MARK-10 Digital Torque Gauges TT03 Series User Guide
- June 6, 2024
- MARK-10
Table of Contents
**Series TT03
DIGITAL TORQUE GAUGES**
User’s Guide
Thank you…
Thank you for purchasing a Mark-10 Series TT03 digital torque gauge, designed
for handheld or test stand used.
With proper usage, we are confident that you will get many years of great
service with this product. Mark-10 instruments are ruggedly built for many
years of service in the laboratory and industrial environments.
This User’s Guide provides setup, safety, and operation instructions.
Dimensions and specifications are also provided. For additional information or
answers to your questions, please do not hesitate to contact us. Our technical
support and engineering teams are eager to assist you.
Before use, each person who is to use a Series TT03 digital torque gauge
should be fully trained inappropriate operation and safety procedures.
OVERVIEW
1.1 List of included items
Qty.
| Part No.|
Description
---|---|---
1| 12-1049| Carrying Case
1| 08-1022| AC adapter body with the US, EU, or UK prong
1| 08-1026| Battery (inside the gauge)
1| –| Certificate of calibration
1| 09-1165| USB cable
–| USB driver, MESUR® Lite software, MESUR®gauge evaluation software, User’s
Guide Download at:
www.mark-10.com/resources
1.2 Safety / Proper Usage
Caution!
Note the torque gauge’s capacity before use and ensure that the capacity is
not exceeded. Producing a torque greater than 150% of the gauge’s capacity can
damage the internal sensor. An overload can occur whether the gauge is powered
on or off.
Typical materials able to be tested include many manufactured items, such as springs, electronic components, fasteners, caps, mechanical assemblies, and many others. Items that should not be used with the gauge include potentially flammable substances or products, items that can shatter in an unsafe manner, and any other components that can present an exceedingly hazardous situation when acted upon by a force.
The following safety checks and procedures should be performed before and during operation:
- Never operate the gauge if there is any visible damage to the AC adapter or the gauge itself.
- Ensure that the gauge is kept away from water or any other electrically conductive liquids at all times.
- The gauge should be serviced by a trained technician only. AC power must be disconnected and the gauge must be powered off before the housing is opened.
- Always consider the characteristics of the sample being tested before initiating a test. A risk assessment should be carried out beforehand to ensure that all safety measures have been addressed and implemented.
- Wear eye and face protection when testing, especially when testing brittle samples that have the potential to shatter under force. Be aware of the dangers posed by potential energy that can accumulate in the sample during testing. Extra bodily protection should be worn if a destructive failure of a test sample is possible.
- In certain applications, such as the testing of brittle samples that can shatter, or other applications that could lead to a hazardous situation, it is strongly recommended that a machine guarding system be employed to protect the operator and others in the vicinity from shards or debris.
- When the gauge is not in use, ensure that the power is turned off.
POWER
The TT03 is powered either by an 8.4V NiMH rechargeable battery or by an AC adapter. Since these batteries are subject to self-discharge, it may be necessary to recharge the unit after a prolonged period of storage. Plug the accompanying charger into the AC outlet and insert the charger plug into the receptacle on the gauge (refer to the illustration below). The battery will fully charge in approximately 8 hours.
Caution!
Do not use chargers or batteries other than supplied or instrument damage
may occur.
If the AC adapter is plugged in, an icon appears in the lower-left corner of
the display, as follows:
If the AC adapter is not plugged in, battery power drainage is denoted in a
five-step process:
- When battery life is greater than 75%, the following indicator is present:
- When battery life is between 50% and 75%, the following indicator is present:
- When battery life is between 25% and 50%, the following indicator is present
- When battery life is less than 25%, the following indicator is present:
- When battery life drops to approximately 2%, the indicator from step 4 will be flashing. Several minutes after (timing depends on usage and whether the backlight is turned on or off), a message appears, “BATTERY VOLTAGE TOO LOW. POWERING OFF”. A 4-tone audio indicator will sound and the gauge will power off.
The gauge can be configured to automatically power off following a period of
inactivity. Refer to the Other Settings section for details.
If battery replacement is necessary, the battery may be accessed by loosening
the two captive screws in the rear half of the housing and separating the two
halves of the housing.
SETUP
3.1 Mechanical Setup
Fig. 3.1
Proper axial loading of Series TT03 torque gauge.
