BRAUN No Touch + Forehead Thermometer BNT400A User Manual
Product description
- Power button
- Scanner
- LCD screen
- Temperature button
- Battery door
- Protective scanner cap
- Silent mode switch
- Guidance light
- Food and bath temperature mode
- AgeSmartTM button
Thank you for purchasing the Braun No touch + forehead thermometer (BNT400).
This thermometer is a high-quality product incorporating the latest technology
and tested in accordance with international standards. With its unique
technology, the product can provide a stable, accurate reading with each
measurement. The thermometer performs a self-test every time it is switched on
to always guarantee the accuracy of measurements. The Braun No touch +
forehead thermometer is intended for the intermittent determination of human
body temperature on people of all ages in a “Forehead” and “No Touch” mode,
using the center of the forehead as the measurement site, in a home use
environment.
Please read these instructions carefully before using this product and keep
the instructions and the thermometer in a safe place.
WARNINGS AND PRECAUTIONS
Keep out of reach of children under 12 years. Never use the thermometer for purposes other than those for which it has been intended. Please follow the general safety precautions when using on children. Never immerse the thermometer into water or other liquids (not waterproof). For cleaning and disinfecting please follow the instructions in the <> section. Do not store this thermometer in temperature extremes below -25 °C or over 55 °C or in excessive humidity (above 95% non-condensing relative humidity).
If thermometer is stored in a location that is cooler or warmer than where it will be used, allow it to acclimate to the room temperature for 10 minutes before taking a measurement. Do not use the thermometer if there are signs of damage on the scanner or on the thermometer itself. If damaged, do not attempt to repair the product.
Never insert a sharp object into the scanner area or any other open surface on
the thermometer. This thermometer consists of high-quality precision parts. Do
not drop the instrument. Protect it from severe impact and shock. Do not twist
the instrument or the measuring sensor.
This thermometer is intended for household use only.
Use of this thermometer is not intended as a substitute for consultation with
your physician.
Temperature elevation may signal a serious illness, especially in neonates and
infants, or in adults who are old, frail, or have a weakened immune system.
Please seek professional advice immediately when a temperature elevation
occurs on persons that are:
- Neonates and infants under 3 months (Consult your physician immediately if the temperature exceeds 37.4 °C)
- Individuals over 60 years of age (Fever may be blunted or absent in older individuals)
- Individuals having diabetes mellitus or a weakened immune system (e.g., HIV positive, cancer chemotherapy, chronic immunosuppressant treatment, splenectomy)
- Individuals who are bedridden (e.g., nursing home patient, stroke, chronic illness, paraplegia, quadriplegia, surgical recovery)
- A transplant recipient (e.g., liver, heart, lung, kidney).
This thermometer is not intended for pre-term babies or small-for
gestationalage babies. Pre-term is defined as babies with a calendar age of
less than 37 weeks. Small-for-gestational-age is defined as a baby born at 37
weeks or later, with a weight below the 10th percentile for babies of the same
gestational age.
This thermometer is not intended to interpret hypothermic temperatures. If the
device displays a temperature of 36.4 °C or less with an LED backlight color
of green or white, and the individual is exhibiting atypical symptoms or
behaviors, contact your physician or health care professional.
DO NOT allow children to take their temperatures unattended.
Please consult your physician if you see symptoms such as unexplained
irritability, vomiting, diarrhea, dehydration, changes in appetite or
activity, seizure, muscle pain, shivering, stiff neck, pain when urinating,
etc., even in the absence of fever.
Even in the absence of fever, those who exhibit a normal temperature may still
need to receive medical attention. People who are on antibiotics, analgesics,
or antipyretics should not be assessed solely on temperature readings to
determine the severity of their illness.
DO NOT modify this equipment without the authorization of the
manufacturer.
Why Braun No touch + forehead?
Measurement in under 2 seconds
The innovative infrared technology allows measurement of forehead temperature
in under 2 seconds whether you use the forehead or no touch option.
Accurate and reliable
Due to the unique technology, the Braun No touch + forehead thermometer
captures the heat naturally given off by the foreheadto calculate body
temperature value. Get the same professional accuracy whether touching your
forehead or holding up to 2.5 cm away.
Easy to use
The Braun No touch + forehead thermometer is non-invasive.
A measurement can be taken even while a child is sleeping.
The Braun No touch + forehead thermometer is safer to use on a child than a
rectal thermometer and is easier to use than other methods.
Safe and hygienic
No touch option helps minimize spreading of germs.
Completely safe for use on children and adults.
How does Braun No touch + forehead work?
The Braun No touch + forehead thermometer measures infrared energy radiated
from the skin at the center. This captured energy is collected through the
lens and converted to an oral equivalent value.
The Braun No touch + forehead thermometer has been clinically tested and
proven to be safe and accurate when used in accordance with its operating
instruction manual.
- A traditional forehead product without any optical system to capture radiated heat.
AgeSmart™ color-coded display
Clinical research shows the definition of fever changes as newborns grow into
infants, infants grow into children and children grow into adults.1
Select the appropriate age setting, take a temperature, and the display will
glow green, yellow, or red to indicate whether the temperature is in a normal,
moderate fever, or high fever range, respectively.
The device is only intended for the measurement of human body temperature. It
is not intended to diagnose or monitor a serious medical condition or disease.
The device also has a food/bath mode.
- Herzog L., Phillips S. Addressing Concerns About Fever. Clinical Pediatrics. 2011; 50 (#5): 383-390.
Temperature taking hints
It is important to know each individual’s normal temperature when they are
well.
This is the only way to accurately diagnose a fever. Take multiple readings
when healthy to determine normal temperature.
A child’s normal temperature can be as high as 37.7 °C or as low as 36.1 °C.
Be sure to note this unit reads 0.5 °C lower than a rectal digital
measurement.
Thermometer needs to acclimate to the room temperature for 10 minutes before
taking a measurement.
The individual being measured should be indoors for 30 minutes before taking a
measurement. External temperature can impact skin temperature.
If the individual is inside but in a room that is more than a few degrees
warmer or cooler than the room the thermometer is in, it will impact the
accuracy of the reading.
ALWAYS hold the thermometer and the forehead steady when taking a
reading.
DO NOT move the thermometer until you hear the final beep. DO NOT
take a measurement while or immediately after nursing a baby.
Individuals should not drink, eat, or be physically active before/while taking
the measurement.
Remove hats and wait 10 minutes before taking a measurement.
Before taking a measurement, remove dirt or hair from the forehead area. Wait
10 minutes after cleaning before taking measurement.
ALWAYS take the temperature exactly as directed. Temperature results may
vary if positioned in the wrong location.
For individuals measuring their own temperature, it is recommended to use the
“forehead” option instead of “no touch.”
In the following situations it is recommended that three temperatures in the same location be taken and the highest one taken as the reading:
- Newborn infants in the first 100 days.
- Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
- When the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the instrument and obtains consistent readings.
General precautions:
- Remove any sweat prior to measuring with a dry cloth.
- Avoid any cooling or warming cloths on the forehead for at least 30 minutes prior to measurement.
- DO NOT take temperature measurements over scar tissue, open sores, or abrasions.
- Keep the measurement sensor and lens clean. Avoid directly touching the sensor or lens.
- ALWAYS store the thermometer with the protective cap in place to prevent dirt and scratches from damaging the device.
- Make sure to remove the protective cap before taking a measurement and to put the cap back on after using the thermometer.
How to use your Braun No touch + forehead
Note: The thermometer and individual should be in the same ambient temperature for at least 10 minutes.
-
Remove protective cap
-
Power on
Press and release the power button once. Backlight will come on and the start- up sequence begins.
NOTE: Make sure to remove the protective scanner cap before taking a measurement.
-
Select age range
Select age with AgeSmartTM button
Device will start cycling through the age ranges, highlighting the
3 ranges you may select; 0-3 months ( ), 3-36 months ( ),
or 36+ months ( ).
To set a preferred age range: (Locking an age setting)
To save a preferred age range, so that the preferred age range will be automatically selected when the thermometer is turned ON, follow these steps:-
Make sure the thermometer is OFF.
-
Press and hold the power button ( ) and AgeSmartTM button ( ) for 4 seconds. The unit will turn on, displaying a yellow backlight and the 3 age range icons will flash.
-
Press the AgeSmartTM ( ) button. The age range selected will flash. Press the AgeSmartTM button until your desired age range is flashing on the display.
-
Once the desired age range is selected, stop pressing the button and wait 3 seconds. The green backlight will illuminate, and the selected age range will turn solid. After 4 seconds, the thermometer will turn off.
-
The preferred age range will be automatically selected when the thermometer is turned ON.
Note: With a preferred age range, an alternate age range can be selected during each use, by pressing the AgeSmartTM button when the age range icon appears, after the thermometer is turned ON. -
To remove the preferred age range, open the battery door, remove the batteries, wait 30 seconds and re-insert the batteries.
IMPORTANT: To ensure accurate readings, you must always take a temperature reading in the age range that corresponds to the individual’s age.
-
-
Position
Position thermometer on or up to 2.5 cm away from the center of the forehead, just between the eyebrows.
