TSI 7515 IAQ-Calc Indoor Air Quality Meter Owner’s Manual

June 5, 2024
tsi

7515 IAQ-Calc Indoor Air Quality Meter
**Owner’s Manual

IAQ-CalcTM Indoor Air Quality Meter**
Model 7515
Operation and Service Manual
P/N 1980571, Revision E April 2022
www.tsi.com

Start Seeing the Benefits of Registering Today!

Thank you for your TSI® instrument purchase. Occasionally, TSI® releases information on software updates, product enhancements and new products. By registering your instrument, TSI® will be able to send this important information to you.
http://register.tsi.com
As part of the registration process, you will be asked for your comments on TSI products and services. TSI’s customer feedback
Copyright
TSI Incorporated / May 2007-2022 / All rights reserved.
Address TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No. 651-490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective February 2015)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator’s manual, to be free from defects in workmanship and material for twenty-four (24) months, or the length of time specified in the operator’s manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications are warranted for 90 days from the date of shipment.
b. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material, under normal use, for 90 days from the date of shipment.
c. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other consumable materials. Only the original manufacturer’s warranty applies.
d. This warranty does not cover calibration requirements, and the seller warrants only that the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory-authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER’S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, THE SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL THE SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL, OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING, OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller’s factory shall be at Buyer’s risk of loss and will be returned, if at all, at Seller’s risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call the Customer Service department at 800-680-1220 (USA) or (1) 651-490-2860 (International).
Trademarks
TSI and TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other countries’ trademark registrations. IAQ-Calc is a trademark of TSI Incorporated.

Chapter 1 Unpacking and Parts Identification

Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI® immediately.

  1. Carrying case
  2. Instrument
  3. Calibration collar
    (This page was intentionally left blank)

Chapter 2 Setting-up

Supplying Power to the Model 7515 IAQ-CalcTM Indoor Air Quality Meter The Model 7515 is powered with four-size AA batteries.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Model 7515 is designed to operate with either alkaline or NiMH rechargeable batteries, although it will not recharge NiMH batteries. Battery life will be shorter if NiMH batteries are used. Carbon- zinc batteries are not recommended because of the danger of battery acid leakage.
Using the Integral Probe
The sensing probe relies on the diffusion of air. For best results, try to keep the sensing probe surrounded by moving air. DO NOT breathe on the probe. Humans exhale CO2 levels exceeding 10,000 ppm and it may take time for the probe to re-stabilize.

Chapter 3 Operation

Keypad Functions

ON/OFF Key| Press to turn the Model 7515 on and off. During the power-up sequence, the display will show the following:  Model Number, Serial Number, Software Revision, and Last Date Calibrated.
---|---
Sample Key| Press and release to start the sample, then press the END key to stop the sample.
Stats Key| Press to view the minimum, maximum, and average of the last sample taken.
CAL Key| Press to begin the field-calibration process. To re-calibrate in the field, first, connect zero gas to the instrument and open the regulator. The instrument will automatically make zero itself after one minute.
After zeroing the sensor, connect span gas to the instrument and open the regulator. The instrument will take one minute to make a reading. Then use the ▲▼ ENTER keys to adjust the concentration displayed on the instrument to match the concentration on the gas cylinder.

Chapter 4 Maintenance

The Model 7515 requires very little maintenance to keep it performing well.
Recalibration To maintain a high degree of accuracy in your measurements, we recommend that you return your Model 7515 to TSI® for annual recalibration. Please contact one of TSI’s offices or your local distributor to make service arrangements and to receive a Service Request number. To fill out an online “Service Request” form, visit TSI®’s website at tsi.com/service.
U.S. & International
TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: 800-680-1220
651-490-2860
Fax: 651-490-3824
The Model 7515 can also be recalibrated in the field using the CALIBRATION menu. These field adjustments are intended to make minor changes in calibration to match a user’s calibration standards. The field adjustment IS NOT intended as a complete calibration capability. For complete, multiple- point calibration and certification, the instrument must be returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft cloth and isopropyl alcohol or a mild detergent. Never immerse the Model 7515. If the enclosure of the Model 7515, it must be replaced immediately to prevent access to hazardous voltage. Storage Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage.

Chapter 5 Troubleshooting

Table 5-1 list the symptoms, possible causes, and recommended solutions for common problems encountered with Model 7515. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI® Incorporated.
Table 5-1: Troubleshooting the Model 7515

Symptom| Possible Causes|

Corrective Action

---|---|---
No Display| Unit is not turned on| Switch unit on.
Low or dead batteries| Replace batteries.
Dirty battery contacts| Clean the battery contacts.
High CO2 levels| Breathing on probe| Shield probe from breath and allow the instrument to re-stabilize.
Instrument Error message appears| Fault in instrument| Factory service is required on the instrument.

WARNING!
Remove the probe from excessive temperature immediately: excessive heat can damage the sensor. Operating temperature limits can be found in Appendix A, Specifications.
Appendix A

Specifications

Specifications are subject to change without notice.
CO2:
Range:……………………………………….0 to 5000 ppm
Accuracy1:………………………………….±3% of reading or ±50 ppm, whichever is greater
Resolution:…………………………………. 1 ppm
Sensor type:…………………………………. Non-Dispersive Infrared (NDIR)
Instrument Temperature Range:

Operating (Electronics): ………………..40 to 113°F (5 to 45°C)
Storage: ………………….-4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient overvoltage category II
External Meter Dimensions:
Meter Weight: ………………………3.3 in. 9.6 in. 1.8 in. (8.4 cm 24.4 cm 4.4 cm)
Weight with batteries:……………………… 0.6 lbs (0.27 kg)
Meter Display Dimensions:
Primary display:……… 4-digit LCD, 0.6 in. (15 mm) digit height
Secondary display:………………. 3.5-digit LCD, 0.3 in. (8 mm) digit height
Power Requirements:
Four AA-size batteries (included)
1 At 77°F (25°C). Add the uncertainty of ±0.2%/°F (±0.2%/°C) away from calibrated temperature

TSI Incorporated ­
Visit our website www.tsi.com for more information.
USA  Tel: +1 800 680 1220
UK Tel: +44 149 4 459200
France Tel: +33 1 41 19 21 99
Germany Tel: +49 241 523030
India Tel: +91 80 67877200
China Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
P/N 1980571 Rev E (US)
©2022 TSI Incorporated
Printed in the U.S.A.