-
Proper alignment
Load must be applied axially with respect to the sensor, as shown in Figure 3.1. If attachments are used, ensure that the sample is acted upon axially with respect to the instrument. Sideloading or off-center loading can damage the instrument, whether power is turned on or off. -
Mounting to a plate
The indicator portion of the gauge can be mounted to a plate with four thumb screws fastened into the appropriate holes in the rear half of the housing.
The torque sensor portion of the gauge (shown in Figure 3.1) can be mounted directly to a Mark-10 torque test stand or tabletop mounting kit without the need for any additional adapters. Flat surfaces are provided to prevent rotation within a fixture.
Refer to the Dimensions section for detailed information.
3.2 Installing the USB driver
If communicating via USB, install the USB driver available at:
www.mark-10.com/resources
Caution!
Install the USB driver before physically connecting the gauge to a PC with
the USB cable.
Further instructions for configuring and using the gauge’s outputs are
provided in the Communications and Outputs section.
HOME SCREEN AND CONTROLS
4.1 Home Screen
No. | Name | Description |
---|---|---|
1 | Measurement direction indicator | – indicates the |
clockwise direction
– indicates the counter-clockwise direction
These indicators are used throughout the display and menu.
2| Peaks| The maximum measured clockwise and counter-clockwise
readings. These readings are reset by pressing ZERO or by powering the
gauge off and on.
3| Primary reading| The currently displayed load reading. See the
Operating Modes section for details.
4| Load bar| Analog indicator to help identify when an overload
condition is imminent. The bar increases either to the right or to the left
from the midpoint of the graph. Increasing to the right indicates clockwise
load, increasing to the left indicates counter-clockwise load. If setpoints
are enabled, triangular markers are displayed for visual convenience. This
indicator reflects the actual load, which may not correspond to the primary
reading (depends on operating mode).
The ZERO key does not reset the load bar. See the Operating Modes
section for details.
5| Units| The current measurement unit. Abbreviations are as
follows:
lbFin — Pound-inch
ozFin — Ounce-inch
kgFmm — Kilogram-millimeter
Ncm — Newton-centimeter
Note: not all sensor models display all the above units. Refer to the
capacity/resolution table for details.
6| Mode| The current measurement mode. Abbreviations are as
follows: RT — Real-Time
PCW — Peak Clockwise
PCCW — Peak Counter-clockwise
See the Operating Modes section for details about each of these modes
7| Battery / AC adapter indicator| Either the AC adapter icon or
battery power icon will be shown, depending on power conditions. Refer to the
Power section for details.
8| High/low limit indicators| Correspond to the programmed set
points. Indicator definitions are as follows:
– the displayed value is greater than the upper load limit
– the displayed value is between the load limits
– the displayed value is less than the lower load limit
9| Setpoints| The programmed load limit values. Typically used for
pass/fail type testing. One, two, or no indicators may be present, depending
on the configuration shown in the Set Points menu item.
4.2 Controls
Primary Label| Primary Function| Secondary Label|
Secondary Function
---|---|---|---
| Powers the gauge on and off. Press briefly to power on, press, and hold to
power off. Active only when the home screen is displayed.| ENTER|
Various uses, as described in the following sections.
ZERO| Zeroes the primary reading and peaks.|
(UP)
| Navigates up through the menu and sub-menus.
MENU| Enters the main menu.| ESCAPE| Reverts one step backward
through the menu hierarchy.
MODE| Toggles between measurement modes.|
__(DOWN)
| Navigates down through the menu and sub-menus.
DATA| Transmits the current reading to an external device via the USB
port.| DIRECTION| Toggles between clockwise and counter-clockwise
directions while configuring set points and other menu functions.
Note: Measurement units are configured through the menu. Refer to Changing The Units section for details.
4.3 Menu navigation basics
Most of the gauge’s various functions and parameters are configured through
the main menu. To access the menu press MENU. Use the UP and DOWN
keys to scroll through the items. The current selection is denoted with clear
text over a dark background. Press ENTER to select a menu item, then use
UP and DOWN again to scroll through the sub-menus. Press ENTER
again to select the sub-menu item.
For parameters that may be either selected or deselected, press ENTER to
toggle between selecting and deselecting. An asterisk (*) to the left of the
parameter label is used to indicate when the parameter has been selected.
For parameters requiring the input of a numerical value, use the UP and
DOWN keys to increment or decrement the value. Press and hold either key
to auto-increment at a gradually increasing rate. When the desired value has
been reached, press ENTER to save the change and revert back to the sub-
menu item, or press ESCAPE to revert back to the sub-menu item without
saving. Press ESCAPE to revert one step back in the menu hierarchy until
back into normal operating mode.
Refer to the following sections for details about setting up particular
functions and parameters.
OPERATING MODES
Caution!
In any operating mode, if the capacity of the instrument has been exceeded by
more than 110%, the display will show “OVER” to indicate an overload. A
continuous audible tone will be sounded (if beeps are enabled) until the
MENU key has been pressed or the load has been reduced to a safe level.]
Three operating modes are possible with the TT03 torque gauge. To cycle
between the modes, press MODE while in the home screen.
5.1 Real-time (RT)
The primary reading corresponds to the live measured reading.
5.2 Peak Clockwise (PCW)
The primary reading corresponds to the peak clockwise reading observed. If the
actual load decreases from the peak value, the peak will still be retained in
the primary reading area of the display. Pressing ZERO will reset the
value.
5.3 Peak Counter-clockwise (PCCW)
Same as above, but for counter-clockwise readings.
CHANGING THE UNITS
The TT03 can display three different measurement units, depending on the model. To change the unit, select Units from the menu. The display will list the available units, for example:
The gauge will always power on with the unit selected.
DIGITAL FILTERS
Digital filters are provided to help smooth out the readings in situations
where there is mechanical interference in the work area or test sample. These
filters utilize the moving average technique in which consecutive readings are
pushed through a buffer and the displayed reading is the average of the buffer
contents. By varying the length of the buffer, a variable smoothing effect can
be achieved. The selection of 1 will disable the filter since the average of a
single value is the value itself.
To access digital filter settings, select Filters from the menu. The display
appears as follows:
Two filters are available:
Current Reading – Applies to the peak capture rate of the instrument.
Displayed Reading – Applies to the primary reading on the display.
Available settings: 1,2,4,8,16,32,64,128,256,512,1024. It is recommended to keep the current reading filter at its lowest value for best performance, and the displayed reading filter at its highest value for best stability.
SETPOINT INDICATORS
8.1 General Information
Setpoints are useful for tolerance checking (pass/fail). Two limits, high and
low, are specified and stored in the non-volatile memory of the instrument and
the primary reading is compared to these limits.
8.2 Configuration
To configure setpoints, select Set Points from the menu. The screen appears as
follows:
Either one, two, or none of the setpoints may be enabled. To toggle between
the clockwise and counter-clockwise directions, press the DIRECTION key.
If two set points have been enabled, they are displayed in the upper left
corner of the display. If only one set point has been enabled, the word “OFF”
appears in place of the value. If no setpoints have been enabled, the upper
left corner of the display will be blank. When set points are enabled, the
following indicators are shown to the left of the primary reading:
– the displayed value is greater than the upper load limit (NO GO HIGH)
– the displayed value is between the limits (GO)
– the displayed value is less than the lower load limit (NO GO LOW)
Note: Setpoint indicators reference the displayed reading, not necessarily the current live load.
COMMUNICATIONS AND OUTPUTS
9.1 Communication Settings
To set up communication settings, select USB Settings from the menu. The
screen appears as follows:
Communication settings are permanently set to the following:
Data Bits: 8
Stop Bits: 1
Parity: None
Other settings are configured as follows:
-
Baud Rate
Select the baud rate as required for the application. It must be set to the same value as the receiving device. -
Data Format
Select the desired data format. The screen appears as follows:
Selection|
Description
---|---
Numeric + Units| Output format includes the value and unit of measure.
Clockwise values have positive polarity, counter-clockwise values have
negative polarity.
Numeric Only| Output format includes the value only. Polarity is the same as
above.
Invert Polarity| Clockwise values have negative polarity, counter-clockwise
values have positive polarity. May be selected in addition to the Numeric +
Units / Numeric Only selection.
Omit Polarity| Both directions are formatted with positive polarity. May be
selected in addition to the Numeric + Units / Numeric Only selection.
9.2 Data Output
Communication with Series TT03 torque gauges is achieved through the micro USB
port located along the left side of the housing, as shown in the illustration
in the Power section. Communication is possible only when the gauge is in the
main operating screen (i.e. not in a menu or configuration area).
