ThermoWorks ThermaData Lite Temperature Loggers Instruction Manual
- June 7, 2024
- ThermoWorks
Table of Contents
ThermoWorks ThermaData Lite Temperature Loggers
Model
ThermaData Lite - White
ThermaData Lite - Black
ThermaData Lite - Yellow
ThermaData Lite - Blue
Order Code
293-010
293-070
293-020
293-050
SPECIFICATIONS
Range 40 to 185°F (40 to 85°C)
Accuracy ±0.9°F (±0.5°C)
Resolution 0.1°F or °C
Memory 16,000 readings, stop when full.
Sample Rate Settable from 6 seconds to 4 hours 15 minutes (255 min)
Sensor Type Thermistor
Battery 3.6V 1/2 AA Lithium / 3 year life typical
Dimensions 2.16 dia. X 1 D inches (55 dia. x 25 D mm), plus eyelet tab
Weight 1.4oz (40g)
ACCESSORIES/RELATED ITEMS
USB-C Interface Cable (812-510)
Waterproof Pouch (830-360)
3.6V 1/2AA Lithium Battery (THS-839-102)
GUARANTEE – This instrument carries a two-year warranty against defects in either components or workmanship. During this period, products that prove to be defective will, at the discretion of ThermoWorks, be either repaired or replaced without charge. This warranty does not apply to probes, where a six- month period is offered. Full details of liability are available within ThermoWorks’ Terms & Conditions of Sale at www.thermoworks.com/product- warranty.
For warranty, service, and technical assistance, please contact ThermoWorks’ Technical Support at 801-756-7705 or email at techsupport@thermoworks.com.
Made in the UK
Operating Instructions
Your new ThermaData Lite logger uses advanced measurement technology for precision temperature recording. Data is collected within the unit’s onboard memory and a PDF report can be downloaded via USB. Settings and more in-depth data analysis can be adjusted via ThermaData Studio for Windows.
INSTALL SOFTWARE – THERMADATA STUDIO
This data logger utilizes a USB-C to USB-A connection cable, which is
purchased separately, to interface with your PC. Along with the hardware,
you’ll need to download and install the free software, ThermaData Studio, from
www.ThermoWorks.com/software. Once you have the software installed, open and
running, and the USB cable connected, you’re ready to plug in your data logger
and start working.
NOTE: ThermaData Studio software is updated regularly. Please visitwww.ThermoWorks.com/software to check for any recent updates.
**CONNECT AND START YOUR LOGGER***
Once connected to the computer with the USB cable, you can begin setting up
your logger. To begin, run ThermaData Studio software and open the ThermaData
tab and click on Instrument Settings. Select your ThermaData Lite from the
drop-down menu, if prompted. From this screen, you can name your logger, set
your temperature alarm limits, change your recording interval, add
instructions to be included in the PDF report, set up a start delay or timed
start, as well as instructing what to do when the memory is full. If you’re
ready to start collecting data, click Send Settings. A message will inform you
that proceeding will clear all data currently on the device. Click OK to
proceed and wait until the software tells you it’s safe to unplug the logger.
When you’re ready to start logging simply press the button on the front of the
logger. It will immediately start recording or the logger will start the delay
you set up in the software.
**RETRIEVE YOUR DATA***
When you’re ready to retrieve your data, connect the logger to the PC using
the USB-C cable. The logger will stop recording and show up in your file
explorer as a Mass Storage Device. Saved on the logger will be a PDF report of
the logging activity. For a more in-depth data analysis, leave the logger
connected to the computer and open ThermaData Studio. On the Home, Tab clicks
Download to retrieve all data points from the logger. Be sure to use the Save
option on the Home Tab to save the data in ThermaData Studio.
STOP AND RESTART YOUR LOGGER
You can manually stop ThermaData Lite by pressing and holding the button on
the front of the logger for 5 seconds. The Red LED will light up for 4
seconds, indicating the logger is stopped. To restart your logger after it has
been stopped or you have downloaded the data, follow the steps in the section
titled CONNECT AND START YOUR LOGGER.
BATTERIES
This data logger comes with a non-rechargeable, preinstalled 1/2 AA 3.6V
lithium battery. When the battery needs to be changed, open the snap-fit cover
using a coin in the slot provided. Replace the battery with a 1/2 AA 3.6V
lithium battery (THS-839-102) available at www.ThermoWorks.com. Failure to use
the same type of battery may result in decreased battery life.
LED LIGHT SEQUENCES
Red & Green together 4 times:
Logger is starting.
Red & Green alternating 2 times:
Delay started.
Red 3 times:
The temperature has exceeded the limits.
Green 3 times:
The temperature has stayed within limits.
Red on for 4 seconds 1 time:
Stop Logging.
Red on for 1 second 2 times:
Logger is stopped.
Red flashing continuously:
Connected to USB.
*Additional details and full operating instructions can be found in the ThermaData Studio software help file.
Documents / Resources
|
ThermoWorks ThermaData Lite Temperature
Loggers
[pdf] Instruction Manual
ThermaData Lite Temperature Loggers, ThermaData Lite White 293-010, ThermaData
Lite Black 293-070, ThermaData Lite Yellow 293-020, ThermaData Lite Blue
293-050
---|---
References
- ThermoWorks | Professional Thermometers from the Temperature Experts
- Professional Thermometers from the Temperature Experts | ThermoWorks
- ThermoWorks Warranty
- Software Packages
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>