BRAUN BST200US Sensian Swipe 5 Thermometer User Manual
- October 30, 2023
- Braun
Table of Contents
- BRAUN BST200US Sensian Swipe 5 Thermometer
- Indications for Use / Intended Use
- Warnings and Precautions
- Why use the Braun Sensian™ Swipe 5 Thermometer?
- Temperature taking hints
- Errors and troubleshooting
- Limited three year warranty
- Product specifications
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
BST200US Sensing Swipe 5 Thermometer
User Manual
BRAUN BST200US Sensian Swipe 5 Thermometer
- Protective scanner cap
- Infrared temperature scanner
- Battery door
- LCD screen
- Temperature button
- Power button
Thank you for purchasing the Braun Sensing™ Swipe 5 Thermometer (BST200). The
Sensing™ Swipe 5 Thermometer is a high quality product incorporating the
latest technology and tested in accordance with international standards. With
its unique technology, the Sensing™ Swipe 5 Thermometer can provide a stable,
heat-interference-free reading with each measurement. The instrument performs
a self-test every time it is switched on to help provide accurate
measurements.
Please read these instructions carefully before using this product and keep
the instructions and the thermometer in a safe place.
Indications for Use / Intended Use
The Braun Sensing™ Swipe 5 Thermometer is a non-sterile, reusable, clinical thermometer intended for the intermittent determination of human body temperature on people of all ages in a home use environment.
Warnings and Precautions
Keep out of reach of children under 12 years.
NEVER use the thermometer for purposes other than those it has been intended
for. Please follow the general safety precautions when using on children.
NEVER immerse the Sensian™ Swipe 5 Thermometer into water or other liquids
(not waterproof ).
For cleaning and disinfecting please follow the instructions in the “Care and
cleaning” section.
DO NOT store this thermometer in temperature extremes below -13 °F or over 131
°F (below -25 °C or over 55 °C) or in excessive humidity (up to 95%
noncondensing relative humidity).
DO NOT use the thermometer if there are signs of damage on the measuring
sensor or on the instrument itself. If damaged, DO NOT attempt to repair the
instrument! Please contact our Consumer Relations Department at:
1-800-327-7226.
NEVER insert a sharp object into the scanner area or any other open surface on
the thermometer.
Let the thermometer and individual being measured acclimate to the temperature
of the room in which the measurement will be taken.
The thermometer needs to adjust to the room temperature for 30 minutes before
taking a measurement.
The individual being measured should be indoors for 30 minutes before taking a
measurement.
External temperature can impact skin temperature.
This Sensian™ Swipe 5 Thermometer consists of high-quality precision parts.
DO NOT drop the instrument. Protect it from severe impact and shock.
Please consult your physician if you see symptoms such as unexplained
irritability, vomiting, diarrhea, dehydration, changes in appetite or
activity, seizure, muscle pain, shivering, stiff neck, pain when urinating,
etc.
Even in the absence of fever, those who exhibit a normal temperature may still
need to receive medical attention. People who are on antibiotics, analgesics,
or antipyretics should not be assessed solely on temperature readings to
determine the severity of their illness.
Temperature elevation may signal a serious illness, especially in neonates and
infants, or in adults who are old, frail, or have a weakened immune system.
Please seek professional advice immediately when there is a temperature
elevation and if you are taking temperature on:
- Neonates and infants under 3 months (Consult your physician immediately if the temperature exceeds 99.4 °F [37.4 °C])
- Individuals over 60 years of age (Fever may be blunted or even absent in elderly Individuals)
- Individuals having diabetes mellitus or a weakened immune system (e.g., HIV positive, cancer chemotherapy, chronic immunosuppressant treatment, splenectomy)
- Individuals who are bedridden due to stroke, chronic illness, surgical recovery, paraplegia, quadriplegia.
- A transplant recipient (e.g., liver, heart, lung, kidney)
It is important to know each individual’s normal temperature when they are well. This is the only way to accurately diagnose a fever. Take multiple readings when healthy to determine normal temperature.
This thermometer is not intended for pre-term babies or small-for-gestational
age newborn (birth to 4 weeks of age) babies. Pre-term is defined as babies
born before 37 weeks of gestation.
Small-for-gestational-age is defined as a newborn baby, born at 37 weeks or
later, with a weight below the 10th percentile for newborn babies of the same
gestational age.
This thermometer is not intended to interpret hypothermic temperatures.
DO NOT allow children to take their temperatures unattended.
DO NOT modify this equipment without authorization of manufacturer.