When the thermometer is ready and on or 2.5 cm away from the forehead a horizontal line of dashes (“—”) will appear on the screen.
For No touch readings, the yellow guidance light will show you where you are aiming.
If the eyebrow area is covered with hair, sweat or dirt, please clean the area beforehand to improve the reading accuracy. It is important to hold the thermometer and the forehead steady during measurement. Movement will impact the temperature reading.
-
Take temperature
Three dashed lines mean the thermometer is ready to take temperature. Press the temperature button. The display will show a dashed line animation while it takes a reading. Once the reading is complete a confirmation beep will be heard and the appropriate fever light color will be displayed on screen.
If the device is positioned too far away from the forehead, it will prompt you to move closer by displaying the characters “0 – 2 cm”. Slowly move the device toward the forehead until the dashed line animation starts and a reading is displayed. -
Read temperature
If temperature is in the normal range, a long beep will sound when the measurement is complete. If the temperature is in the fever range, 10 short beeps will sound to indicate the measurement is complete.| | |
---|---|---|---
Age range
| Green Normal temperature| Yellow Moderate fever| Red High fever
*0-3 Months**
| ≥35.8 – ≤37.4 °C| |
> 37.4 °C
3-36 Months| ≥35.4 – ≤37.6 °C| > 37.6 – ≤38.5 °C|
> 38.5 °C
36 M onths – adult
| ≥35.4 – ≤37.7 °C| > 37.7 – ≤39.4 °C|
> 39.4 °C
-
0 months is defined as having a gestational age of greater than or equal to 37 calendar weeks at birth
- To turn off
Press the power button to turn off. Device will also shut off automatically after 30 seconds of no use.
- To turn off
How to use the Braun No touch setting to take a food/bath temperature
-
Remove protective cap
-
Power on
Press and release the power button once. Backlight will come on and the start- up sequence begins.
NOTE: Make sure to remove the protective scanner cap before taking a measurement.
-
Select food & bath temperature mode
Press and release the food & bath temperature mode button. A beep will sound, the food & bath icon ( ) will be displayed, the screen backlight will turn blue and three dashes will show on the screen indicating the thermometer is ready to take a temperature.
-
Position
Position device up to 6 cm away from the bath water or food. Note: When measuring food or bath temperature the thermometer must be aimed directly at the food or bath water. DO NOT aim at sides of container holding the liquids or food. DO NOT touch liquid or food with the thermometer. Keep the sensor dry and clean. If the thermometer becomes dirty, please follow the <> section to clean.
-
Take temperature
Press the temperature button (you can press and hold the button or press and release it.) The display will show a dashed line animation while it takes a reading. Once the reading is complete, the display will show the temperature reading. -
Read temperature
Remove the device and read the temperature
Changing the sound mode
The Braun No touch + forehead thermometer allows you to silence the beeps on
the thermometer to avoid waking a sleeping child.
To activate the silence feature, slide the switch to the silent mode position
and take a temperature reading. An icon will appear on the screen to indicate
that the thermometer is in silent mode.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this
thermometer is used according to the use instructions, periodic re-adjustment
is not required. If at any time you question the accuracy, please contact
customer service.
The LOT and SN of your device are located inside the battery compartment.
Manufacturing date is given by the LOT number. The first three (3) digits
represent the Julian date that the product was manufactured and the next two
(2) digits represent the last two numbers of the calendar year the product was
manufactured.
The last identifiers are the letters that represent the manufacturer.
An example: LOT 11614tav, this product has been manufactured on the 116th day
of the year 2014.
Error message
| Situation|
Solution
---|---|---
| When 20% of battery
life is left, the display flashes the low battery warning symbol; however, the
device can still work until the battery life has 0% left.| Replace batteries
| If the steady battery
icon is the only symbol shown on the display, the device cannot work. The
battery should be replaced immediately| Replace batteries.
| This message displays
when the measured temperature is lower than 34 °C or higher than 43 °C or when
the room temperature is outside the operating range of 15 °C – 40 °C .| Re-
measure the temperature, carefully following the instructions in <> section.
| System error – self-
check display flashes continuously and will not be followed by the ready beep
and the ready symbol. If error persists, If error still persists,| Wait 1
minute until the thermometer turns off automatically, then turn on again. …
reset the thermometer by removing the batteries and putting them back in. …
please contact customer service.
| Blank display.
Thermometer does not have power.| Please check if the batteries have been
loaded correctly. Also check polarity (<+> and <–>) of batteries. Contact
customer service if thermometer still does not function.
