TSI 4146 Primary Calibrators User Guide

June 5, 2024
tsi

PRIMARY CALIBRATORS MODEL 4046/4146
USER’S GUIDE
**P/N 1980512, REVISION F MARCH 2016

**

START SEEING THE BENEFITS OF REGISTERING TODAY’
Thank you for your TSI instrument purchase. Occasionally, TSI releases information on software updates, product enhancements, and new products. By registering your instrument, TSI will be able to send this important information to you.
http://register.tsi.com
As part of the registration process, you will be asked for your comments on TSI products and services. TSI’s customer feedback program gives customers like you a way to tell us how we are doing.
UNDERSTANDING, ACCELERATED
TSI Incorporated Visit our website www.tsi.com for more information.
USA Tel: +1 800 874 2811
India Tel: +9180 67877200
UK Tel: +44 149 4 459200
China Tel: +86 10 8219 7688
France Tel: +33 1 41 19 21 99
Singapore Tel: +65 6595 6388
Germany Tel: +49 241 523030
02015 T51 Incorporated Printed in U.S.A.
Copyright TSI Incorporated / 2004-2016 / All rights reserved.
Address TSI Incorporated, 500 Cardigan Road, Shoreview, MN 55126 USA
WARNING
TSI calibrators employ a heated platinum sensor. They should not be used in the presence of flammable or explosive gases or mixtures.
caution
TSI calibrators are not medical devices under FDA 510k and in no situation should they be utilized for human respiration measurements.
LIMITATION OF WARRANTY AND LIABILITY
(effective February 2015) (For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator’s manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator’s manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other consumable materials. Only the original manufacturer’s warranty applies;
e. This warranty does not cover calibration requirements, and the seller warrants only that the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO THE SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. THE BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT THE SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER’S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, THE SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL THE SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL, OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING, OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller’s factory shall be at Buyer’s risk of loss and will be returned, if at all, at Seller’s risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI’s Customer Service department at 800-874-2811 (USA) or (001 651) 490-2811 (International) or visit www.tsi.com.
Trademarks
TSI and TSI logo are registered trademarks of TSI Incorporated.

Unpacking and Parts Identification

Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components in Table 1. If any parts are missing or damaged, notify TSI immediately.
Table 1. List of Components (see figures 1-1 and 1-2)

Qty Item Description Part/ Model
1 Calibrator 4046 or 4146
1 Filter Model 4046: 22 mm 1S0 taper

Model 4146: 63 mm HEPA filter with 0.375-inch barbed fittings

|
| 1602292
| 1602342
1| Tubing Kit

Model 4046 with connectors and battery pack mounting lugs

Model 4146 with connectors and battery pack mounting lugs

|
| TK4000
| TK4100
1| Battery Pack| 4199
1| Soft Carrying Case|
| Model 4046| 1319403
| Model 4146| 1319404
1| Model 4046/4146 User’s Guide| 1980512
1| Dampening Module (Model 4146 only)| 1040229

Table 2. Optional Accessories

**Qty** **Item Description** Part/ Model
1 AC Adapter

115 V, North America, ungrounded 100–240 V, NEMA 5-15 plug, grounded 100–240 V, Europlug, CEE 7/16, grounded 100–240 V, Great Britain, grounded, fused 100–240 V, Australia/NZ| **** 2613033
8918-NA
8918-EC
8918-GB
8918-AT

  1. Calibrator

  2. Soft-Sided Carrying Case

  3. Mounting Lugs

  4. Tubing

  5. 6 AA-size Batteries

  6. Adapter

  7. User’s Guide

  8. High-Efficiency Inlet Filter

  9. Calibrator

  10. Soft-Sided Carrying Case

  11. Mounting Lugs

  12. Latex Tubing

  13. 6 AA-size Batteries

  14. High-Efficiency Inlet Filter

  15. User’s Guide

  16. Adapters

  17. Dampening Module

Chapter 2

Setting-Up

Assembly Instructions for the Calibrators The calibrator can be quickly attached to the battery pack by screwing the mounting lugs into the threaded holes in the base of the calibrator. The lugs should then be placed over the receiving holes in the top of the battery pack and snapped gently into place. Insert the connector from the battery pack to the receptacle in the back of the calibrator. The calibrator is now ready to use.
Supplying Power
The calibrator can be powered in one of two ways: through the power jack using the supplied 4199 battery pack or using an optional AC adapter. The DC power input connector is shown below along with the power requirements.
Power Supply: 7.5 VDC ± 1.5 V, 300 mA maximum
Connecting Filter and Flow Tubes
The Model 4046/4146 has a thermal sensor exposed to the airflow that must be protected from foreign matter and particles. TSI supplies a filter that should be connected to the inlet of the calibrator; however, any filter will work as long as it has a minimum efficiency of 99.9%.
Caution
Always use a filter on the inlet of the calibrator. Failure to filter the airflow may change the calibration and/or permanently damage the sensor.
Note: Flow direction is identified by the large arrow printed on the bottom side of the calibrator and on the bottom of the battery pack.
Attach the filter to the inlet of the calibrator using supplied tubing and/or adapters. Connecting a tube to the outlet of the calibrator will create backpressure. See Appendix A for calibrator accuracy specifications when operating at various pressures. In general, minimize back pressure on the calibrator by using shorter lengths of tubing to maintain the highest level of accuracy.
Sampling Pump with internal dampening chamber

