C wide range working temperature
  • Protections: Short circuit / Overload / Over voltage
  • LED indicator for power on
  • Various DC plug quick adapter accessories are available (Plug kit sold sperately, please refer to: https://www.meanwell.com/upload/pdf/DC_plug.pdf)
  • 3 years warranty

Applications

  • Blood glucose meter
  • Blood pressure meter
  • Nebulizer
  • Inhaler
  • Portable medical device

GTIN CODE

Description

  • GSM36E is a highly reliable, 36W wall-mounted style single-output green medical adaptor series. This product is equipped with a 2-pin (no FG) standard European power plug, adopting the input range from 80 VAC to 264 VAC. The entire series supplies different output voltages between 5VDC and 48VDC that can satisfy the demands for various kinds of miniature medical devices. The circuitry design meets the international medical standards (2 x MOPP), having an ultra-low leakage current (<50A), and fitting the medical devices in direct electrical contact with the patients.
  • With an efficiency up to 88% and extremely low no-load power consumption below 0.1W, GSM36E is compliant with EU ErP and meets the Code of Conduct(CoC) Version 5; the supreme feature allows the adaptor to save energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock. GSM36E is approved with international medical safety certificates.

Model Encoding

SPECIFICATION

ORDER NO.| GSM36E05- P1J| GSM36E07- P1J| GSM36E09- P1J| GSM36E12- P1J| GSM36E15- P1J| GSM36E18- P1J| GSM36E24- P1J| GSM36E48- P1J
---|---|---|---|---|---|---|---|---
**OUTPUT| SAFETY MODEL NO.| GSM36E05| GSM36E07| GSM36E09| GSM36E12| GSM36E15| GSM36E18| GSM36E24| GSM36E48
DC VOLTAGE Note.2| 5V| 7.5V| 9V| 12V| 15V| 18V| 24V| 48V
RATED CURRENT| 4.5A| 4.32A| 4A| 3A| 2.4A| 2A| 1.5A| 0.75A
CURRENT RANGE| 0 ~ 4.5A| 0 ~ 4.32A| 0 ~ 4A| 0 ~ 3A| 0 ~ 2.4A| 0 ~ 2A| 0 ~ 1.5A| 0 ~ 0.75A
RATED POWER (max.)| 22.5W| 32.4W| 36W| 36W| 36W| 36W| 36W| 36W
RIPPLE & NOISE (max.) **Note.3| 80mVp-p| 80mVp-p| 80mVp-p| 120mVp-p| 120mVp-p| 150mVp-p| 180mVp-p| 240mVp-p
VOLTAGE TOLERANCE Note.4| ±6.0%| ±5.0%| ±5.0%| ±3.0%| ±3.0%| ±3.0%| ±2.0%| ±2.0%
LINE REGULATION Note.5| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
LOAD REGULATION| ±6.0%| ±5.0%| ±5.0%| ±3.0%| ±3.0%| ±3.0%| ±2.0%| ±2.0%
SETUP, RISE TIME Note.6| 500ms, 30ms / 230VAC           1000ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)| 16ms / 230VAC           10ms / 115VAC at full load
**INPUT| VOLTAGE RANGE Note.7| 80 ~ 264VAC  113 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 80%| 83%| 84%| 86%| 87%| 87%| 87%| 88%
AC CURRENT (Typ.)| 0.9A / 115VAC           0.45A / 230VAC
INRUSH CURRENT (Typ.)| 55A / 230VAC            30A / 115VAC
LEAKAGE CURRENT(max.)| Touch current < 50μA/264VAC
****
PROTECTION| OVERLOAD| 105 ~ 170% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| 5.25 ~ 7.5V| 7.88 ~ 10.5V| 9.45 ~ 13V| 12.6 ~ 17.2V| 15.75 ~ 20.25V| 18.9 ~ 25.2V| 25.2 ~ 32.4V| 50.4 ~ 64.8V
Protection type : Shut down o/p voltage, re-power on to recover
****
ENVIRONMENT| WORKING TEMP.| -25 ~ +60℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0~40℃)
VIBRATION**| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes






**SAFETY & EMC **(Note. 8)

| SAFETY STANDARDS| IEC 60601-1:2005+A1+A2; IEC 60601-1-11:2015+A1, TUV BS EN/ EN 60601-1:2006+A1+A12+A2; BS EN/ EN 60601-1-11:2015+A1, EAC TP TC 004 approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
**EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| EN55011 (CISPR11)| Class B
Radiated emission| EN55011 (CISPR11)| Class B
Harmonic current| EN61000-3-2| Class A
Voltage flicker| EN61000-3-3| —–
****
EMC IMMUNITY| EN60601-1-2, EN61204-3
Parameter| Standard| Test Level / Note**
ESD| EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )

Table 9, 9~28V/m( 385MHz~5.78GHz )

EFT bursts| EN61000-4-4| Level 3, 2KV
Surge susceptibility| EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| EN61000-4-6| Level 3, 10V
Magnetic field immunity| EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,

100% interruptions 250 periods

**OTHERS| MTBF| 3777.8K hrs min.       Telcordia SR-332 (Bellcore) ; 651.4K hrs min.       MIL-HDBK-217F (25℃)
DIMENSION| 795433mm (LWH)
PACKING| 230g ; 60pcs / 14.8Kg / CARTON
CONNECTOR| PLUG| See page 4~5 ; Other type available by customer requested
CABLE| See page 4~5 ; Other type available by customer requested
****
NOTE**| 1.  All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.

2.  DC voltage: The output voltage set at point measure by plug terminal & 50% load.

3.  Ripple & noise are measured at 20MHz by using a 12″ twisted pair terminated with a 0.1μF & 47μF capacitor.

4.  Tolerance: includes set up tolerance, line regulation, load regulation.

5.  Line regulation is measured from low line to high line at rated load.

6.  Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.

7.  Derating may be needed under low input voltage. Please check the derating curve for more details.

8.  The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”

(as available on https:// www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Derating Curve

Static Characteristics

Mechanical Specification

DC output plug

Standard plug: P1J

DC plug changeable through:

  1. Customization of the standard part with an optional DC plug according to the table (MOQ applicable)
  2. Quick adapter accessory (sold separately without MOQ) Please refer to the below table and online selection guide: https://www.meanwell.com/upload/pdf/DC_plug.pdf

Example quick adapter accessory:

Optional DC plug: (Available in customized cable or quick adapter)

Downloaded from Arrow.com.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

MEAN WELL User Manuals

Related Manuals