Digital Manometer Wohler DM 602 Digital Manometer

General Information

1.1 Start-up when delivered with optional battery pack

Wohler DM 602 Digital Manometer - fig

The bottom part of the battery module Wöhler M 602, lever marked by an arrow

If your unit is operated with a rechargeable battery and not with batteries (optional), the transport lock is set during delivery so that the unit cannot be switched on.

  • In this case, open the battery compartment on the back of the unit.
  • Move the small black lever at the bottom of the battery module from the off position to the on position.
    This procedure is only necessary before using the unit for the first time.

1.2 Manual information

This operation manual allows you to safely work with the Wöhler DM 602. Please keep this manual for your information.
The Wöhler DM 602 Pressure Gauge should be employed by professionals for its intended use only.
Liability is void for any damages caused by not following this manual.

1.3 Notes

WARNING!
Not following this warning can cause injury or death.
ATTENTION!
Not following this note can cause permanent damage to the device.
NOTE!
Useful information
1.4 Intended Use
The Wöhler DM 602 is used for differential pressure and temperature measurements.
In particular, it can be used to set the gas pressure on gas fireplaces, measure the fan differential pressure, and measure the chimney draught.

1.5 Basic equipment

Unit Scope of supply
Wöhler DM 602 Wöhler DM 602 Pressure Gauge

2 batteries 1,5 V
| Hand strap
| Measuring Hose 1,5 m

1.6 Hand strapWohler DM 602 Digital Manometer - Hand
strap

Fixing the hand strap under the battery compartment

If you want to use the hand strap included in the delivery, mount it as follows:

  • Remove the battery cover on the back of the unit.
  • Hang the loop on the small hook below the batteries.
  • Close the battery cover.

1.7 Information on disposal

Electronic equipment does not belong in domestic waste, but must be disposed of in accordance with the applicable statutory provisions.
You may hand in any defective batteries taken out of the unit to our company as well as to recycling places of public disposal systems or to selling points for new batteries or storage batteries.

1.8 Manufacturer

Wöhler Technik GmbH
Wöhler-Platz 1
33181 Bad Wünnenberg
Tel.: +49 2953 73-100
E-Mail: info@woehler.de

Specifications

2.1 Measured values
Differential Pressure

Description Data
Default unit Pa, bar, or in, depending on the device version
Range ± 3.000 hPa
Permissible overload ± 3.200 hPa
Precision < 3 % of reading, when < ± 2 hPa better than ± 0,06 hPa
Resolution 0,01 hPa

when –210 hPa to
+210 hPa, otherwise 0,1 hPa

Internal temperature measurement

Description Data
Default unit °C or °F, depending on the device version
Range -20 °C to 60°C
Precision < ± 1 °C
Resolution 0,1°C

External Temperature Measurement
(e.g. with temperature measuring clamp )

Description Data
Range -20,0 °C to +800,0 °C
Precision < ± 1 °C

when -20°C to 67°C, otherwise 1,5 % of reading according to EN 50379-1
Resolution| 0,1°C

2.1 Technical Data

Description Data
Power Supply 2 batteries 1,5 V, AA LR6
Power supply (option) Lithium polymer battery 3,7 V, 700 mAh, Charging via

USB-C cable
Battery Charging Cycles (option)| After 500 charging cycles, at least 70 % of the capacity is still available.
Battery charging time (option)| 4 h
Operating time of the rechargeable battery (option)| 38 h
Auto Off Function| After 1 hour (can be deactivated in the settings)
Storage Temperature| -20 °C to + 50 °C
Operating Temperature| +5 °C to 40 °C for compliance with the accuracies
Weight| 140 g
Dimension| 170 x 47 x 29 mm

Component explanation

Wohler DM 602 Digital Manometer - fig1Figure 3: Wöhler DM 603 Digital Manometer

Button (1) Press Button Funktionen
__ 3 seconds on/off
Measuring mode briefly zero the pressure sensor
double click print measuring data on thermoprinter
Settings mode briefly select unit / solution / display lighting
double click confirm setting
Number Function
--- ---
1 Button
2 USB-C- charging port (option, only with rechargeable battery)
3 Display
4 Pressure socket (right) Main socket for differential pressure measurement

(positive)
5 Pressure socket (left)| Reference socket for differential pressure measurement (negative)
6| Jack for external temperature probe or clamp NiCr-Ni (Type K)
7| The battery compartment on the backside of the unit
8| IR Port (Thermoprinter)

Wöhler DM 602 external temperature sensor

Temperature measurement connection:
For external temperature measurement, any type of K the thermocouple can be connected, e.g. a temperature clamp or a temperature probe.

  • Insert the temperature plug into the temperature socket on the underside of the unit (Abb. 3, part 6).

Switch on unit

Figure 5: Button

WARNING!
Due to the compressibility of gases, the accident prevention regulations “Working on Gas Installations” and the rules and regulations “Technical Rules for Gas Installations”must  be observed for all pressure tests with air or inert gas for safety reasons.

  •  Switch on the unit: Press and hold the button for 2 seconds.
  • Switch off the unit: Press and hold the button for 3 seconds.