Care and cleaning
Use an alcohol swab or cotton swab moistened with alcohol (70% isopropyl) to
clean the thermometer casing and the measuring probe. Ensure that no liquid
enters the interior of the thermometer. Never use abrasive cleaning agents,
thinners or benzene for cleaning and never immerse the instrument in water or
other cleaning liquids. Wait 10 minutes after cleaning before taking a
temperature measurement.
Make sure to replace the protective cap after use to prevent scratches or
damage from occurring to the thermometer.
Never insert a sharp object into the scanner area or any other open surface on
the thermometer.
Replacing the batteries
The Braun No touch + forehead thermometer comes with 2 AAA batteries. Replace
with 2 new AAA batteries when the flashing battery symbol appears on the LCD
screen. To change the batteries, slide open the battery cover and remove
batteries.
Replace the batteries being sure to align properly as indicated inside the
battery compartment. Remove the battery from the product if it is not required
for extended periods of time in order to avoid damage to the thermometer
resulting from a leaking battery.
To protect the environment, dispose of empty batteries at appropriate
collection sites according to national or local regulations.
Limited warranty
This product comes with limited warranty commencing on the date of purchase
(See warranty card for details). Within the warranty period we will eliminate,
free of charge, any defects in the appliance resulting from faults in
materials or workmanship, by replacing the complete appliance.
This warranty is applicable only for the appliance supplied by the appointed
distributor. This warranty does not cover: damage due to improper use, normal
wear or use as well as defects that have a negligible effect on the value or
operation of the appliance. The warranty becomes void if repairs are
undertaken by unauthorized persons and if original Braun parts are not used.
Product specifications
Type
|
No touch + forehead (BNT 400)
---|---
Measuring range
|
34 °C – 43 °C
Resolution
|
0.1 °C
Laboratory accuracy
|
0.2 °C for 35 °C – 42 °C
outside that range ±0.3 °C (Ambient temperature: 15 °C – 40 °C) This
thermometer displays a calculated oral equivalent estimate. Clinical accuracy
characteristics and procedures are available upon request.
Display
|
Liquid crystal display, 4 digits plus special icons
Acoustic
| Audio:
Normal temp range = Green temp range: 1 long beep for 0.4 second duration
Fever = Red or Yellow temp range: 10 short beeps for 0.2 second duration
Operating temperature
| 15 °C – 40 °C
Automatic switch-off|
Approx. 30 seconds after last measurement has been taken
Weight
| 100g. (with batteries), 82.5g. (w/o batteries)
Service life|
5 years
Long term storage ranges
Storage/transport temperature
| : -25 °C – 55 °C
Humidity|
15 – 95% non-condensing
Battery
| (2) AAA Batteries – at least 500 measurements
Pressure|
700-1060 hPA (0.7-1.06 atm)
This infrared thermometer meets requirements established in ASTM Standard E
1965-98 (for the thermometer system). Full responsibility for the conformance
of the product to the standard is assumed by Kaz Europe Sàrl, A Helen of Troy
Company, Place Chauderon 18, CH-1003 Lausanne, Switzerland
ASTM laboratory accuracy requirements for the thermometer only in the display
range of 37 °C to 39 °C for infrared thermometers is ±0.2 °C, whereas for
mercury-in-glass and electronic thermometers, the requirement per ASTM
Standards E 667-86 and E 1112-86 is ±0.1 °C .
This device conforms to the following standards:
EN 60601-1:2014 Medical electrical equipment. General requirements for basic
safety and essential performance.
ASTM E1965-98:2016 – Standard Specification for Infrared Thermometers for
Intermittent Determination of Patient Temperature.
IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General
requirements for basic safety and essential performance – Collateral Standard:
Electromagnetic disturbances – Requirement and tests.
NOTE: Do not use this device in the presence of electromagnetic or other
interference outside the normal range specified in IEC 60601-1-2:2014.
EN 1041: 2008 Information supplied by the manufacturer of medical devices. IEC
60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements
for basic safety and essential performance – Collateral standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment.
Equipment with type BF applied parts
See Instruction for use
Operating temperature
Storage temperature
Keep it dry
Internally Powered Equipment Continuous Operation
IP22: Protected against solid foreign objects of 12.5 mm diameter and greater.
Protected against vertically falling water drops when the device is tilted up
to 15 degrees.
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC.
For detailed description of EMC requirements please contact customer service.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Remove the battery from the instrument if it is not required for extended
periods of time in order to avoid damage to the thermometer resulting from a
leaking battery.
Please do not dispose of the product in the household waste at the end of its
useful life.
To protect the environment, dispose of empty batteries at appropriate
collection sites according to national or local regulations.
Guidance and manufacturer’s declaration — electromagnetic immunity
The BNT400 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the BNT400 should ensure that it is used in
such an environment.
Phenomenon
| Basic EMC
standard or test method| IMMUNITY TEST LEVELS
HOME HEALTHCARE ENVIRONMENT
ELECTROSTATIC DISCHARGE| IEC 61000-4-2|
L 8 kV contact
2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated RF EM fields a)
| IEC 61000-4-3| 10 V/m f)
80 MHz — 2,7 GHz b) 80 % AM at 1 kHz c)
Proximity fields from RF wireless
communications equipment
| IEC 61000-4-3|
See RF wireless communication equipment immunity table below
RATED power frequency magnetic fields d) e)| IEC 61000-4-8|
30 A/m g)
50 Hz or 60 Hz
- The interface between the PATIENT physiological signal simulation, if used, and the ME EQUIPMENT or ME SYSTEM shall be located within 0,1 m of the vertical plane of the uniform field area in one orientation of the ME EQUIPMENT or ME SYSTEM.
- ME EQUIPMENT and ME SYSTEMS that intentionally receive RF electromagnetic energy for the purpose of their operation shall be tested at the frequency of reception. Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS. This test assesses the BASIC SAFETY and ESSENTIAL PERFORMANCE of an intentional receiver when an ambient signal is in the pass band. It is understood that the receiver might not achieve normal reception during the test.
- Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS.
- Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive components or circuitry.
- During the test, the ME EQUIPMENT or ME SYSTEM may be powered at any NOMINAL input voltage, but with the same frequency as the test signal (see Table 1).
- Before modulation is applied.
- This test level assumes a minimum distance between the ME EQUIPMENT or ME SYSTEM and sources of power frequency magnetic field of at least 15 cm. If the RISK ANALYSIS shows that the ME EQUIPMENT or ME SYSTEM will be used closer than 15 cm to sources of power frequency magnetic field, the IMMUNITY TEST LEVEL shall be adjusted as appropriate for the minimum expected distance.
Guidance and manufacturer’s declaration — electromagnetic emissions
The BNT400 equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of the BNT400 should ensure that it
is used in such an environment.
Emissions Test| Compliance| Electromagnetic environment — guidance
RF Emissions CISPR 11| Group 1| The ME equipment uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11| Class B| Complies
Harmonic emissions IEC 61000-3-2| Not Applicable| The ME equipment is solely
battery powered.
Voltage fluctuations/ flicker emissions| Not Applicable
Guidance and manufacturer s declaration — RF wireless communication
equipment immunity
Test frequency
(MHz)| Band a) (MHz)| Service ‘)| Modulation b)|
Maximum power
(W)| Distance
(m)| Immunity test level
(V/m)
385| 380 —390| TETRA 400| Pulse modulation b)
18 Hz| 1,8| 0.3| 27
450| 430 — 470| GMRS 460, FRS 460| FM C)
-.: 5 kHz deviation
1 kHz sine| 2| 0.3| 28
710| 704 — 787| LTE Band 13, 17| Pulse modulation b)
217 Hz| 0,2| 0,3| 9
745
780
810| 800 — 960| GSM 800/900, TETRA 800. iDEN 820, CDMA 850. LTE
Band 5| Pulse modulation b)
18 Hz| 2| 0.3| 28
870
930
1720| 1 700 —
1 990| GSM 1800; CDMA 1900; GSM 1900; DECT;
LTE Band 1, 3, 4, 25; UMTS| Pulse modulation b)
217 Hz| 2| 0,3| 28
1845
1970
2450| 2 400 —
2 570| Bluetooth,
WLAN,
802.11 b/g/n, RFID 2450, LTE Band 7| Pulse modulation b)
217 Hz| 2| 0,3| 28
5240| 5 100 —
5 800| WLAN 802.11 a/n| Pulse modulation b)
217 Hz| 0,2| 0,3| 9
5500
5785
a)For some services, only the uplink frequencies are included.
b)The carrie shall be modulated using a 50 0/ duty cycle square wave signal.
c) As an alte native to FM modulation, 50 % pulse modulation at 18 Hz may be
used because while it does not
represent actual modulation, it would be worst case.
Certain trademarks used under license from The Procter & Gamble Company or its affiliates.
AgeSmartTM is a trademark of Helen of Troy Limited and/or its affiliates.
© 2021 All rights reserved. Made in China.
Distributed for:
Kaz Europe Sàrl
A Helen of Troy Company
Place Chauderon 18
CH-1003 Lausanne
Switzerland
www.BraunHealthcare.com
Patents: www.kaz.com/patents/braun
Printed in China
BNT400KO
PN: A002077R2
08DEC20