TSI 9545 VelociCalc Air Velocity Meter Instruction Manual

June 4, 2024
tsi

9545 VelociCalc Air Velocity Meter

VelociCalc® Air Velocity Meter
Model 9545/9545-A
Operation and Service Manual
P/N 1980564, Revision F July 2022
www.tsi.com

Start Seeing the Benefits of Registering Today!
Thank you for your TSI® instrument purchase. Occasionally, TSI® releases information on software updates, product enhancements and new products. By registering your instrument, TSI® will be able to send this important information to you.
http://register.tsi.com
As part of the registration process, you will be asked for your comments on TSI products and services. TSI’s customer feedback program gives customers like you a way to tell us how we are doing.

Copyright TSI Incorporated / 2007-2022 / All rights reserved.
Address TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No. 651-490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective February 2015) (For country- specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator’s manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator’s manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components when indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on any fuses, batteries or other consumable materials. Only the original manufacturer’s warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER’S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF

THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller’s factory shall be at Buyer’s risk of loss, and will be returned, if at all, at Seller’s risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call Customer Service department at 800-680-1220 (USA) or (1) 651-490-2860 (International).
Trademarks TSI, the TSI logo, and VelociCalc are registered trademarks of TSI Incorporated in the United States and may be protected under other country’s trademark registrations. LogDat2 is a trademark of TSI Incorporated.

CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ………… 1
CHAPTER 2 SETTING-UP ……………………………………………………… 3 Supplying Power to the Model 9545/9545-A ……………………………. 3 Installing the Batteries ……………………………………………………… 3 Using the Optional AC Adapter …………………………………………. 3 Using the Telescoping Probe ………………………………………………… 3 Extending the Probe ………………………………………………………… 3 Retracting the Probe………………………………………………………… 4 Connecting to a Computer…………………………………………………….. 4
CHAPTER 3 OPERATION………………………………………………………. 5 Keypad Functions ………………………………………………………………… 5 Common Terms …………………………………………………………………… 5 Menus ………………………………………………………………………………… 6 DISPLAY SETUP ……………………………………………………………………… 6 SETTINGS …………………………………………………………………………….. 6 FLOW SET UP ………………………………………………………………………… 6 ACTUAL/STANDARD SETUP……………………………………………………….. 6 DATA LOGGING ……………………………………………………………………… 7 Measurements ………………………………………………………………… 7 Log Mode/Log Settings ……………………………………………………. 7 Delete Data…………………………………………………………………….. 7 % Memory ……………………………………………………………………… 7 LogDat2TM Downloading Software …………………………………………. 8
CHAPTER 4 MAINTENANCE …………………………………………………. 9 Recalibration ……………………………………………………………………….. 9 Cases…………………………………………………………………………………. 9 Storage ………………………………………………………………………………. 9
CHAPTER 5 TROUBLESHOOTING……………………………………….. 11
APPENDIX A SPECIFICATIONS …………………………………………… 13
iii

(This page intentionally left blank) iv

Chapter 1 Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI® Incorporated immediately.
1. Carrying case 2. Instrument 3. USB cable 4. Downloading software (available on TSI® website:
http://www.tsi.com/SoftwareDownloads).
1

(This page intentionally left blank)

2

Chapter 1

Chapter 2
Setting-up
Supplying Power to the Model 9545/9545-A The TSI® Model 9545/9545-A VelociCalc® Air Velocity Meter is powered with four size AA batteries.
Installing the Batteries Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Model 9545/9545-A is designed to operate with either alkaline or NiMH rechargeable batteries, although it will not recharge NiMH batteries. Battery life will be shorter if NiMH batteries are used. Carbon-zinc batteries are not recommended because of the danger of battery acid leakage.
Using the Optional AC Adapter When using the AC adapter, the batteries (if installed) will be bypassed. Be sure to provide the correct voltage and frequency, which is marked on the back of the AC adapter. The AC adapter will not recharge the batteries.
Using the Telescoping Probe The telescoping probe contains the velocity, temperature, and humidity sensors. When using the probe, make sure the sensor window is fully exposed and the orientation dimple is facing upstream.
NOTICE For temperature measurements, make sure that at least 3 inches (7.5 cm) of the probe is in the flow to allow the temperature sensor to be in the air stream.
Extending the Probe To extend the probe, hold the handle in one hand while pulling on the probe tip with the other hand. DO NOT hold the cable while extending the probe as this prevents the probe from extending.
3

Retracting the Probe To retract the probe, hold the handle in one hand while pulling gently on the probe cable with the other hand.
Connecting to a Computer Use the Computer Interface USB Cable provided with the Model 9545/9545-A to connect the instrument to a computer for downloading stored data.
For more information on how to download stored data see Chapter 3 section titled LogDat2TM Downloading Software.
CAUTION This symbol is used to indicate that the data port of the Model 9545/9545-A IS NOT intended for connection to a public telecommunications network. Connect the USB data port only to another USB port.

