MeanWell RPT-75 Series 72W Reliable Triple Output Medical Grade Owner’s Manual

July 13, 2024
MeanWell

MeanWell RPT-75 Series 72W Reliable Triple Output Medical Grade Owner’s Manual
MeanWell RPT-75 Series 72W Reliable Triple Output Medical
Grade

Features

  • 5″×3″ compact size
  • Medical safety approved (2 x MOPP) according to ANSI/A A MI ES 60601-1 and IEC/BS EN/EN 60601-1
  • Suitable for BF application with appropriate system consideration
  • 72W convection, 100W force air
  • EMI class B for class I configuration
  • Extremely low leakage current
  • Protections: Short circuit / Overload / Over voltage
  • Lifetime > 140K hours
  • 3 years warranty

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors Sleep apnea devices

GTIN CODE

MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description

RPT-75 is a 72W highly reliable PCB type medical power supply with a high power density on the 5″ by 3″ footprint. It accepts 90~264VAC input and offers triple output voltages. RPT-75 is able to be used for Class I system design. The extremely low leakage current is less than 150μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011.

Model Encoding

Model Encoding

SPECIFICATION

MODEL RPT- 75A RPT- 75B RPT- 75C
**OUTPUT** OUTPUT NUMBER CH1
CH2 CH3 CH1 CH2
DC VOLTAGE 5V 12V -5V
RATED CURRENT 6A 3A 0.5A
CURRENT RANGE 0.6 ~ 8A 0.2 ~ 4A 0.1 ~ 1A
~ 1A 0.6 ~ 8A 0.1 ~ 3A 0.1 ~ 1A
RATED POWER 68.5W 72W 72W
PEAK LOAD (23.5CFM) 93W 100W 100W
RIPPLE & NOISE (max.) Note.2 80mVp-p 120mVp-p
80mVp-p 80mVp-p 120mVp-p 80mVp-p
VOLTAGE ADJ. RANGE CH1:4.75 ~ 5.5V
VOLTAGE TOLERANCE Note.3 ±2.0% ±6.0% ±5.0%
±5.0% ±2.0% ±8.0% ±5.0%
LINE REGULATION ±0.5% ±1.0% ±1.0%
±1.0% ±1.0%
LOAD REGULATION ±1.5% ±3.0% ±1.0%
±3.0% ±1.0%
SETUP, RISE TIME 500ms, 30ms/230VAC       500ms, 30ms/115VAC at

full load
HOLD UP TIME (Typ.)| 90ms/230VAC        20ms/115VAC at full load
**INPUT| VOLTAGE RANGE| 90 ~ 264VAC     127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY(Typ.)| 76%| 77%| 77%
AC CURRENT (Typ.)| 1.5A/115VAC       1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC      50A/230VAC
LEAKAGE CURRENT ****Note.4| Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC
****
PROTECTION| OVERLOAD| 140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| Ch1: 5.7 ~ 6.8V
Protection type : Shut down o/p voltage, re-power on to recover
****
ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 45℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.5| 3000 meters
SAFETY & EMC (Note 8)| SAFETY STANDARDS| IEC60601-1, UL ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved, EAC TP TC 004,TUV BS EN/EN60601-1 approved
ISOLATION LEVEL| Primary-Secondary:2xMOPP,  Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC  I/P-FG:2KVAC  O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
****
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
****
EMC IMMUNITY| BS EN/EN55035,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,100% interruptions 250 periods
****
OTHERS| MTBF| 2305.8K hrs min.    Telcordia SR-332 (Bellcore) ; 521.3K hrs min.  MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 12776.231mm or 5″ 3″ 1.22″ inch
PACKING| 0.25Kg; 63pcs/17.3Kg/1.28CUFT
NOTE**|

  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. Ripple & noise are measured at 20 MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
  3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4.  Touch current was measured from primary input to DC output.
  5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft)
  6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
  7. Heat Sink HS1,HS2,HS3 can not be shorted.
  8. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

MODEL RPT- 75D RPT- 7503
**OUTPUT** OUTPUT NUMBER CH1
CH2 CH3 CH1
DC VOLTAGE 5V 24V
RATED CURRENT 5A 1.5A
CURRENT RANGE 0.6 ~ 7A 0.1 ~ 2A

1.5A
RATED POWER| 73W| 61.8W
PEAK LOAD (23.5CFM)| 95W| 81.1W
RIPPLE & NOISE (max.) Note.2| 80mVp-p| 200mVp-p| 120mVp-p| 80mVp-p| 120mVp-p| 120mVp-p
VOLTAGE ADJ. RANGE| CH1:4.75 ~ 5.5V| ——
VOLTAGE TOLERANCE Note.3| ±2.0%| ±8.0%| ±8.0%| ±4.0%| ±6.0%| +10,-6%
LINE REGULATION| ±0.5%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.5%
LOAD REGULATION| ±1.5%| ±3.0%| ±3.0%| +3,-4%| +5,-4%| ±6.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC       500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)| 90ms/230VAC        20ms/115VAC at full load
**INPUT| VOLTAGE RANGE| 90 ~ 264VAC     127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY(Typ.)| 79%| 74%
AC CURRENT (Typ.)| 1.5A/115VAC       1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC      50A/230VAC
LEAKAGE CURRENT ****Note.4| Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC
****
PROTECTION| OVERLOAD| 140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| Ch1: 5.7 ~ 6.8V| Ch1: 3.8 ~ 4.5V
Protection type : Shut down o/p voltage, re-power on to recover
****
ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 45℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.5| 3000 meters
****
SAFETY & EMC (Note 8)| SAFETY STANDARDS| IEC60601-1, UL ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved, EAC TP TC 004,TUV BS EN/EN60601-1 approved
ISOLATION LEVEL| Primary-Secondary:2xMOPP,  Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC  I/P-FG:2KVAC  O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
****
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
****
EMC IMMUNITY| BS EN/EN55035,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,100% interruptions 250 periods
****
OTHERS| MTBF| 2305.8K hrs min.    Telcordia SR-332 (Bellcore) ; 521.3K hrs min.  MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 12776.231mm or 5″ 3″ 1.22″ inch
PACKING| 0.25Kg; 63pcs/17.3Kg/1.28CUFT
****
NOTE**|

  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
  3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. Touch current was measured from primary input to DC output.
  5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
  7. Heat Sink HS1,HS2,HS3 can not be shorted.
  8. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Block Diagram
Block Diagram
Derating Curve
Debating Curve
AMBIENT TEMPERATURE
(C)
Output Derating VS Input Voltage

Output Input Voltage
INPUT VOLTAGE (VAC) 60Hz

Mechanical Specification

Top View
Mechanical Overview
Side View
AC Input Connector (CN1) : JSTB3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N JST VHRor equivalent JST SVH-21T-P1.1or equivalent
2 No Pin
3 AC/L

DC Output Connector (CN2) : JST B8P-VH orequivalent

Pin No. Assignment Mating Housing Terminal
1,2 V1 JST VHRor equivalent JST SVH-21T-P1.1or equivalent
3,4,5

COM

6,7| V2
8| V3

  :Grounding Required

  1. H S 1, 1 . H S 1 , H S 2 , H S 3 H S 2, H S 3 cannot be shorted.
  2. M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1,M2 and chassis grounding.

Installation Manual

Please referto : http://www.meanwell.com/manual.htm|
Symbols Safety
ANSI/AAMI ES60601-1 BS EN/ENB0601-1 IEC60801-1 TPTCO04,
Blood Icons
QR Code

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals