sauermann TM 110 Temperature Transmitter User Guide

June 5, 2024
sauermann

sauermann-TM-110-Temperature-Transmitter-logo

sauermann TM 110 Temperature Transmitter

sauermann-TM-110-Temperature-Transmitter-pro

Dimensions

sauermann-TM-110-Temperature-
Transmitter-1

Stainless steel probe technical specifications (duct model): Ø6 mm contact tip; available lengths: 100 mm, 150 mm, 200 mm, 250 mm or 300 mm.

Symbols used

For your safety and in order to avoid any damage of the device, please follow the procedure described in this document and read carefully the notes preceded by the following symbol:

The following symbol will also be used in this document, please read carefully the information notes indicated after this symbol:

Connections

sauermann-TM-110-Temperature-
Transmitter-2

  1.  Terminal block (only for model with terminal block)
  2.  Switch
  3.  LCC-S software connection
  4.  Output terminal block
  5.  Power supply terminal block
  6.  Cable gland

Connection of a Pt100 probe on the terminal block

Electrical connections as per NFC15-100 standard

This connection must be made by a qualified and trained technician. To make the connection, the  transmitter MUST NOT BE ENERGIZED.

  • For TM110 – AO and TM110 AN models with – 0-10 V output – active, 4 wires:sauermann-TM-110-Temperature-Transmitter-4
  • To make a 3-wire connection, BEFORE POWERING UP THE TRANSMITTER please connect the output ground to the input ground. See drawing below.sauermann-TM-110-Temperature-Transmitter-5
  • For TM110 – PO and TM110 – PN models with 4-20 mA output – passive:sauermann-TM-110-Temperature-Transmitter-6

Settings and use of the transmitter

Configuration

CAUTION: the transmitter MUST NOT BE ENERGIZED before a configuration. Then, proceed to the required settings, with the DIP switches (as shown on the drawings hereafter). Once the transmitter configured, it can be powered up.To configure the transmitter, unscrew the 4 screws of the casing then open it. The switches allow the different settings are accessible.sauermann-TM-110
-Temperature-Transmitter-7

Measuring range setting

To set a measuring range, put the switches 1, 2 and 3 of the measuring ranges as shown in the following table.sauermann-TM-110-Temperature-
Transmitter-8

Unit of measurement setting
To set the unit of measurement, put the on-off switch 4 as shown in the following table.sauermann-TM-110-Temperature-
Transmitter-9

Configuration via LCC-S software (option)

An easy and friendly configuration with the software! It is possible to configure intermediate ranges. sauermann-TM-110-Temperature-
Transmitter-10The minimum difference between the high range and the low range is 20. Example: for a 0-100 °C transmitter, the minimum delta of the range is 20 °C. It is then possible to configure the transmitter from 0 to +100 °C, or from 0 to +20 °C, etc. To access the configuration via software: Set the DIP switches as shown beside. Connect the cable of the LCC-S to the connection of the transmitter. The configuration of the parameters can be done either with the DIP switch or via software (you can not combine both solutions).

Mounting

sauermann-TM-110-Temperature-
Transmitter-11

To install the transmitter, mount the ABS plate on the wall (drilling: Ø6 mm, screws and pins are supplied). Insert the transmitter on the fixing plate (see A on the drawing beside). Rotate the housing in clockwise direction until you hear a “click” which confirms that the transmitter is correctly installed. The mounting plate is not available for ambient models. 4 fixing holes inside the back housing can be used to install the transmitter on the required location.

Maintenance: please avoid any aggressive solvent. Please protect the transmitter and its probes from any cleaning product containing formalin, that may be used for cleaning rooms or ducts.

Precautions for use: please always use the device in accordance with its intended use and within parameters described in the technical features in order not to compromise the protection ensured by the device.

References