Active Sampling Instrument with internal dampening chamber
Figure 2-2 Typical Setup Configurations

Use the 6 in. (15 cm) tubing to connect the dampening chamber inline between the Model 4146 Calibrator and the instrument. For best results, position Dampening Module with openings to the side, off the work surface.

Operation

Overview The Model 4046/4146 Calibrator measures the volumetric flow rate inside the flow tube. All measurements made by Models 4046/4146 are NIST traceable.
ON/OFF Switch
Slide the switch to the ON position. The LCD displays volumetric flow in units of liters per minute (L/min).
Flow Rate Measurement
Flow rate data will be displayed on the Model 4046/4146 through the LCD.
Volumetric flow is displayed in units of liters per minute (L/min).
Setting Flow Rates and Calibrating Instruments
The Model 4046/4146 Calibrator may be used to set flow rates and calibrate sampling pumps or other instruments which draw an active flow sample. Be sure to follow all instrument manufacturer’s flow settings and calibration procedures.

  1. Start up the Model 4046/4146 Calibrator and let it warm up for 1 minute prior to setting any flow rates or conducting any calibrations.
  2. Configure your sampling with the desired sampling media and sampling train to be used.
  3. Connect the outlet of the Model 4046/4146 Calibrator to the inlet of the sample media (used with sampling pumps) or to the inlet of the active sampling instrument using the tubing supplied
  4. Follow the instrument manufacturer’s flow setting and/or calibration procedures. The volumetric flow in units of liters per minute (L/min) will be continuously displayed on the LCD of the calibrator.

Dampening Module
The Dampening Module is included for use with systems that do not have internal dampening chambers. The Dampening Module (P/N 1040229) can be added inline between the end of the sampling train and the Model 4146 calibrator. The Dampening Module reduces flow pulsation with an insignificant effect on the flow rate or backpressure.
All diaphragm pumps produce flow pulsations during operation. Most commercially available 3 L/min and 5 L/min sampling pumps are designed with internal dampening chambers to minimize flow pulsations.
Some common air sampling instruments that may not have internal flow pulsation dampening chambers are listed below. Contact your instrument manufacturer if you have questions. This list is just an example and is not intended to be all-inclusive.

  • Low flow personal sampling pumps
  • Confined space meters
  • Photometers
  • Optical Particle
  • Counters (OPCs)
  • Condensation Particle Counters (CPCs)
  • Photo Ionization Detectors (PIDs)
  • Flame Ionization Detectors (FIDs)

Note:
The TSI DustTrakTM and SidePakTM Sampling Pumps contain internal dampening chambers. The Dampening Module is not needed to measure flow rates with these instruments.
The TSI SidePakTM AM510 Personal Aerosol Monitor does not contain an internal dampening chamber. TSI recommends using the Dampening Module when using the Model 4146 Calibrator to measure AM510 flow rates.

Maintenance

Flow Sensor Periodically inspects the flow sensor by looking into the outlet of the calibrator. Remove dust, particles, and fibers from the sensor with clean, dry compressed air. The flow sensor may break if touched. Never run liquids through the calibrator and never touch the sensor with a brush. Dust or other deposits on the flow sensor will degrade the 4046/4146’s flow accuracy.
Caution
The calibrator must be switched off for cleaning. Only use clean, dry, compressed air when attempting to remove contamination from the sensor.
Re-certification To maintain a high degree of confidence in the measurements made by the Model 4046/4146, TSI recommends that the instrument be returned to TSI every 12 months for re-certification. The calibrator will be recalibrated and returned with a certificate of calibration to US National Institute of Standards Technology (NIST) traceable standards.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft cloth dipped in isopropyl alcohol or mild detergent. Never submerge the calibrator or allow liquids to enter the flow tube.
Storage
When storing the calibrator, always cover the ends of flow tubes with the caps provided to prevent dust or other foreign matter from entering the tube.
(This page was intentionally left blank)

Troubleshooting

Table 3 lists the symptoms, possible causes, and recommended solutions for common problems encountered with Model 4046/4146 Calibrators. If the symptom is not listed, or if none of the solutions solves the problem, please contact TSI Technical Support at 1-800874-2811 or 651-490-2811.
Table 3. Troubleshooting

Symptom Possible Causes Corrective Action
No display. Unit not switched on. Switch on the unit.
No power to the instrument. Check power connection.
The temperature reads high at low or zero flows. The temperature sensor is
being heated from the flow sensor. The temperature value will track the

actual air temperature once the flow rate through the calibrator exceeds 1 Std L/min.
Flow readings fluctuate greatly.| The flow is fluctuating.| Improve inlet conditions or increased display averaging
time.
The display shows flow over the range with no flow passing through the flow tube.| The sensor may be damaged or broken.| Return calibrator to TSI for service.

Technical Contacts

If you have any difficulty installing the Model 4046/4146, or if you have technical or application questions about this instrument, contact an applications engineer at TSI Incorporated, at 651-490-2811 or contact answers@tsi.com.
If the Model 4046/4146 fails, or if you are returning it for service, visit our website at http://rma.tsi.com or contact TSI at:
TSI Incorporated 500 Cardigan Road Shoreview, MN 55126 USA
Phone: +1-800-874-2811
(USA) or +1 651-490-2811
E-mail: answers@tsi.com
TSI GmbH Neuköllner
Strasse 4 52068
Aachen
GERMANY
Telephone: +49 241-52303-0
Fax:+49 241-52303-49
E-mail:tsigmbh@tsi.com
Web: www.tsiinc.de
TSI Instruments Ltd.
Stirling Road
Cressex Business Park High
Wycombe, Bucks HP12 3ST
UNITED KINGDOM
Telephone: +44 (0) 149 4 459200
Fax: +44 (0) 149 4 459700
E-mail:tsiuk@tsi.com
Web: www.tsiinc.co.uk

Specifications

| **Specifications*
---|---
Flow Measurement
Measurement Range Accuracy Air| ** Model 4046:     2.5 to 300 L/min
Model 4146:     0.01 to 20 L/min
Model 4046:      ±2% of reading or 0.05 std. L/min, whichever is greater
Model 4146:     ±2% of reading or 0.005 std. L/min, whichever is greater
at standard conditions (21.1°C and 101.3 kPa) See notes 1 through 5 below.
Instrument Temp. Range Operation, Ambient Storage, Ambient| 0 to 50°C
-20 to 60°C
Physical Dimensions
External Dimensions Tube Adapters (Inlet & Outlet
Weigh Flow Body Material| Model 4046:     7.2 in. × 2.5 in. × 2.1 in.(18.3 cm × 6.4 cm × 5.3 cm)
Model 4146:     5 in. × 2 in. × 1.25 in. (12.7 cm × 5 cm × 3.2 cm)
Model 4046:     3.0

Model 4046:     22 mm male ISO taper Model 4146:     0.375 inch O.D. straight

Model 4046:     1.2 lbs (0.5 kg)
Model 4146:     1.7 lbs (0.8 kg)
Polycarbonate
Warm-up Time| 1 min
Power| Battery pack
7.5 VDC ± 1.5 V, 300 mA maximum

Notes:

  1. Accuracy is stated at standard conditions of 21.1°C and 101.3 kPa.
  2. Accuracy stated with air temperature and flow body temperature within ±10°C of one another.
  3. Accuracy stated measuring dry air (less than 10% R.H.).
  4. Includes ±0.5% of reading repeatability.
  5. The volumetric flow rate is calculated from the mass flow measurement. Add an additional 0.25% reading to the flow accuracy to account for the uncertainty in measuring air temperature and pressure.
    *Specifications are subject to change without notice.

Primary vs. Secondary Standard

By definition, a primary standard is one whose value is without relation or comparison to any other reference. There are three primary standards–mass, length, and time. There is no primary standard for airflow rate since the measurement is dependent on all three primary standards of mass, length, and time. Hence, all airflow calibrators are secondary standards in the strictest sense of the definition.
In the industrial hygiene arena, the terms “primary” and “secondary” have become generally accepted terms over time and should not be confused with actual primary and secondary standards as prescribed by the National Institute of Science and Technology (NIST). Some manufacturers of calibrators, by their own definition, refer to their products as “primary” standards referring to the manner they are applied and generally accepted for use in the industry. Other manufacturers adequately refer to their instruments as “calibrators”. There are no industry standards established for such designations based on the measurement technology being employed.
What is important is accuracy, repeatability, and regular, documented verification to a higher accepted standard such as NIST. Both primary and secondary calibrators must be verified on a regular basis and that testing recorded. Some guidelines recommend an annual calibration to a higher accepted standard while others suggest as frequently as every 30 days or before a significant survey. It is important to have a certificate of traceability to NIST or other industry-recognized testing organizations indicating that an instrument has been calibrated against some mother device verified by NIST.(This page was intentionally left blank)

TSI Incorporated
– Visit our website www.tsi.com for more information.
USA Tel: +1 800 874 2811
India Tel: +91 80 67877200
UK Tel: +44 149 4 459200
China Tel: +86 10 8219 7688
France Tel: +33 1 41 19 21 99
Germany Tel: +49 241 523030
Singapore Tel: +65 6595 6388
P/N 1980512 Rev. F
©2016 TSI Incorporated
Printed in U.S.A.
1980512

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals