WOHLER GS 300 Propane Gas Sniffer User Manual

June 5, 2024
WOHLER

WOHLER LOGO Manual Gas SnifferWOHLER GS 300 Propane Gas
Sniffer

General Information

Operation Manual Information

This operation manual allows you to work safely with the Wöhler GS 300 Gas Sniffer. Please keep this manual for your information.
The Wöhler GS 300 should be employed by professionals for its intended use only.
Liability is void for any damages caused by not following this manual.

Notes

WARNING!
Not following this warning can cause injury or death.
ATTENTION!
Not following this note can cause permanent damage to the device.
NOTE!
Useful information

Intended Use

The Wöhler GS 300 Gas Sniffer is a gas leak detector for tightness control and for the leakage testing of gas tubes. The meter shows the concentration of methane and propane on a bar graph display from 0 ppm to 1200 ppm. It is suitable for the leakage detection of gas lines carrying combustible gases. The meter is not suitable for the measurement of the exact gas concentration. The meter can be used indoors only. Do not use the meter for any other use than set out in this manual.

Technical Data

Description Data
Sensor Semiconductor
Range Methan 0 to 1,200 ppm
 Propane 0 to 1,200 ppm
Alarm intermittent beep and vibration alarm
Response Time 10 seconds, approx.
Warm-up time 60 seconds, approx.
Environmental conditions: Operating Temperature -5 to 45 °C
Storage Temperature -20 to 45 °C
Rel. Humidity <85%
Auto-Off-Function After 20 min.
LC display 31 x 51 mm
Flashlight 2 LEDs
Power supply 4 mignon cells AA
Current consumption battery life ca. 100 mA
Dimensions in general 14 hours (continuous working), depending on the type
Flexible Probe 190 x 55 x 40 mm
Connections 440 mm

Important information

WARNING!
The Wöhler GS 300 may never be used in an environment where there is a risk of explosion. If the environment is suspected to conglomerate potentially explosive gases in no case you must use the Wöhler GS 300. Batteries may not be taken from the device and/or be changed in such an environment.
ATTENTION!
Always store the device in a closed case at clean air. The sensor will be damaged, if it is exposed to extreme conditions for a long time, e.g. a high degree of relative humidity, extreme temperatures or pollution.
ATTENTION!
Avoid exposure to silicone adhesives, e.g. in cleaning tools. Silicone vapors will cause damage to the sensor.
ATTENTION!
High-density exposure to corrosive materials such as H2S, SOx, Cl2, HCl, etc. may cause corrosion or breakage of the lead wires or heater material.
ATTENTION!
Protect the sensor against salt water vapor and water. Sensor drift may occur when the sensor is contaminated by alkaline metals, especially salt water spray.
Light condensation under indoor usage should not be a problem for sensors. However, if water condenses on the sensor‘s surface for a while, the sensor characteristics may still drift.
ATTENTION!
If water freezes on the sensing surface, the sensor would crack and alter characteristics.
ATTENTION!
Sensor performance may be also affected if exposed to a high-density gas for a long time, regardless of the powering condition.

NOTE!
his meter cannot work well in zero or low oxygen atmosphere. This meter requires to operate under around 21% ambient oxygen environment in order to function properly.

Product

Meter

WOHLER GS 300 Propane Gas Sniffer - fig

Fig. 8: Gas Leak Detector

  1. Flexible Probe

  2. On/Off-key

  3. Mode-key: Selection Propane / Methane

  4. Alarm-key: Selection beep / vibration alarm / beep and vibration alarm / no alarm

  5. Light-sensor

  6. Earphone jack

  7. Mains connection

  8. Flashlight

  9. Battery Icon
    ScreenWOHLER GS 300 Propane Gas Sniffer - Screen Fig. 9: Screen of the Gas Leak Detector
    ATTENTION!
    Change batteries as soon as the battery icon appears

  10. Vibration alarm

  11. Acoustical alarm (beep)

  12. without function

  13.  currently measured gas methane or propane

  14. Calibration time, Countdown

  15. Bar graph display

Operation

Switching the meter on and off

NOTE!
Switch on the meter in a noncontaminated environment, e.g. outside.

  • Press the ON/OFF key.
    After that, the Firmware Version will shortly appear on the screen.
    The warm-up period will be 60 seconds. The warm-up period is automatically extended as follows if the meter has not been used for a longer  period of time:
Last use of the meter Warm-up period
≤ 6 days 60 s
˃ 6 days 90 s
˃ 16 days 120 s
˃ 31 days 150 s
˃ 51 days 180 s
180 s
  • After the warm-up period, the measured value of the selected gas will appear on screen (methane or propane).
  • To switch off the meter, keep the On/Off key depressed for 3 seconds.
    The meter will switch off after 3 seconds.

NOTE!
If the Wöhler GS 300 has not been used for a long time, we recommend repeating the warm-up/calibration phase several times to achieve higher sensitivity and accuracy.
• To do this, switch on the meter 2 or 3 times and off again after approx. 3 minutes each.

