MEAN WELL GSM90A Series 90W AC-DC Reliable Green Medical Adaptor User Manual
- September 26, 2024
- MEAN WELL
Table of Contents
GSM90A Series 90W AC-DC Reliable Green Medical Adaptor
“`html
Specifications:
- Model: GSM90A series
- DC Voltage: 12V, 15V, 19V, 24V, 48V
- Rated Power: 90W
- Output Ripple & Noise: 120mVp-p
- Efficiency: 88% – 91%
- Safety Standards: IEC 60601-1:2005+A1+A2, TUV
BS EN/EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2
CAN/CSA C22.2 No. 60601-1:2014+A2,EAC TP TC 004 approved
Product Usage Instructions:
1. Safety Precautions:
Before using the Reliable Green Medical Adaptor, please read the
user manual carefully and adhere to the following safety
precautions:
-
Ensure the input voltage matches the device
specifications. -
Avoid water exposure to prevent electric shock.
-
Do not attempt to disassemble or modify the adaptor.
2. Connection:
Follow these steps to connect the adaptor to your device:
-
Identify the appropriate DC voltage required for your
device. -
Plug the adaptor into a compatible power outlet.
-
Connect the DC plug to your device securely.
3. Operation:
To operate the adaptor correctly:
-
Check for proper connection and ensure the LED indicator is
lit. -
If the LED indicator is off, check the power source and
connections.
4. Storage and Maintenance:
When not in use, store the adaptor in a cool, dry place away
from direct sunlight. Avoid bending or twisting the cables to
prevent damage.
FAQ:
Q: What should I do if the adaptor overheats during use?
A: If the adaptor overheats, disconnect it from the power source
immediately and allow it to cool down before reconnecting.
Q: Can I use this adaptor with devices that require different
voltage outputs?
A: No, it is essential to match the DC voltage output of the
adaptor with your device’s requirements to prevent damage.
“`
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
User’s Manual
Video
++
Features
https://www.meanwell.com/serviceGTIN.aspx
https://www.meanwell.com/upload/pdf/DC_plug.pdf
{
Downloaded from Arrow.com.
File Name:GSM90A-SPEC 2024-04-15
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
SPECIFICATION
ORDER NO.
GSM90A12-P1M
GSM90A15-P1M
GSM90A19-P1M
GSM90A24-P1M
GSM90A48-P1M
SAFETY MODEL NO.
GSM90A12
GSM90A15
GSM90A19
GSM90A24
GSM90A48
DC VOLTAGE
Note.2 12V
15V
19V
24V
48V
RATED CURRENT
6.67A
6A
4.74A
3.75A
1.87A
CURRENT RANGE
0 ~ 6.67A
0 ~ 6A
0 ~ 4.74A
0 ~ 3.75A
0 ~ 1.87A
RATED POWER (max.)
80W
90W
90W
90W
90W
OUTPUT
RIPPLE & NOISE (max.) Note.3 120mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
LINE REGULATION Note.5 ±1.0%
LOAD REGULATION
±5.0%
120mVp-p ±5.0% ±1.0% ±5.0%
120mVp-p ±4.0% ±1.0% ±4.0%
180mVp-p ±3.0% ±1.0% ±3.0%
200mVp-p ±2.5% ±1.0% ±2.5%
SETUP, RISE TIME Note.6 1000ms, 50ms / 230VAC 1000ms, 50ms / 115VAC at full load
HOLD UP TIME (Typ.)
40ms / 230VAC
VOLTAGE RANGE
Note.7 80 ~ 264VAC
25ms / 115VAC at full load
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.91 / 230VAC PF>0.95 / 115VAC at full load
INPUT
EFFICIENCY (Typ.) AC CURRENT (Typ.)
88% 1.3A / 115VAC
89% 0.6A / 230VAC
89%
90%
91%
INRUSH CURRENT (Typ.) Cold start 30A/115VAC 60A / 230VAC LEAKAGE CURRENT(max.) Earth leakage current < 125A/264VAC , Touch current < 100A/264VAC
OVERLOAD PROTECTION
OVER VOLTAGE
110 ~ 150% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed 105 ~ 135% rated output voltage Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE WORKING TEMP.
