Thermo Scientific Multidrop 384 Liquid Dispensing User Manual
- June 3, 2024
- Thermo Scientific
Table of Contents
Multidrop 384 Liquid Dispensing
Thermo Scientific Multidrop® 384
User Manual
Rev. 3.4
Thermo Scientific Multidrop® 384
User Manual
Rev. 3.4, Cat. no. 1507010
Thermo Scientific Multidrop 384, Cat. no. 5840150 or 5840157 User Manual Rev.
3.4; October 2008, Cat. no. 1507010
CONTENTS
1 SAFETY SYMBOLS AND MARKINGS………………………………………………………………………………. 6 2 ABOUT THE
USER MANUAL ………………………………………………………………………………………….. 7 3 INTRODUCTION TO THE
MULTIDROP 384………………………………………………………………………. 8 4 FUNCTIONAL DESCRIPTION
………………………………………………………………………………………… 10 5
INSTALLATION…………………………………………………………………………………………………………….. 15 6 ROUTINE
OPERATION………………………………………………………………………………………………….. 21 7
MAINTENANCE…………………………………………………………………………………………………………….. 36 8 TECHNICAL
SPECIFICATIONS ……………………………………………………………………………………… 60 9
TROUBLESHOOTING……………………………………………………………………………………………………. 66 10 WARRANTY
CERTIFICATE …………………………………………………………………………………………… 71 11 ORDERING INFORMATION
…………………………………………………………………………………………… 72 12 GLOSSARY AND ABBREVIATIONS
………………………………………………………………………………. 75 13
INDEX…………………………………………………………………………………………………………………………… 77 14 TABLE OF FIGURES
…………………………………………………………………………………………………….. 81 APPENDIX 1. THERMO SCIENTIFIC
MULTIDROP 384 BRIEF USER’S GUIDE………………………… 83 APPENDIX 2. THERMO SCIENTIFIC
MULTIDROP 384 FEEDBACK FORM……………………………… 84 APPENDIX 3.
ADDRESSES………………………………………………………………………………………………….. 85
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
2
TABLE OF CONTENTS
1 SAFETY SYMBOLS AND MARKINGS………………………………………………………………………………. 6
2 ABOUT THE USER MANUAL ………………………………………………………………………………………….. 7
3 INTRODUCTION TO THE MULTIDROP 384………………………………………………………………………. 8
3.1 INTENDED USE ……………………………………………………………………………………………………………. 8 3.2 PRINCIPLE OF
OPERATION……………………………………………………………………………………………… 8 3.3 ADVANTAGES OF USING THE
MULTIDROP 384 ……………………………………………………………………. 9
4 FUNCTIONAL DESCRIPTION ………………………………………………………………………………………… 10
4.1 INSTRUMENT LAYOUT………………………………………………………………………………………………….. 10 4.1.1 Front
view ………………………………………………………………………………………………………… 10 4.1.2 Back
view…………………………………………………………………………………………………………. 13
4.2 MULTIDROP 384 MICROPLATE DISPENSER……………………………………………………………………….. 14
5 INSTALLATION…………………………………………………………………………………………………………….. 15
5.1 INSTALLATION CHECK LIST……………………………………………………………………………………………. 15 5.2 WHAT TO
DO UPON DELIVERY……………………………………………………………………………………….. 15
5.2.1 How to unpack ………………………………………………………………………………………………….. 15 5.2.2 Checking
delivery for completeness…………………………………………………………………….. 15 5.2.3 Checking for
damage during transport …………………………………………………………………. 16 5.2.4 Environmental
requirements……………………………………………………………………………….. 16 5.2.5 Things to avoid
…………………………………………………………………………………………………. 16 5.2.6 Technical
prerequisites………………………………………………………………………………………. 16 5.3 SETUPS BEFORE YOU PUT
THE INSTRUMENT INTO OPERATION ……………………………………………… 17 5.3.1 Installation
procedure ………………………………………………………………………………………… 17 5.3.2 How to ensure startup
……………………………………………………………………………………….. 18 5.4 OPERATIONAL CHECK
…………………………………………………………………………………………………. 19 5.5 OPERATING PRECAUTIONS AND
LIMITATIONS BEFORE OPERATION ………………………………………… 20
6 ROUTINE OPERATION………………………………………………………………………………………………….. 21
6.1 OPERATIONAL CHECK LIST …………………………………………………………………………………………… 21 6.2
SWITCHING ON ………………………………………………………………………………………………………….. 21 6.3 HOW TO INSERT
THE PRIMING VESSEL AND THE DISPENSING CASSETTE …………………………………. 22 6.4 HOW TO
INSERT AND ADJUST THE PLATE ADAPTER ……………………………………………………………. 25 6.5 HOW TO USE
THE CONTROL PANEL ………………………………………………………………………………… 30
6.5.1 Control panel indicators……………………………………………………………………………………… 30 6.5.2
Control panel thumbwheels (digital switches) ……………………………………………………….. 30 6.5.3
Control panel keys…………………………………………………………………………………………….. 31 6.6 HOW TO SELECT
THE 96 / 384-WELL PLATE……………………………………………………………………… 31 6.7 HOW TO
PRIME………………………………………………………………………………………………………….. 32 6.8
DISPENSING……………………………………………………………………………………………………………… 32 6.8.1 How to dispense
automatically ……………………………………………………………………………. 32 6.8.2 How to dispense manually
into the specified column(s) …………………………………………. 33 6.8.3 How to dispense
multiple volumes ………………………………………………………………………. 33
6.8.3.1 Dispensing tenfold and hundredfold volumes…………………………………………………………………..
33
6.8.4 After use ………………………………………………………………………………………………………….. 34 6.9
SHUTDOWN………………………………………………………………………………………………………………. 35 6.10 GUIDELINES FOR
USE………………………………………………………………………………………………. 35
7 MAINTENANCE…………………………………………………………………………………………………………….. 36
7.1 REGULAR AND PREVENTIVE MAINTENANCE………………………………………………………………………. 36 7.1.1
General ……………………………………………………………………………………………………………. 36 7.1.2 Immediate
………………………………………………………………………………………………………… 36
7.2 HOW TO WASH THE DISPENSING CASSETTE ……………………………………………………………………… 37 7.3 HOW
TO AUTOCLAVE THE DISPENSING CASSETTE ……………………………………………………………… 37
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
3
7.4 HOW TO REPLACE THE FUSES ………………………………………………………………………………………. 38 7.5 HOW TO
RECALIBRATE THE DISPENSING CASSETTE……………………………………………………………. 40
7.5.1 Volumetric calibration ………………………………………………………………………………………… 40 7.5.2
Gravimetric calibration……………………………………………………………………………………….. 42 7.6 HOW TO
CHANGE THE TUBING SET…………………………………………………………………………………. 45 7.6.1 Changing the
tip band ……………………………………………………………………………………….. 45 7.6.2 Fitting the new tubing
set step by step …………………………………………………………………. 46 7.7 DISPOSAL OF MATERIALS
…………………………………………………………………………………………….. 57 7.8 DECONTAMINATION PROCEDURE
…………………………………………………………………………………… 57 7.9 HOW TO PACK FOR
SERVICE…………………………………………………………………………………………. 58 7.10 SERVICE
CONTRACTS………………………………………………………………………………………………. 58 7.11 DISPOSAL OF THE
INSTRUMENT …………………………………………………………………………………. 58
8 TECHNICAL SPECIFICATIONS ……………………………………………………………………………………… 60
8.1 GENERAL SPECIFICATIONS …………………………………………………………………………………………… 60 8.2 SAFETY
SPECIFICATIONS……………………………………………………………………………………………… 61
8.2.1 Live parts …………………………………………………………………………………………………………. 61 8.3 IN
CONFORMITY WITH THE REQUIREMENTS ………………………………………………………………………. 62 8.4 REMOTE
CONTROL TO MULTIDROP 384 ………………………………………………………………………….. 63
8.4.1 Hardware for the computer port ………………………………………………………………………….. 63 8.4.2
Multidrop 384 PC interface …………………………………………………………………………………. 63
9 TROUBLESHOOTING……………………………………………………………………………………………………. 66
9.1 TROUBLESHOOTING GUIDE…………………………………………………………………………………………… 66 9.2 FAQS ABOUT
THE MULTIDROP 384 ……………………………………………………………………………….. 67 9.3
HAZARDS…………………………………………………………………………………………………………………. 68
9.3.1 Electrical ………………………………………………………………………………………………………….. 68 9.3.2
Mechanical……………………………………………………………………………………………………….. 68 9.3.3
Environmental…………………………………………………………………………………………………… 68 9.3.4 Defects and
abnormal stresses…………………………………………………………………………… 69 9.4 SERVICE REQUEST PROTOCOL
……………………………………………………………………………………… 69 9.4.1
Decontamination……………………………………………………………………………………………….. 69
9.4.1.1 Certificate of Decontamination ……………………………………………………………………………………… 70
10 WARRANTY CERTIFICATE………………………………………………………………………………………… 71
10.1 WARRANTY CERTIFICATE …………………………………………………………………………………………. 71
11 ORDERING INFORMATION………………………………………………………………………………………… 72
11.1 11.2 11.3 11.4 11.5 11.6 11.7 11.8
INSTRUMENTS ……………………………………………………………………………………………………….. 72 DISPENSING CASSETTES ………………………………………………………………………………………….. 72 SPARE PARTS FOR DISPENSING CASSETTES …………………………………………………………………. 73 PRIMING VESSELS (CONSUMABLES) ……………………………………………………………………………. 73 LIST OF ACCESSORIES……………………………………………………………………………………………… 73 LIST OF PLATES (CONSUMABLES) ……………………………………………………………………………….. 73 LIST OF SPARE PARTS ……………………………………………………………………………………………… 74 LIST OF RECOMMENDED SPARE PARTS ………………………………………………………………………… 74
12 GLOSSARY AND ABBREVIATIONS …………………………………………………………………………… 75
12.1 12.2
KEYWORDS FOR WEB PAGES …………………………………………………………………………………….. 76 LITERATURE ………………………………………………………………………………………………………….. 76
13 INDEX……………………………………………………………………………………………………………………….. 77
14 TABLE OF FIGURES………………………………………………………………………………………………….. 81
APPENDIX 1. THERMO SCIENTIFIC MULTIDROP 384 BRIEF USER’S GUIDE………………………… 83
APPENDIX 2. THERMO SCIENTIFIC MULTIDROP 384 FEEDBACK FORM……………………………… 84
APPENDIX 3. ADDRESSES………………………………………………………………………………………………….. 85
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
4
Copyright
Copyright © 2008 Thermo Fisher Scientific Corporation. All rights reserved.
First edition published in 1987. Printed in Finland. Reproduction of the
accompanying user documentation in whole or in part is prohibited.
Patents
This product is protected by the following patent: US 4995432.
Trademarks
Immulon, Microfluor and Multidrop are registered trademarks of Thermo Fisher
Scientific. Microlite is a trademark of Thermo Fisher Scientific.
All other trademarks and registered trademarks are the property of their
respective holders.
Disclaimer
Thermo Fisher Scientific reserves the right to change its products and
services at any time to incorporate technological developments. This manual is
subject to change without prior notice as part of a continuous product
development.
Although this manual has been prepared with every precaution to ensure
accuracy, Thermo Fisher Scientific assumes no liability for any errors or
omissions, nor for any damages resulting from the application or use of this
information. This manual supersedes all previous editions.
No liability for consequential damages
Thermo Fisher Scientific shall not be liable for any damages whatsoever
arising out of the use or inability to use this product.
Contact information
Thermo Fisher Scientific Oy P.O. Box 100, FI-01621 Vantaa, Finland Tel.
+358-9-329 100, Fax +358-9-3291 0415 www.thermo.com
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
5
Chapter 1: Safety Symbols and Markings
1 SAFETY SYMBOLS AND MARKINGS
These symbols are intended to draw your attention to essential information and
alert you to the presence of hazards as indicated. SAFETY SYMBOLS USED IN THE
MULTIDROP 384
Power ON
Power OFF
WEEE symbol This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC.
WARNING MARKINGS USED IN THE DOCUMENTATION Caution: risk of electric shock.
Caution: biohazard risk.
Caution: risk of personal injury to the operator or a safety hazard to the
surrounding area. Caution: risk of damage to the instrument, other equipment
or loss of performance or function in a specific application.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
6
Chapter 2: About the User Manual
2 ABOUT THE USER MANUAL
This user manual has been written for the end user (for example, laboratory
technician) and provides information on the Thermo Scientific Multidrop 384
microplate dispenser. This manual contains the installation and operating
instructions for the Multidrop 384 microplate dispenser.
Read the manual in its entirety prior to operating the instrument.
This user manual has been designed to give you the information you need to:
· Review safety precautions · Install the Multidrop 384 · Carry out dispensing
procedures · Recalibrate the dispensing cassette · Perform basic maintenance
procedures · Troubleshoot the instrument performance
This user manual also describes features and specifications of the Multidrop
384 hardware and onboard software.