Selecting the measuring mode

Click the mode key to switch between the measuring modes methane and propane.
M (methane) or P (Propane) will be displayed in the upper right corner of the display.

Alarm setting

Fig. 10:
Icons vibration alarm (1) and beep (2)

• Press the alarm key several times until the required alarm setting is displayed. A press on the alarm key will successively activate the vibration alarm and the beep together, the beep only, the vibration alarm only or it will turn off the alarm.
The correspondent icons will appear on the screen. The meter beeps every one second to assure the meter is in normal working status (if the beep has been activated). The beep frequency will speed up while the detected gas concentration becomes higher. While you use the earphone, the meter will automatically mute but you can hear the beep through the earphone.

Leakage Checking

• Approaching the sensor to the pipe, slide slowly the sensor along the pipe; repeat the procedure from the other side of the pipe.
When the sensor is near to a leak, the bar graph display will show the gas concentration and the beep frequency or the vibration frequency will increase.

Flashlight

In a dark environment, the display illumination and the flashlight will switch on automatically.
NOTE!
It is not possible to switch on or off the display illumination and the flashlight manually.

After the leakage testing

NOTE!
If connecting lines, tools, and the meter are stored in the same case, the gas may evaporate from the tools. The meter will detect this gas even after several days.

Changing the batteries

 Fig. 11: Battery compartment on the backside of the meter

When the battery voltage is low, a battery icon will appear on the screen.
ATTENTION!
Change batteries as soon as the battery icon appears

  • Replace the batteries by 4 new mignon cells as follows:
  • Open the battery compartment on the back of the unit. Replace the batteries. Observe the polarity indicated in the battery compartment.

NOTE!
If the batteries have been changed, the warm-up time after switching on is 3 minutes.

Charging the rechargeable batteries

Fig. 12: Power supply pack 9 V

It is also possible to operate the meter with 4 rechargeable batteries AA.
• For charging the rechargeable batteries connect the meter to the power supply via the power supply pack. Plug the jack of the power supply unit into the main connection at the bottom of the instrument, see fig. 1, part 7. When charging is in progress the battery icon will blink.
When charging is complete the battery symbol will disappear from the screen.

WARNING!
Risk of electrical shock!
Never touch the power supply pack with wet hands!
Protect the power supply pack against water and moisture!
Do not unplug the recharger by pulling the cable!
Do not use the power supply pack when the voltage requirements of the recharger and the supply do not match!
The batteries can be recharged even while they are located in the meter.

NOTE!
It is possible to operate the meter when charging is in progress.
ATTENTION!
Ensure rechargeable batteries are equipped in the meter before charging. Never try to charge disposable batteries. Use 4 AA rechargeable batteries only.
NOTE!
To ensure maximum capacity use batteries of the same age and condition only. Always replace a full set of batteries when necessary.

Calibration

We recommend sending the meter to Wöhler for calibration every year.

Warranty and Service

Warranty

Each Wöhler GS 300 Gas Sniffer will be tested in all functions and will leave our factory only after extensive quality control testing. If used properly, the warranty period for the Wöhler GS 300 Gas Sniffer will be twelve months from the date of sale. Batteries are not covered by this warranty. Service by nonauthorized personnel or making modifications to the analyzer voids any warranty.

Service

Wöhler has built our reputation on excellence in customer service. Therefore, of course, we are readily available to assist you after the warranty period ends.

  • Send us the device and we will repair it and return it to you with our package service.
  • Immediate help is provided by our technical staff over the telephone.

Declaration of conformity

The product
Produktname: Wöhler GS 300 Leakage Tester

complies with the key safety requirements set down in the guidelines of the Council for the Harmonization of the Legal Requirements of the Member States in relation to electromagnetic compatibility (2014/30/EU).
The following standards were availed of to evaluate the product in respect of electromagnetic compatibility:
EN 61326 : 1997 + A1 : 1998 + A2 : 2001 + A3 : 2003
EN 61000-4-2: 1995+A1:1998+A2 : 2001
EN 61000-4-3: 2006

Bad Wünnenberg, 28.03.2022

Dr. Stephan Ester, Geschäftsführer/Managing Director

Accessories

Headphone with volume control Order no. 55145
Power supply pack 9 V Order no. 4281
Rechargeable batteries NiMH Order no. 9407
Information on disposal

Electronic equipment does not belong to domestic waste, but must be disposed of in accordance with the applicable statutory provisions.
You may hand in any defective batteries taken out of the unit to our company as well as to recycling places of public disposal systems or to selling points for new batteries or storage batteries.

Direction

Wöhler Technik GmbH
Wöhler-Platz 1
33181 Bad Wünnenberg
Tel.: +49 2953 73-100
Fax: +49 2953 73-96100
E-Mail: info@woehler.com

Points of sale and service

USA
Wohler USA Inc.
208 S Main Street
Middleton, MA 01949
Tel.: +1 978 750 9876
www.wohlerusa.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals