MEANWELL RPS-60-12 60W Reliable Green Medical Power Supply Instruction Manual
- June 13, 2024
- MeanWell
Table of Contents
60W Reliable Green Medical Power Supply
R P S – 6 0series
RPS-60-12 60W Reliable Green Medical Power Supply
https://www.meanwell.com/Upload/PDF/Medical_EN.pdf
Features
- 4″%x2″ compact size
- IT & Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1, IEC/BS EN/EN60601-1 and IEC/BS EN/EN/UL 62368-1
- Suitable for BF application with appropriate system consideration
- Cooling by free air convection
- EMI class B forclass I configuration
- No load power consumption<0.75W
- Protections: Short circuit / Overload / Over voltage
- Operating altitude up to 3000 meters
- 3 years warranty
Applications
- Oralirrigator
- Hemodialysis machine
- Medical computer monitors
- Sleep apnea devices
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Description
RPS-60is a 60W highly reliable green PCB type medical power supply with a high power density on the 4” by 2” footprint. It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 86% and the extremely low no load power consumption is down below 0.75W. RPS-60 is able to be used for Class I (with FG) system design. The extremely low leakage current isless than 130xA. In addition, it conforms to international IT and medical regulations (2*MOPP) and EMC BS EN/EN55022/BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.
Model Encoding
SPECIFICATION
MODEL | RPS-60-3.3 | RPS-60-S | RPS-60-12 | RPS-60-15 | RPS-66-24 | RPS-60-48 |
---|---|---|---|---|---|---|
OUTPUT | DC VOLTAGE | 3.3V | 5V | 12V | 15V | 24V |
RATED CURRENT | 10A | 10A | 5A | 4A | 2.5A | 1.25A |
CURRENT RANGE | 0 ~ 11A | 0 ~ 11A | 0 ~5.5A | 0 ~ 4.4A | 0 ~ 2.75A | 0~ 1.375A |
RATED POWER | 33W | 50W | 60W | 60W | 60W | 60W |
PEAK LOAD(10sec.) Note.2 | 36.3W | 55W | 66W | 66W | 66W | 66W |
RIPPLE & NOISE (max.) Note.3 | 60mVp-p | 60mVp-p | 60mVp-p | lOOmVp-p | lOOmVp-p |
lOOmVp-p
VOLTAGE ADJ. RANGE| 3.1~ 3.6V| 4.75 ~ 5.5V| 11.4 ~13.2V| 13.5 ~ 16.5V| 22.8~
27.6V| 45.6 ~ 52.8V
VOLTAGE TOLERANCE NoteA| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)| 60ms/230VAC 12ms/115VAC at ful load
INPUT| VOLTAGE RANGE| 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 74%| I79%| 184%| 185%| 187%| 186%
AC CURRENT (Typ.)| 1.8A/115VAC 1 N230VAC
INRUSH CURRENT (Typ.)| COLD START 60A/230VAC 30A1115VAC
LEAKAGE CURRENT(max.)Note.5| Earth leakage current < 1300/264VAC , Touch
current <1001/N264VAC
PROTECTION| OVER LOAD| 115 -150% rated output power
Protection type : hiccup mode, recovers automaticaly after fault condition is
removed
OVER VOLTAGE| 3.8 ~ 5V| I 5.7 ~6.8V| 1 13.8~16.2V| 1 17.2 ~ 20.3V| 1 28.4 ~
32.4V| 1 55.2 ~ 64.8V
Protection type : Shutdown oft) voltage, re-power on to recover
EIMROINENI| WORKING TEMP.| -20 ~+70°C (Refer to “Derating Curve)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~+85°C , 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ± 0.03%f C (0 ~ 50° 0)
VIBRATION| 10 ~ 500Hz, 2G 10mini1cycle, period for 60min. each along X, Y, Z
axes
OPERATING ALTITUDE Motel.| 3000 meters
SAFETY &
EMC (Note8)| SAFETY STANDARDS| I EC,62368-1, U L62368-1, TUV BS EN/EN62368-1,
IEC60601-1, TUV BS EN/EN60601-1, UL ANSI /AAMI ES60601-1 (3.1 version),
CAN/CSA-C22.2 No. 60601-1:14 – Edition 3, EAC TP TC 004 approved; Design refer
to BS EN/EN60335-1
ISOLATION LEVEL| Primary-Secondary: 2xMOPP, Prinary-Earth:lxMOPP, Secondary-
Earth:lxMOPP
WITHSTAND VOLTAGE| I/P-0/P:4KVAC I/P-FG2KVAC 0/P-FG;1.5KVAC
ISOLATION RESISTANCE| I/P-0/P, I/P-FG, 0/P-FG:100M Ohms /500VDC/25°C/70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011(CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EWE N61000-3-2| Class A
Voltage flicker| BS EWE N61000-3-3| —-
EMC IMMUNRY| BS EN/E1455035 . BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EWE N61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EWE N61000-4-3| Level 3, 10V/m(80MEtz-2.7GHz )
Table 9, 9-28V1m( 385MHz-5.78GEtz )
EFT bursts| BS EWEN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EWE N61000-4-5| Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility| BS EWE N61000.4-6| Level 3,10V
Magnetic field immunity| BS EWE N61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EWE N61000-4-11| 100% dip 1
pLOods.30%dip25periods, 1CO% int nwices 250 periods
OTHERS| MTBF| 5153.0K hrs min. Telcordia SR-332 (Bellcore) ; 353.6K hrs min.
MIL-HDBK-217F (25t )
DIMENSION (1.11rH)| 101.6’50.8’29mm or 4″ 2″ 1.141″ inch
PACKING| 0.15Kg;96pcs/15.4Kg/0.89CUFT
NOTE| 1.All parameters NOT speda h/ mentioned are measured at 230VAC input,
rated load and 25°C of ambient temperature.
2.33% Duty cyde maximum within every 30 seconds. Average output power should
not exceed the rated power.
3.Ripple & noise are measured at 20MHz of bandwidth by using a lr twisted
pair-wire terminated with a 0.14 & 47pf parallel capacibr.
4.Tolerance : includes set up tolerance, line regulation and load regulation.
5.Touch current was measured from primary input to DC Output.
6.The ambient temperature derating of 3.5.01000m with fanless models and of 5t
/1000m with fan models for operating altitude higher than 200006500ff)
7.Heat Sink HS1,HS2 can not be shorted.
8.The power supply is considered a component which will be installed into a
final equipment. All the EMC tests are been executed by mounting the unit on a
360mm’360mm metal plate with lmm of thickness. The final equipment must be re-
confirmed that it still meets EMC directives. For guidance on how to perform
these EMC tests, please refer to -EMI testing of component power supplies.”(as
available on http://www.meanwell.com)
※ Product Liability Disdaimer : For detailed information, please refer to
https://www.meanwell.00miserviceDisclaimeraspx
Block Diagram
Output Derating
Output Derating VS Input Voltage
Mechanical Specification
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/N | JST VHR or equivalent | JST SVH-21T-P1.1 or equivalent |
2 | No Pin | ||
3 | AC/L |
DC Output Connector (CN2) : JST B4P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1,2 | +V | JST VHR | |
or equivalent | JST SVH-21T-P1.1 |
or equivalent
3,4| -V
: Grounding Required
1.HS1,HS2 cannot be shorted.
2.M1 is safety ground. For better EMC performance,
Please secure an electrical connection between
M1,M2 and chassis grounding.
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
File Name:RPS-60-SPEC 2022-09-20