Shut down o/p voltage, re-power on to recover -30 ~ +70 (Refer to “Derating Curve”)
WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT VIBRATION
20% ~ 90% RH non-condensing -40 ~ +85, 10 ~ 95% RH non-condensing ±±00.0.033%%/ /((00~~4400) 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.8 3000 meters
SAFETY STANDARDS
IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2 CAN/CSA C22.2 No. 60601-1:2014+A2,EAC TP TC 004 approved
ISOLATION LEVEL
Primary-Secondary: 2xMOPP, Primary-Earth: 1xMOPP
WITHSTAND VOLTAGE ISOLATION RESISTANCE
EMC EMISSION
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:SHORT
I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH
Parameter
Standard
Conducted emission
BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22, CAN ICES-3(B)/NMB-3(B)
Radiated emission
BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22, CAN ICES-3(B)/NMB-3(B)
Harmonic current
BS EN/EN61000-3-2
Test Level / Note Class B
Class B Class A
SAFETY &
EMC (Note 9)
Voltage flicker
BS EN/EN61000-3-3
BS EN/EN60601-1-2, BS EN/EN61204-3
Parameter
Standard
—-Test Level / Note
EMC IMMUNITY
ESD
RF field susceptibility
EFT bursts Surge susceptibility Conducted susceptibility
BS EN/EN61000-4-2
BS EN/EN61000-4-3
BS EN/EN61000-4-4 BS EN/EN61000-4-5 BS EN/EN61000-4-6
Level 4, 15KV air ; Level 4, 8KV contact Level 3, 10V/m( 80MHz~2.7GHz ) Table
9, 9~28V/m( 385MHz~5.78GHz ) Level 3, 2KV
Level 3, 1KV/Line-Line , 2KV/Line-FG Level 3, 10V
Magnetic field immunity
BS EN/EN61000-4-8
Level 4, 30A/m
MTBF
Voltage dip, interruption
BS EN/EN61000-4-11
2538.4K hrs min. Telcordia SR-332 (Bellcore) ; 387.5K hrs min.
100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods
MIL-HDBK-217F (25)
OTHERS DIMENSION
1456032mm (LWH)
PACKING
0.45Kg; 30pcs/14.5Kg/0.9CUFT
PLUG CONNECTOR CABLE
See page 4~5 ; Other type available by customer requested See page 4~5 ; Other type available by customer requested
NOTE
1. All parameters are specified at 230VAC input, rated load, 25 70% RH
ambient. 2. DC voltage: The output voltage set at point measure by plug
terminal & 50% load. 3. Ripple & noise are measured at 20MHz by using a 12″
twisted pair terminated with a 0.1F & 47F capacitor. 4. Tolerance: includes
set up tolerance, line regulation, load regulation. 5. Line regulation is
measured from low line to high line at rated load. 6. Length of set up time is
measured at first cold start. Turning ON/OFF the power supply may lead to
increase of the set up time.
7. Derating may be needed under low input voltages. Pleas check the derating
curve for more details. 8. The ambient temperature derating of 3.5/1000m with
fanless models and of 5/1000m with fan models for operating altitude higher
than 2000m(6500ft). 9. The power supply is considered as an independent unit,
but the final equipment still need to re-confirm that the whole system
complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer
to “EMI testing of component power supplies.” (as available on
https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf ) Product Liability
DisclaimerFor detailed information, please refer to
https://www.meanwell.com/serviceDisclaimer.aspx
Downloaded from Arrow.com.
File Name:GSM90A-SPEC 2024-04-15
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
Derating Curve
Static Characteristics
100
80
60 50 40
20
-30 0
10 20 30 40
50 60
70 (HORIZONTAL)
AMBIENT TEMPERATURE ()
100 90 80 70 60 50 40
80 90 100 110 120 140 160 180 200 220 240 264
INPUT VOLTAGE (VAC) 60Hz
LOAD (%) 60
LOAD (%)
Mechanical Specification
32
145
AC Inlet IEC320-C14
DC output plug Standard plug: P1M
P1M 5.5
2.5
POWER LED 11±0.5mm
Case No. GS90A Unit:mm
Ul1185 14AWG 1000±50mm for 12 ~ 15V UL1185 16AWG 1200±50mm for 19 ~ 48V
70±10mm
Pin Assignment
C”+”
Outside
Inside
-V connected to AC FG(standard) -V not connected to AC FG(optional)
Downloaded from Arrow.com.
File Name:GSM90A-SPEC 2024-04-15
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
DC plug changeable through: (1) Customization of the standard part with an
optional DC plug according to the table (MOQ applicable) (2) Quick adapter
accessory (sold separately without MOQ) Please refer to below table and online
selection guide : https://www.meanwell.com/upload/pdf/DC_plug.pdf
Example quick adapter accessory:
Optional DC plug: (Available in customized cable or quick adapter)
Tuning Fork Style
Type No.
A B
Barrel Style
C (Straight) C
(Right-angled)
P1I P1L P1J P1JR P1IR P1LR P1MR
Type No.
C
P2I
P2J
P2L
A
(Straight)
P2M
B
C
P2IR
P2JR
(Right-angled)
P2LR P2MR
Lock Style
Type No.
A
Floating Locking
C
P2S(S761K)
P2K(761K)
B
P2C(S760K)
SWITCHCRAFT original or equivalent P2D(760K)
Min. Pin Style
Type No.
A
C
P3A
B EIAJ equivalent
P3B P3C
Center Pin Style
Type No.
A
C
P4A
B
P4B
D
EIAJ equivalent
P4C
A OD 5.5 5.5 5.5 5.5 5.5 5.5 5.5 A OD 5.5 5.5 5.5 5.5 5.5 5.5 5.5 5.5 A OD
5.53 5.53 5.53 5.53 A OD
2.35
4.0
4.75
A OD
5.5
6.5
7.4
B ID 2.1 2.5 2.1 2.1 2.1 2.5 2.5 B ID 2.1 2.1 2.5 2.5 2.1 2.1 2.5 2.5 B ID
2.03 2.54 2.03 2.54 B ID
0.7
1.7
1.7
BC ID L
3.4 11.0
4.4 11.0
5.1 11.0
C L 9.5 9.5 11.0 11.0 9.5 9.5 11.0 C L 9.5 11.0 9.5 11.0 9.5 11.0 9.5 11.0 C L 12.06 12.06 9.52
Quick Adapter Accessory Available
(Current rating: 7.5A max.)
None
None
9.52 C L
11.0
None
11.0
11.0
D
Center Pin
1.0
None
1.4
0.6
Downloaded from Arrow.com.
File Name:GSM90A-SPEC 2024-04-15
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
Min. DIN 3 Pin with Lock (male)
1 3
2
KYCON KPPX-3P equivalent
Min. DIN 4 Pin with Lock (male)
23 14
KYCON KPPX-4P equivalent
Min. DIN 4 Pin with Lock (female)
23 14
KYCON KPJX-CM-4S equivalent
DIN 5 Pin (male)
13 425
Stripped and tinned leads
Type No. R6B
Type No. R7B
Type No. R7BF
Type No. R1B
Type No.
Pin Assignment
PIN No.
Output
1
+Vo
2
-Vo
3
+Vo
Pin Assignment
PIN No.
Output
1
+Vo
2
-Vo
3
-Vo
4
+Vo
Pin Assignment
PIN No.
Output
1
+Vo
2
-Vo
3
-Vo
4
+Vo
Pin Assignment
PIN No.
Output
1
-Vo
2
-Vo
3
+Vo
4
-Vo
5
+Vo
Pin Assignment
PIN No.
Output
L (red)
1 2
1
+Vo
L1 (black) by customer
Length of Land L1 by request
(MW’s standard length, L: 25 mm, L1: 5 mm)
( NOTE: The wire color is for reference only, please refer to the actual
product)
2
-Vo
Quick Adapter Accessory None Available
(Current rating: 7.5A max.)
None
None
None
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
Downloaded from Arrow.com.
File Name:GSM90A-SPEC 2024-04-15
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>