4

Chapter 2

Chapter 3 Operation

Keypad Functions
ON/OFF Key
Arrow () Keys (Enter) Key Arrow ( or ) and Menu Soft Keys

Press to turn the Model 9545/9545-A on and off. During the power up sequence the display will show the following: Model Number, Serial Number, Software Revision and Last Date Calibrated.
Press to scroll through choices while setting a parameter.
Press to accept a value or condition.
Press arrow keys to change choices while setting a parameter. Press the Menu soft key to select the Menu selections, which are Display Setup, Settings, Flow Setup, Actual/Std Set up, Data Logging and Calibration.

Common Terms
In this manual there are several terms that are used in different places. The following is a brief explanation of the meanings of those terms.

Sample Test ID
Time Constant

Consists of all of the measurement parameters stored at the same time.
A group of samples. The statistics (average, minimum, maximum, and count) are calculated for each test ID. The maximum number of test IDs is 100.
The time constant is an averaging period. It is used to dampen the display. If you are experiencing fluctuating flows, a longer time constant will slow down those fluctuations. The display will update every second, but the displayed reading will be the average over the last time constant period. For example, if the time constant is 10 seconds, the display will update every second, but the displayed reading will be the average from the last 10 seconds. This is also referred to as a “moving average”.

5

Menus
DISPLAY SETUP
Display setup menu is where you will setup the desired parameters to be displayed on the running screen. With a parameter highlighted you can then use the ON soft key to have it show up on the running screen or select the OFF soft key to turn off the parameter. Use PRIMARY soft key to have a parameter show up on the running screen in a larger display. Only one parameter can be selected as a primary, and up to 2 secondary parameters can be selected at one time.
SETTINGS
Settings menu is where you can set the general settings. These include Language, Beeper, Select Units, Time Constant, Contrast, Set Time, Set Date, Time Format, Date Format, Number Format, Backlight and Auto Off. Use the or soft keys to scroll through the settings for each option and use the key to accept settings.
FLOW SET UP
In Flow Setup mode, there are 4 types: Round Duct, Rectangle Duct, Duct Area, and Horn. Use the or soft keys to scroll through the types and then press the key to accept the desired type. To change the value, highlight the Enter Settings option and press the key. Use the instructions below to change the value of the type that you selected.
NOTICE
The horn numbers are the models of the horns. For example, 100 refers to a horn model number AM 100. Only horns with Model numbers as follows can be used with this function: AM 100, AM 300, AM 600, and AM 1200. If a horn model number is chosen, the instrument will return to measuring mode and use a preprogrammed curve to calculate flow rate from velocity.
ACTUAL/STANDARD SETUP
Choose Actual/Standard measurements and parameters in the Act/Std Setup menu. Within this menu, the user can also select Standard Temperature, Standard Pressure and a source for the actual temperature. The actual barometric pressure must be entered to convert air velocity and volume measurements to actual conditions.

6

Chapter 3

DATA LOGGING

Measurements
Measurements to be logged are independent of measurements on the display, and must therefore be selected under DATA LOGGING Measurements.

Log Mode/Log Settings
You can set Log Mode to Manual, Auto-save or Cont-key.
· Manual mode does not automatically save data, but instead prompts the user to save a sample.
· In Auto-save mode, the user manually takes samples that are automatically logged.
· In Cont-key mode, the user starts taking readings and logging by pressing the key. The instrument will continue taking measurements until the key is pressed again.
· Auto-save and Cont-Key modes have the following additional Log Settings:

Mode

Log Settings

Auto-save

Log Interval

Cont-key

Log Interval

· Pressing the keys simultaneously will lock the keypad to

prevent unauthorized adjustments to the instruments. To unlock the keypad, press the keys simultaneously.

Delete Data Use this to delete all data, delete test or delete sample.

% Memory
This option displays the memory available. Delete All, under Delete Data, will clear memory and reset the memory available.

Operation

7

LogDat2TM Downloading Software The Model 9545/9545-A comes with special software called LogDat2TM Downloading Software, which is designed to provide you with maximum flexibility and power.
LogDat2TM software and updates are available from the TSI® website at http://www.tsi.com/SoftwareDownloads. Follow online instructions to download software.
To download data from the Model 9545/9545-A, connect the supplied computer interface USB cable to the Model 9545/9545-A and to a computer USB port. Then run the LogDat2TM software. Within the LogDat2TM software, either select the tests to be downloaded or double-click on a test to open it.

8

Chapter 3

Chapter 4
Maintenance
The Model 9545/9545-A requires very little maintenance to keep it performing well.
Recalibration To maintain a high degree of accuracy in your measurements, we recommend that you return your Model 9545/9545-A to TSI® Incorporated for annual recalibration. Please contact one of TSI’s offices or your local distributor to make service arrangements and to receive a “Service Request” number. To fill out an online “Service Request” form, visit TSI’s website at tsi.com/service.
U.S. & International TSI Incorporated 500 Cardigan Road Shoreview MN 55126-3996 Tel: 800-680-1220
651-490-2860 Fax: 651-490-3824
The Model 9545/9545-A can also be recalibrated in the field using the CALIBRATION menu. These field adjustments are intended to make minor changes in calibration to match a user’s calibration standards. The field adjustment IS NOT intended as a complete calibration capability. For complete, multiple- point calibration and certification, the instrument must be returned to the factory.
Cases If the instrument case or storage case needs cleaning, wipe it off with a soft cloth and isopropyl alcohol or a mild detergent. Never immerse the Model 9545/9545-A. If the enclosure of the Model 9545/9545-A or the AC adapter becomes broken, it must be replaced immediately to prevent access to hazardous voltage.
Storage Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage.
9

(This page intentionally left blank)

10

Chapter 4

Chapter 5

Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Model 9545/9545-A. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI® Incorporated.

Table 5-1: Troubleshooting the Model 9545/9545-A

Symptom No Display

Possible Causes Unit not turned on Low or dead batteries

Dirty battery contacts

Velocity reading fluctuates / unstable

Fluctuating flow

Instrument Error Memory is full message appears

Fault in instrument

Corrective Action
Switch unit on.
Replace batteries or plug in AC adapter.
Clean the battery contacts.
Reposition probe in less-turbulent section or use longer time constant.
Download data if desired, then DELETE ALL memory.
Factory service required on instrument.

WARNING
Remove the probe from excessive temperature immediately: excessive heat can damage the sensor. Operating temperature limits can be found in Appendix A, Specifications.

11

(This page intentionally left blank)

12

Chapter 5

Appendix A

Specifications

Specifications are subject to change without notice.

Velocity Probe:

Range:

0 to 6000 ft/min (0 to 30 m/s)

Accuracy1&2: ±3% of reading or ±3 ft/min (±0.015 m/s),

whichever is greater

Resolution: 1 ft/min (0.01 m/s)

Duct Size: Range:

1.0 to 500 inches in increments of 0.1 in. (2.5 to 1270 cm in increments of 0.1 cm)

Volumetric Flow Rate:

Range:

Actual range is a function of actual velocity, and

duct size

Temperature from Velocity Probe:

Range:

14 to 140°F (-10 to 60°C)

Accuracy3: ±0.5°F (±0.3°C)

Resolution: 0.1°F (0.1°C)

Relative Humidity from Velocity Probe:

Range:

5 to 95% RH

Accuracy4: ±3% RH

Resolution: 0.1% RH

Wet Bulb Temperature from Velocity Probe:

Range:

40 to 140°F (5 to 60°C)

Resolution: 0.1°F (0.1°C)

Instrument Temperature Range:

Operating (Electronics): 40 to 113°F (5 to 45°C)

Operating (Probe): 14 to 140°F (-10 to 60°C)

Storage:

-4 to 140°F (-20 to 60°C)

Instrument Operating Conditions When Using the Optional AC Adapter: Altitude up to 4000 meters Relative humidity up to 80% RH, non-condensing Pollution degree 1 in accordance with IEC 664 Transient over voltage category II

13

Data Storage Capabilities:

Range:

12,700+ samples and 100 test IDs (one sample can

contain fourteen measurement types)

Logging Interval:

Intervals:

1 second to 1 hour

Time Constant: User selectable

Response Time: Velocity: Temperature: Humidity:

200 msec 2 minutes (to 66% of final value) <1 minute (to 66% of final value)

External Meter Dimensions: 3.3 in. 7.0 in. 1.8 in. (8.4 cm 17.8 cm 4.4 cm)

Meter Probe Dimensions:

Probe length:

40 in. (101.6 cm)

Probe diameter of tip: 0.28 in. (7.0 mm)

Probe diameter of base: 0.51 in. (13.0 mm)

Articulating Probe Dimensions:

Articulating section length:

6.0 in (15.24 cm)

Diameter of articulating knuckle: 0.38 in (9.5 mm)

Meter Weight: Weight with batteries: 0.6 lbs (0.27 kg)

Power Requirements: Four AA-size batteries (included) or AC adapter (optional) 9 VDC, 300 mA min.

1 Temperature compensated over an air temperature range of 40 to 150°F (5 to 65°C).
2 The accuracy statement of ±3.0% of reading or ±3 ft/min (±0.015 m/s), whichever is greater, begins at 30 ft/min through 6000 ft/min (0.15 m/s through 30 m/s).
3 Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F (0.03°C/°C) for change in instrument temperature.
4 Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C) for change in probe temperature. Includes 1% hysteresis.

14

Appendix A

Notes

Specifications

15

Notes

16

Appendix A

(This page intentionally left blank)

Specifications

17

TSI Incorporated ­ Visit our website www.tsi.com for more information.

USA

Tel: +1 800 680 1220

UK

Tel: +44 149 4 459200

France Tel: +33 1 41 19 21 99

Germany Tel: +49 241 523030

India

Tel: +91 80 67877200

China

Tel: +86 10 8219 7688

Singapore Tel: +65 6595 6388

P/N 1980564 Rev F

©2022 TSI Incorporated

Printed in U.S.A.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals