MeanWell RPT-75 Series 72W Reliable Triple Output Medical Grade Owner’s Manual
- July 13, 2024
- MeanWell
Table of Contents
MeanWell RPT-75 Series 72W Reliable Triple Output Medical Grade Owner’s
Manual
Features
- 5″×3″ compact size
- Medical safety approved (2 x MOPP) according to ANSI/A A MI ES 60601-1 and IEC/BS EN/EN 60601-1
- Suitable for BF application with appropriate system consideration
- 72W convection, 100W force air
- EMI class B for class I configuration
- Extremely low leakage current
- Protections: Short circuit / Overload / Over voltage
- Lifetime > 140K hours
- 3 years warranty
Applications
- Oral irrigator
- Hemodialysis machine
- Medical computer monitors Sleep apnea devices
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Description
RPT-75 is a 72W highly reliable PCB type medical power supply with a high power density on the 5″ by 3″ footprint. It accepts 90~264VAC input and offers triple output voltages. RPT-75 is able to be used for Class I system design. The extremely low leakage current is less than 150μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011.
Model Encoding
SPECIFICATION
MODEL | RPT- 75A | RPT- 75B | RPT- 75C |
---|---|---|---|
**OUTPUT** | OUTPUT NUMBER | CH1 | |
CH2 | CH3 | CH1 | CH2 |
DC VOLTAGE | 5V | 12V | -5V |
RATED CURRENT | 6A | 3A | 0.5A |
CURRENT RANGE | 0.6 ~ 8A | 0.2 ~ 4A | 0.1 ~ 1A |
~ 1A | 0.6 ~ 8A | 0.1 ~ 3A | 0.1 ~ 1A |
RATED POWER | 68.5W | 72W | 72W |
PEAK LOAD (23.5CFM) | 93W | 100W | 100W |
RIPPLE & NOISE (max.) Note.2 | 80mVp-p | 120mVp-p | |
80mVp-p | 80mVp-p | 120mVp-p | 80mVp-p |
VOLTAGE ADJ. RANGE | CH1:4.75 ~ 5.5V | ||
VOLTAGE TOLERANCE Note.3 | ±2.0% | ±6.0% | ±5.0% |
±5.0% | ±2.0% | ±8.0% | ±5.0% |
LINE REGULATION | ±0.5% | ±1.0% | ±1.0% |
±1.0% | ±1.0% | ||
LOAD REGULATION | ±1.5% | ±3.0% | ±1.0% |
±3.0% | ±1.0% | ||
SETUP, RISE TIME | 500ms, 30ms/230VAC 500ms, 30ms/115VAC at |
full load
HOLD UP TIME (Typ.)| 90ms/230VAC 20ms/115VAC at full
load
**INPUT| VOLTAGE RANGE| 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY(Typ.)| 76%| 77%| 77%
AC CURRENT (Typ.)| 1.5A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC 50A/230VAC
LEAKAGE CURRENT ****Note.4| Earth leakage current < 150μA/264VAC ,
Touch current < 100μA/264VAC
****PROTECTION| OVERLOAD| 140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| Ch1: 5.7 ~ 6.8V
Protection type : Shut down o/p voltage, re-power on to recover
****ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to
“Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 45℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.5| 3000 meters
SAFETY & EMC (Note 8)| SAFETY STANDARDS|
IEC60601-1, UL ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1:14 – Edition 3
approved, EAC TP TC 004,TUV BS EN/EN60601-1 approved
ISOLATION LEVEL| Primary-Secondary:2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC /
25℃/ 70% RH
****EMC EMISSION| Parameter| Standard| Test
Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
****EMC IMMUNITY| BS EN/EN55035,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz
)Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods,100% interruptions 250 periods
****OTHERS| MTBF| 2305.8K hrs min. Telcordia SR-332 (Bellcore) ;
521.3K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 12776.231mm or 5″ 3″ 1.22″ inch
PACKING| 0.25Kg; 63pcs/17.3Kg/1.28CUFT
NOTE**|
- All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
- Ripple & noise are measured at 20 MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
- Tolerance : includes set up tolerance, line regulation and load regulation.