This thermometer is intended for household use only.
Use of this Sensing™ Swipe 5 Thermometer is not intended as a substitute for
consultation with your physician or pediatrician.
Why use the Braun Sensian™ Swipe 5 Thermometer?
Measurement in 3 seconds
The innovative infrared technology allows measurement of forehead temperature
in 3 seconds.
Accurate and reliable
Thanks to Braun’s Temples wipe™ Technology, the Sensing™ Swipe 5 Thermometer
captures the heat naturally given off by the forehead and the temple to
calculate oral equivalent temperature.
Easy to use
The Sanxian™ Swipe 5 Thermometer is non-invasive.
Design is easy to hold and use.
The Sensing™ Swipe 5 Thermometer is safer to use on a child than a rectal
thermometer and is easier to use than other methods.
Safe and hygienic
Helps minimize spreading of germs.
Completely safe for use on children and adults.
Auto-Display memory
The last reading is automatically displayed for 2 seconds when the unit is
switched ON.
How the Braun Sensing™ Swipe 5 Thermometer works
The Sanxian™ Swipe 5 Thermometer measures infrared energy radiated from the
skin above the eyebrow area, the temple, and the surrounding tissue.
Temperature readings obtained by scanning the temple area will provide the
greatest accuracy.
The Sensing™ Swipe 5 Thermometer has been clinically tested and proven to be
safe and accurate when used in accordance with its operating instruction
manual.
Normal temperature is a range. The range of normal also varies from person to
person and fluctuates throughout the day. It is, therefore, important to
determine your normal temperature range.
The best way to determine your normal temperature range is to use the
thermometer when feeling well. Record readings twice a day (early morning and
late afternoon). Take the average of the two temperatures to calculate normal
oral equivalent temperature.
A child’s normal temperature can be as high as 99.9°F or as low as 97.0°F.
Remeasure with a standard digital thermometer for confirmation, especially on
infants.
Color-coded fever guidance
Fever guidance helps you to better understand the meaning of your child’s
temperature with the color indicated on the display. The screen displays green
for a normal temperature, yellow for a fever, and red for a high fever.
Color Range | Reading | Meaning |
---|---|---|
Green | 96.4 °F – 99.4 °F | |
(35.8 °C – 37.4 °C) | Normal | |
Yellow | >99.4 °F – 101.3 °F | |
(>37.4 °C – 38.5 °C) | Fever | |
Red | >101.3 °F | |
(>38.5 °C) | High Fever |
How to use your Braun Sensian™ Swipe 5 Thermometer
First time use
Insert 2 AAA batteries (see page 8 for instructions).
Remove 98.6 °F label from display.
-
Power on
Remove protective cap. Press and release the power button once. Backlight will come on and the warm up sequence starts.
The last reading is automatically displayed for 2 seconds.
When device displays STEADY “00.0” on the screen, it is ready to take a measurement. -
Position
Place the thermometer against the skin, on the forehead just above the eyebrow. Be sure to hold the thermometer flush against the forehead.
NOTE: ALWAYS remove the protective scanner cap before taking a temperature reading. -
Take temperature
Press the “Temperature button” and slowly swipe diagonally down the forehead to the temple and then back. Be sure to hold the thermometer flush against the forehead for the duration of the measurement.
When the temperature button is pressed a dashed line will appear on the screen and a soft beeping is heard to let you know the thermometer is working. DO NOT remove the thermometer from the forehead until you hear the long, 2-second confirmation beep. -
Read temperature
At the confirmation beep, remove the thermometer from the forehead and read the temperature. The screen will illuminate the appropriate color for the temperature reading. Green for a normal temperature, yellow for a fever, and red for a high fever.
To repeat Go to step 3. -
To turn off
Press the power button. To conserve battery power, the unit will automatically shut down after 60 seconds of no use.
Temperature taking hints
Before having temperature measured, individuals should be indoors for at least
30 minutes.
Hold the thermometer flush against the forehead and swipe slowly from above
the eyebrow down to the temple, and back again. Swiping too slow or too fast
may impact your reading.
DO NOT remove the thermometer from the forehead before hearing the final beep.
Nursing mothers should not take a temperature measurement on themselves or
have their temperature measured by another individual while or immediately
after nursing a baby, as nursing can affect body temperature.
DO NOT use the Sensing™ Swipe 5 Thermometer in high humidity environments.
Individuals should not drink, eat, or be physically active before/while taking
the measurement.
Before placing the thermometer sensor onto the forehead, remove dirt, hair, or
sweat.
Remove blankets and pillows away from the head and face prior measurement.
Use an alcohol swab to carefully clean the sensor and wait for 5 minutes
before taking a measurement on another Individual. Wiping the forehead with a
warm or cool cloth may impact your reading. It is advised to wait 15 minutes
before taking a reading.
ALWAYS take the temperature in the same location, since the temperature
readings may vary from different locations on the forehead.
In the following situations, it is recommended that three temperatures be
taken and the highest one used as the reading:
- Infants within the first 100 days of their birth.
- Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
- When the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the instrument and obtains consistent readings.
Memory mode
The Braun Sanxian™ Swipe 5 Thermometer includes a memory feature to recall the
last temperature reading. The last temperature taken is stored in its memory
and will be automatically displayed for 2 seconds when it is turned on.
How to change temperature between °F and °C and How to mute.
This thermometer has the ability to switch temperature units from
Fahrenheit to Celsius, as well the ability to mute the device. When the device
is in mute mode, all audible sounds will be disabled. The mute icon will be
displayed on the LCD at all times when the device is in mute mode.
- While the device is off, press and hold the button for 6 seconds until SET appears on the screen.
- Release the measurement button when the desired temperature scale and mute option are shown. The selected icons will flash on the screen to confirm the new setting, then the thermometer will turn off automatically.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this
thermometer is used according to the use instructions, periodic re-adjustment
is not required. If at any time you question the accuracy of temperature
measurements, please contact Consumer Relations.
Care and cleaning
Use an alcohol swab or cotton swab moistened with alcohol (70% Isopropyl) to
clean the thermometer casing and the measuring probe. Ensure that no liquid
enters the interior of the thermometer. NEVER use abrasive cleaning agents,
thinners or benzene for cleaning and NEVER immerse the instrument in water or
other cleaning liquids. Take care not to scratch the surface of the lens by
replacing the protective cap after use.
Replacing the batteries
The Sensing™ Swipe 5 Thermometer comes with 2 AAA batteries.
To install the batteries, slide open the battery door as shown. Insert 2 AAA
batteries, correctly positioning with positive “+” and negative “-” as shown.
Replace the battery door and close securely.
Replace the batteries when the flashing battery symbol appears on the LCD
screen. Dispose of batteries according to local waste management directions.
Do not mix alkaline, standard, or rechargeable batteries.
To protect the environment, dispose of empty batteries at appropriate
collection sites according to national or local regulations. Keep out of reach
of young children, elderly, and pets.
Errors and troubleshooting
Error message | Situation | Solution |
---|---|---|
Error symbol: during taking measurement |
A “Err” sign appears when the temperature measured is BELOW 93.2 °F or 34 ºC.
The mark “Err” appears if the operating temperature is ABOVE 108 °F or 42.2
ºC.| When you receive this symbol, press the power button to reset the
thermometer before taking a new measurement.
| When ambient temperature is too high or too low to take an accurate reading,
device will provide an “Err” message with flashing up/down arrow icon. The
error message will appear after start up and memory recall.| When you receive
this symbol, press the power button to reset the thermometer before taking a
new measurement. If the ambient temperature is not within the allowed range
of 59 ºF – 104.0 ºF (15 ºC – 40 ºC), you will not be able to take a reading.
| Low battery
When 20% of battery life is left, the display FLASHES the low battery warning
symbol, however, the device can still work until the battery life has 0%
left.| Replace batteries. See instructions on page 11.
| Dead battery indication If the steady battery icon is the only symbol shown
on the display, the device cannot work.| Replace batteries. See instructions
on page 11.
| Blank display Thermometer does not have power| Please check if the batteries
have been loaded correctly. Also check polarity (<+> and <–>) of batteries.
| All segments display or Blank display System error.| Replace batteries. See
instructions on page 11.
If error persists, call consumer relations.
Limited three year warranty
You should first read all Instructions before attempting to use this
product.
A. This 3 year limited warranty applies to repair or replacement of product
found to be defective in material or workmanship. This warranty does not apply
to damage resulting from commercial, abusive, unreasonable use or supplemental
damage. Defects that are the result of normal wear and tear will not be
considered manufacturing defects under this warranty. KAZ IS NOT LIABLE FOR
INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY NATURE, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ON THIS PRODUCT IS
LIMITED IN DURATION TO THE DURATION OF THIS WARRANTY.
Some jurisdictions do not allow the exclusion or limitation of incidental or
consequential damages or limitations on how long an implied warranty lasts,
so the above limitations or exclusions may not apply to you. This warranty
gives you specific legal rights, and you also may have other rights which
vary from jurisdiction to jurisdiction. This warranty applies only to the
original purchaser of this product from the original date of purchase.
B. At its option, Katz will repair or replace this product if it is found to
be defective in material or workmanship.
C. This warranty does not cover damage resulting from any unauthorized
attempts to repair or from any use not in accordance with the instruction
manual.
Call us toll-free at 1-800-327-7226 or visit: www.BraunHealthcare.com
/contact-us Or visit our website
at: www.BraunHealthcare.com Please be sure
to specify a model number.
NOTE: IF YOU EXPERIENCE A PROBLEM, PLEASE CONTACT CONSUMER RELATIONS FIRST OR
SEE YOUR WARRANTY. DO NOT RETURN THE PRODUCT TO THE ORIGINAL PLACE OF
PURCHASE. DO NOT ATTEMPT TO OPEN THE THERMOMETER HOUSING YOURSELF, DOING SO
MAY VOID YOUR WARRANTY AND CAUSE DAMAGE TO THE PRODUCT OR PERSONAL INJURY.
Product specifications
Type: | Sensian™ Swipe Thermometer |
---|---|
Measuring range: | 93.2°F to 108°F (34.0°C to 42.2°C) |
Resolution: | 0.1°F (0.1°C) |
Accuracy (Laboratory) | 80.6≥T≥96.8 – ±0.5F (27.0≥T≥36.0 ±0.3C) |
96.8< T≥ 102.2 ±0.4 F (36.0<T≥39.0 ±0.2C)
102.2<T≥ 109.4 ±0.5F (39.0< T≥ 43.0 ±0.3C)
Clinical accuracy and procedures are available upon request.
Display: | Liquid Crystal Display, 4 digits plus special icons Audio: |
---|---|
Acoustic: | Single long beep for 2 sec duration |
Operating ambient temperature: | 59°F to 104°F (15°C to 40°C) |
Automatic Switch-off: | Approx. 60 seconds after last measurement has been |
taken
Weight:| 99.5 grams (3.51 oz.) with batteries 77.1 grams (2.72 oz.) without
batteries
Long term storage ranges Storage/transport temperature:| –13°F to 131°F (–25°C
to 55°C)
Humidity:| 15–95% non condensing
Battery:| (2) AAA Batteries – 1,000 continuous readings without replacing
battery(s) after 2 years storage
Pressure:| 700-1060 haps (0.7-1.06 ATM)
This thermometer is an adjusted mode thermometer that provides an oral equivalent reading.
IP22: Protected against solid foreign objects of 12.5 mm diameter and larger.
Protected against water drops falling vertically when the device is tilted up
to 15°.
This infrared thermometer meets requirements established in ASTM Standard E
1965-98 (for the thermometer system). Full responsibility for the conformance
of the product to the standard is assumed by Kaz USA, Inc., a Helen of Troy
Company, 400 Donald Lynch Blvd., Suite 300, Marlborough, MA 01752.
ASTM laboratory accuracy requirements in the display range of 37 to 39°C (98
to 102°F) for IR thermometers is ±0.2°C (±0.4°F), whereas for mercury-in-glass
and electronic thermometers, the requirement per ASTM Standards E667-86 and
E1112-86 is ±0.1°C (±0.2°F).
This device conforms to the following standards:
ASTM E1965-98 – Standard Specification for Infrared Thermometers for
Intermittent Determination of Patient Temperature.
ISO 80601-2-56 – Medical electrical equipment – Part 2-56: Particular
requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement.
ANSI AAMI IEC 60601-1 – Medical electrical equipment – Part 1: General
requirements for basic safety and essential performance.
IEC 60601-1-2 – Medical electrical equipment – Part1-2: General requirements
for basic safety and essential performance – Collateral Standard:
Electromagnetic disturbances – Requirements and tests.
NOTE: DO NOT use this device in the presence of electromagnetic or other
interference outside the normal range specified in IEC 60601-1-2.
IEC 60601-1-11 – Medical electrical equipment – Part 1-11: General
requirements for basic safety and essential performance — Collateral standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment.
ISO 10993-1 – Biological evaluation of medical devices – Part 1: Evaluation
and testing within a risk management process.
ISO 15223-1 – Medical devices – Symbols to be used with medical device labels,
labeling and information to be supplied – Part 1: General requirements.
Internally Powered Equipment
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC.
For detailed description of EMC requirements please contact Consumer
Relations.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Guidance and manufacturer’s declaration —electromagnetic immunity
The BNT400 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the BNT400 should ensure that it is used in such
an environment.
Phenomenon| Basic EMC standard or test method| IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC DISCHARGE| IEC 61000-4-2| ± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated RF EM fields a)| IEC 61000-4-3| 10 V/m f)
80 MHz — 2,7 GHz b) 80 % AM at 1 kHz c)
Proximity fields from RF wireless communications equipment| IEC 61000-4-3| See
RF wireless communication equipment immunity table below
RATED power frequency magnetic fields d) e)| IEC 61000-4-8| 30 A/m g)
50 Hz or 60 Hz
a) The interface between he PATIENT physiological signal simulation, if used,
and the ME EQUIPMENT or
ME SYSTEM shall be located within 0,1 m of the vertical plane of the uniform
field area in one orientation of the ME EQUIPMENT Or ME SYSTEM.
b) ME EQUIPMENT and ME SYSTEMS that intentionally receive RE electromagnetic
energy for the purpose of their
operation shall be tested at the frequency of reception. Testing may be
performed at other modulation frequencies identified by the RISK MANAGEMENT
PROCESS. This test assesses the BASIC SAFETY and ESSENTIAL PERFORMANCE of an
intentional receiver when an ambient signal is in the passband. It is
understood that the receiver might not achieve normal reception during the
test.
c) Testing may be performed at other modulation frequencies identified by the
RISK MANAGEMENT PROCESS.
d) Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive
components or circuitry.
e)During the test, the ME EQUIPMENT or ME SYSTEM may be powered at any NOMINAL
input voltage, but with the same frequency as the test signal (see Table 1).
I) Before modulation is applied.
,D) This test level assumes a minimum distance between the ME EQUIPMENT or ME
SYSTEM and sources of power frequency magnetic field of at least 15 cm. If the
RISK ANALYSIS shows that the ME EQUIPMENT Or ME SYSTEM will be used closer
than 15 cm to sources of power frequency magnetic field, the IMMUNITY TEST
LEVEL shall be adjusted as appropriate for the minimum expected distance.
Guidance and manufacturer’s declaration —electromagnetic emissions
The BNT400 equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of the BNT400 should ensure that it
is used in such an environment.
Emissions Test| Compliance| Electromagnetic environment— guidance
RF Emissions CISPR 11| Group 1| The ME equipment uses RF energy only for its
internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions CISPR 11| Class B| Complies
Harmonic emissions IEC 61000-3-2| Not Applicable| The ME equipment is solely
battery powered. Not
Voltage fluctuations/ flicker emissions| Applicable
Guidance and manufacturer’s declaration — RF wireless communication equipment
immunity
Test frequency (MHz)| Band a) (MHz)| Service a)| Modulation b)| Maximum power
(W)| Distance (m)| IMMUNITY TEST LEVEL (Wm)
385| 380 —390| TETRA 400| Pulse
modulation b)
18 Hz| 1,8| 0,3| 27
450| 430 — 470| GMRS 460,
FRS 460| FM c)
-I 5 kHz deviation
1 kHz sine| 2| 0,3| 28
710| 704 — 787| LTE Band 13,
17| Pulse
modulation b)
217 Hz| 0,2| 0,3| 9
745
780
810| 800 — 960| GSM 800/900,
TETRA 800,
ident 820,
CDMA 850,
LTE Band 5| Pulse
modulation b)
18 Hz| 2| 0,3| 28
870
930
1720| 1 700 —
1 990| GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS| Pulse
modulation b)
217 Hz| 2| 0,3| 28
1845
1970
2450| 2 400 —
2 570| Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7| Pulse
modulation b)
217 Hz| 2| 0,3| 28
5240| 5 100 —
5 800| WLAN 802.11
a/n| Pulse
modulation b)
217 Hz| 0,2| 0,3| 9
5500
5785
a)For some services, only the uplink frequencies are included.
b)The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be worst
case.
Contact Us
Questions or Comments?
Call: 1-800-327-7226
Visit: BraunHealthcare.com/contact-us
www.BraunHealthcare.com
Certain trademarks used under license from The Procter & Gamble Company or its
affiliates.
Sensing™ is a trademark of Helen of Troy Limited.
© 2023. All rights reserved.
Kaz, USA, Inc., a Helen of Troy Company
400 Donald Lynch Blvd., Suite 300
Marlborough, MA 01752
USA/MEX
Patents/Patents:
www.BraunHealthcare.com/us_en/patents
Made in China.