TSI 7525 IAQ-Calc Indoor Air Quality Meter Owner’s Manual
- June 5, 2024
- tsi
Table of Contents
TSI 7525 IAQ-Calc Indoor Air Quality Meter
Setting-up
Supplying Power to the Model 7525 IAQ-Calc™ Indoor Air Quality Meter
The Model 7525 can be powered in one of two ways: four size AA batteries or
the optional AC adapter.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on the inside of
the battery compartment. The Model 7525 is designed to operate with either
alkaline or NiMH rechargeable batteries, although it will not recharge NiMH
batteries. Battery life will be shorter if NiMH batteries are used. Carbon-
zinc batteries are not recommended because of the danger of battery acid
leakage.
Using the Optional AC Adapter
When using the AC adapter, the batteries (if installed) will be bypassed. Be
sure to provide the correct voltage and frequency, which is marked on the back
of the AC adapter. The AC adapter is not a battery charger.
Using the Probe
The sensing probe relies on the diffusion of air. For best results, try to
keep the sensing probe surrounded by moving air. Do not breathe on the probe.
Humans exhale CO2 levels exceeding 10,000 ppm and it may take time for the
probe to re-stabilize. Use the probe holder to support the probe when in
continuous data logging mode.
Connecting to a Computer
Use the USB Cable provided with the Model 7525 to connect the instrument to a
computer. For more information on how to
download stored data see Chapter 3 section titled LogDat2™ Downloading
Software.
C A U T I O N
This symbol is used to indicate that the data port of the Model 7525 IS NOT
intended for connection to a public telecommunications network. Connect the
USB data port only to another USB port.
Operation
Keypad Functions
ON/OFF Key| Press to turn the Model 7525 on and off. During the power-up
sequence, the display will show the following: Model Number, Serial Number and
Software Revision.
---|---
Arrow ( ) Keys| Press to scroll through choices while setting a
parameter. Pressing the keys simultaneously will lock the keypad to
prevent unauthorized adjustments to the instruments. To unlock the keypad,
press the keys simultaneously.
(Enter) Key| Press to accept a value or condition.
Arrow ( ) and Menu Soft Keys| Press arrow keys to change choices
while setting a parameter. Press the Menu soft key to select the Menu
selections, which are Display Setup, Settings, Barometric Pressure, Data
Logging, Applications, and Calibration.
Common Terms
In this manual there are several terms that are used in different places. The
following is a brief explanation of the meanings of those terms.
Sample| Consists of all of the measurement parameters stored at the same
time.
---|---
Test ID| A group of samples. The statistics (average, minimum, maximum,
and count) are calculated for each test ID. The maximum number of test IDs is
100.
Time Constant| The time constant is an averaging period. It is used to
dampen the display. If you are experiencing fluctuating flows, a longer time
constant will slow down those fluctuations. The display will update every
second, but the displayed reading will be the average over the last time
constant period. For example, if the time constant is 10 seconds, the display
will update every second, but the displayed reading will be the average from
the last 10 seconds. This is also referred to as a “moving average”.
Logging Interval| The logging interval is a frequency period that the
instrument will log readings. For example, if the logging interval is set to
30 minutes, each sample will be the average of the last 30 minutes.
---|---
Menus
DISPLAY SETUP
The display setup menu is where you will setup the desired parameters to be
displayed on the running screen. With a parameter highlighted you can then use
the ON soft key to have it show up on the running screen or select the OFF
soft key to turn off the parameter. Use the PRIMARY soft key to have a
parameter show up on the running screen in a larger display. Only one
parameter can be selected as a primary, and up to 2 secondary parameters can
be selected atone time.
SETTINGS
Settings menu is where you can set the general settings. These include
Language, Beeper, Select Units, Time Constant, Contrast, Set Time, Set Date,
Time Format, Date Format, Number Format, Backlight and Auto Off. Use the soft
keys to scroll through the settings for each option and use the key to accept
settings.
BAROMETRIC PRESSURE
The barometric Pressure menu is where you set the barometric pressure, which
can affect the CO2 measurement.
DATA LOGGING
Measurements
Measurements to be logged are independent of measurements on the display, and
must therefore be selected under DATA LOGGING → Measurements.
Log Mode/Log Settings
You can set Log Mode to Manual, Auto-save, or Cont-key.
- Manual mode does not automatically save data, but instead prompts the user to save a sample.
- In Auto-save mode, the user manually takes samples that are automatically logged.
- In Cont-key mode, the user starts taking readings and logging by pressing the key. The instrument will continue taking measurements until the key is pressed again.
- Auto-save and Cont-Key modes have the following additional Log Settings:
Mode| Log Settings
---|---
Auto-save| Log Interval
Cont-key| Log Interval - Pressing the keys simultaneously will lock the keypad to prevent unauthorized adjustments to the instruments. To unlock the keypad, press the keys simultaneously.
Delete Data
Use this to delete all data, delete test or delete sample.
% Memory
This option displays the memory available. Delete All, under Delete Data, will clear memory and reset the memory available.
APPLICATIONS
You can choose % Outside Air in the Applications menu. After choosing this application, take measurements or enter data for each line.
LogDat2™ Downloading Software
The Model 7525 comes with special software called LogDat2™ Downloading
Software, which is designed to provide you with maximum flexibility and power.
To install this software on your computer, follow the instructions on the
label of the LogDat2™ CD-ROM.
You can also download LogDat2™ software from the tsi.com website in the Service/Support page under Software. Once downloaded onto your PC, click the application file to begin installation. Follow the on-screen instructions to install the software. To download data from the Model 7525, connect the supplied computer interface USB cable to the Model 7525 and to a computer USB port. Then run the LogDat2™ downloading software. Within the LogDat2™ software, either select the tests to be downloaded or double-click on a test to open it.
Maintenance
The Model 7525 requires very little maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, TSI® recommends
that you return your Model 7525 to TSI® for annual recalibration. Please
contact one of TSI’s offices or your local distributor to make service
arrangements and to receive a Service Request number. To fill out an online
“Service Request” form, visit TSI® Incorporated’s website at tsi.com/service.
U.S. & International
TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: 800-680-1220
651-490-2860 Fax:
651-490-3824
The Model 7525 can also be recalibrated in the field using the CALIBRATION
menu. These field adjustments are intended to make minor changes in
calibration to match a user’s calibration standards. The field adjustment IS
NOT intended as a complete calibration capability. For complete, multiple-
point calibration and certification, the instrument must be returned to the
factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the Model 7525.
If the enclosure of the Model 7525 or the AC adapter becomes broken, it must
be replaced immediately to prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month to prevent
damage due to battery leakage.
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for
common problems encountered with the Model 7525. If your symptom is not
listed, or if none of the solutions solves your problem, please contact TSI®.
Table 5-1: Troubleshooting the Model 7525
Symptom | Possible Causes | Corrective Action |
---|---|---|
No Display | Unit not turned on | Switch unit on. |
Low or dead batteries | Replace batteries or plug in AC adapter. | |
Dirty battery contacts | Clean the battery contacts. | |
Humidity reading near zero or not believable | Probe exposed to intense light |
Shade the probe while taking samples.
High CO2 levels| Breathing on probe| Shield probe from breath and allow
instrument to re- stabilize.
No response to keypad| Keypad locked out| Unlock keypad by pressing keys
simultaneously.
Instrument Error message appears| Memory is full| Download data if desired
then
DELETE ALL memory.
Fault in instrument| Factory service required on instrument.
W A R N I N G !
Remove the probe from excessive temperature immediately: excessive heat can
damage the sensor. Operating temperature limits can be found in Appendix A,
Specifications.
Specifications
Specifications are subject to change without notice.
CO2:
-
Range: 0 to 5000 ppm
-
Accuracy1: ±3% of reading or ± 50 ppm, whichever is greater
-
Resolution: 1 ppm
-
Sensor type: Non-Dispersive Infrared (NDIR)
Temperature: -
Range: 32 to 140F (0 to 60C)
-
Accuracy2: ±1.0F (±0.6C)
-
Resolution: 0.1F (0.1C)
-
Response time: 30 seconds (90% of final value, air velocity at 2 m/s)
-
Display units: F or C (user selectable)
-
Type: Thermistor
Humidity: -
Range: 5 to 95% RH
-
Accuracy3: ±3% RH (includes ±1% hysteresis)
-
Resolution: 0.1% RH
-
Response time: 20 seconds (for 63% of final value)
-
Sensor type: Thin-film capacitive
Instrum ent Temperature Range: -
Operating (Electronics): 40 to 113°F (5 to 45°C)
-
Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions: -
Altitude up to 4000 meters
-
Relative humidity up to 80% RH, non-condensing
-
Pollution degree 1 in accordance with IEC 664
-
Transient over voltage category II
Data Storage Capabilities: -
Range: 30,300 data points with key (3) measured parameters enabled
Logging Interval: -
Intervals: 1 second up to 1 hour (user selectable)
Time Constant: -
Intervals: User selectable
External Meter Dimensions: -
3.3 in. 7.0 in. 1.8 in. (8.4 cm 17.8 cm 4.4 cm)
Meter Probe Dimensions: -
Probe Length: 7.0 in. (17.8 cm)
Probe Diameter of Tip: 0.75 in. (1.9 mm)
Meter Weight: -
Weight With Batteries: 0.6 lbs (0.27 kg)
Power Requirements: -
Four AA-size batteries (included) or AC adapter (optional) 9 VDC, 300 mA, 4-18 watts (input voltage and frequency vary depending on which adapter is used)
- Add ±0.2%/○F (±0.2%/○C) away from calibration temperature.
- At 77°F (25°C). Add the uncertainty of ±0.2%/°F (±0.2%/°C) away from calibrated temperature
- At 77°F (25°C). Add the uncertainty of ±0.03% RH/°F (±0.03% RH/°C) away from calibrated temperature.
TSI Incorporated – Visit our website www.tsi.com for
more information.
UK Tel: +44 149 4 459200
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>