WOHLER GS 300 Propane Gas Sniffer User Manual
- June 5, 2024
- WOHLER
Table of Contents
Manual Gas Sniffer
General Information
Operation Manual Information
This operation manual allows you to work safely with the Wöhler GS 300 Gas
Sniffer. Please keep this manual for your information.
The Wöhler GS 300 should be employed by professionals for its intended use
only.
Liability is void for any damages caused by not following this manual.
Notes
WARNING!
Not following this warning can cause injury or death.
ATTENTION!
Not following this note can cause permanent damage to the device.
NOTE!
Useful information
Intended Use
The Wöhler GS 300 Gas Sniffer is a gas leak detector for tightness control and for the leakage testing of gas tubes. The meter shows the concentration of methane and propane on a bar graph display from 0 ppm to 1200 ppm. It is suitable for the leakage detection of gas lines carrying combustible gases. The meter is not suitable for the measurement of the exact gas concentration. The meter can be used indoors only. Do not use the meter for any other use than set out in this manual.
Technical Data
Description | Data |
---|---|
Sensor | Semiconductor |
Range Methan | 0 to 1,200 ppm |
Propane | 0 to 1,200 ppm |
Alarm | intermittent beep and vibration alarm |
Response Time | 10 seconds, approx. |
Warm-up time | 60 seconds, approx. |
Environmental conditions: Operating Temperature | -5 to 45 °C |
Storage Temperature | -20 to 45 °C |
Rel. Humidity | <85% |
Auto-Off-Function | After 20 min. |
LC display | 31 x 51 mm |
Flashlight | 2 LEDs |
Power supply | 4 mignon cells AA |
Current consumption battery life | ca. 100 mA |
Dimensions | in general 14 hours (continuous working), depending on the type |
Flexible Probe | 190 x 55 x 40 mm |
Connections | 440 mm |
Important information
WARNING!
The Wöhler GS 300 may never be used in an environment where there is a risk of
explosion. If the environment is suspected to conglomerate potentially
explosive gases in no case you must use the Wöhler GS 300. Batteries may not
be taken from the device and/or be changed in such an environment.
ATTENTION!
Always store the device in a closed case at clean air. The sensor will be
damaged, if it is exposed to extreme conditions for a long time, e.g. a high
degree of relative humidity, extreme temperatures or pollution.
ATTENTION!
Avoid exposure to silicone adhesives, e.g. in cleaning tools. Silicone vapors
will cause damage to the sensor.
ATTENTION!
High-density exposure to corrosive materials such as H2S, SOx, Cl2, HCl, etc.
may cause corrosion or breakage of the lead wires or heater material.
ATTENTION!
Protect the sensor against salt water vapor and water. Sensor drift may occur
when the sensor is contaminated by alkaline metals, especially salt water
spray.
Light condensation under indoor usage should not be a problem for sensors.
However, if water condenses on the sensor‘s surface for a while, the sensor
characteristics may still drift.
ATTENTION!
If water freezes on the sensing surface, the sensor would crack and alter
characteristics.
ATTENTION!
Sensor performance may be also affected if exposed to a high-density gas for a
long time, regardless of the powering condition.
NOTE!
his meter cannot work well in zero or low oxygen atmosphere. This meter
requires to operate under around 21% ambient oxygen environment in order to
function properly.
Product
Meter
Fig. 8: Gas Leak Detector
-
Flexible Probe
-
On/Off-key
-
Mode-key: Selection Propane / Methane
-
Alarm-key: Selection beep / vibration alarm / beep and vibration alarm / no alarm
-
Light-sensor
-
Earphone jack
-
Mains connection
-
Flashlight
-
Battery Icon
Screen Fig. 9: Screen of the Gas Leak Detector
ATTENTION!
Change batteries as soon as the battery icon appears -
Vibration alarm
-
Acoustical alarm (beep)
-
without function
-
currently measured gas methane or propane
-
Calibration time, Countdown
-
Bar graph display
Operation
Switching the meter on and off
NOTE!
Switch on the meter in a noncontaminated environment, e.g. outside.
- Press the ON/OFF key.
After that, the Firmware Version will shortly appear on the screen.
The warm-up period will be 60 seconds. The warm-up period is automatically extended as follows if the meter has not been used for a longer period of time:
Last use of the meter | Warm-up period |
---|---|
≤ 6 days | 60 s |
˃ 6 days | 90 s |
˃ 16 days | 120 s |
˃ 31 days | 150 s |
˃ 51 days | 180 s |
180 s |
- After the warm-up period, the measured value of the selected gas will appear on screen (methane or propane).
- To switch off the meter, keep the On/Off key depressed for 3 seconds.
The meter will switch off after 3 seconds.
NOTE!
If the Wöhler GS 300 has not been used for a long time, we recommend repeating
the warm-up/calibration phase several times to achieve higher sensitivity and
accuracy.
• To do this, switch on the meter 2 or 3 times and off again after approx. 3
minutes each.
Selecting the measuring mode
Click the mode key to switch between the measuring modes methane and propane.
M (methane) or P (Propane) will be displayed in the upper right corner of the
display.
Alarm setting
Fig. 10:
Icons vibration alarm (1) and beep (2)
• Press the alarm key several times until the required alarm setting is
displayed. A press on the alarm key will successively activate the vibration
alarm and the beep together, the beep only, the vibration alarm only or it
will turn off the alarm.
The correspondent icons will appear on the screen. The meter beeps every one
second to assure the meter is in normal working status (if the beep has been
activated). The beep frequency will speed up while the detected gas
concentration becomes higher. While you use the earphone, the meter will
automatically mute but you can hear the beep through the earphone.
Leakage Checking
• Approaching the sensor to the pipe, slide slowly the sensor along the pipe;
repeat the procedure from the other side of the pipe.
When the sensor is near to a leak, the bar graph display will show the gas
concentration and the beep frequency or the vibration frequency will increase.
Flashlight
In a dark environment, the display illumination and the flashlight will switch
on automatically.
NOTE!
It is not possible to switch on or off the display illumination and the
flashlight manually.
After the leakage testing
NOTE!
If connecting lines, tools, and the meter are stored in the same case, the gas
may evaporate from the tools. The meter will detect this gas even after
several days.
Changing the batteries
Fig. 11: Battery compartment on the backside of the meter
When the battery voltage is low, a battery icon will appear on the screen.
ATTENTION!
Change batteries as soon as the battery icon appears
- Replace the batteries by 4 new mignon cells as follows:
- Open the battery compartment on the back of the unit. Replace the batteries. Observe the polarity indicated in the battery compartment.
NOTE!
If the batteries have been changed, the warm-up time after switching on is 3
minutes.
Charging the rechargeable batteries
Fig. 12: Power supply pack 9 V
It is also possible to operate the meter with 4 rechargeable batteries AA.
• For charging the rechargeable batteries connect the meter to the power
supply via the power supply pack. Plug the jack of the power supply unit into
the main connection at the bottom of the instrument, see fig. 1, part 7. When
charging is in progress the battery icon will blink.
When charging is complete the battery symbol will disappear from the screen.
WARNING!
Risk of electrical shock!
Never touch the power supply pack with wet hands!
Protect the power supply pack against water and moisture!
Do not unplug the recharger by pulling the cable!
Do not use the power supply pack when the voltage requirements of the
recharger and the supply do not match!
The batteries can be recharged even while they are located in the meter.
NOTE!
It is possible to operate the meter when charging is in progress.
ATTENTION!
Ensure rechargeable batteries are equipped in the meter before charging. Never
try to charge disposable batteries. Use 4 AA rechargeable batteries only.
NOTE!
To ensure maximum capacity use batteries of the same age and condition only.
Always replace a full set of batteries when necessary.
Calibration
We recommend sending the meter to Wöhler for calibration every year.
Warranty and Service
Warranty
Each Wöhler GS 300 Gas Sniffer will be tested in all functions and will leave our factory only after extensive quality control testing. If used properly, the warranty period for the Wöhler GS 300 Gas Sniffer will be twelve months from the date of sale. Batteries are not covered by this warranty. Service by nonauthorized personnel or making modifications to the analyzer voids any warranty.
Service
Wöhler has built our reputation on excellence in customer service. Therefore, of course, we are readily available to assist you after the warranty period ends.
- Send us the device and we will repair it and return it to you with our package service.
- Immediate help is provided by our technical staff over the telephone.
Declaration of conformity
The product
Produktname: Wöhler GS 300 Leakage Tester
complies with the key safety requirements set down in the guidelines of the
Council for the Harmonization of the Legal Requirements of the Member States
in relation to electromagnetic compatibility (2014/30/EU).
The following standards were availed of to evaluate the product in respect of
electromagnetic compatibility:
EN 61326 : 1997 + A1 : 1998 + A2 : 2001 + A3 : 2003
EN 61000-4-2: 1995+A1:1998+A2 : 2001
EN 61000-4-3: 2006
Bad Wünnenberg, 28.03.2022
Dr. Stephan Ester, Geschäftsführer/Managing Director
Accessories
Headphone with volume control | Order no. 55145 |
---|---|
Power supply pack 9 V | Order no. 4281 |
Rechargeable batteries NiMH | Order no. 9407 |
Information on disposal
Electronic equipment does not belong to domestic waste, but must be disposed
of in accordance with the applicable statutory provisions.
You may hand in any defective batteries taken out of the unit to our company
as well as to recycling places of public disposal systems or to selling points
for new batteries or storage batteries.
Direction
Wöhler Technik GmbH
Wöhler-Platz 1
33181 Bad Wünnenberg
Tel.: +49 2953 73-100
Fax: +49 2953 73-96100
E-Mail: info@woehler.com
Points of sale and service
USA
Wohler USA Inc.
208 S Main Street
Middleton, MA 01949
Tel.: +1 978 750 9876
www.wohlerusa.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>