- Touch current was measured from primary input to DC output.
- The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft)
- Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
- Heat Sink HS1,HS2,HS3 can not be shorted.
- The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
MODEL | RPT- 75D | RPT- 7503 |
---|---|---|
**OUTPUT** | OUTPUT NUMBER | CH1 |
CH2 | CH3 | CH1 |
DC VOLTAGE | 5V | 24V |
RATED CURRENT | 5A | 1.5A |
CURRENT RANGE | 0.6 ~ 7A | 0.1 ~ 2A |
1.5A
RATED POWER| 73W| 61.8W
PEAK LOAD (23.5CFM)| 95W| 81.1W
RIPPLE & NOISE (max.) Note.2| 80mVp-p| 200mVp-p|
120mVp-p| 80mVp-p| 120mVp-p| 120mVp-p
VOLTAGE ADJ. RANGE| CH1:4.75 ~ 5.5V| ——
VOLTAGE TOLERANCE Note.3| ±2.0%| ±8.0%| ±8.0%| ±4.0%| ±6.0%|
+10,-6%
LINE REGULATION| ±0.5%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.5%
LOAD REGULATION| ±1.5%| ±3.0%| ±3.0%| +3,-4%| +5,-4%| ±6.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC 500ms, 30ms/115VAC at
full load
HOLD UP TIME (Typ.)| 90ms/230VAC 20ms/115VAC at full
load
**INPUT| VOLTAGE RANGE| 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY(Typ.)| 79%| 74%
AC CURRENT (Typ.)| 1.5A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC 50A/230VAC
LEAKAGE CURRENT ****Note.4| Earth leakage current < 150μA/264VAC ,
Touch current < 100μA/264VAC
****PROTECTION| OVERLOAD| 140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| Ch1: 5.7 ~ 6.8V| Ch1: 3.8 ~ 4.5V
Protection type : Shut down o/p voltage, re-power on to recover
****ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to
“Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 45℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.5| 3000 meters
****SAFETY & EMC (Note
8)| SAFETY STANDARDS| IEC60601-1, UL ANSI/AAMI ES60601-1,
CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved, EAC TP TC 004,TUV BS
EN/EN60601-1 approved
ISOLATION LEVEL| Primary-Secondary:2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC /
25℃/ 70% RH
****EMC EMISSION| Parameter| Standard| Test
Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
****EMC IMMUNITY| BS EN/EN55035,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz
)Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods,100% interruptions 250 periods
****OTHERS| MTBF| 2305.8K hrs min. Telcordia SR-332 (Bellcore) ;
521.3K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 12776.231mm or 5″ 3″ 1.22″ inch
PACKING| 0.25Kg; 63pcs/17.3Kg/1.28CUFT
****NOTE**|
- All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
- Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
- Tolerance : includes set up tolerance, line regulation and load regulation.
- Touch current was measured from primary input to DC output.
- The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
- Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
- Heat Sink HS1,HS2,HS3 can not be shorted.
- The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Block Diagram
Derating Curve
AMBIENT TEMPERATURE (C)
Output Derating VS Input Voltage
INPUT VOLTAGE (VAC) 60Hz
Mechanical Specification
Top View
Side View
AC Input Connector (CN1) : JSTB3P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/N | JST VHRor equivalent | JST SVH-21T-P1.1or equivalent |
2 | No Pin | ||
3 | AC/L |
DC Output Connector (CN2) : JST B8P-VH orequivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1,2 | V1 | JST VHRor equivalent | JST SVH-21T-P1.1or equivalent |
3,4,5 |
COM
6,7| V2
8| V3
:Grounding Required
- H S 1, 1 . H S 1 , H S 2 , H S 3 H S 2, H S 3 cannot be shorted.
- M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1,M2 and chassis grounding.
Installation Manual
Please referto :
http://www.meanwell.com/manual.htm|
ANSI/AAMI ES60601-1 BS EN/ENB0601-1 IEC60801-1 TPTCO04,