Chapter 6 ROUTINE OPERATION explains the dispensing principles and procedures.
In Chapter 9 TROUBLESHOOTING you will find a problem-solving guide. The user
should be familiar with the contents of Chapter 7 on maintenance.
For warranty and ordering information refer to Chapters 10 WARRANTY
CERTIFICATE and 11 ORDERING INFORMATION.
In an effort to produce useful and appropriate documentation, we appreciate
your comments on this document to your local Thermo Fisher Scientific
representative.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
7
Chapter 3: Introduction to the Multidrop 384
3 INTRODUCTION TO THE MULTIDROP 384
There are two models of the Multidrop 384: · 5840150 Multidrop 384, 220240 V
50/60 Hz · 5840157 Multidrop 384, 110120 V 50/60 Hz
Each Multidrop 384 comes with one dispensing cassette: · 24072670 Standard
tube dispensing cassette (with 40 cm tubing set)
3.1 Intended use
The Multidrop 384 microplate dispenser is intended for professional research
use by trained personnel. The instrument is intended for automated dispensing
of up to eight different reagents simultaneously into microplates and strips
in 96 or 384-well plate format. Use for self-testing is excluded.
3.2 Principle of operation
The Multidrop 384 is an automatic, programmable, eight-channel microplate
dispenser. It has a peristaltic pump for rapid and continuous dispensing of
liquids into 96 and 384-well microplates. It can dispense 1 to 8 different
reagents from external liquid reservoirs into different rows and can be used
in most microtitration techniques, for example, in the fields of bacteriology,
cell biology, hematology, immunology, toxicology and virology. It is
particularly useful in the pharmaceutical industry. The Multidrop 384 is
lightweight, transportable and compact on a laboratory bench. It can dispense
20 µl into the entire 384-well microplate in 20 seconds or 20 µl into the
entire 96-well microplate in 5 seconds.
Fig. 3.1 Multidrop 384 microplate dispenser
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
8
Chapter 3: Introduction to the Multidrop 384
3.3 Advantages of using the Multidrop 384
The Multidrop 384 microplate dispenser provides several advantages relating
mainly to the principle of operation in that it has:
· Superior accuracy and precision · A flexible volume range · High-speed
continuous dispensing · Robot compatibility and versatile remote control
commands · An autoclavable, detachable and calibrateable dispensing cassette
The Multidrop 384 is an automated microplate dispenser designed for high-
throughput screening and microvolume dispensing into 384 and 96-well plates.
The Multidrop 384 can dispense up to eight different reagents with high
accuracy, precision and speed. It fills a 384-well plate with 20 µl/well in 20
seconds or a 96-well plate in 5 seconds. With a volume range of 5 100 µl for
384-well plates and 5 395 µl for 96-well plates, the Multidrop 384 offers
superior flexibility for a wide range of applications.
All reagent lines can be backflushed to the reagent bottle, minimizing the
loss of expensive reagents. From 1 to 24 columns can be selected for 384-well
plates and from 1 to 12 columns for 96-well plates.
To ensure sterility and avoid cross contamination, the dispensing cassette is
detachable and autoclavable.
The Multidrop 384 can be used as a standalone instrument or added to robotic
systems via an RS-232 serial port.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
9
4 FUNCTIONAL DESCRIPTION
4.1 Instrument layout 4.1.1 Front view
15
12
11
10
9 13 14
Chapter 4: Functional Description
4
1
3
2
16
5
8 7
6
1 Instrument cover 2 Control panel
3 Thumbwheels for selecting the volume (volume µl) and number of columns (col)
4 Plate switch (plate) for selecting the 96/384-well plate (= rocker switch)
5 Plate carrier 6 Plate adapter 7 Plate adapter lock 8 Transfer rails 9
Priming vessel 10 Lower part slots 11 Upper part slots 12 Rest position slots
of the upper part 13 Pump rotor 14 Rotor needles 15 Rotor cover not pulled
over the rotor 16 Pump body
Fig. 4.1 Multidrop 384 front view (A)
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
10
Chapter 4: Functional Description
2 1 2
3
1 Mains switch (ON/OFF) 2 Instrument cover retaining screws 3 Cooling-air
outlet
Fig. 4.2 Multidrop 384 front view (B)
1 2 3
4
1 Rotor cover pulled over the rotor 2 Standard tube dispensing cassette
attached 3 Plate adapter 4 Plate carrier
Fig. 4.3 Multidrop 384 front view (C)
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
11
Chapter 4: Functional Description
3
2
4
1
5
8 6
7
1 Tips (8) 2 Lower part of dispensing cassette with tips 3 Tension limiting
wires (2) 4 Upper part of dispensing cassette with tubes 5 Sight strips (8) 6
Tubes (A H) 7 Tubing weight, eight tubes 8 Calibration screw cover
Fig. 4.4 Close-up of the Multidrop 384 standard tube dispensing cassette
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
12
4.1.2 Back view
Chapter 4: Functional Description
1
3
2
1 Mains input socket 2 Fuses in the fuse holder cap 3 RS-232C serial interface
connector
Fig. 4.5 Multidrop 384 rear view
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
13
Chapter 4: Functional Description
4.2 Multidrop 384 microplate dispenser
The Multidrop 384 microplate dispenser (Fig. 4.6) automates dispensing into
microplates. The microprocessor-based system permits volumes between 5 100 µl
for 384-well plates and 5 395 µl for 96-well plates to be selected (in 5 µl
increments) and then dispensed simultaneously into eight wells (one column of
a 96-well microplate). Setting the volume to be dispensed and the number of
columns per plate (maximum 12 columns in 96-well plates and 24 columns in
384-well plates) is carried out using thumbwheels.
The movement of the microplate is automatic:
In the 96-well plate mode, the columns with 8 wells are filled starting from
column 1 up to the column number set, after which the plate carrier will
return to the home position. In the 384-well plate mode, the columns with 16
wells must be filled in two phases: first every other row (A, C, …, M, O) is
filled starting from column 1 up to the set column number. Then the plate
carrier shifts sideways so that the remaining rows (B, D, …, N, P) are filled
in the opposite order, from the set column number back to column 1.
Normally, the microplate moves through the instrument automatically, but a
semiautomatic mode is
possible by using the step and drop keys, which control dispensing into any
column of the plate.
The dispensing cassette contains eight individual tubes. The dispensing
cassette is detachable. Each reagent can have a separate dispensing cassette.
Alternatively, the same dispensing cassette can be used with several reagents
using the backflush feature. Refer to section 11.2 Dispensing cassettes.
The latest technology has been used in the manufacture of the dispensing
cassette which includes an injection molded manifold. All dispensing cassettes
can be autoclaved and Thermo Fisher Scientific guarantees them up to 50
autoclavings. After this amount it is recommended you replace the cassette.
To avoid damaging the tubes, follow the dispensing cassette maintenance,
washing, autoclaving and storage instructions carefully (see the instructions
in sections 6.8.4 After use, 7.2 How to wash the dispensing cassette and 7.3
How to autoclave the dispensing cassette).
Fig. 4.6 Multidrop 384 microplate dispenser
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
14
Chapter 5: Installation
5 INSTALLATION
5.1 Installation check list
The chapter on installation will contain an outline of the points mentioned in the check list below.
Tick Item
T
Unpack the Multidrop 384 instrument carefully. Refer to 5.2.1. Keep the original packaging
and packing material for future transportation.
T
Check the delivery for completeness. Refer to 5.2.2.
T
Check for damage during transport. Refer to 5.2.3.
T
Place the dispenser on a normal laboratory bench taking into account both the environmental
and technical prerequisites. Refer to 5.2.4 and 5.2.6. Leave sufficient clearance on both sides
and at the rear of the unit.
T
Install the instrument. Refer to 5.3.1.
T
Connect the mains supply cable to the mains input socket. See 5.3.2.
T
Carry out the operational check. Refer to 5.4.
5.2 What to do upon delivery
5.2.1 How to unpack
Move the unpacked instrument to its site of operation. Unpack the Multidrop
384 instrument and accessories carefully with the arrows on the transport
package pointing upwards. The following notes and instructions are sent with
the instrument and are immediately available when you open the package:
· Warranty Certificate card · packing instructions/packing list · Multidrop
384 User Manual
Caution: Do not touch or loosen any screws or parts other than those specially
designated in the instructions. Doing so might cause misalignment and will
invalidate the instrument warranty.
Retain the original packaging and packing material for future transportation. The packaging is designed to assure safe transport and minimize transit damage. Use of alternative packaging materials may invalidate the warranty. Also retain all instrument-related documentation provided by the manufacturer for future use.
5.2.2 Checking delivery for completeness
Check the enclosed packing list against order. If any parts are missing,
contact Thermo Fisher Scientific Oy or your local Thermo Fisher Scientific
representative.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
15
Chapter 5: Installation
5.2.3 Checking for damage during transport
Visually inspect the transport package, the instrument and the accessories for
any possible transport damage.
If the carton has been damaged in transit, it is particularly important that
you retain it for inspection by the carrier in case there has also been damage
to the instrument.
Neither the manufacturer nor its agents can be held responsible for any damage
incurred in transit, but the manufacturer will make every effort to help
obtain restitution from the carrier. Upon receipt of the carrier’s inspection
report, arrangements will be made for repair or replacement.
Visually check all interconnections in the basic instrument. Check that there
are no loose parts inside the instrument.
If any parts are damaged, contact Thermo Fisher Scientific Oy or your local
Thermo Fisher Scientific representative.
5.2.4 Environmental requirements
When you set up your Multidrop 384, avoid sites of operation with excess dust,
vibrations, strong magnetic fields, direct sunlight, draft, excessive moisture
or large temperature fluctuations.
· Make sure the working area is flat, dry, clean and vibration-proof and leave
additional room for accessories, cables, reagent bottles, etc.
· Make sure the ambient air is clean and free of corrosive vapors, smoke and
dust. · Make sure the ambient temperature range is between +10°C (50°F) and
+40°C (104°F). · Make sure relative humidity is between 10% and 80% (non-
condensing).
Leave sufficient space (at least 10 cm) on both sides and at the back of the
unit to allow adequate air circulation.
The Multidrop 384 does not produce operating noise at a level that would be
harmful. No sound level measurements are required after installation.
Warning: Do not operate the instrument in an environment where potentially
damaging liquids or gases are present.
5.2.5 Things to avoid
Do not smoke, eat or drink while using the Multidrop 384. Wash your hands
thoroughly after handling test fluids. Observe normal laboratory procedures
for handling potentially dangerous samples. Use proper protective clothing.
Use disposable gloves. Be sure the working area is well-ventilated.
Never spill fluids in or on the equipment.
5.2.6 Technical prerequisites
Place the instrument on a normal laboratory bench. The net weight of the
entire equipment is approx. 6.2 kg [13.7 lbs.] and the dispensing cassette 124
g [0.3 lb.].
The instrument operates at voltages of 100 240 Vac and a frequency range of
50 60 Hz.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
16
Chapter 5: Installation
5.3 Setups before you put the instrument into operation 5.3.1 Installation
procedure
The following procedure may be carried out by any properly skilled person
using standard tools. In case the local voltage does not correspond to the
label at the rear of the instrument on top of the mains input socket (Fig.
4.5), follow the next installation procedures.
Warning: Ensure the mains switch (Fig. 4.2) on the rear of the left side panel
is in the OFF position. Also ensure the mains supply cable is disconnected
(Fig. 5.3). 1. Remove the instrument cover retaining screws (Fig. 5.1 and Fig.
4.2) and lift up the cover.
Cover retaining screw on both sides
Fig. 5.1 Location of the cover retaining screws 2. Check that the voltage
selector switch (Fig. 5.2) is in the correct position, that is, the voltage
selected corresponds to the local voltage. Voltage
Fig. 5.2 Location of the voltage selector
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
17
Chapter 5: Installation
3. Visually check that you did not leave any loose parts inside the
instrument. 4. Close the cover and fasten the cover retaining screws.
5.3.2 How to ensure startup
1. Warning: Ensure the mains switch (Fig. 4.2) on the rear of the left side
panel is in the OFF position.
2. Connect the mains supply cable to the mains input socket (Fig. 4.3 and
Fig. 5.3) at the bottom of the back panel. If you need to use any other type
of mains supply cable than supplied, use only cables certified by the local
authorities.
3. Connect the instrument to a correctly installed line power outlet that has
a protective conductor that is grounded.
Warning: Never operate your instrument from a power outlet that has no ground
connection. Never use a mains supply cable other than the Thermo Scientific
mains supply cable designed for your region.
Fig. 5.3 Connecting the mains supply cable
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
18
Chapter 5: Installation
5.4 Operational check
The following procedure should be completed without the dispensing cassette to confirm the correct functioning of the dispenser prior to normal use.
1
Connect the mains supply cable (Fig. 5.3). Switch the instrument on using the mains switch (Fig. 4.2).
· If the instrument starts:
– The green power LED will light up
– The plate carrier will return to the home position
· If the instrument does not start:
)Pull the rotor cover over the rotor )Check the mains cable connection
2 Press prime/drop to prime
– The pump rotor will rotate
· If the pump rotor does not rotate:
)Pull the rotor cover over the rotor
· If the red error LED lights up: )Insert the priming vessel, press the stop
key and prime again
3 Press start
– Starts the dispensing cycle
· If the dispensing cycle does not start:
)Pull the rotor cover over the rotor
· If the red error LED lights up: )Press the stop key, then press the prime
key and finally press the start key again )Insert the priming vessel, press
the stop key, then press the prime key and finally press the
start key again
4 plate (plate switch)
– Dispenses liquid into the 384 or 96-well microplates
– Try the alternative plate formats by pressing the start key
5 step
– Advances the microplate by one column · If the microplate does not advance:
)Pull the rotor cover over the rotor
6 After step press prime/drop
– To dispense a preselected volume into the column selected by steps – The
plate carrier will return to the home position after reaching the last column
After step press start (partial filling of the tubing)
– To dispense the selected volume automatically from the set column to the
last column
7 empty
– The pump rotor will rotate
· If the pump rotor does not rotate:
)Pull the rotor cover over the rotor
8 stop
– Terminates the operation and returns the plate to the home position For more
information on the control panel, refer to section 6.5 How to use the control
panel.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
19
Chapter 5: Installation
5.5 Operating precautions and limitations before operation
1. Read this manual in its entirety, as it contains information necessary to
ensure safe operation. 2. Always ensure that the local supply voltage in the
laboratory conforms to that specified on the type
label on the back of the instrument.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
20
Chapter 6: Routine Operation
6 ROUTINE OPERATION
6.1 Operational check list
The chapter on routine operation will contain an outline of the points mentioned in the check list below.
Tick Item
T
Switch on the instrument. Refer to 6.2.
T
Insert the priming vessel and the dispensing cassette. Refer to 6.3.
T
Use the plate adapter for the plate carrier to function. Refer to 6.4.
T
Pull the rotor cover over the rotor. Refer to 6.3.
T
Familiarize yourself with the control panel. Refer to 6.5.
T
Press the plate switch (plate) to position 96 or 384 according to the plate type you use.
Refer to 6.5.2.
T
Prime all the eight channels so that there are no air bubbles in the tubes. Refer to 6.7.
T
Carry out dispensing either manually or automatically. Refer to 6.8.
T
After all the plates have been dispensed, press the empty key to return the reagent or liquid
from the tubes to the reservoir. Refer to 6.8.4.
T
Wash the dispensing cassette by priming it with deionized distilled water when not in use.
Refer to 6.8.4 and 7.2.
T
After washing, store the dispensing cassette in the rest position. Refer to 6.8.4 and 7.2.
T
Maintain the instrument regularly. Refer to Chapter 7.
Note: Keep the dispensing cassette in the rest position when the dispensing cassette is not in use.
6.2 Switching on
Switch on the instrument by using the mains switch on the left-hand side of
the instrument (Fig. 4.2).
The green power LED will light up.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
21
Chapter 6: Routine Operation
6.3 How to insert the priming vessel and the dispensing cassette
1. Ensure that the priming vessel is inserted correctly into its slot on the
left of the plate carrier (Fig. 6.1). You have fastened it correctly when it
snaps slightly and is inserted into the sensor slot.
Priming vessel sensor slot
Fig. 6.1 Inserting the priming vessel into place 2. Take the lower part of the
cassette into your right hand with the dispensing tips pointing down
and the upper part into your left hand (Fig. 6.2).
Fig. 6.2 Inserting the standard tube dispensing cassette
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
22
Chapter 6: Routine Operation 3. Carefully place the eight tubes below the pump
rotor and insert the lower part of the cassette into
the lower part slots of the pump body (Fig. 6.3). Check that the tubes are
freely placed below the rotor and the tension limiting wires below the rotor
shaft.
Lower part slots
Fig. 6.3 Inserting the lower part of the dispensing cassette into its slots 4.
Take a firm grip on the upper part of the cassette with your right hand and
carefully pull the tubes
around the rotor until the upper part reaches and fits into the upper part
slots of the pump body (Fig. 6.4 and Fig. 4.1).
Fig. 6.4 Inserting the upper part of the dispensing cassette
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
23
Chapter 6: Routine Operation
Note: The tension limiting wires are designed to prevent the user from pulling
the tubes too much as excessive tension may damage the tubes or change the
calibration of the cassette. 5. Double-check that both the upper and the lower
parts are properly placed into their corresponding slots (Fig. 6.3 and Fig.
6.5).
Upper part slots
Fig. 6.5 Lower and upper parts of the dispensing cassette inserted evenly into
their slots 6. Ensure that all the tubes are evenly placed on the rotor
needles, four tubes on each half of the
rotor. 7. Ensure also that the tension limiting wires have a loose fit around
the rotor shaft. 8. PULL THE ROTOR COVER OVER THE ROTOR (Fig. 6.6). The plate
carrier will then return to
the furthest right of the instrument, that is, the home position. The
microswitch controls that the cover is properly placed over the rotor. The
Multidrop 384 is now ready for priming.
Fig. 6.6 Pulling the rotor cover over the rotor
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
24
Chapter 6: Routine Operation
Note: If the rotor cover is not pulled over the rotor, the instrument will not
function.
Note: It is recommended you change the dispensing cassette or tubing set after
dispensing about 3000 5000 96-well plates, depending on the dispensing
frequency and the liquids used. Note: Keep the dispensing cassette in the rest
position when the dispensing cassette is not in use.
6.4 How to insert and adjust the plate adapter
The Multidrop 384 requires a plate adapter for the plate carrier to function.
Two different plate adapters are included (Fig. 6.7).
The PLATE ADAPTER, PLATE HEIGHT MAX. 15 MM, is suitable for plates
approximately 15 mm high, for example, 96well plates and some 384-well plates.
Fig. 6.7 Two different plate adapters
The PLATE ADAPTER, PLATE HEIGHT MAX. 11 MM, is suitable for lower plates, for example, 384-well plates.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
25
Chapter 6: Routine Operation
The plate adapter is inserted into the plate carrier in the following way: 1.
Open the plate adapter lock located in the front left corner of the plate
carrier (Fig. 6.8). Select
the plate adapter according to the height of the plate.
Fig. 6.8 Opening the plate adapter lock
2. Take the plate adapter in your left hand and lower the right end. Place
the plate adapter into the right end corner of the plate carrier (Fig. 6.9).
Plate adapter lock Fig. 6.9 Placing the right end of the plate adapter
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
26
Chapter 6: Routine Operation 3. Then press the left end of the plate adapter
onto the plate carrier (Fig. 6.10).
Fig. 6.10 Pressing the left end of the plate adapter 4. Lock the plate adapter
into place with the plate adapter lock located in the front left corner of the
plate carrier (Fig. 6.11).
Fig. 6.11 Closing the plate adapter lock
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
27
Chapter 6: Routine Operation 5. To adjust the plate size, loosen first the two
attached plastic-covered screws on the right of the
plate carrier (Fig. 6.12).
Fig. 6.12 Loosening the screws on the right of the plate carrier 6. Then
loosen the two attached plastic-covered screws at the front of the plate
adapter (Fig. 6.13).
Fig. 6.13 Loosening the screws at the front of the plate adapter
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
28
Chapter 6: Routine Operation
7. Insert the microplate onto the plate adapter. Push the microplate at the
same time towards the left-hand top corner of the plate adapter while you
fasten the front screws (Fig. 6.14).
Fig. 6.14 Fastening the front screws of the plate adapter
8. Then fasten the screws to the right of the plate carrier, pushing the
microplate simultaneously towards the left-hand corner of the plate adapter
(Fig. 6.15). If the instrument is used for robotics, leave sufficient moving
space for the microplate according to the robotic specifications.
Fig. 6.15 Fastening the screws to the right of the plate carrier
Note: Ensure the dispensing cassette is attached and the rotor cover has been
pulled over the rotor.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
29
Chapter 6: Routine Operation
6.5 How to use the control panel
The keypad has an embossed, tactile membrane with five keys. There are two
control panel indicators (LEDs) in all. The control panel is shown in Fig.
6.16.
Fig. 6.16 Multidrop 384 control panel
6.5.1 Control panel indicators
power
The green power LED indicates that the dispenser is powered.
error
The red error LED shows when the dispenser has some kind of error.
6.5.2 Control panel thumbwheels (digital switches)
plate
volume µl col
96/384-well switch: sets the plate carrier movement and dispensing for 96 or
384well plates.
Reagent volume thumbwheels: sets the volume to be dispensed per well.
Column thumbwheels: sets the number of columns per plate, up to 12 with 96well
plates or up to 24 with 384-well plates.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
30
Chapter 6: Routine Operation
6.5.3 Control panel keys
start prime/ drop
empty step stop
Starts the dispensing cycle.
Primes the cassette if the tips are above the priming vessel. Primes
continuously as long as the user presses the key. Dispenses the adjusted
volume if the tips are above the wells.
Empties the liquid from the tubes to the liquid reservoir.
Advances the microplate by one column.
Terminates the operation and returns the plate to the home position.
6.6 How to select the 96 / 384-well plate
· Press the plate switch (plate) down into position 96 or 384 according to the
plate type to be used
(Fig. 6.17).
· Select the number of columns (col) between 1 and 24 in the 384 mode.
· Select the number of columns (col) between 1 and 12 in the 96 mode (Fig.
6.17). Note that if you set the column selection above 12 in the 96 mode, the
red error light will light up and the
dispensing operation will be terminated.
Fig. 6.17 Multidrop 384 control panel
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
31
Chapter 6: Routine Operation
6.7 How to prime
Note: Ensure that the instrument is sufficiently level on the table to prevent
liquids from overspilling when priming.
Note: Ensure that the liquid temperature is close to the calibration
temperature of 22°C. Differences greater than ± 2°C will cause an error in the
results.
Note: Silicon tubing (silicon rubber) is resistant to a number of chemical
liquids. Most water solutions typically used in a medical laboratory can be
dispensed with the Multidrop 384. Some strong substances may, however, harm
the silicone tubing and cause swelling or breaking. If the substance has even
a slight effect on silicone, the calibration of the cassette may be affected
and the cassette may become unusable.
1. Place the eight tube ends into the same liquid reservoir or into different
reservoirs taking care that all tube ends are below the liquid level. Use the
tubing weight provided if possible.
2. Check that the priming vessel is firmly inserted next to the plate
carrier. The Multidrop 384 will not operate if the priming vessel is not
completely inserted into its slot.
3. Press the prime/drop key until all eight channels are dispensing
continuously into the priming
vessel and there are no more air bubbles in the tubes. It is recommended you
fill the priming vessel two or three times with the reagent to be used before
starting the actual dispensing in order to bring the tubes into working
condition.
Note: If you press the prime/drop key at the same time as the rotor cover is
pulled over
the rotor, the program enters into the volumetric calibration mode.
Note: You can empty the contents of the priming vessel back into the reagent
reservoir by pulling the priming vessel out of its slot if necessary. It is
recommended you do this if you use expensive reagents. Therefore, wash the
priming vessel always after each use to avoid any cross contamination. If you
use several liquids at the same time in different rows, you can order a
special eight-cavity priming vessel for this purpose to prevent mixing the
liquids. Also a self-draining (flow-through) priming vessel is available.
Refer to section 11.4 Priming vessels (consumables).
6.8 Dispensing
6.8.1 How to dispense automatically
1. Set the required reagent volume per well. The permitted volume range is 5
100 µl for 384-well plates and 5 395 µl for 96-well plates.
2. Set the required number of columns to be dispensed per plate.
3. Place the microplate onto the plate carrier equipped with a suitable plate
adapter. Refer to section 6.4 How to insert and adjust the plate adapter.
4. Press the prime/drop key to prime the tubes. Refer to section 6.7 How to
prime.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
32
Chapter 6: Routine Operation
5. Press the start key. The Multidrop 384 will now automatically dispense the
preselected volume
of reagent into the preselected number of columns. When dispensing has been
completed, the plate carrier will automatically return to its home position.
6. Now you can dispense any number of plates by only changing the plate and
pressing start.
Note: Check that there is always enough liquid in the reservoir and ensure
that all the tube ends are below the liquid level. Use the tubing weight
supplied.
Note: Check after each dispensing that the priming vessel has not filled up.
Empty it if necessary, otherwise the liquids may spill over during operation.
Note that the Multidrop 384 always primes itself before each plate filling to
ensure maximum accuracy and precision. It is recommended you empty the priming
vessel after dispensing 30 40 plates. Alternatively, use the self-draining
priming vessel.
6.8.2 How to dispense manually into the specified column(s)
1. Set the desired reagent volume. The permitted volume range is 5 100 µl in
the 384-well plate mode and 5 395 µl in the 96-well plate mode.
2. Place the microplate onto the plate carrier equipped with a suitable plate
adapter.
3. Use the step key to move the plate carrier column by column until you
reach the desired
column.
4. Press the prime/drop key. The Multidrop 384 will now dispense the
preselected volume into
the column desired. The plate carrier will return to the home position after
reaching the last
column. If you use the start key, the instrument will automatically dispense
the selected volume
from the set position to the last column selected.
Note: If you have pressed either the stop or empty key, the Multidrop 384
refuses to start dispensing before you have pressed the prime/drop key once
again.
6.8.3 How to dispense multiple volumes
Ten- or hundredfold volumes can be dispensed compared to the volume chosen.
6.8.3.1 Dispensing tenfold and hundredfold volumes
1. Press the stop and prime/drop keys simultaneously.
2. Having done that first, release the stop key and then the prime/drop key.
Note that the
Multidrop 384 is in the volumetric calibration mode.
3. Now you can push the plate carrier into the desired dispensing position.
4. Set one tenth of the volume required on the thumbwheel. Then press the
start key and a tenfold
volume will be dispensed.
5. Set one hundredth of the volume required on the thumbwheel. Then press the
step key and a
hundredfold volume will be dispensed.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
33
Chapter 6: Routine Operation
6.8.4 After use
After all the plates have been dispensed, press the empty key to return the
reagent or liquid from the
tubes to the reservoir. This saves expensive reagents.
If there is only a short break in dispensing, no washing or cleaning of the
tubes is necessary.
However, after pressing the empty key, release the tubing tension by pulling
the upper part of the
cassette from its slots and place it into its rest position slots located on
the left-hand side of the rotor (Fig. 6.18).
If there is a longer break (for example, a day), wash the dispensing cassette
by priming it with deionized distilled water or with special washing detergent
if necessary. In the end, rinse the washing detergent away with deionized
distilled water.
After washing, store the dispensing cassette in the rest position (Fig. 6.18).
Note: Keep the dispensing cassette in the rest position when the dispensing
cassette is not in use.
Fig. 6.18 Rest position of the dispensing cassette
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
34
Chapter 6: Routine Operation
6.9 Shutdown
Remove any microplates still on the plate carrier. Dispose of all microplates
and strips as biohazardous waste.
After all the plates have been dispensed, press the empty key to return the
reagent or liquid from the
tubes to the reservoir. This saves expensive reagents. If there is only a
short break in dispensing, no washing or cleaning of the tubes is necessary.
However, after pressing the empty key, release the tubing tension by pulling
the upper part of the
cassette from its slots and place it into its rest position slots located on
the left-hand side of the rotor (Fig. 6.18). If there is a longer break (for
example, a day), wash the dispensing cassette by priming it with deionized
distilled water or with special washing detergent if necessary. After washing,
store the dispensing cassette in the rest position (Fig. 6.18). Switch the
Multidrop 384 off by pressing the mains switch (Fig. 4.2) at the end of the
left side panel of the instrument into the OFF position. Wipe the instrument
surfaces with a soft cloth or tissue paper moistened with deionized distilled
water, a mild detergent (SDS, sodium dodecyl sulfate) or soap solution. If you
have spilt infectious agents on the dispenser, disinfect with 70% alcohol or
some other disinfectant (see section 7.8 Decontamination procedure).
6.10 Guidelines for use
· When the error LED is on, the start or step keys will not function.
· To clear the error and continue the dispensing procedure, press the stop key
and wait until the
plate carrier has returned to the home position. The dispensing procedure can
then be restarted.
· If you press the stop key, all operations will stop. The plate carrier is
automatically returned to
the home position. The Multidrop 384 will now require priming again.
· Preventative maintenance should be followed closely to keep the microplate
dispenser in the best condition for maximum reliability. A poorly maintained
instrument will not give the best results.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
35
Chapter 7: Maintenance
7 MAINTENANCE
7.1 Regular and preventive maintenance
7.1.1 General
Routine and service procedures must be performed by the user to prevent
unnecessary wear or hazards and are described below at the frequency with
which they should be applied. Abrasive cleaning agents are not recommended,
because they are likely to damage the paint finish. It is recommended you
clean the case of the instrument periodically to maintain its good appearance.
A soft cloth dampened in a warm, mild detergent solution will be sufficient.
Clean the outside of the instrument and the plate carrier with clean low-
pressure compressed air or a cloth dampened with water or a mild detergent
when necessary.
Painted surfaces can be cleaned with most laboratory detergents. Dilute the
cleaning agent as recommended by the manufacturer. Do not expose painted
surfaces to concentrated acids or alcohols for prolonged periods of time as
damage may occur.
Clean the control panel with a mild laboratory detergent. Plastic covers and
surfaces can be cleaned with a mild laboratory detergent or alcohol.
If any surfaces have been contaminated with biohazardous material, a mild
sterilizing solution should be used.
The dispensing cassette is autoclavable. Do not autoclave any other part of
this instrument.
7.1.2 Immediate
Although the Multidrop 384 is constructed from high-quality materials, you
must immediately wipe away spilled saline solutions, solvents, acids or
alkaline solutions from outer surfaces to prevent damage and wipe with
deionized distilled water.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
36
Chapter 7: Maintenance
7.2 How to wash the dispensing cassette
Wash the tubes by priming them first with tap water and then with deionized
distilled water. Make sure that all the tubes are washed properly. If cleaning
with water is insufficient, use a mild laboratory detergent and then prime
with large amounts of tap water followed by deionized distilled water. The
dispensing cassette can be dried at room temperature. After washing, store the
dispensing cassette in the rest position (Fig. 6.18). If the instrument is not
used for a couple of hours, insert the dispensing cassette into the rest
position. Refer to section 6.8.4 After use. Handle the dispensing cassette
with great care so that you do not damage the tubes and dispensing tips. These
factors will affect the service life of the dispensing cassette.
7.3 How to autoclave the dispensing cassette
The standard tube dispensing cassette (Cat. no. 24072670) can be autoclaved 50
times with the same tubes. The autoclaving conditions of the dispensing
cassette are as follows: 1 bar pressure at 121°C for 20 minutes.
Note: After autoclaving, the dispensing cassette must cool down to room
temperature before use.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
37
Chapter 7: Maintenance
7.4 How to replace the fuses
1. Switch off (Fig. 4.2) and unplug the instrument from the mains input
socket (Fig. 4.5). 2. Remove the fuse holder cap with a screwdriver or similar
tool. Press both ends of the fuse holder
cap with the tool (Fig. 7.1 and Fig. 7.2).
Fig. 7.1 Removing the fuse holder cap (A)
Fig. 7.2 Removing the fuse holder cap (B)
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
38
3. Check the fuses in the fuse holder cap (Fig. 7.3).
Chapter 7: Maintenance
Fig. 7.3 Fuse holder cap with fuses
4. Replace the blown fuse with the same type and push the fuse holder cap
back into its position (Fig. 7.4).
Fig. 7.4 Inserting the fuse holder cap 5. Reconnect to the mains input socket and switch the power on.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
39
Chapter 7: Maintenance
7.5 How to recalibrate the dispensing cassette
The dispensing cassette has been calibrated by the manufacturer with deionized
distilled water at 22°C ± 2°C. It is recommended you recalibrate the cassette
after dispensing about 1000 96-well plates or after every three months. It is
also recommended you recalibrate the dispensing cassette if you have forgotten
the cassette in the Multidrop 384 under tension for prolonged periods, for
example, a weekend or longer.
Two different calibration methods are described below, that is, volumetric
calibration and gravimetric calibration.
7.5.1 Volumetric calibration
For volumetric calibration, use the following steps:
1. Remove the priming vessel and microplate.
2. Install the dispensing cassette into the Multidrop 384 if it is not
already installed. Place the dispensing cassette into the rest position (Fig.
6.18). Refer to section 6.3 How to insert the priming vessel and the
dispensing cassette.
3. Install the tube of the calibration device into the tip of the channel to
be recalibrated (Fig. 7.5), for example, A.
Fig. 7.5 Placing the tube onto the tip of channel A
4. Insert the dispensing cassette into the working position so that the lower
and upper parts of the dispensing cassette are inserted into their slots (Fig.
6.5).
5. Pull the rotor cover over the rotor.
6. Place the tube of channel A into the reservoir with water at room
temperature (working temperature).
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
40
Chapter 7: Maintenance
7. Open the calibration screw cover on the upper part of the cassette (Fig.
7.7 and Fig. 4.4). 8. Set the volume to 100 µl.
9. Press the stop and prime/drop keys at the same time. After that first
release the stop key and then the prime/drop key. This brings the Multidrop
384 into the volumetric calibration
mode.
10. Press the start key as many times until the water enters the calibration
pipette. Note: Take care not to press the step key accidentally or otherwise
10 ml of water will be
dispensed.
11. Adjust the water level in the pipette to the lower mark by pulling or
pushing the syringe piston (Fig. 7.6).
Upper mark
Lower mark
Fig. 7.6 Adjusting the liquid level first to the lower mark and then to the
upper mark
12. Press the start key and wait until the water level stops. It should be at
the upper mark, 1000 µl
(± 1.5%) (Fig. 7.6).
13. If the volume is higher than 1000 µl, turn the calibration screw (Fig.
7.7) clockwise with a screwdriver; one turn for each 10 µl (1% from the volume
dispensed); if the volume is lower than 1000 µl, turn in the opposite
direction.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
41
Calibration screw cover opened
Chapter 7: Maintenance
Fig. 7.7 Screwing the calibration screws according to the calibration results
14. Press the empty key and adjust the water level in the pipette to the
lower mark and repeat Steps
10 to 13 until the result is 1 ml. Three accurate calibrations must be
obtained. 15. Change the tube to the next channel and repeat Steps 10 to 14.
16. After all the channels have been recalibrated, press the stop key to
reset the Multidrop 384 into
normal operation and reinstall the priming vessel. 17. Close the calibration
screw cover.
7.5.2 Gravimetric calibration
For gravimetric calibration, use the following steps: 1. Install the
dispensing cassette into the working position so that the lower and upper
parts of the
dispensing cassette are inserted into their slots (Fig. 6.5). 2. Pull the
rotor cover over the rotor. 3. Ensure that the priming vessel is installed.
Refer to section 6.3 How to insert the priming vessel
and the dispensing cassette. 4. Ensure that you have the plate adapter for 15
mm high plates installed. Refer to section 6.4 How
to insert and adjust the plate adapter.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
42
Chapter 7: Maintenance
Note: Keep the dispensing cassette in the rest position when the dispensing
cassette is not in use.
5. Place the tube of channel A into the reservoir with water at room
temperature (working temperature).
6. Use 8 x 12 well microstrips and adjust the Multidrop 384 for 100 µl in 10
columns and set the
plate switch (plate) for 96-well plates (Fig. 7.8).
Fig. 7.8 8 x 12 well microstrips adjusted onto the plate adapter
7. Press the prime/drop key to prime the tubes. Refer to section 6.7 How to
prime.
8. Open the calibration screw cover on the upper part of the cassette. 9.
Tare an analytical balance (resolution 0.01 mg) with one empty, dry 1 x 12
well microstrip. 10. Place the microstrip into the microplate frame on the row
corresponding to the channel to be
recalibrated (from A to H). Press the start key.
11. Weigh the filled microstrip (Fig. 7.9). The balance should now read 1000
mg ± 20 mg.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
43
Chapter 7: Maintenance
Fig. 7.9 Removing the filled microstrip to be weighed
12. If the balance reads less than 980 mg, turn the calibration screw
counterclockwise: one turn for each 10 mg (1% from the volume dispensed). If
the balance reads more than 1020 mg, turn in the opposite direction (Fig.
7.10). Recheck the delivery after each adjustment using Steps 9 to 11. Three
acceptable calibrations must be obtained.
Fig. 7.10 Screwing the calibration screw according to the calibration results 13. Continue in the same way with each channel until all eight channels have been recalibrated. 14. Close the calibration screw cover.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
44
Chapter 7: Maintenance
7.6 How to change the tubing set 7.6.1 Changing the tip band
1. Place the dispensing cassette onto the laboratory bench with the
dispensing tips pointing down. 2. Unfasten the four cover retaining screws of
the lower part of the dispensing cassette (Fig. 7.11).
Fig. 7.11 Removing the cover plate of the lower part of the dispensing
cassette
3. Remove the tip manifold, the tube fasteners and the rest of the tubing
from the lower part of the dispensing cassette. Then detach the tip band from
the tubing by pulling the tubes away from the tip band (Fig. 7.12).
Fig. 7.12 Removing the old tip band
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
45
Chapter 7: Maintenance 4. Replace the tip band (Fig. 7.13). The tips of the
tip band must not be clogged in any way but be
clean to function correctly (Fig. 7.14).
Fig. 7.13 Tip band removed Insert tube here
Tip
Fig. 7.14 New tip band
7.6.2 Fitting the new tubing set step by step
1. First place the new tubing set onto the laboratory bench with the tip band
attached (Fig. 7.15). Ensure that you have saved the eight calibration screws
and the tubing weight from the previous tubing set.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
46
Chapter 7: Maintenance
Tubing identification plate with tube lot number
Fig. 7.15 New tubing set with the tip band attached 2. Remove the four cover
retaining screws on the lower part of the dispensing cassette, the tubes
with the tube fasteners and the tip manifold with the tips, the three cover
retaining screws on the upper part of the dispensing cassette, the two tension
limiting wires, the calibration screws and the tubes with the tube fasteners.
Save the calibration screws and the tubing weight for further use when
installing the new tubing set. 3. Insert the tip manifold so that the tips go
through the holes in the lower part of the dispensing cassette (Fig. 7.16) and
guide the tubes through the eight openings in the side wall by pressing the
small cubic tube fasteners into place (Fig. 7.17). Ensure that each tube is in
its designated opening (Fig. 7.18).
Tip band
Tube fasteners
Fig. 7.16 Inserting the tip manifold into the lower part of the dispensing cassette
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
47
Chapter 7: Maintenance
Fig. 7.17 Pressing the tube fasteners into place
Fig. 7.18 New tubing inserted into the lower part of the dispensing cassette
4. Attach the tension limiting wires into their slots (Fig. 7.19).
Slots for tension limiting wires
Fig. 7.19 Tension limiting wires attached
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
48
Chapter 7: Maintenance 5. Fasten the four cover retaining screws of the lower part of the dispensing cassette (Fig. 7.20).
Fig. 7.20 Replacing the lower part cover of the dispensing cassette
6. Once the lower part of the dispensing cassette has been fitted, place the
remaining parts belonging to the upper part of the dispensing cassette onto
the laboratory bench (Fig. 7.21).
Fig. 7.21 Parts belonging to the upper part of the dispensing cassette
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
49
Chapter 7: Maintenance 7. First place the plastic sight window with the
grooves pointing upwards onto the bottom cover of
the upper part of the dispensing cassette, if it is not already there (Fig.
7.22).
Fig. 7.22 Placing the sight window 8. After that press the white tubing
identification plate into place on the right-hand side of the bottom
cover of the upper part of the dispensing cassette (Fig. 7.23).
Fig. 7.23 Pressing the tubing identification plate into place
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
50
Chapter 7: Maintenance 9. Then press each tube fastener into place in its groove on the plastic sight window (Fig. 7.24).
Fig. 7.24 Pressing the tube fasteners into place
10. Fasten the calibration screws from the previous tubing into their places
using a hexagonal screwdriver (Fig. 7.25).
Fig. 7.25 Fastening the calibration screws
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
51
Chapter 7: Maintenance 11. Insert the tension limiting wires into their slots
into the upper part of the dispensing cassette
(Fig. 7.26). Slots for tension limiting wires
Fig. 7.26 Calibration screws fastened 12. Replace the upper part cover of the
dispensing cassette carefully (Fig. 7.27).
Fig. 7.27 Replacing the cover of the upper part of the dispensing cassette 13.
Ensure you do not get the tubing trapped or bent between the bottom and upper
cover of the
upper part of the dispensing cassette (Fig. 7.28).
Fig. 7.28 Ensuring the tubing does not get trapped
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
52
Chapter 7: Maintenance 14. Fasten the three cover retaining screws of the
upper part of the dispensing cassette, leaving the
front screw slightly open (Fig. 7.29).
Fig. 7.29 Fastening the cover retaining screws 15. Insert the calibration
screw cover easier by lifting or bending the upper cover slightly (Fig. 7.30).
Fig. 7.30 Inserting the plate on the calibration screws
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
53
Chapter 7: Maintenance 16. Fasten the front cover retaining screw (Fig. 7.31).
Continued below.
Fig. 7.31 Lower and upper part of the dispensing cassette fitted
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
54
Chapter 7: Maintenance 17. Turn the dispensing cassette so that the eight
sight strips are visible (Fig. 7.32). Look at the eight
sight strips in the upper part of the dispensing cassette. You will notice
that the tubes are not precalibrated. First open the calibration screw cover.
For precalibration of the dispensing cassette, adjust the tube fasteners in
the sight strips roughly to the middle using a hexagonal screwdriver (Fig.
7.33).
Sight
Fig. 7.32 Dispensing cassette before precalibration
Fig. 7.33 Precalibrating the dispensing cassette
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
55
Chapter 7: Maintenance
18. When the tube fasteners have all been adjusted roughly to the middle of
the sight strip, close the calibration screw cover (Fig. 7.34).
Fig. 7.34 Dispensing cassette precalibrated
19. Insert the eight tubes into the tubing weight either at this stage or
earlier (Fig. 7.35). The new tubing set has now been successfully fitted.
Fig. 7.35 Inserting the tubes into the tubing weight
20. Calibrate the dispensing cassette according to the instructions in
section 7.5 How to recalibrate the dispensing cassette.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
56
Chapter 7: Maintenance
7.7 Disposal of materials
Refer to local regulations for the disposal of infectious material.
The samples can be potentially infectious. Dispose of all used microplates,
microstrips, disposable gloves, syringes, disposable tips, etc. as
biohazardous waste.
7.8 Decontamination procedure
If there is any risk of contamination with biohazardous material, the procedure recommended below or some other corresponding efficient decontamination procedure must be performed.
You must perform the complete decontamination procedure before shipping the instrument to Thermo Fisher Scientific Oy, for example, for repair, or relocating the instrument from one laboratory to another. Refer to section 7.9 How to pack for service. Decontamination is not required for the proper functioning of the instrument.
Examples of decontaminants
· Ethanol · Virkon solution · Glutaraldehyde solution · Chloramine T · Microcide SQ
70% 1 3% 4%
1:64
IF LOCAL OR LABORATORY REGULATIONS PRESCRIBE REGULAR DISINFECTION, IT IS NOT ADVISABLE TO USE FORMALDEHYDE, SINCE EVEN SMALL TRACES OF FORMALDEHYDE AFFECT THE ENZYME BEING USED IN EIA TESTS IN A NEGATIVE WAY RESULTING IN BAD TEST RESULTS.
Always use disposable gloves and protective clothing and operate in a well- ventilated area.
Do not autoclave any other parts of the instrument except the dispensing cassette.
1. Prepare the decontaminant: for example, 1 3% Virkon solution or 200 ml 4%
glutaraldehyde solution (or another agent recommended by your safety officer).
2. Empty the reagent container.
3. Switch the power OFF and disconnect the mains supply cable (Fig. 4.2).
4. Disinfect the outside of the instrument using a wad of cotton wool/cloth
soaked in 70% ethanol.
5. Place the plate adapter and dispensing cassette, for example, in 1 3%
Virkon solution or 200 ml 4% glutaraldehyde solution for 10 min and rinse them
with deionized distilled water.
6. Place the entire instrument in a large plastic bag. Ensure that the
reagent containers are disconnected and open.
7. Place a wad of cotton wool soaked in the prepared solution into the bag.
Ensure that the wad does not make contact with the instrument.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
57
Chapter 7: Maintenance
8. Close the bag firmly and leave the instrument in the bag for at least 24
hours.
9. Remove the instrument from the bag.
10. Clean the instrument using a mild detergent.
11. Remove any stains using 70% ethanol.
12. Flush the fluid path of the instrument with deionized distilled water
using the priming procedure.
13. The dispensing cassette must be autoclaved.
14. Also wash and disinfect the reagent containers.
15. After performing this decontamination procedure, enclose a signed and
dated Certificate of Decontamination both inside the transport package and
attached to the outside of the package.
7.9 How to pack for service
When you ship the instrument for service remember to:
· Inform about the use of hazardous materials. · Decontaminate the instrument
beforehand. · Pack the instrument according to the enclosed packing
instructions. · Use the original packaging to ensure that no damage will occur
to the instrument during shipping.
Any damage will incur additional labor charges. · Enclose a dated and signed
Certificate of Decontamination (see section 9.4.1.1) both inside and
attached to the outside of the package, in which you return your instrument
(or other items). · Enclose the return authorization number (RGA) given by the
Thermo Fisher Scientific
representative. · Indicate the fault after you have been in touch with your
local Thermo Fisher Scientific
representative or the Thermo Fisher Scientific technical service department.
See section 8.1 for details on storage and transportation temperatures.
7.10 Service contracts
It is recommended you maintain and service the instrument regularly every
twelve months on a contract basis by the manufacturer’s trained service
engineers. This will ensure the product is properly maintained and gives
trouble-free service. Contact the Thermo Fisher Scientific technical service
department for more details.
7.11 Disposal of the instrument
1. Decontaminate the instrument prior to disposal. See sections 7.8, 9.4.1
and 9.4.1.1 on decontamination. Also wash and disinfect the dispensing
cassettes prior to disposal.
2. Follow laboratory and country-specific procedures for biohazardous or
radioactive waste disposal. Dispose of the instrument according to the
legislation stipulated by the local authorities concerning take-back of
electronic equipment and waste. The proposals for the procedures vary by
country.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
58
Chapter 7: Maintenance
WEEE symbol Thermo Fisher Scientific has contracted with one or more
recycling/disposal companies in each EU Member State (European Country), and
this product should be disposed of or recycled through them. Further
information on Thermo Fisher Scientific’s compliance with these Directives,
the recyclers in your country, and information on Thermo Scientific products
which may assist the detection of substances subject to the RoHS Directive are
available at www.thermo.com/WEEERoHS.
3. Regarding the original packaging and packing materials, please use the
recycling operators known to you.
4. For more information, contact your Thermo Fisher Scientific
representative.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
59
Chapter 8: Technical Specifications
8 TECHNICAL SPECIFICATIONS
8.1 General specifications
Overall dimensions
310 mm (12.2 in.) (W) x 320 mm (12.6 in.) (D) x 155 mm (6.1 in.) (H)
Weight (total)
Instrument 6.2 kg [13.7 lbs.]
Dispensing cassette 124 g [0.3 lb.]
Mains power supply
100 120 Vac, 50/60 Hz, nominal (operating range 90 130 Vac) 200 240 Vac, 50/60 Hz, nominal (operating range 180 256 Vac)
Power consumption 50 VA max.
Heat dissipation Fuses
171 BTU max.
2 x 300 mA for 200 240 V, UL 198G Time Delay, 5 x 20 mm 2 x 600 mA for 100 120
V, UL 198G Time Delay, 5 x 20 mm
Operating conditions +10°C +40°C (indoor use only)
Transportation conditions Storage conditions
-40°C +70°C, packed in transport packaging -25°C +50°C, packed in transport packaging
Autoclaving conditions of the dispensing cassette
pressure: 1 bar temperature: 121°C time: 20 min autoclavings: 50 times
Control panel
Keys: start, prime/drop, empty, step, stop Selection switches: plate (= 96/384), volume µl, col (= column)
Interface
Serial RS-232C
384-well dispensing
96-well dispensing
Dispensing volume Dispensing speed
5 100 µl in 5 µl increments 20 s/20 µl into 384 wells
5 395 µl in 5 µl increments 5 s/20 µl into 96 wells
Dispensing accuracy Dispensing precision
± 3% or ± 1 µl, whichever is greater 5 µl: ± 3% or ± 1 µl, whichever is
greater 20 µl: ± 2% (typical) 100 µl: ± 1% (typical)
Valid when counted for the entire plate
5 µl: CV 10% or SD 0.6 µl, whichever is greater 10 395 µl: CV 3.0% or SD 0.4
µl, whichever is greater
5 µl: 10% (typical) 20 µl: 1.5% (typical) 100 µl: 1% (typical) Valid when
counted for the entire plate
- using the calibration procedure in section 7.5 the accuracy can be improved to ± 1.5%
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
60
Chapter 8: Technical Specifications
8.2 Safety specifications
8.2.1 Live parts
The instrument is safe to operate with the covers fitted and these must not be
removed during operation. The covers protect the user from live parts and they
should only be removed after switching the instrument off and disconnecting
the mains supply cable, and only by suitably qualified maintenance and repair
personnel.
Voltages dangerous for human beings are present in this instrument. Before
removing any covers, disconnect the instrument from the power supply.
The Multidrop 384 fulfils the following requirements:
EN 61010-1:1993 + A2:1995/IEC 61010-1:1990 + A1:1992 + A2:1995 taking into account US and CA National differences EN 61010-1:2001 (Ed. 2)
The safety specifications are also met under the following environmental conditions in addition to or in excess of those stated in the operating conditions:
Altitude Temperature Mains supply fluctuations Installation category (overvoltage category) Pollution degree
Up to 2000 m +10°C +40°C ± 10% from nominal II according to IEC 60664-1 (Note
- 2 according to IEC 60664-1 (Note 2)
Notes
1. The installation category (overvoltage category) defines the level of transient overvoltage, which the instrument is designed to withstand safely. It depends on the nature of the electricity supply and its means of overvoltage protection. For example, in CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains, such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V for a 230 V supply and 1500 V for a 120 V supply.
2. The pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2 assumes that normally only nonconductive pollution, such as dust, occurs with the exception of occasional conductivity caused by condensation.
Both of these affect the dimensioning of the electrical insulation within the instrument.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
61
Chapter 8: Technical Specifications
8.3 In conformity with the requirements
The Multidrop 384 bears the following markings:
Type 832 100 120 Vac, 200 240 Vac 50/60 Hz, 50 VA CE mark CSA monogram
The Multidrop 384 conforms to the following requirements:
2006/95/EC (Low Voltage Directive) 2004/108/EC (Electromagnetic Compatibility Directive, EMC) FCC Part 15, Subpart B/Class B 2002/96/EC (Waste of Electrical and Electronic Equipment)
Safety performance:
EN 61010-1:1993 + A2:1995/IEC 61010-1:1990 + A1:1992 + A2:1995 taking into account US and CA National differences EN 61010-1:2001 (Ed. 2)
EMC performance:
EN 50081-1:1992
Generic emission standard. Residential, commercial and light industry.
EN 50082-1:1997
Generic immunity standard. Residential, commercial and light industry.
EN 61326-1:1997 + A1:1998 Product family standard.
+ A2:2001 + A3:2003
Test standards and
Performance limits
EN 55022:1998 EN 61000-3-2:1995 + A1:1998 + A2:1998 + A13:1997 + A14:2000 EN 61000-3-3:1995 ANSI C63.4:1992
Class B, 150 kHz 1 GHz Class A Class B, 450 kHz 1 GHz
EN 61000-4-2:1995 + A1:1998 EN 61000-4-3:1996 + A1:1998 EN 61000-4-4:1995 EN 61000-4-5:1995 EN 61000-4-6:1996 EN 61000-4-8:1994 EN 61000-4-11:1994
4 kV CD, 8 kV AD, Criteria B 3 V/m, 80 MHz 2 GHz, Criteria A 1 kV, Criteria B 2 kV line to ground, 1 kV line to line, Criteria B 3 Vrms, 150 kHz 80 MHz, Criteria A 3 A/m, Criteria A 30%/10 ms, Criteria B 60%/100 ms, Criteria C 100%/20 ms, Criteria B > 95%/5 s, Criteria C
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
62
Chapter 8: Technical Specifications
8.4 Remote control to Multidrop 384
The interface system consists of the serial RS-232C computer port which is
used for computer control.
8.4.1 Hardware for the computer port
The character length is 10 bits with 1 start bit, 8 data bits, 1 stop bit and
no parity. The format cannot be changed.
The following signals are connected to the computer port DB-25 connector at
the rear of the instrument:
DB-25 CONNECTOR Pin no. 2 3 4 7 20
Signal name TxD RxD RTS GND DTR
Signal description
2: 3: 7: 4/20:
TxD, transmit data (output) RxD, receive data (input) GND, signal ground RTS/DTR (handshake lines)
The handshake lines: RTS/DTR are not in use in the standard Multidrop 384.
8.4.2 Multidrop 384 PC interface
Baudrate Parity Data bits Stop bits Handshake
9600 None 8 1 XON/XOFF
Commands
Commands to the Multidrop 384 have two to three fields as described in the
following syntax. The first letter identifies the command. Then a numeric
argument field is optional for some and required for other commands. The
numeric field immediately follows the uppercase letter without any separating
characters. The last character of a command is a linefeed character
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
63
Chapter 8: Technical Specifications
In the following list of commands, the command arguments in square brackets are optional.
D
Dispense the volume set by the ‘V’ command to the entire plate. Primes 10 µl into the priming vessel before dispensing.
E
Empty the pump. A volume of 880 µl is pumped backwards.
G
Dispense the plate using parameters from the plate switch and the
volume/columns
thumbwheels. This command has the same effect as pressing the start key. This
command has been added to version 1.7.
M[c]
Dispense the given number of columns starting from the current column. If the column count is not given, one column is dispensed. The dispensing volume is set by the ‘V’ command. After the command is completed, the dispensing tips remain at the last column dispensed. If this command is received when the plate is in the home position, dispensing starts from column 1. ER3 is reported if more columns are requested than left over from the current column to the last column of the plate.
N
Report the internal software version of the instrument.
O
Drive the plate out to the priming position.
P[v]
Prime the given volume. For a 96-well plate the volume must be in the range of 5 to 1000 µl. For a 384-well plate the range is 5 to 100 µl. The volume must be dispensed in 5 µl increments. If the volume is not given, 200 µl is primed. If the plate is not in the home position when the command is received, it is first driven into home position.
Q
Reset the instrument. There is no response to this command.
S[c]
Drive the requested column under the dispensing tips. Valid columns are 1 to 12 for a 96-well plate and 1 to 24 for a 384-well plate. If the column is not given, the steps are one column forward or the plate returns to home position after the last column.
Tt
Set the plate type assumed for all other commands except the `G’ command. At startup the default type is read from the plate switch.
t = 0 96-well plate t = 1 384-well plate This command has been added to version 1.7.
V
Report the internal software version of the instrument just like the `N’ command.
VER
Vv
Sets the volume for dispensing. For a 96-well plate the volume must be in the range of 5 to 1000 µl. For a 384-well plate the range is 5 to 140 µl. The volume must be dispensed in 5 µl increments.
Zs
Shake the plate the desired amount of time. The time is seconds and may range from 1 to 60.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
64
Chapter 8: Technical Specifications
Responses and error codes
The instrument sends a response to each command after the command has been
executed. The only exception is the ‘Q’ command, for which there is no
response.
The response to an ‘N’ command is as follows: Mdrop384 Vr.l[-b]
where: r is the release number; l is the level number, and b is the optional
branch number.
The response to other commands, if the command is executed successfully, is as
follows: OK
If an error is detected, the response is one of the following: ER3
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
65
Chapter 9: Troubleshooting
9 TROUBLESHOOTING
9.1 Troubleshooting guide
The problems covered below are considered as faults that require repair or corrective work. If the installation procedure is carefully followed, no faults should arise. However, if problems do occur or reoccur, contact authorized technical service immediately.
Error No power to the instrument
The keys do not function
Cause The instrument has not been connected to the mains.
A fuse has blown.
Drop on the end of the The tip is defective or clogged. tip exceptionally large
The tube is defective.
Large air bubbles in the tubes Poor dispensing accuracy and precision
The dispensing cassette is not properly installed. The instrument is not
primed.
The dispensing cassette is not properly installed. The tubes are worn out.
There is friction on one of the rotor needles. Bearing malfunction.
Liquid being dispensed does not fall into the required well of the microplate
Plate carrier moves incorrectly
The dispensing cassette is not properly installed.
The tip is defective or clogged.
There is foreign matter in the way of the carrier.
The drive band of the drive wheel is worn out or broken.
The drive wheel is loose.
Instrument does not function
The error LED
operates when trying
to prime
The drive wheel is not properly adjusted.
The priming vessel is not present. There is dirt in the priming vessel sensor.
Action Connect the instrument to the mains.
Replace the blown fuse, refer to section 7.4. Check that the rotor cover is
over the rotor.
Check that the priming vessel is properly inserted into its slot. Clear the
tip with a needle or replace the tip strip.
Replace the tubes.
Reinstall the cassette.
Prime, refer to section 6.7.
Reinstall the cassette.
Replace the tubes.
The rotor needles should spin without any friction. If even a slight friction
is present, contact authorized technical service. Reinstall the cassette.
Clear the tip with a needle or replace the tip strip. Remove the foreign
matter.
Contact authorized technical service.
Contact authorized technical service.
Contact authorized technical service. Contact authorized technical service.
Insert the priming vessel.
Clean the priming vessel sensor.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
66
Chapter 9: Troubleshooting
9.2 FAQs about the Multidrop 384
Q1: Can the same dispensing cassette be used in all Multidrop models? A1: No,
use only dispensing cassettes designed for Multidrop 384. It is not possible
to use small tube dispensing cassettes.
Q2: How many times can the dispensing cassette be autoclaved? A2: Fifty times.
Q3: What are the autoclaving conditions for the dispensing cassette? A3: The
autoclaving conditions are as follows: 1 bar pressure at 121°C for 20 min.
Q4: What is the dead volume of the standard 40 cm tubing set? A4: The dead
volume is about 5.6 ml. Note that you do not waste the reagent due to the
special backflush feature of the Multidrop 384, with which you can empty your
reagents from the tubing back into the original container.
Q5: What liquid is recommended to be used for washing the Multidrop 384 tubes?
A5: It is recommended you use deionized distilled water and/or detergent
solutions, for example, 0.2 1% Tween followed by deionized distilled water.
Q6: What is the material of the Multidrop 384 priming vessel? A6: The
Multidrop 384 priming vessel is made of black polypropylene (PP).
Q7: What is the material of the dispensing cassette tip band? A7: The tip band
(Fig. 7.14) material is polypropylene (PP). The melting temperature of the
material is 160 165°C.
Q8: Can ultraviolet light be used for sterilization? A8: Ultraviolet rays can
be used to sterilize objects, for example, the surface of the Multidrop 384
instrument and other areas where destruction of bacteria is necessary.
Q9: When should I calibrate the dispensing cassette? A9: The dispensing
cassette needs to be calibrated when the silicone tubes have been changed. It
is also recommended you recalibrate the dispensing cassette after dispensing
about 1000 96-well plates or after every three months. The dispensing cassette
must also be recalibrated if it has been forgotten under tension for prolonged
periods, for example, a weekend or longer.
Q10: What is the dead volume of the 2 m tubing set? A10: The dead volume is
about 24 ml.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
67
Chapter 9: Troubleshooting
9.3 Hazards
This instrument is designed to provide full user protection. When correctly
installed, operated and maintained, the instrument will present no hazards to
the user.
The following recommendations are given to ascertain added user safety.
9.3.1 Electrical
Ensure that the mains supply cable supplied with the unit is always used. If a
correct type of mains cable is not provided, use only cables certified by the
local authorities.
The mains plug should only be inserted into a socket outlet provided with a
protective ground contact. The protective action must not be negated by use of
an extension cable without a protective ground wire.
When the instrument is connected to the mains supply, the opening of covers or
removal of components is likely to expose live parts. The instrument should be
disconnected from all voltage sources by disconnecting the mains supply cable,
before it is opened for any adjustment, replacement, maintenance or repair
purposes.
Any adjustment, maintenance or repair of the opened instrument under voltage
should be avoided, if possible, but if unavoidable, should only be carried out
by a skilled technician aware of the hazard.
Voltages dangerous for human beings are present in this instrument. Before
removing any covers, disconnect the instrument from the power supply.
The same precautions applicable when using any electrical equipment should
certainly be observed with this instrument. Do not touch switches or
electrical outlets with wet hands. Switch the instrument off before
disconnecting it from the mains supply.
9.3.2 Mechanical
The dispensing cassette is a free-moving mechanical device controlled by the
instrument and not directly by the operator. It is designed for hands-off’ operation and should be used as such. Never reach into the work space, while the instrument is performing a dispensing cycle. If it is necessary to stop the operation of the equipment, press the stop key on the control panel. 9.3.3 Environmental Infectious samples and corrosive fluids are commonly used with this equipment. The
hands-off’ nature of the system allows the user to dispense into the
reaction wells without getting into direct contact with these fluids. However,
the wells that have been in contact with potentially hazardous fluids must be
handled before and after the dispensing process, and this should be done with
utmost care. Hand and eye protections should always be worn as well as
corrosive resistant laboratory coats.
Observe normal laboratory procedures for handling potentially hazardous
samples.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
68
Chapter 9: Troubleshooting
9.3.4 Defects and abnormal stresses
Whenever it is likely that the protection against safety hazards has been
impaired, the instrument should be made inoperative and be secured against any
unintended operation.
The protection is likely to be impaired if, for example, the instrument:
shows visible damage;
fails to perform the intended functions;
has been subjected to prolonged storage under unfavorable conditions, or
has been subjected to severe transport stresses.
9.4 Service request protocol
If the Multidrop 384 requires service, contact your local Thermo Fisher
Scientific representative or the Thermo Fisher Scientific technical service
department. Do not under any circumstances send the instrument for service
without any prior contact. It is imperative to indicate the fault and nature
of the required service. This will ensure a faster return of the instrument to
the customer.
The Thermo Fisher Scientific representative or distributor will take care of
sending a complaint form (that is, the Warranty Claim Technical Sheet) to the
Thermo Fisher Scientific technical service department. The Warranty Claim
Technical Sheet contains a more detailed description of the fault, symptom or
condition. Give all the necessary information to the distributor, who will
fill in and forward the Warranty Claim Technical Sheet to the Thermo Fisher
Scientific technical service department.
Check section 7.9 How to pack for service. You will find instructions on how
to proceed before shipping the instrument for service.
Check that any necessary decontamination procedure has been carried out before
packing. See sections 9.4.1 and 9.4.1.1 on decontamination. Refer also to
section 7.8 Decontamination procedure. Ensure that the Certificate of
Decontamination as well as the return authorization number (RGA) are sent with
the instrument.
The Thermo Fisher Scientific technical service department will keep you up to
date with the progress of service and provide you with any further details you
might need, for example, on maintenance, serviceability, troubleshooting and
replacement.
9.4.1 Decontamination
Decontamination should be performed in accordance with normal laboratory
procedures. Any decontamination instructions provided with the reagents used
should be followed.
A decontamination procedure is only recommendable when infectious substances
have been in direct contact with any part(s) of the instrument.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
69
Chapter 9: Troubleshooting
9.4.1.1 Certificate of Decontamination
The decontamination procedure is required prior to shipping the instrument to Thermo Fisher Scientific Oy, for example, for repair. If, for any reason, the instrument is shipped back to Thermo Fisher Scientific Oy, it must be accompanied by a dated and signed Certificate of Decontamination, which must be attached to the outside of the package containing the instrument. See section 7.8 Decontamination procedure.
Failure to confirm decontamination will incur additional labor charges or at worst the items will be
returned for proper cleaning.
Before returning any instrument(s) or item(s), ensure that they are fully decontaminated. Confirm A or
B
status:
Name:
Address:
Tel./Fax:
Name:
Serial no.:
A) I confirm that the returned items have not been contaminated by body fluids, toxic, carcinogenic or radioactive materials or any other hazardous materials. B) I confirm that the returned items have been decontaminated and can be handled without exposing the personnel to health hazards.
Materials used in the unit:
Specific information about contaminants:
Chemicals + Biological ·
Radioactive *)
Decontamination procedure1:
Date and place:
Signature:
Name (block capitals):
*) The signature of a Radiation Safety Officer is also required when the unit has been used with radioactive materials. This unit is certified by the undersigned to be free of radioactive contamination.
Date and place:
Signature:
Name (block capitals):
PHOTOCOPIABLE
1 Please include decontaminating solution used.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
70
Chapter 10: Warranty Certificate
10 WARRANTY CERTIFICATE
10.1 Warranty Certificate
Thermo Fisher Scientific Microplate Instrumentation Business products are
fully guaranteed against defective parts and materials, including defects
caused by poor workmanship, for a period of one year from the date of
delivery. Thermo Fisher Scientific will repair or replace defective parts or
materials during the term of warranty at no extra charge for materials and
labor provided that the products were used and maintained in accordance with
Thermo Fisher Scientific’s instructions. The warranty is invalid if products
have been misused or abused. For the warranty to be effective, the product
must have been purchased either directly from Thermo Fisher Scientific or from
an authorized Thermo Fisher Scientific distributor. The guarantee is not
transferable to a third party without prior written approval from Thermo
Fisher Scientific. This guarantee is subject to the following exclusions:
· Any defects caused by normal wear and tear. · Defects caused by fire,
lightning, flood, earthquake, explosion, sabotage, war, riot, or any
other occurrence of the type listed above. · Refurbished products that are
subject to different warranty conditions. THIS WARRANTY IS IN LIEU OF ALL
OTHER EXPRESSED OR IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. The
seller is not liable for any loss or damage arising out of or in connection
with the use of the product or other indirect damages. These warranty terms
and conditions can be obtained from your local Thermo Fisher Scientific
dealer.
Note: Consumables are not included in the warranty.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
71
Chapter 11: Ordering Information
11 ORDERING INFORMATION
11.1 Instruments
Code
Instrument
5840150 Multidrop 384, 220 240 V 50/60 Hz 5840157 Multidrop 384, 110 120 V 50/60 Hz
11.2 Dispensing cassettes
Code
Dispensing cassette
24072670 24072671 24072672 24072677
24070300 24070307
Standard tube dispensing cassette (with 40 cm tubing set) Standard tube dispensing cassette 5-pack Standard tube dispensing cassette 10-pack Long standard tube dispensing cassette (length of tubing 50 200 cm in 10 cm increments, please specify the length when ordering) Dispensing cassette Machined covers (with 40 cm tubing set) (Fig. 11.1) Special dispensing cassette Multidrop, Machined covers (length of tubing 50 200 cm in 10 cm increments, please specify the length when ordering)
Fig. 11.1 Dispensing cassette Machined covers
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
72
Chapter 11: Ordering Information
11.3 Spare parts for dispensing cassettes
Code
Item
24070290 24070297
1046200 1190780 1002430 2805980
Tubing set for dispensing cassette (with 40 cm tubing set) Special tubing set for dispensing cassette, special tubing set Multidrop (length of tubing 50 200 cm in 10 cm increments, please specify the length when ordering) Tube tip manifold Tubing weight, eight tubes Tubing weight, one tube Tube holder assembly
11.4 Priming vessels (consumables)
Code
Priming vessel
1046240 1046241 10260315
Priming vessel, one cavity Priming vessel, eight cavities Self-draining priming vessel
11.5 List of accessories
Code 1507010 2406910 1610641 24071790 24071800 2305290
Item Multidrop 384 User Manual Volumetric calibration device Dust cover Plate adapter (max. 15 mm height) Plate adapter (max. 11 mm height) Serial cable F9/F25
11.6 List of plates (consumables)
Code
3355 9502227 7571 9502887 7605 9502867 6310
95040000 95040020
Item 96-well microplates Immulon 1B Plate, Flat bottom 96 Well Plate UB, Flat bottom Microlite 1 + Plate, Flat bottom White 96 Well Plate UB, Flat bottom Microfluor 1 Black Plate, Flat bottom Black 96 Well Plate UB, Flat bottom Immulon 1 B 1×12 Strip Plate, Flat bottom 384-well microplates 384 Round Well Plate White 384 Round Well Plate
Qty
50 pcs/box 50 pcs/box 50 pcs/box 50 pcs/box 50 pcs/box 50 pcs/box 100 pcs/box
50 pcs/box 50 pcs/box
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
73
Chapter 11: Ordering Information
11.7 List of spare parts
Code
Item
Qty
1210810 Fuse 300 mA 10 pcs/box 1210850 Fuse 600 mA 10 pcs/box
11.8 List of recommended spare parts
Cat. no. 10478101 1093270 1100420 1190760 1210810 1210850 2003350 2004330 2004340 2304720 2304860
Item Carriage Tooth belt Transfer rail Tension limiting wire Fuse 0.6 A slow 5 x 20 (115 V) Fuse 0.3 A slow 5 x 20 (230 V) MUKEY PCB MUDROP-03 PCB MUBCD-02 PCB (384) Microswitch assembly Flat cable
1 2 unit(s) / year 1 4 1 1
1
10 units / year 1 4 2 6 1 1 1 1 3 1 1
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
74
Chapter 12: Glossary and Abbreviations
12 GLOSSARY AND ABBREVIATIONS
dispense
To distribute (pipette) liquid into the wells of the preselected strips or microplate.
dispensing cassette
The dispensing cassette contains eight individual tubes. The dispensing
cassette is detachable. Each reagent can have a separate dispensing cassette.
Alternatively, the same dispensing cassette can be used with several reagents
using the backflush feature. Refer to section 11.2 Dispensing cassettes.
The dispensing cassettes contain either Machined or Extruded covers. The
tubing length also varies from 40 cm to a selectable length of 50 to 2000 cm.
All dispensing cassettes can be autoclaved up to 50 times.
To avoid damaging the tubes, follow the dispensing cassette maintenance,
washing, autoclaving and storage instructions carefully. Refer to sections
6.8.4 After use, 7.2 How to wash the dispensing cassette and 7.3 How to
autoclave the dispensing cassette.
EIA
enzyme immunoassay
empty
The operation of discharging the contents of the tubing of fluid.
EN
European Norm
EU
European Union
FAQ
frequently asked questions
home position The plate carrier is positioned to the furthest right of the transfer rails (Fig. 4.1).
IEC
International Electrotechnical Commission
plate adapter
The Multidrop 384 requires a plate adapter for the plate carrier to function.
Two different plate adapters are included: the PLATE HEIGHT MAX. 15 MM plate
adapter is suitable for plates approximately 15 mm high and the PLATE HEIGHT
MAX. 11 MM plate adapter is suitable for lower plates.
Insert the plate adapter correctly. Refer to section 6.4 How to insert and
adjust the plate adapter.
PP
polypropylene
prime
The operation of filling a pump intake with fluid to expel the air.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
75
Chapter 12: Glossary and Abbreviations
12.1 Keywords for web pages
automated dispensers automated fluid dispensing automated liquid handling
dispensing drug discovery EIA ELISA enzyme immunoassay immunoassay liquid
handling microplate microplate dispenser Microtiter plate
microwell plate multidispense multidrop Multidrop Multidrop 384 multidrops plate robotic platform sandwich assay separation technique(s) solid phase assay(s) Thermo Fisher Scientific Thermo Scientific
12.2 Literature
Editorial (2004): Multidrop bulk reagent dispenser family from Thermo
Electron; precise dispensing in the microliter area allowing assay
miniaturization. Innovations in Pharmaceutical Technology 4 (14), 36.
Ennis, C. and Tanner, A. (Fall 1999): Analysis of CHO cell requirements for
assay miniaturization in high-throughput screening. The HTS Forum 8, 5 pages.
Falconer M. et al. (2002): High-Throughput Screening for Ion Channel
Modulators. Journal of Biomolecular Screening 7 (5), 460 465.
Hudson Control Group (Jan 2001): Integrating the Thermo Labsystems Multidrop
with the PlateCraneTM. IB104A, 3 pages.
Kiger A. A. et al. (2003): A functional genomic analysis of cell morphology
using RNA interference. Journal of Biology 2, 27.
Tanaka M. et al. (2005): An unbiased cell morphology-based screen for new,
biologically active small molecules. PLoS Biology 3 (5), e128.
Yarrow J. C. et al. (2004): A high-throughput cell migration assay using
scratch wound healing, a comparison of image-based readout methods. BMC
Biotechnology 4, 21.
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
76
Chapter 13: Index
13 INDEX
A
Abbreviations……………………………………………………………………………………………………………………….. 75 Accessories
………………………………………………………………………………………………………………. 15, 16, 73 Addresses
…………………………………………………………………………………………………………………………… 85 Advantages of using the
Multidrop 384……………………………………………………………………………………… 9 Application
…………………………………………………………………………………………………………… 5, 6, 9, 76, 84
B
Bottle ……………………………………………………………………………………………………………………………………. 9
C
Cable ………………………………………………………………………………….. 15, 17, 18, 19, 57, 61, 68, 73, 74, 81
Calibration ……………………………12, 24, 32, 33, 40, 41, 42, 43, 44, 46, 47, 51, 52, 53,
55, 60, 73, 81, 82
screw cover………………………………………………………………………………………. 12, 41, 42, 43, 44, 53, 55
Channels ……………………………………………………………………………………………………….. 40, 42, 43, 44, 81 Check
…………………………………………………………….15, 16, 17, 18, 19, 21, 23, 24, 32, 33, 39, 66, 69, 83
Clean ……………………………………………………………………………………….. 16, 34, 35, 36, 37, 46, 58, 66, 70
Col ………………………………………………………………………………………………… 14, 19, 30, 31, 33, 60, 64, 83
Columns …………………………………………………………………………………… 10, 14, 19, 30, 31, 33, 60, 64, 83
Consumables ……………………………………………………………………………………………………………. 32, 71, 73 Control
panel ………………………………………………………………….. 10, 19, 21, 30, 31, 36, 60, 68, 81, 82, 83
indicators …………………………………………………………………………………………………………………………. 30
keys…………………………………………………………………………………………………………………………………. 31 Cover retaining
screws………………………………………………………….. 11, 17, 18, 45, 47, 49, 53, 54, 81, 82
D
Dead volume ……………………………………………………………………………………………………………………….. 67 Decontamination
……………………………………………………………………………………………………….. 58, 69, 70
Certificate of…………………………………………………………………………………………………………… 58, 69, 70
Decontamination procedure ……………………………………………………………………………… 35, 57, 58, 69, 70
Dimensions………………………………………………………………………………………………………………………….. 60
Directives…………………………………………………………………………………………………………………………….. 62
Disclaimer……………………………………………………………………………………………………………………………… 5 Dispenser
………………………………………………………………………………………. 7, 8, 9, 14, 15, 19, 30, 35, 76 Dispensing.
7, 8, 9, 11, 12, 14, 16, 19, 21, 22, 23, 24, 25, 29, 30, 31, 32, 33, 34, 35,
36, 37, 40, 42, 43,
45, 47, 48, 49, 50, 52, 53, 54, 55, 56, 57, 58, 60, 64, 66, 67, 68, 72, 73,
75, 76, 81, 82, 83 accuracy …………………………………………………………………………………………………………………………..
60 cassette.. 7, 8, 9, 11, 12, 14, 16, 19, 21, 22, 23, 24, 25, 29, 34, 35, 36,
37, 40, 42, 43, 45, 47, 48, 49,
50, 52, 53, 54, 55, 56, 57, 58, 60, 66, 67, 68, 72, 73, 75, 81, 82, 83
precision ………………………………………………………………………………………………………………………….. 60 volume
……………………………………………………………………………………………………………………….. 60, 64 Disposal of
instrument…………………………………………………………………………………………………………… 58 Disposal of materials
…………………………………………………………………………………………………………….. 57
E
Empty ………………………………………………….19, 21, 31, 32, 33, 34, 35, 42, 43, 57, 60, 63, 64,
67, 75, 83 Environmental requirements …………………………………………………………………………………………………..
16 Error……………………………………………………………………………………………………. 19, 30, 31, 32, 35, 65, 66
code ………………………………………………………………………………………………………………………………… 65 Extruded covers
……………………………………………………………………………………………….. 8, 12, 72, 75, 81
F
FAQ ………………………………………………………………………………………………………………………………. 67, 75 Frequently asked
questions …………………………………………………………………………………………………… 67 Fuse holder
cap……………………………………………………………………………………………………. 13, 38, 39, 81
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
77
Chapter 13: Index
Fuses……………………………………………………………………………………………………….. 13, 38, 39, 66, 74, 81
G
Gravimetric calibration……………………………………………………………………………………………………… 40, 42
Guidelines for use ………………………………………………………………………………………………………………… 35
H
Hazard …………………………………………………………………………………………………………………………….. 6, 68 Home position
………………………………………………………………………………… 14, 19, 24, 31, 33, 35, 64, 75 How to
autoclave the dispensing cassette…………………………………………………………………….. 14, 37, 75 How
to change the tip band …………………………………………………………………………………………………… 45 How to
disinfect the instrument ………………………………………………………………………………. 35, 57, 69, 70 How to
dispense automatically……………………………………………………………………………………………….. 32 How to
dispense manually into the specified column(s) …………………………………………………………….. 33
How to dispense multiple volumes………………………………………………………………………………………….. 33 How to
fit the new tubing set ………………………………………………………………………………………………….. 46 How to
install the Multidrop 384 …………………………………………………………………… 15, 17, 19, 20, 21, 61
How to pack for service ………………………………………………………………………………………………. 57, 58, 69 How
to prime ………………………………………………………………………………………………………………….. 32, 43 How to
recalibrate the dispensing cassette ………………………………………………………………………… 40, 56 How to
replace the fuses……………………………………………………………………………………………………….. 38 How to select
the 96/384 well plate…………………………………………………………………………………………. 31 How to
shutdown…………………………………………………………………………………………….. 14, 34, 35, 37, 75 How to switch
on ………………………………………………………………………………………………………………….. 21 How to troubleshoot the
instrument ………………………………………………………………………………………… 66 How to use the control
panel…………………………………………………………………………………………….. 19, 30 How to wash the dispensing
cassette …………………………………………………………………………… 14, 37, 75
I
Indicators …………………………………………………………………………………………………………………………….. 30 Installation
………………………………………………………………………………….. 7, 15, 16, 17, 40, 42, 47, 61, 66
check list ………………………………………………………………………………………………………………………….. 15 Instrument
layout………………………………………………………………………………………………………………….. 10 Intended
use………………………………………………………………………………………………………………………….. 8
K
Keys………………………………………………………………………………………………. 14, 30, 31, 33, 35, 41, 60, 66
Keywords for web pages ……………………………………………………………………………………………………….. 76
L
LED……………………………………………………………………………………………………………….. 19, 21, 30, 35, 66 Literature
…………………………………………………………………………………………………………………………….. 76 Live parts
…………………………………………………………………………………………………………………………….. 61 Lower
part……………………………………………………………………………. 10, 12, 22, 23, 24, 45, 47, 48, 49, 81
M
Machined covers …………………………………………………………………………………………………………….. 72, 82 Mains
…………………………………………………..11, 13, 15, 17, 18, 19, 21, 35, 38, 39, 57, 60, 61, 66,
68, 81
switch ……………………………………………………………………………………………………. 11, 17, 18, 19, 21, 35 Maintain
………………………………………………………………………………………………………………. 21, 36, 58, 83 Maintenance
…………………………………………………………………….. 7, 14, 21, 35, 36, 58, 61, 68, 69, 75, 83
Microplate dispenser ………………………………………………………………………………….. 7, 8, 9, 14, 35, 76, 81
Microstrip …………………………………………………………………………………………………………. 5, 43, 44, 57, 81
Microswitch…………………………………………………………………………………………………………………….. 24, 74
Mode……………………………………………………………………………………………………………… 14, 31, 32, 33, 41 Multidrop
384….7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 24, 25, 30, 31, 32, 33, 35, 36, 40,
41, 42, 43, 61, 62,
63, 67, 69, 72, 73, 74, 75, 76, 81, 82, 83
O
Operating precautions and limitations……………………………………………………………………………………… 20
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
78
Chapter 13: Index
Operational check ……………………………………………………………………………………………………… 15, 19, 21
list……………………………………………………………………………………………………………………………………. 21
Ordering information……………………………………………………………………………………………………………….. 7
P
Packing…………………………………………………………………………………………………………………….. 15, 58, 69
instructions………………………………………………………………………………………………………………….. 15, 58
list……………………………………………………………………………………………………………………………………. 15 materials
………………………………………………………………………………………………………………………….. 58
Plate …. 7, 8, 9, 10, 11, 14, 19, 21, 22, 24, 25, 26, 27, 28, 29, 30, 31, 32,
33, 34, 35, 36, 40, 41, 42, 43, 44, 45, 50, 53, 55, 57, 60, 64, 66, 67, 73,
74, 75, 76, 81, 82, 83 adapter………………………………………..10, 11, 21, 25, 26, 27, 28, 29,
32, 33, 42, 43, 57, 73, 75, 81, 83 adapter lock
……………………………………………………………………………………………………… 10, 26, 27, 81 carrier………………………….10,
11, 14, 19, 21, 22, 24, 25, 26, 27, 28, 29, 30, 32, 33, 35, 36, 66, 75, 81
switch ……………………………………………………………………………………………………. 10, 19, 21, 31, 43, 64 type
…………………………………………………………………………………………………………………. 21, 31, 64, 83
Polypropylene tubing……………………………………………………………………………………………………….. 67, 75
Precalibration ……………………………………………………………………………………………………………. 55, 56, 82 Prime ….
10, 19, 21, 22, 24, 31, 32, 33, 34, 35, 37, 40, 41, 42, 43, 58, 60, 64, 65,
66, 67, 73, 75, 81, 83 Prime/drop …………………………………………………………………………………………… 19, 32,
33, 41, 43, 60, 83 Priming vessel ……………………………………………10, 19, 21, 22, 31, 32, 33,
40, 42, 65, 66, 67, 73, 81, 83 Protocol
………………………………………………………………………………………………………………………………. 69 Pump body
…………………………………………………………………………………………………………………….. 10, 23 Purge
………………………………………………………………………………………………………………………………….. 75
R
Remote control …………………………………………………………………………………………………………………. 9, 63 Rest
position………………………………………………………………………… 10, 21, 25, 34, 35, 37, 40, 43, 81, 83
Robotic ………………………………………………………………………………………………………………………. 9, 29, 76
Robotics…………………………………………………………………………………………………………………………. 29, 76 Rocker switch
………………………………………………………………………………………………………. 10, 21, 31, 43
Rotor……………………………………………………10, 11, 19, 21, 23, 24, 25, 29, 32, 34, 35, 40, 42,
66, 81, 83
cover ……………………………………………………………….. 10, 11, 19, 21, 24, 25, 29, 32, 40, 42, 66,
81, 83 needles …………………………………………………………………………………………………………………. 10, 24, 66 shaft
…………………………………………………………………………………………………………………………… 23, 24 Routine
………………………………………………………………………………………………………………………….. 21, 36 Run
…………………………………………………………………………………………………………………………………….. 83
S
Samples……………………………………………………………………………………………………………………. 16, 57, 68
Service……………………………………………………………………………………………………… 36, 37, 57, 58, 66, 69
contracts ………………………………………………………………………………………………………………………….. 58 pack for
…………………………………………………………………………………………………………………. 57, 58, 69 request protocol
………………………………………………………………………………………………………………… 69
Shutdown…………………………………………………………………………………………………………………………….. 35 Sight
strips………………………………………………………………………………………………………………… 12, 55, 56 Sight window
…………………………………………………………………………………………………………….. 50, 51, 81 Silicone tubing
…………………………………………………………………………………………………………… 32, 47, 67
Slots……………………………………………………………………………………………………………………. 22, 32, 65, 66 Software
………………………………………………………………………………………………………………………….. 7, 64 Spare parts
…………………………………………………………………………………………………………………….. 73, 74
recommended…………………………………………………………………………………………………………………… 74 Specifications
………………………………………………………………………………………………………… 7, 29, 60, 61 general
…………………………………………………………………………………………………………………………….. 60 safety
………………………………………………………………………………………………………………………………. 61 Startup
……………………………………………………………………………………………………………………… 18, 21, 64 Step
………………………………………………………………………………………………. 14, 19, 31, 33, 35, 41, 46, 60 Switching
on ………………………………………………………………………………………………………………………… 18
Symbols………………………………………………………………………………………………………………………………… 6
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
79
Chapter 13: Index
T
Table of Figures …………………………………………………………………………………………………………………… 81 Tension
limiting wires ………………………………………………………………………. 12, 23, 24, 47, 48, 52, 74, 81
Thumbwheels ………………………………………………………………………………………………………………………. 33
Tip…………………………………………………………………………………………………. 40, 45, 46, 47, 66, 67, 73, 81
band …………………………………………………………………………………………………………… 45, 46, 47, 67, 81 manifold
…………………………………………………………………………………………………………… 45, 47, 73, 81 Transfer rails
…………………………………………………………………………………………………………….. 10, 74, 75 Transport
……………………………………………………………………………………………………….. 15, 16, 58, 60, 69 damage
………………………………………………………………………………………………………………………. 15, 16 package
………………………………………………………………………………………………………………… 15, 16, 58
Troubleshooting………………………………………………………………………………………………………………. 66, 69 guide
……………………………………………………………………………………………………………………………….. 66
Tubes……………………………………………………………..32, 33, 40, 42, 43, 45, 47, 48, 51, 55, 56, 66,
73, 81 Tubing……………………8, 12, 19, 25, 32, 33, 34, 35, 45, 46, 47, 48, 50, 51,
52, 56, 67, 72, 73, 75, 81, 82 identification plate
………………………………………………………………………………………………………… 50, 81 set
…………………………………………………………………………………….. 8, 25, 45, 46, 47, 56, 67, 72, 73, 81
weight…………………………………………………………………………………………. 12, 32, 33, 46, 47, 56, 73, 82
U
Unpack ……………………………………………………………………………………………………………………………….. 15 Upper
part……………………….10, 12, 22, 23, 24, 34, 35, 40, 41, 42, 43, 47, 49, 50, 52, 53,
54, 55, 81, 82 User manual ……………………………………………………………………………………………………………….. 7, 15,
73
V
View……………………………………………………………………………………………………………………. 10, 11, 13, 14
front………………………………………………………………………………………………………………………. 10, 11, 81 internal
…………………………………………………………………………………………………………………………….. 14
rear…………………………………………………………………………………………………………………………….. 13, 81
Voltage …………………………………………………………………………………………………….. 16, 17, 20, 62, 68, 81
selector ………………………………………………………………………………………………………………………. 17, 81
Volumetric calibration …………………………………………………………………………………. 32, 33, 40, 41, 60, 73
W
Warnings ………………………………………………………………………………………………………………….. 16, 17, 18 Warranty
…………………………………………………………………………………………………………………….. 7, 15, 71 Weight
……………………………………………………………………………………… 12, 16, 32, 33, 46, 47, 56, 60, 73
Thermo Scientific Multidrop® 384 User Manual; Rev. 3.4, Cat. no. 1507010
80
Chapter 14: Table of Figures
14 TABLE OF FIGURES
FIG. 3.1 MULTIDROP 384 MICROPLATE DISPENSER ………………………………………………………………………….. 8
FIG. 4.1 MULTIDROP 384 FRONT VIEW (A)……………………………………………………………………………………. 10 FIG.
4.2 MULTIDROP 384 FRONT VIEW (B)……………………………………………………………………………………. 11 FIG. 4.3
MULTIDROP 384 FRONT VIEW (C) ……………………..
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>