GSM90A Series 90W AC to DC Reliable Green Medical Adaptor
Product Information
The 90W AC-DC Reliable Green Medical Adaptor is part of the
GSM90A series and is a medical safety-approved power unit. It has a
3-pole AC inlet IEC320-C14 and is suitable for use with a mobile
clinical workstation. The adapter has an extremely low leakage
current and no-load power consumption of 130 K hours. It also comes
with a 3-year warranty.
The adapter has various DC plug quick adapter accessories
available, which are sold separately. It has an efficiency of up to
91% and overload protection with hiccup mode, which recovers
automatically after a fault condition is removed. It also has
over-voltage protection with shut down output voltage and re-power
on to recover feature. The adapter has a wide operating temperature
range of -30 ~ +70 (Refer to Derating Curve) and can operate at
altitudes of up to 3000 meters.
The adapter is safety-approved by IEC60601-1, TUV BS
EN/EN60601-1, ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22.2 No.
60601-1:14 – Edition 3, EAC TP TC 004 approved. It has
primary-secondary isolation level of 2xMOPP and primary-earth
isolation level of 1xMOPP. The adapter also has EMC emission and
immunity standards such as BS EN/EN55011 (CISPR11), FCC PART 15 /
CISPR22, CAN ICES-3(B)/NMB-3(B).
The adapter has a dimension of 1456032mm (LWH), and it
weighs 0.45Kg. The packing includes 30pcs/14.5Kg/0.9CUFT.
Product Usage
To use the 90W AC-DC Reliable Green Medical Adaptor, follow
these steps:
-
Insert the 3-pole AC inlet IEC320-C14 into the appropriate
power source. -
Connect the appropriate DC plug quick adapter accessory to the
adapter, which is sold separately. Refer to
https://www.meanwell.com/upload/pdf/DC_plug.pdf for available
options. -
Connect the other end of the DC plug quick adapter accessory to
the device that requires power. -
Switch on the power source.
Note: Ensure that the power source is within the input voltage
range of the adapter. The adapter has an input current of 1.3A /
115VAC and 0.6A / 230VAC. It also has an inrush current of cold
start 30A/115VAC and 60A / 230VAC.
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
User’s Manual
Video
EN 60601-1 IEC 60601-1
8
BS EN/EN60601-1 ANSI/AAMI ES60601-1 IEC60601-1 TPTC004
++
Features
Applications
3 pole AC inlet IEC320-C14, Classpower unit
Mobile clinical workstation
Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1 Extremely low leakage current No load power consumption<0.15W Energy efficiency level VI and meet CoC Version 5
Oral irrigator Portable hemodialysis machine Breath Machine Medical computer monitor
-30~+70 wide range working temperature Protections: Short circuit / Overload / Over voltage / Over temperature LED indicator for power on
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Lifetime > 130 K hours
Various DC plug quick adapter accessory available
(Plug kit sold sperately, please refer to : https://www.meanwell.com/upload/pdf/DC_plug.pdf )
3 years warranty
Description
GSM90A is a highly reliable, 90W desktop style single-output green medical
adaptor series. This product is equipped with a 3-pin (with FG) standard
IEC320-C14 power plug, adopting the input range from 80VAC to 264VAC. The
entire series supplies different output voltages between 12VDC and 48VDC that
can satisfy the demands for various kinds of medical electrical devices. The
circuitry design meets the international medical standards (2*MOPP), having an
ultra low leakage current (<100 A), fitting the medical devices in direct
electrical contact with the patients. With the efficiency up to 91% and the
extremely low no-load power consumption below 0.15W, GSM90A is compliant with
USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP,and
meet Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor
to save the energy when it is either under the operating mode or the standby
mode. The entire series utilizes the 94V-0 flame retardant plastic case.
GSM90A is approved with the international medical safety certificates.
Model Encoding GSM90 A 12 – P1M
{ DC plug type P1M: Standard model,2.5×5.5x11mm,C+,tuning fork type Other options available by customer requested (see Page 4~5) Output voltage IEC320-C14 AC inlet Rated wattage Series name
File Name:GSM90A-SPEC 2022-11-21
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
SPECIFICATION
ORDER NO.
GSM90A12-P1M
GSM90A15-P1M
GSM90A19-P1M
GSM90A24-P1M
GSM90A48-P1M
SAFETY MODEL NO.
GSM90A12
GSM90A15
GSM90A19
GSM90A24
GSM90A48
DC VOLTAGE
Note.2 12V
15V
19V
24V
48V
RATED CURRENT
6.67A
6A
4.74A
3.75A
1.87A
CURRENT RANGE
0 ~ 6.67A
0 ~ 6A
0 ~ 4.74A
0 ~ 3.75A
0 ~ 1.87A
RATED POWER (max.)
80W
90W
90W
90W
90W
OUTPUT
RIPPLE & NOISE (max.) Note.3 120mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
LINE REGULATION Note.5 ±1.0%
LOAD REGULATION
±5.0%
120mVp-p ±5.0% ±1.0% ±5.0%
120mVp-p ±4.0% ±1.0% ±4.0%
180mVp-p ±3.0% ±1.0% ±3.0%
200mVp-p ±2.5% ±1.0% ±2.5%
SETUP, RISE TIME Note.6 1000ms, 50ms / 230VAC 1000ms, 50ms / 115VAC at full load
HOLD UP TIME (Typ.)
40ms / 230VAC
VOLTAGE RANGE Note.7 80 ~ 264VAC
25ms / 115VAC at full load
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.91 / 230VAC PF>0.95 / 115VAC at full load
INPUT
EFFICIENCY (Typ.)
88%
89%
89%
90%
91%
AC CURRENT (Typ.)
1.3A / 115VAC 0.6A / 230VAC
INRUSH CURRENT (Typ.) Cold start 30A/115VAC 60A / 230VAC LEAKAGE CURRENT(max.) Earth leakage current < 125A/264VAC , Touch current < 100A/264VAC
OVERLOAD PROTECTION
OVER VOLTAGE
110 ~ 150% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed 105 ~ 135% rated output voltage Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE WORKING TEMP.
Shut down o/p voltage, re-power on to recover -30 ~ +70 (Refer to “Derating Curve”)
WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT VIBRATION
20% ~ 90% RH non-condensing -40 ~ +85, 10 ~ 95% RH non-condensing ±±00.0.033%%/ /((00~~4400) 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.8 3000 meters
SAFETY STANDARDS
IEC60601-1, TUV BS EN/EN60601-1, ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 – Edition 3, EAC TP TC 004 approved
ISOLATION LEVEL
Primary-Secondary: 2xMOPP, Primary-Earth: 1xMOPP
WITHSTAND VOLTAGE
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:SHORT
ISOLATION RESISTANCE EMC EMISSION
I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH
Parameter
Standard
Conducted emission
BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22, CAN ICES-3(B)/NMB-3(B)
Radiated emission
BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22, CAN ICES-3(B)/NMB-3(B)
Test Level / Note Class B Class B
Harmonic current
BS EN/EN61000-3-2
Class A
SAFETY &
EMC (Note 9)
Voltage flicker
BS EN/EN61000-3-3
BS EN/EN60601-1-2, BS EN/EN61204-3
Parameter
Standard
—-Test Level / Note
EMC IMMUNITY
ESD
RF field susceptibility EFT bursts Surge susceptibility
BS EN/EN61000-4-2
BS EN/EN61000-4-3 BS EN/EN61000-4-4 BS EN/EN61000-4-5
Level 4, 15KV air ; Level 4, 8KV contact Level 3, 10V/m( 80MHz~2.7GHz ) Table
9, 9~28V/m( 385MHz~5.78GHz ) Level 3, 2KV
Level 3, 1KV/Line-Line , 2KV/Line-FG
Conducted susceptibility
BS EN/EN61000-4-6
Level 3, 10V
Magnetic field immunity
BS EN/EN61000-4-8
Level 4, 30A/m
Voltage dip, interruption
BS EN/EN61000-4-11
100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods
MTBF
2538.4K hrs min. Telcordia SR-332 (Bellcore) ; 387.5K hrs min. MIL-HDBK-217F (25)
OTHERS DIMENSION
1456032mm (LWH)
PACKING
0.45Kg; 30pcs/14.5Kg/0.9CUFT
PLUG CONNECTOR
CABLE
See page 4~5 ; Other type available by customer requested See page 4~5 ; Other type available by customer requested
NOTE
1. All parameters are specified at 230VAC input, rated load, 25 70% RH
ambient. 2. DC voltage: The output voltage set at point measure by plug
terminal & 50% load. 3. Ripple & noise are measured at 20MHz by using a 12″
twisted pair terminated with a 0.1f & 47f capacitor. 4. Tolerance: includes
set up tolerance, line regulation, load regulation. 5. Line regulation is
measured from low line to high line at rated load. 6. Length of set up time is
measured at first cold start. Turning ON/OFF the power supply may lead to
increase of the set up time. 7. Derating may be needed under low input
voltages. Pleas check the derating curve for more details. 8. The ambient
temperature derating of 3.5/1000m with fanless models and of 5/1000m with fan
models for operating altitude higher than 2000m(6500ft). 9. The power supply
is considered as an independent unit, but the final equipment still need to
re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer
to “EMI testing of component power supplies.” (as available on
http://www.meanwell.com) Product Liability DisclaimerFor detailed information,
please refer to https://www.meanwell.com/serviceDisclaimer.aspx
File Name:GSM90A-SPEC 2022-11-21
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
Derating Curve
Static Characteristics
100
80
60 50 40
20
-30 0
10 20 30 40
50 60
70 (HORIZONTAL)
AMBIENT TEMPERATURE ()
100 90 80 70 60 50 40
80 90 100 110 120 140 160 180 200 220 240 264
INPUT VOLTAGE (VAC) 60Hz
LOAD (%) 60
LOAD (%)
Mechanical Specification
32
145
AC Inlet IEC320-C14
DC output plug Standard plug: P1M
P1M 5.5
2.5
POWER LED 11±0.5mm
Case No. GS90A Unit:mm
Ul1185 14AWG 1000±50mm for 12 ~ 15V UL1185 16AWG 1200±50mm for 19 ~ 48V
70±10mm
Pin Assignment
C”+”
Outside
Inside
-V connected to AC FG(standard) -V not connected to AC FG(optional)
File Name:GSM90A-SPEC 2022-11-21
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
DC plug changeable through: (1) Customization of the standard part with an
optional DC plug according to the table (MOQ applicable) (2) Quick adapter
accessory (sold separately without MOQ) Please refer to below table and online
selection guide : https://www.meanwell.com/upload/pdf/DC_plug.pdf
Example quick adapter accessory:
Optional DC plug: (Available in customized cable or quick adapter)
Tuning Fork Style
Type No.
A B
Barrel Style
C (Straight) C
(Right-angled)
P1I P1L P1J P1JR P1IR P1LR P1MR
Type No.
C
P2I
P2J
P2L
A
(Straight)
P2M
B
C
P2IR
P2JR
(Right-angled)
P2LR P2MR
Lock Style
Type No.
A
Floating Locking
C
P2S(S761K)
P2K(761K)
B
P2C(S760K)
SWITCHCRAFT original or equivalent P2D(760K)
Min. Pin Style
Type No.
A
C
P3A
B EIAJ equivalent
P3B P3C
Center Pin Style
Type No.
A
C
P4A
B
P4B
D EIAJ equivalent
P4C
A
B
C
OD
ID
L
Quick Adapter Accessory
5.5
2.1
9.5
5.5
2.5
9.5
5.5
2.1 11.0
Available
5.5
2.1
11.0 (Current rating: 7.5A max.)
5.5
2.1
9.5
5.5
2.5
9.5
5.5
2.5 11.0
A
B
C
OD
ID
L
5.5
2.1 9.5
5.5
2.1 11.0
5.5
2.5 9.5
5.5
2.5 11.0
5.5
2.1 9.5
None
5.5
2.1 11.0
5.5
2.5 9.5
5.5
2.5 11.0
A
B
C
OD
ID
L
5.53
2.03 12.06
5.53
2.54 12.06
5.53
2.03 9.52
None
5.53
2.54 9.52
A
B
C
OD
ID
L
2.35
0.7 11.0
4.0
1.7 11.0
None
4.75
1.7 11.0
AB OD ID
CD L Center Pin
5.5 3.4 11.0 1.0
None
6.5 4.4 11.0 1.4
7.4 5.1 11.0 0.6
File Name:GSM90A-SPEC 2022-11-21
90W AC-DC Reliable Green Medical Adaptor
GSM90A series
Min. DIN 3 Pin with Lock (male)
12 3
KYCON KPPX-3P equivalent
Min. DIN 4 Pin with Lock (male)
23 14
KYCON KPPX-4P equivalent
Min. DIN 4 Pin with Lock (female)
23 14
KYCON KPJX-CM-4S equivalent
DIN 5 Pin (male)
13 425
Stripped and tinned leads
Type No. R6B
Type No. R7B
Type No. R7BF
Type No. R1B
Type No.
Pin Assignment
PIN No.
Output
1
+Vo
2
-Vo
3
+Vo
Pin Assignment
PIN No.
Output
1
+Vo
2
-Vo
3
-Vo
4
+Vo
Pin Assignment
PIN No.
Output
1
+Vo
2
-Vo
3
-Vo
4
+Vo
Pin Assignment
PIN No.
Output
1
-Vo
2
-Vo
3
+Vo
4
-Vo
5
+Vo
Pin Assignment
PIN No.
Output
L (red)
1 2
1
+Vo
L1 (black) by customer
Length of Land L1 by request
(MW’s standard length, L: 25 mm, L1: 5 mm)
( NOTE: The wire color is for reference only, please refer to the actual
product)
2
-Vo
Quick Adapter Accessory None Available
(Current rating: 7.5A max.)
None
None
None
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
File Name:GSM90A-SPEC 2022-11-21
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>