The current reading is transmitted from the gauge when the DATA key is
pressed. Series TT03 gauges may also be controlled by an external device
through the USB channel by sending ASCII command ‘?’ (no quotes). The command
must be terminated by a CR (Carriage Return) character, 0x0D, or a CR-LF
(Carriage Return – Line Feed) pair, where the Line Feed, 0x0A, is ignored.
9.3 Command Responses
In response to the reading request command ‘?’ the gauge will return a string
with the load data, followed by a space, then the load unit (if enabled under
the USB Settings → Data format sub-menu). It will be terminated by a CR-LF
pair.
Example return strings:
-18.78 lbFin
4.285 Nm
The number of digits after the decimal point is dependent on the gauge’s
capacity and resolution.
Polarity may be inverted or omitted, as shown in the table above.
Any detected errors are reported back by means of error code *10 (illegal
command).
Series TT03 gauges can send individual data points or continuous data to
MESUR™ gauge and MESUR™ Lite software.
CALIBRATION
10.1 Initial Physical Setup
The sensor should be mounted vertically to a test stand or fixture rugged
enough to withstand a load equal to the full capacity of the instrument.
Vertical orientation is preferable to avoid side loading, which can affect the
readings. Suitable certified calibration equipment is required, and caution
should be taken while handling such equipment.
The illustration below depicts a recommended vertical setup:
10.2 Calibration Procedure
-
Select Calibration from the menu. The display appears as follows: The sensor can be calibrated at up to 10 points in each direction. Enter the number of calibration points for each direction. At least one point must be selected for each direction.
Note: To achieve the accuracy specification of ±0.5% of full scale, it is recommended to calibrate the sensor at 5 or more even increments in both the clockwise and counter-clockwise directions. For example, a sensor with a capacity of 50 lb-in should be calibrated at 10, 20, 30, 40, and 50 lb-in loads in each direction. -
To escape the Calibration menu at any time, press ESCAPE. The display appears as follows: Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will return to the menu without saving changes.
-
After the number of calibration points has been entered, press ENTER. The display appears as follows:
-
Place the torque sensor vertically in a fixture free from vibration, then press ZERO. The gauge will calculate offsets, and the display appears as follows: If failed:
-
The following screen appears after the offsets have been calculated: Keep the sensor in a vertical position, as explained in the previous sub-section. Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply any calibration loads. Then press ENTER.
-
The display appears as follows: Optionally exercise the sensor several times (at full scale, if possible), then press ENTER.
-
The display appears as follows: Apply torque equal to the full scale of the instrument, then press ENTER.
-
After displaying “PLEASE WAIT…” the display appears as follows: Remove the torque applied in Step 7, leave the fixtures in place, then press ZERO.
-
The display appears as follows: Use the UP and DOWN keys to adjust the torque value as required. The torque values default to even increments, as indicated by the previously entered number of data points described in Step
1. Then press ENTER.
Repeat the above step for the number of data points selected. -
After all the clockwise calibration points have been completed, the display appears as follows: Press ENTER.
-
At the completion of the counter-clockwise calibration, the display appears as follows: To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select “EXIT W/O SAVING”.
-
Any errors are reported by the following screens: Displayed at the start of calibration if a disallowed unit is selected. Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again. The calibration load does not match the set value. The entered calibration point is too close to the previous point.
OTHER SETTINGS
11.1 Automatic Shutoff
The gauge may be configured to automatically power off following a period of
inactivity while on battery power. Inactivity is defined as the absence of any
key presses or load changes of 100 counts or less. To access these settings,
select Automatic Shutoff from the menu. The display appears as follows:
Select Disabled to disable automatic shutoff. Select Enabled to enable
it. The length of time of inactivity is programmed in minutes via the Set
Minutes parameter. Available settings: 5-30, in 5-minute increments.
Note: If the AC adapter is plugged in, the gauge will ignore the
Automatic Shutoff setting and remain powered on until the POWER key is
pressed.
11.2 Backlight
Several initial settings are available, upon powering on the gauge. To access
these settings, select Backlight from the menu. The display appears as
follows:
Selection|
Description
---|---
Off| Backlight to be off upon powering on the gauge.
On| Backlight to be on upon powering on the gauge.
Auto| Backlight to be on upon powering gauge but will shut off after a
period of inactivity (as defined in the Automatic Shutoff sub-section).
The backlight will turn on again when the activity resumes. The length of time
of inactivity is programmed in minutes via the Set Minutes parameter.
Available settings: 1-10 , in 1-minute increments.
Note: If the AC adapter is plugged in, the gauge will ignore these settings and keep the backlight on. Selecting the On or Off setting in the Backlight menu will manually turn the backlight on or off as if the Backlight key were pressed.
11.3 LCD Contrast
The contrast of the display may be adjusted. Select LCD Contrast from the
menu. The screen appears as follows:
Press ENTER to modify the contrast. Select a value from 0 to 25, 25 producing the most contrast.
11.4 Tones
Audible tones can be enabled for all keypresses and alerts, such as overload,
setpoint value reached, etc. The SetPoint alert can be configured to be either
a momentary tone or a continuous tone (until the load is restored to a value
between the setpoints). To configure the functions for which audible tones
will apply, select Tones from the menu. The screen appears as follows:
11.5 Initial Mode
This section is used to configure the initial mode upon powering on the gauge.
To access this parameter, select Initial Mode from the menu. The screen will
display the available modes. An example is as follows:
The default value is Real-Time.
11.6 Restore Default Settings
Default factory settings can be restored by selecting Restore Defaults from
the menu. The settings may be found in the Specifications section. The screen
appears as follows:
11.7 Information / Welcome Screen
The following screen is displayed at power-up and can be accessed at any time
by selecting Information from the menu:
SPECIFICATIONS
12.1 General
Accuracy: | ±0.5% of the full scale |
---|---|
Sampling rate: | 2,000 Hz |
Power: | AC or rechargeable battery. A low battery indicator appears when |
the battery level is low, and the gauge powers off automatically when power
reaches a critical stage.
Battery life:| The backlight on up to 7 hours of continuous use
Backlight off: up to 24 hours of continuous use
Measurement units:| lbFin, ozFin, kgFmm, Ncm (depending on model)
USB output:| Individual data point output by pressing DATA button.
Configurable up to 115,200 baud.
Safe overload:| 150% of full scale (display shows “OVER” at 110% and
above)
Weight:| 1.8 lb [0.8 kg]
Included accessories:| Carrying case, AC adapter, battery, USB cable,
NIST-traceable certificate of calibration
Environmental requirements:| 40 – 100°F, max. 96% humidity, non-
condensing
Warranty:| 3 years (see the individual statement for further details)
Literature & Software:| Download
at:www.mark-10.com/resources
12.2 Factory Settings
Parameter|
Setting
---|---
Setpoints|
Upper| Disabled (defaults to 80% of full scale, clockwise, when enabled)
Lower| Disabled (defaults to 40% of full scale, clockwise, when enabled)
Filters|
Current| 8
Displayed| 128
Backlight| Auto
Minutes| 1
USB Output| Disabled
Baud Rate| 9,600
Data Format| Numeric + units
Automatic Shutoff| Enabled
Minutes| 5
Tones|
Keys| Enabled
Alerts| Enabled
Set Points| Momentary
Initial Mode| Real-Time
Units| Depending on the gauge model
12.3 Capacity & Resolution
Model
| oz-in| lbFin| KGF|
Name
---|---|---|---|---
MTT03-10Z| 10 x 0.01| –| 7 x 0.005| 7 x 0.005
MTT03-20Z| 20 x 0.02| –| 14 x 0.01| 14 x 0.01
MTT03-50Z| 50 x 0.05| –| 36 x 0.05| 36 x 0.05
MTT03-12| 192 x 0.2| 12 x 0.01| –| 135 x 0.1
MTT03-50| 800 x 0.5| 50 x 0.05| –| 570 x 0.5
MTT03-100| 1600 x 1| 100 x 0.1| –| 1150 x 1
12.4 Dimensions
IN [MM]
Mark-10 Corporation has been an innovator in the force and torque measurement fields since 1979. We strive to achieve 100% customer satisfaction through excellence in product design, manufacturing, and customer support. In addition to our standard line of products, we can provide modifications and custom designs for OEM applications. Our engineering team is eager to satisfy any special requirements. Please contact us for further information or suggestions for improvement.
Force and torque measurement engineered better
Mark-10 Corporation
11 Dixon Avenue
Copiague, NY 11726 USA
1-888-MARK-TEN
Tel: 631-842-9200
Fax: 631-842-9201
Internet: www.mark-10.com
E-mail: info@mark-10.com