Aerpro UFR105 Infrared Forehead Thermometer Instruction Manual

June 6, 2024
Aerpro

UFR105 Infrared Forehead Thermometer

thermometerAerpro APIRT04
Infrared Forehead Thermometer
Please read the manual carefully before you use the unit, and keep for future reference.

ARTG Ref No: 333301
Model:UFR105

▲Your new Infrared forehead thermometer uses advanced infrared (IR) technology to measure temperature instantly and accurately on the forehead or object.
▲Easy to use and less measurement time
This thermometer does not need to contact body or object to ensure the safety and hygiene, Its ergonomic design makes this thermometer be simple and very easy to check the temperature. It only takes 1 second to take measurement and reading.
▲Body and object modes
This thermometer supports tomeasure temperature of body and object: Measure
ing range of object mode is 0.0 to 100.0°C (32.0°F~199.9°F ). That means except
to take body temperature, it also supports to take the surface temperature of
object as below,
Surface temperature of milk in the bottle;
Surface temperature of baby bath;
Environment temperature.
▲Alarm sounds
When body temperature is over 37.5°C, an alarm sounds for 10 times,
▲Memory function
It stores up to 20 sets recent measurement data.
▲Auto power off
Automatically power off if left idle for Approx 6020s,
▲Measurement
Measurement time interval in 1 second and the measurement distance is within
1~3cm,
Please read the manual carefully before you use the unit, and keep for future
referene.
▲lntended use
The Infrared Thermometer is intended for the intermittent measurement and
monitoring of human body temperature from forehead or object, The device is
indicated for use by people of all ages at homecare and in hospital,

Safety Information

To assure the correct use of the product, basic safety measures should always be followed including the warning and the caution listed in the instruction manual :

Symbol descriptions

The following symbols may appear in this manual, on the label,on the device, or on it’s accessories. Some of the symbols represent standards and compliances associated with
the device and its use.

As a Object Thermometer

object moves1). Press the START/STOP button, all symbols appear on the
display, you can hear 1 short beep, then unit °C/°F on the
display start to flash,
2). Press the Mode button to select Object Mode, the tempera
ture unit flashes.
3). Move the probe close to the object, make sure the probe is
flat and take measurement with a distance between 1~3cm,
Press the START/STOP button, The meter will take measure-
ment.
4), The measurement will be finished in 1 second, when it has
been completed, you can hear 1 short beep, in the meantime,
the reading will appears on the display.

Note:

  • As the object scan temperature is likely to be affected by the surroundings, the reading shall be taken as a reference only.
  • If the probe is placed at an angle close the object scan, the reading will be affected by surrounding temperature.
  • Take care Polished metal like a stainless steel kettle or saucepan will reflect the surrounding temperature of the room rather than their own emitted temperature. You may get a reading of 38°c (100°F) on the side of a pot on the stove and think it is safe to touch but you will burn your fingers as it could be boiling at 100° (212°F) The same is true for very cold materials as well so please take care when using this device on reflective surfaces”

Memory-recall of measurements

This Non-contact forehead thermometer automatically stores 20 sets measurements value, the oldest record will be replaced by the latest measurement value when more than 20 sets. Read memory record

Press the °C/°F Button to enter the memory mode when power off. Each time you
press the Mode button, a number (from 1 to 20) will be displayed along with the
symbol (MJ, 1 second later, the measurement will be shown, as well as subse-
quent measurements can be display one after the other by pressing the °C/°F
Button each time.

Note:

The reading number “1” is the latest measurement, “20”is the oldest measurement. When the oldest measurement is read, pressing the °C/°F Button again will read the reading “1” again.
WARNING: This alert identifies hazards that may cause serious personal

-Body

injury or death

First Choice

CAUTION: This alert identifies hazards that may cause minor personal injury, product damage, or property damage.

Celsius/°C

Fahrenheit/°F

MODEO

.Type BF applied part

Infrared Forehead Thermometer Instruction manual

  1. Auto-sense Probe
  2. Display Screen
  3. Mode Button
  4. °C/°F Button
  5. START/STOP Button 6.Battery cover

Proper Use Of The Unit

Manufacturer

DU!

SN

Specifies serial number

EC REP

Authorized Representative in the European Community CE Mark: conforms to essential requirements of the Medical Device Directive 93/42/EEC.

DISPOSAL: Do not dispose this product as unsorted municipal waste. Collection of such waste separately for special treatment is necessary.

Direct current

Follow instructions for use

Pre-measurement About Normal Body Temperature&Fever The temperature in the forehead and temple area differs from the internal temperature, which is taken orally or rectally. Vasoconstriction, an effect which constricts the blood vessels and cools the skin, can occur during the early stages of a fever. In this case, the temperature measured by the Infrared thermometer may be unusually low. If the measurement therefore does not match the patient’s own perception or is unusually low, re peat the measurement every 15 minutes. As a reference, you can also measure the internal body temperature using a conventional oral or rectal thermometer. Body temperature can vary from one individual/person to next. It also varies by location on the body and time of day. Below shows the statistical normal ranges from different sites. Please keep in mind that temperatures measured from different sites, even at the same time, should not be directly compared. Fever indicates that the body temperature is higher than normal. This symptom may be caused by infection, overdressing or immunization. Some people may not experience fever even when they are ill.

These include, but are not limited to, infants younger than 3 months old, persons with compromised immune systems, persons taking antibiotics, steroids or antipyretics (aspirin, ibuprofen, acetaminophen), or persons with certain chronic illnesses. Please consult your physician when you feel ill even if you do not have fever.

RONA

Table*1 Body Site Normal Temperature Range

Normal Temperature Range

Body Site –

Oral

0.6°C(T°F) or more above or below 37°C(98.6°F) 0.3°C to 0.6 °C(0.5°F to 1°F) higher than oral temperature

Rectal/ear

Axillary (armpit)

0.3°C to 0.6 (0.5% to 1°F)

lower than oral temperature Note: Body Temperature at WebMD: website: http://firstaid.webmd.com

CAUTION: Consult accompanying documents

Battery Installation A This thermometer is not intended to substitute for a consultation with your Remove the battery cover from the battery compartment, insert the battery.

physician. The forehead scan temperature serves as a reference only. It

1. Remove the battery cover from the battery compart cannot be a judgment on fever.

ment as the arrow direction accordingly. Introduction

A Basic safety precautions should always be observed, especially when the

2. Insert 2 AAA powerful batteries into the compartment

thermometer is used on or near children and disabled persons. Your new Infrared forehead thermometer uses advanced infrared (IR) technol- .

and ensure each battery is in the proper direction, A Please place the device unreachable by young.

Positive (+) and Negative (-) are displayed on the back ogy to measure temperature instantly and accurately on the forehead or object. I

AAvoid direct sunlight.

of battery cover. A Easy to use and less measurement time

ADo not touch the lens.

3. Close the battery cover. This thermometer does not need to contact body or object to ensure the safety |

ANo modification of this device is allowed. and hygiene. Its ergonomic design makes this thermometer be simple and very easy to check the temperature. It only takes 1 second to take measurement and I

AThe swallowing of small parts like packing bag, battery, battery cover and so reading.

on may cause the suffocation. .

Low battery and replacement A Body and object modes

T A Please do not use a dilution agent, alcohol or petrol to clean the unit. Please – + -Atreat is gently and prevent the falling from a high place.

When power on, the low battery symbol will display once the unit start to This thermometer supports to measure temperature of body and object. Measur– + –

– – – – –

work, and you must replace with new batteries, otherwise the unit can’t work. ing range of object mode is 0.0 to 100.0°C (32.0°F~199.9°F ). That means except

A Please do not immersed it in liquid. to take body temperature, it also supports to take the surface temperature of

Never leave battery in the battery compartment for a long time without use, as I Battery type and replacement object as below,

they may leak and cause damage to the unit.

Please use 2 x AAA identical 1.5V alkaline batteries. Surface temperature of milk in the bottle;

T A Please take off the battery if you do not intend to use within 3 months. Replace

Do not use the batteries beyond their expiry date. Surface temperature of baby bath;

w ith new batteries if the unit display a low battery symbol.

Please remove the batteries if you do not need to use for long time. Environment temperature.

ADo not mix the old and new batteries together. AAlarm sounds

ADo not use during the transportation.

WARNING

Dispose of the battery in accordance with all federal, state and local laws. To When body temperature is over 37.5oC, an alarm sounds for 10 times.

A WARNING:

avoid fire and explosion hazard, do not burn or incinerate the battery. A Memory function

Do not dispose of electrical appliances as unsorted municipal waste, use | It stores up to 20 sets recent measurement data.

separate collection facilities. Contact you local government for information | AAuto power off

regarding the collection systems available. If electrical appliances are i Setting Mode Automatically power off if left idle for Approx 60+20s.

disposed of in landfills or dumps, hazardous substances can leak into the A Measurement

groundwater and get into the food chain, damaging your health and well-be- I How To Set Measurement time interval in 1 second and the measurement distance is within

1-3cm.

1. Mode setting:

Classification Please read the manual carefully before you use the unit, and keep for future

Press START/STOP Button, the LCD backlight will be on. After the display all

  1. Internally powered equipment;

shows for one second, the measurement preparation screen is displayed. Press reference.

MGA Button Seart the houditorohet me gurament mode ancianter 2. Type BF applied part; .

– Mode Button to select the body or object measurement mode and enter the Alntended use :

I temperature measurement mode.

3. Protection against ingress of water or Particulate matter: IP21; The Infrared Thermometer is intended for the intermittent measurement and I monitoring of human body temperature from forehead or object. The device is

4. Not category APIAPG equipment; indicated for use by people of all ages at homecare and in hospital.

5. Mode of operation: Continuous operation.

I The user must check that the equipment functions safely and see that it is in proper

working condition before being used. Safety Information

Body mode

Object mode Product Structure

Note: The Body Mode is used to measure the forehead temperature, then the Object Mode To assure the correct use of the product, basic safety measures should always is used to measure the object temperature. be followed including the warning and the caution listed in the instruction -Display

1 manual :

2

1.Body temperature indicator

2. Unit setting: 2.Object temperature indicator

In shutdown state, press °C/°F Button to enter switch temperature unit (°C/°F), Symbol descriptions

3.Memory mode

and press the °C/°F Button to select the unit. After the selection is completed, the The following symbols may appear in this manual, on the label,on the device, or on it’s

4. Temperature reading

current temperature unit is automatically confirmed. accessories. Some of the symbols represent standards and compliances associated with

5.Temperature unit the device and its use.

6. Low battery indicator

As a Body Thermometer 1).Press the START/STOP button, all symbols appear on the display, you can hear 1 short beep, then unit °C/°F on the display start to flash. 2). Press the Mode button to select Body Mode ,the tempera ture unit flashes. 3).Move the probe close to the forehead make sure the probe is flat and take measurement with a distance between 1~3cm. Press the START/STOP button, The meter will take measurement.

4). The measurement will be finished in 1 second, when it has been completed you can hear 1 short beep, in the meantime, the reading will appears on the display.

Note: off the reading is237 .5°C(99.5°F)and<43°C(109.4°F), the display will show together with ten short beeps.

•Only after the unit °C/°F start to flash again, it can continue to take measure ment.

•As the forehead measurement temperature is likely to be affected by sweat, oil and the surroundings, the reading shall be taken as a reference only. olf the probe is placed at an angle close to the forehead measurement, the reading will be affected by surrounding temperature.

•Babies’ skin reacts very quickly in the ambient temperature. Therefore, do not take their temperature with the non-contact thermometer during/after breastfeed ing, because the skin temperature maybe lower than the internal body tempera ture.

I

MODE TOO8

Table 2

Low battery, replace all the wom batteries with new ones.

Guidance and manufacturer’s declaration – electromagnetic Immunity

Erp

Thermometer system fails or affected by electric magnetic field

Immunity Test

IEC 60601-1-2 Test level

Compliance level

As a Object Thermometer 1). Press the START/STOP button, all symbols appear on the display, you can hear 1 short beep, then unit °C/°F on the display start to flash. 2). Press the Mode button to select Object Mode, the tempera ture unit flashes. 3). Move the probe close to the object, make sure the probe is flat and take measurement with a distance between 1-3cm. Press the START/STOP button, The meter will take measure ment.

4). The measurement will be finished in 1 second, when it has been completed, you can hear 1 short beep, in the meantime, the reading will appears on the display.

A Please contact the distributor if you can not solve the problem, do not disassemble the unit

by yourself!

Electrostatic discharge (ESD) IEC 61000-4-2

+8 kV contact +2 kV, +4 kV, 18 kV, 115 kV air

+8 kV contact +2 kV, +4 kV, +8 kV, 215 kV air

Care And Maintenance

N/A

Warranty Information

A The unit is guaranteed to be free of defects in workmanship and materials under normal use for a period of 1 Year from the date of purchase.

A For repair under this warranty. Our authorized service agent must be advised of the fault within the period of the warranty. This warranty only covers parts and labor service under normal operations. Any defect resulting from natural causes, eg. flood, hurricane etc, is not covered in this guarantee. This guarantee also does not cover damage incurred by use of the unit not in accordance with the instructions, accidental damage, or being tampered or serviced by unauthorized service agents.

A The following will be excluded from this warranty-if the thermometer has been misused, abused, or neglect in following the manual’s instructions on purpose

and unauthorized repair or modifications.

A The device requires no calibration.

The device is not repairable and contains no user serviceable parts.

Electrical fast transient/burst IEC 61000-4-4

Power supply lines: +2 kV input/output lines: +1 kV

NA

Surge IEC 61000-4-5

line(s) to line(s): +1 kV. line(s) to earth: +2 kV. 100 kHz repetition frequency

Care for the main unit

Keep the unit in the storage case when not in use.

•Clean the unit with a soft dry cloth. Do not use any abrasive or volatile cleaners.

•Never immerse the unit or any of its’ component in water. Maintenance

NA

Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11

0% 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 3150 0% 1 cycle And 70% 25/30 cycles Single phase: at 0 0% 300 cycle

Do not clean the unit with naphtha, thinner or gasoline etc.

Store the unit in a clean and dry location. Do not subject the unit to extreme hot or cold temperature, humidity and direct sunlight.

Note:

•As the object scan temperature is likely to be affected by the surroundings, the reading shall be taken as a reference only.

• If the probe is placed at an angle close the object scan, the reading will be affected by surrounding temperature.

• Take care “Polished metals like a stainless steel kettle or saucepan will reflect the surrounding temperature of the room rather than their own emitted temperature. You may get a reading of 38°c (100°F) on the side of a pot on the stove and think it is safe to touch but you will burn your fingers as it could be boiling at 100o (212°F) The same is true for very cold materials as well so

please take care when using this device on reflective surfaces”

Power frequency magnetic fiald FC 61000-4-8

30 A/m 50Hz/60Hz

30 A/m 50Hz/60HZ

t

EMC

N/A

Thin

Conduced RF IEC61000-4-6

150KHz to 80MHz: 3Vrms 6 Vrms (in ISM and amateur radio bands)80% Am at 1kHz

Naphtha

Radiated RF IEC61000-4-3

Remove the batteries if the unit will not be used in 3 months or longer.

Do not use the unit under the influence of electromagnetic interference (nearby cellphones, microwave etc.)

10 V/m 80 MHz – 2,7 GHz 80% AM at 1 kHz

10 V/m 80 MHz – 2,7 GHz 80 % AM at 1 kHz

I

Memory-recall of measurements

NOTE UT is the AC mains voltage prior to application of the test level.

Table 3

Guidance and manufacturer’s declaration – electromagnetic Immunity

Service

Band (MHz)

·

Modulation Modulation | Distance IMMUNITY

(m) TEST

LEVEL(V/m)

Test

Frequency Radiated RF

(MHz) IEC61000-4

385

TETRA 400

1,8

0.3

27

Note: We will not be responsible for any quality problem if you do not care and maintain the product as instructed.

380 – 390

This Non-contact forehead thermometer automatically stores 20 sets measure ments value, the oldest record will be replaced by the latest measurement value when more than 20 sets. Read memory record Press the °C/°F Button to enter the memory mode when power off. Each time you press the Mode button, a number (from 1 to 20) will be displayed along with the symbol M, 1 second later, the measurement will be shown, as well as subse quent measurements can be display one after the other by pressing the °C/°F Button each time. Note:

The reading number “1” is the latest measurement, “20”is the oldest measure ment. When the oldest measurement is read, pressing the °C/°F Button again will

read the reading “1” again.

– – – – –

– – – – – – – – – – – – – Mi ☺ M2 M2

Pulse modulation

18 Hz

2

380 – 390

0.3

28

GMRS 460, FRS 460

l

Specification

(Test specifica tions for ENCLO SURE – PORT

710

+ +

Description –

704 – 707 —

LTE Band 13, 17— —

IEC 60601-1-2: 2014 ME EQUIPMENT and ME SYSTEMS identification, marking and documents for Class B product Instructions for use

The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environ- ments and so on. Warning: Don’t use near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high. Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.” _ -. Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Infra-red Ear Thermometer (TE-66, TE-68, TE-82), including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. If any: a list of all cables and maximum lengths of cables (if applicable), transducers and other ACCESSORIES that are replaceable by the RESPONSI BLE ORGANIZATION and that are likely to affect compliance of the ME EQUIPMENT or ME SYSTEM with the requirements of Clause 7 (EMIS SIONS) and Clause 8 (IMMUNITY). ACCESSORIES may be specified either generically (e.g. shielded cable, load impedance) or specifically (e.g. by MANU FACTURER and EQUIPMENT OR TYPE REFERENCE). If any: the performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a description of what the OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost or degraded due to EM DISTURBANCES (the defined term “ESSENTIAL PERFORMANCE” need not be used).

Non-contact infrared Forehead Thermometer –

LCD digital display

FM # 5 kHz deviation

1 kHz sine Pulse modulation 217 Hz

Pulse modulation

18 Hz

745 780

Display

2

0.3

28

IMMUNITY

810

800 – 960

Measuring localization

365

361

to

Measurement range

870

Forehead and object surface Body mode 32.0°C-43.0°C(89.6°F-109.4°F); Object mode0.0-100.0°C(32.0°F-199.9°F); °C/°F

GSM 800/900, TETRA 800, IDEN 820, CDMA 850, LTE Band 5

RF wireless

930

Temperature unit

communica

1720

Display resolution

0.3

0.1°C/0.1°F

28

2

1 700 – 1 990

tions

Pulse modulation 217 Hz

Memory-clear Of Measurements

Accuracy

+0.2°C/+0.4°F

equipment)

1845

GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS

Memory function

20 sets memory of measurement values

– –

1970

If you are sure that you want to remove all stored memories permanently. Switch the unit off, Long press the °C/°F Button for 10 seconds until CLr appears. CLr will flashes for 4 times to clear all the memories along with 4 short beeps.

Beep alarm

2450

2

0.3

L

28

1 short beep when power on and start measurement; 1 long beep when measurement reading is below 37.5°C/99.5°F: 10 short beeps when measurement reading is greater than or equal to 37.5°C/99.5°F; 3 short beeps when system fails 2 x AAA alkaline battery

2 400 2 570

Bluetooth, WLAN, Pulse 802.11

modulation b/g/n, RFID 2450, 217 Hz LTE Band 7

Power source

5240

0.2

| 0.3

– –

Automatic power-off

In 60+20s

5 100-1 5 800

WLAN 802.11

a/n

Pulse modulation 217 Hz

5500

Main unit weight

5785

Exceptional Situation

Main unit size

Battery life

Support Information/Authorized agent

Accessories

Symboj – + – – – – – – Correction

Approx. 100g (batteries not included) L190mmXW55mmXH41mm Could be used for 300 times for normal condition Instruction manual Body mode 10.0°C~40.0°C/50.0°F-104.0°F Object mode 5°C-40.0°C/41.0°F-104.0°F Temperature:-20.0°C-50.0°C/-4.0°F~122.0°F Humidity:15%-95%RH avoid crash, sun burn or rain during transportation five years

-| –

– +

Operating environment

Imported by TDJ Australia PTY LTD.

ARTG Ref No: 333301

I

In Body Mode, Measured temperature is above Measur ing range of 43 °C/109.4°F,it is shown on the LCD.

Storage and shipping environment Expected service life

Technical description – – – – – – – – – – – – – – – – – – 1.All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life. 2. Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity.

Please retain your proof of purchase receipt in order to make a claim about faulty products or services. If you need assistance setting up or using your Aerpro product, or make a claim now or in the future, call Aerpro Support. Australia

TEL: 03 – 8587 8898 FAX: 03 – 8587 8866 Mon-Fri 9am – 5pm AEST

In Body Mode, Measured temperature is below the Measuring range of 32°C/89.6°F, it is shown on the LCD.

Table 1

Guidance and manufacturer’s declaration – electromagnetic emissions

Clinical Measurement Accuracy And Safety Verification:

The product has passed clinical trials. The measured results of the infrared forehead thermometer was compared with the measured results of Mercury thermometers, the deviation average At = 0.011°C not exceeding 0.3°C; the clinical repeatability of the infrared forehead thermometer ŜR =0.100°C, not exceeding 0.3°C. The measured results up to the laboratory standard and the clinical standard. Therefore, the deviation average and the clinical repeatability of the infrared forehead thermometer are complied with the regulatory requirement ISO 80601-2-56. The conclusions are drawn from the clinical trials, the accuracy and safety are complied with the regulatory requirement.

If you would like to download a digital copy of this manual, or other Aerpro manuals/software, please visit the http://aerpro.com website and search for the product model number for more information, accessories and products.

Emissions test

In Object Mode, Measured temperature is above Measur ing range 100.0°C/199.9°F,or environmental temperature is above system Operating range 40°C/104.0°F, it is shown on the LCD.

Compliance

RF emissions CISPR 11

Group 1

This manual is considered correct at time of printing but is subject to change. For latest manuals and updates refer to the website.

RF emissions CISPR 11

Class B

Harmonic emissions IEC 61000-3-2

N/A

In Object Mode, Measured temperature is below Measur ing range 0.0°C/32.0°F,or environmental temperature is below system operating range 5°C/41.0°F, it is shown on

the LCD.

NA

Voltage fluctuations/ flicker emissions IEC 61000-3-3

**CE..

REV:0.0