Thermo Scientific HERATHERM Service Software Owner’s Manual

June 8, 2024
Thermo Scientific

Thermo Scientific logo HERATHERM Service Software
Owner’s Manual

HERATHERM Service Software

Instructions to use the Phoenix program to disable voltage test

Application:

The intent of this instruction is to disable the voltage self-test in the Heratherm ovens and incubators.
The voltage test is a safety function that protects the unit from inadvertent connection to the wrong power source where the device would produce the E014 error code. In some cases, harmonic distortion and/or line noise can cause the E014 error when the supply voltage is correct causing nuisance alarms. This alarm can be mitigated by disabling the voltage self-test. Once the test is disabled it cannot be reenabled without replacing the main control board.
If the incorrect voltage is connected after the test is disabled:
After disabling the test, you must never connect a 120VAC unit to a 240VAC source. Connecting a 120VAC unit to a 240VAC source will cause damage to components.
In the case of connecting a 240VAC unit to a 120VAC source you will likely get poor performance from the device.
Prior to going onsite:
The technician must obtain a USB serial adapter cable to connect between a Laptop/Tablet and the DB9-pin communication socket on rear of equipment.
We recommend USB to Serial Part Number 350185, which is a TRIPPLITE U209-000-R. The Driver for this USB cable Must be loaded into the Laptop/Tablet prior to going on site.
Download the Phoenix service software into the Laptop/Tablet prior to performing this service.
See the link below to download the Phoenix service software and Overview.
Knowledge Base Article 348716 – Heratherm Service Software
At the device
With the unit connect to t mains power turn the unit on by pressing and holding the power button.
Allow the device to go through the self-test; If the E014 is present you can silence the alarm and proceed with the process.
Connect the RS232 to USB adapter to the RS232 port on the back on the unit and the USB port on your Laptop/TabletThermo Scientific HERATHERM Service
Software - Figure 1Instructions for navigating the Phoenix program:
Start by opening the Phoenix program. The program will open to the MAIN tab.
You will get the System – Message File Phoenix.cft does not exist.
Simply clear the message by clicking OK. Thermo Scientific HERATHERM Service
Software - Figure 2Select the GENERAL tab and you will go the PRESETTING tab. If there is no equipment connected you will get some interface errors. Adjust the Baud rate to 57600 and select the correct com port for your set up, use device manager to determine the COM port you are connected to. Select change and you should recive a message stating communications OK. Thermo Scientific HERATHERM Service Software - Figure
3Navigate to the PASSWORD tab then enter the Service password – elektronik. After clicking the ENTER button you will see Password o.k!!! Thermo Scientific HERATHERM Service Software -
Figure 4Navigate to the WATCH ALL VALUES and ensure that the Temp reading is live. Next navigate to the ERROR LOGGER tab and review existing errors. This allows the program to sustain live communication with the Equipment.
You will then go to the TEST COM tab. First clear the error list with request: ! and address: 3300. Thermo Scientific HERATHERM Service Software - Figure
8After clicking send you should receive confirmation the string was sent.Thermo Scientific HERATHERM Service
Software - Figure 5Next disable the voltage test using request: ! , address: 3220 , attachment: ff. Thermo Scientific
HERATHERM Service Software - Figure 7After clicking send you should receive confirmation the string was sent. Thermo Scientific HERATHERM Service Software - Figure
6Next power the unit down using the ON/OFF button on the display then disconnect from the power source to fully remove power from the board. After waiting 30 seconds reconnect to the power source and power the unit up using the ON/OFF button. The unit should cycle through its normal self-test without the E014 message returning.

Thermo Scientific logoRevision B 9/21/2021