Laserliner XP-083.036A MultiMeter Pocket

Laserliner XP-083.036A
MultiMeter Pocket-PRODUCCTFunction/Application

Multimeter for taking measurements in the range of overvoltage category CAT III up to max. 1000 V. The meter can be used to measure DC and AC voltages and direct and alternating currents within the speci ed ranges, as well as for continuity and diode testing. The meter is also fitted with a non-contact voltage detector and has an integrated LED FLashlight.

Symbols

  • Hazardous electrical voltage warning: Unprotected live components inside the device housing may pose a risk of electric shock.
  • Danger area warning
  • Protection class II: The test device has reinforced or double insulation.
  • Overvoltage category III: Equipment in fixed installations and for applications where specific requirements with regard to the reliability and availability of equipment have to be met, e.g. circuit-breakers in fixed installations and devices used in industrial applications which are permanently connected to the fixed installation.

Safety instructions

  • Make sure that you always select the correct connections, the correct rotary switch position and the correct range for the measurement to be taken.
  • Before measuring or checking the resistance, continuity, diodes or capacitance, disconnect the power supply to the electric circuit. Check that all high-voltage capacitors are discharged.
  • Isolate the device from all current sources before opening the battery compartment cover.
  • If possible, do not work alone.
  • If you have to take hold of the measuring spikes, do so by the grip sections only. Do not touch the measuring contacts whilst the measurement is being taken.
  • If the device comes into contact with moisture or another conductive residue, work must not be carried out under voltage. At and above voltages of 25 V AC/60 V DC, the presence of moisture creates the risk of life-threatening electric shocks.
  • Clean and dry the device before use. When using the device outdoors, make sure that the weather conditions are appropriate and/or that suitable protection measures are taken.
  • If you are working with voltages higher than 25 V AC/60 V DC, exercise extreme caution. Touching the electrical conductors at such voltages poses a risk of life-threatening electric shocks.
  • Do not use the device in environments in which there are conductive particles or where the occurrence of moisture (in the form of condensation, for example) can create transient conductivity.
  • The device must only be used in accordance with its intended purpose and within the scope of the speciAL cations.
  • If you are taking measurements in the hazardous vicinity of electrical installations, do not work alone and seek guidance from an electrically skilled person before starting work.
  • Before taking any measurements, make sure that both the area to be tested (e.g. a line), the test device and the accessories used (e.g. connection cable) are in proper working order. Test the device by connecting it to known voltage sources (e.g. a 230 V socket in the case of AC testing or a car battery in the case of DC testing). Stop using the device if one or a number of its functions fails.
  1. Rotary switch to set the measuring function
  2. ON/OFF FLashlight
  3. ON/OFF FLashlight
  4. LCD
  5. Sensor (non-contact voltage detector)
  6. Indicator (non-contact voltage detector)
  7. Holder for test prods
  8. Hold actual measured value
  9. Measuring contacts: red „+“, black „–“
  10. red „+“, black „–“
  • Measured value display (3.5 segments, 1999 digits)
  • Negative measured values
  • Direct (DC) or alternating (AC) variables
  • Automatic range selection
  • Diode test
  • Continuity test
  • Actual measurement is retained
  • Low battery charge
  • measurement units: mV, V, mA, Ohm, kOhm, MOhm
  • Display: O.L: Open line/over: measuring circuit not closed or measuring range exceededLaserliner XP-083.036A MultiMeter Pocket-4

AUTO OFF function

In order to preserve the batteries, the meter switches off automatically if it is left idle for 15 minutes.

  1. Insertion of batteries

  2. Attaching the test prodsLaserliner XP-083.036A MultiMeter Pocket-7 In order to avoid the risk of injury, the test prods must always be kept in the holder on the rear of the meter when not in use and during transport.

  3. Voltage measurement DC/ACLaserliner XP-083.036A MultiMeter Pocket-8Laserliner XP-083.036A MultiMeter Pocket-9 To take a voltage measurement, turn the rotary switch to position „V“ and press the „Mode“ button to set the voltage mode (AC, DC). Then connect the measuring contacts to the object to be tested. The measured value acquired and the polarity appear on the display.

  4. Resistance measurement To measure the resistance, turn the rotary switch to position „“. Then connect the measuring contacts to the object to be tested. The measured value acquired appears on the display. If „O.L.“ appears on the display instead of a measured value, either the measuring range has been exceeded or the measuring circuit is not closed or has been interrupted. Resistance can only be measured correctly in isolation; therefore, the components might need to be disconnected from the remainder of the circuit. When measuring resistance, to avoid the risk of the results of a measurement being distorted, there must be no traces of dirt, oil, solder spray or other contamination on the test prods.

  5. Continuity testLaserliner XP-083.036A MultiMeter Pocket-13 To test continuity, turn the rotary switch to position and press the „Mode“ switch twice to activate the „Continuity test“ function. Then connect the measuring contacts to the object to be tested. A measured value of < 150 ohms is recognized as continuity; this is confirmed by an audible signal. If „O.L.“ appears on the display instead of a measured value, either the measuring range has been exceeded or the measuring circuit is not closed or has been interrupted.

  6. Diode testLaserliner XP-083.036A MultiMeter Pocket-15 To test the diode, turn the rotary switch to position and press the „Mode“ switch once to activate the „Diode test“ function. Then connect the measuring contacts to the diode. The measured value acquired appears on the display. If „O.L.“ appears on the display instead of a measured value, the diode has either been tested in the reverse direction or is faulty.

  7. Current measurement DC/ACLaserliner XP-083.036A MultiMeter Pocket-16 To take a current measurement in the range from 0 to 200 mA, turn the rotary switch to position „mA“ and press the „Mode“ button to set the voltage mode (AC, DC). To take a current measurement in the range from 0 to 200 A, turn the rotary switch to position and press the „Mode“ button to set the voltage mode (AC, DC). Disconnect the circuit before connecting the meter. Then connect the measuring contacts to the object to be tested. The measured value acquired and the polarity appear on the display. Disconnect the circuit again before disconnecting the meter.
    No currents above 200 mA may be measured in the A/mA range! Currents above this level trigger the automatic fuse in the meter.

  8. Voltage detection, non-contact (AC warning) The non-contact voltage detector integrated in the meter is able to detect AC voltages from 100 V to 600 V. Even when the device is switched off, live cables or cable breaks still have to MODE HOLD be reckoned with. Run the voltage
    sensor along the object to be tested (5 – 10 mm). The display lights up if AC voltage is detected.
    Non-contact voltage detection is no substitute for conventional voltage testing. As the device detects an electrical field, it will react even to static charge.

Voltage detection, single-pole phase test

Connect the red test prod to the phase or neutral conductor. The red LED only lights up if the phase conductor is live. This function also works if the device is switched off. When the single-pole phase test is carried out on the outer conductor, the indicator function may be adversely affected under certain conditions (e.g. when insulating personnel protective equipment is used or at insulated locations).
The single-pole phase test is not suitable for checking for zero voltage. To do this, you need to carry out a two-pole phase test.

Flashlight function

To switch on the flashlight, press and hold down the corresponding button. The light switches itself off automatically as soon as the button is released.

Automatic fuse

The meter is equipped with an electronic auto-reset fuse in all ranges and can intercept switching errors under normal operating conditions. If the electronic fuse trips, de-energise the electrical circuit and disconnect the meter. Rectify the switching error. The device will resume normal operation when restarted.

CalibratioN

The meter needs to be calibrated and tested on a regular basis to ensure it produces accurate measurement results. We recommend carrying out calibration once a year.

Technical Data

(Subject to technical changes without notice. Rev18W48)

Technical Data

(Subject to technical changes without notice. Rev18W48)


Distance measurement
Precision (typical)*| ± 2 mm
Inside measurement range| 0,1 m – 50 m
Angle measurement**
Measuring range| ± 90°
Resolution| 0,1°
Precision| 0,1°
Laser class| 2 < 1 mW
Laser wavelength| 650 nm

Operating conditions

| -10°C … 40°C, max. humidity 80% rH, no condensation, max. working altitude 2000 m above sea level
Storage conditions| -20°C … 70°C, max. humidity 80% rH

Radio module operating data

| Bluetooth LE 4.x interface; Frequency band: ISM band 2400–2483.5 MHz, 40 channels; Transmission power: max. 10 mW; Bandwidth: 2 MHz;

Bit rate: 1 Mbit/s; Modulation: GFSK/FHSS

Automatic switch-off| 30-sec laser / 10 min device
Power supply| 2 x AAA 1.5 Volt batteries
Technische Daten (Subject to technical changes without notice)

Dimensions (W x H x D)| 43 x 110 x 27 mm
Weight (incl. batteries)| 100 g

  • measuring distance up to 10 m with strongly reflective target surface and at room temperature. The measurement deviation may increase by ± 0.2 mm/m for greater distances and under unfavourable measuring conditions such as strong sunlight or weakly reflective target surfaces.
    ** at max. 5000 lux

EU directives and disposal

This device complies with all necessary standards for the free movement of goods within the EU.
This product is an electric device and must be collected separately for disposal according to the European Directive on waste electrical and electronic equipment.
Further safety and supplementary notices at: http://laserliner.com/info?an=ACH

SERVICE
Umarex GmbH & Co KG
– Laserliner –
Möhnestraße 149, 59755 Arnsberg, Germany Tel.: +49 2932 638-300, Fax: +49 2932 638-333 info@laserliner.com
Umarex GmbH & Co KG Donnerfeld 2 59757 Arnsberg, Germany
Tel.: +49 2932 638-300, Fax: -333 www.laserliner.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Laserliner User Manuals

Related Manuals