ThermoWorks PT-100 Precision & Reference Thermometers Instruction Manual

June 7, 2024
ThermoWorks

ThermoWorks PT-100 Precision & Reference Thermometers Instruction Manual
ThermoWorks PT-100 Precision & Reference
Thermometers

MODEL

| ORDER CODE
---|---
Precision|

222-553

Precision Plus

| 222-550
Reference|

222-555

SPECIFICATIONS

Precision / Precision Plus Thermometers

  • Range: –328 to 1562°F (–200 to 850°C)
  • Accuracy (Precision): ±0.4°F (±0.2°C) meter only
  • Resolution (Precision): 0.1°
  • Sensor (Precision): 3 or 4-wire RTD Class A, 100ohms
  • Accuracy (Precision Plus): ±0.1°F (±0.05°C) meter only
  • Resolution (Precision Plus): 0.01° Auto-Ranging to 0.1° above 199.9°
  • Sensor (Precision Plus): 3 or 4-wire RTD 1/10 DIN, 100 ohms
  • Probe: Interchangeable:
  • Auto-Off: After 10 min. to save battery life
  • Battery: 3 x AAA, 2,000 hrs.
  • Operating Range: –4 to 122°F (–20 to 50°C)
  • Certificate: NIST-traceable certificate included

Reference Thermometer

  • Range: –328 to 400°F (–200 to 200°C)
  • Accuracy: ±0.05°F (±0.03°C)
  • Resolution: 0.01°
  • Probe: 4W RTD 1/10 DIN, fixed
  • Battery: 3 x AAA, 2,000 hrs.
  • Operating Range: –4 to 122°F (–20 to 50°C)
  • Certificate: UKAS Accredited Certificate included

ACCESSORIES/RELATED ITEMS

  • Standard Protective Boot (THS-830-227)

  • Rubber Boot w/ Probe Holder (THS-830-255)
    Rubber Boot w/ Probe Holder

  • SST Wall Bracket w/Probe Holder ( THS-832-050)

  • Duracell Procell AAA Battery, 4-pack (PC2400)

INSTRUMENT OPERATION

Connect a probe via the socket located on the top of the instrument (where applicable). Screw the collar of the plug, finger-tight to ensure a secure connection (do not over tighten). The instrument is switched on by pressing the ON/OFF button. Apply the tip of the probe to the substance, medium, or surface to be  measured. The instrument reading may take a few seconds to stabilize, depending on the nature of the measurement and sensitivity of the probe.

HOLD (PRECISION, PRECISION PLUS ONLY)

Press the HOLD button to freeze the display. Press again to continue measuring.

°C/°F – Press and hold the ON/OFF button for 5 seconds upon start up and the displayed temperature will change from °F to °C.

MAX/MIN (PRECISION, PRECISION PLUS ONLY)

Press the MAX/MIN button to display the values.
The instrument retains the max/min values when switched off.To reset, press the MAX/MIN button for three seconds, ‘rst’ will be displayed.

TRIM (PRECISION, PRECISION PLUS ONLY)

Press the MAX/MIN and ON/OFF buttons at start-up for five seconds to enter (Offset is displayed). Press HOLD (up) or MAX/MIN (down) to set the trim value (maximum of ±3.6°F). Press the ON/OFF button to save the value and exit Trim mode.

AUTO-OFF

The unit will switch off automatically after 10 minutes. To disable auto off press the HOLD button while switching on the instrument (this is reset when the instrument is switched off).

OPEN CIRCUIT/FAULTY PROBE

If the probe has developed an ‘Open Circuit’ fault the instrument will display ‘Err’

SENSORS

This instrument should only be used with PT100 (platinum resistance probes, 0°C = 100 ohms) BS EN 60751 fitted with suitable miniature plug (where applicable).

BATTERY REPLACEMENT

The instrument will display ‘LO BAT’ to warn that the batteries will need changing soon. The instrument continues to measure accurately but after further usage the instrument will display ‘flat bat’ and shutdown.
Replace the batteries with three AAA batteries located under the rear cover that is held in place by a single screw.

INSTRUMENT CLEANING

Clean the instrument regularly with an anti-bacterial probe wipe to avoid potential food-borne bacteria growth.

WARNING

IPA and other solvents may cause damage to the case and screen of this instrument.

GUARANTEE

This instrument carries a two-year warranty against defects in either components
or workmanship. During this period, products that prove to be defective will, at the discretion of ThermoWorks, be either repaired or replaced without charge. This warranty does not apply to probes, where a six-month period is offered. Full details of liability are available within ThermoWorks’ Terms & Conditions of Sale at www.thermoworks.com/product- warranty

For warranty, service, and technical assistance, please contact ThermoWorks’ Technical Support at 801-756-7705 or email at techsupport@thermoworks.com.

Probe Wiring Diagram
PT100 Binder Plug Wiring
Precision / Precision Plus Thermometers

Pin

| Function
---|---
1|

2

| +
3|

– (3 Wire)

4

| – (4 Wire)
5|

Shield

  • Pin numbering as viewed from rear
    Pin numbering as viewed from rear

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

ThermoWorks User Manuals

Related Manuals