After switching on, the overall display lights up briefly and then the start screen lights up for approx. 7 seconds.
During this time, the unit carries out a pressure zeroing.

ATTENTION!
During zeroing, no hose may be connected or no differential pressure applied, as the unit stabilizes and determines its zero point.
After zeroing the measuring display appears.

Display

In measuring mode, the main display shows the currently measured pressure value and below the temperature value.
Top line:
Battery level indicator:

  • 3 dashes: high battery voltage
  • Flashing: very low battery voltage

Middle line:
Actual measured differential pressure
Bottom line:
Internally measured temperature or externally measured temperature (with a connected temperature sensor or clamp)
The differential pressure is always shown in the display without (+) or (-).
Instead, there is a plus sign (+) under the pressure port through which the higher value is measured and a minus sign (-) under the pressure port with the lower value.

Note!
We recommend connecting the overpressure or underpressure to be measured to the right-hand connection and leaving the left-hand connection open as a reference connection.

Wöhler DM 602 external temperature sensor

Temperature measurement connection:
For external temperature measurement, any type of K thermocouple can be connected, e.g. a temperature clamp or a temperature probe.

  • Insert the temperature plug into the temperature socket on the underside of the unit.

Settings

  • Immediately after switching on, as soon as the battery symbol has appeared in the display and before the measured value display appears, double-click the button.

This will take you to the setting mode.

6.1 Selecting the pressure unit

The display shows „Press unit“.

  • To select the pressure unit, press the button briefly.
    The pressure units bar, bar, WC, psi, mmH2O, mH2O, Pa, hPa, kPa, and mPa appear one after the other.

  • Confirm the selected unit with a double click.

6.2 Selecting the pressure solution

The display shows „reso“ (resolution).
You can now set the displayed resolution for the pressure. The resolution refers to the previously selected pressure unit. There are 3 different resolutions to choose from.

  • Press the button briefly to select the desired resolution.
  • Confirm the selection with a double click.

6.3 Selecting the temperature unit
The display shows „Temp. unit“.
You can now select between the temperature units °C or °F.

  • To change between °C and °F press the button briefly.
  • Confirm the selection with a double click.

6.4 Switching the display lighting on/off

The display shows „Light on“.

  • Press the button briefly to switch the display lighting on and off again.
  •  Confirm the selection with a double click.

6.5 Auto-Off Function

“Auto Off” appears on the display.
The unit switches off automatically after one hour if no key has been pressed. This auto-off function can be deactivated at this point.

  • Press the operating key briefly to switch between “Auto Off” (automatic switch-off activated) and “Ever ON” (automatic switch-off deactivated).
  • Confirm your selection with a double click.
    The display then switches back to the start screen for a few seconds and then to the measuring mode.

Changing the batteries

  • Switch off the unit.
  • Remove the battery cover from the backside of the unit.
  • Replace the batteries with 2 new 1.5 V AA LR6 batteries.
  • Observe the correct polarity according to the labeling in the battery compartment.

Maintenance

Interval Maintenance work
When necessary (user) Clean the housing with a damp cloth and mild detergent.
1x per year (factory) Send the unit to one of our service centers for in-

section and calibration.

Accessories

Draught probe with hose and cone Art. no. 4578

Service work on gas pipes

Pressure Measurement Set for Wöhler Digital Manometer Art. no. 56126
High-pressure plug ½ “ and 1“ Art. no. 7213
Hose set with measuring hose Ø 8 mm for measurements up to1.5 bar Art. no.

11341

Temperature

Clamp temperature sensor type K K Art. no. 62616

Printer

Wöhler TD 100 Thermal Printer Art. no. 4160
Thermal Paper, 10 rolls Art. no. 4145

Warranty and Service

10.1 Warranty

Each Wöhler Digital Manometer DM 602 will be tested in all functions and will leave our factory only after extensive quality control testing. The final control will be recorded in detail in a test report and delivered with any unit.
If used properly, the warranty period for the Wöhler DM 602 will be twelve months from the date of sale. Not covered by the warranty are batteries and defects of the pressure sensor caused by overpressure.
Service by non-authorized personnel or making modifications to the unit void any warranty.

10.2 Service

Wöhler has built our reputation on excellence in customer service. Therefore, of course, we are readily available to assist you after the warranty period ends.

  • Send us the device, and we will repair it and return it to you with our package service.
  • Immediate help is provided by our technical staff over the telephone.

Declaration of conformity

The manufacturer:
WÖHLER Technik GmbH
Wöhler-Platz 1, D-33181 Bad Wünnenberg  declares that the product:

Product Name: Wöhler DM 602 Digital Manometer

complies with the key safety requirements set down in the guidelines of the Council for the Harmonization of the Legal Requirements of the Member States in relation to the
electromagnetic compatibility 2014/53/EU and 2014/30/EU.
The following standards were availed to evaluate the product in respect of electromagnetic compatibility:

EN 55011:2016 + A1:2017
EN 55032:2015
EN 61000-4-2:2009
EN 61000-4-3:2006 + A1:2008 + A2:2010
Bad Wünnenberg, 06.07.2022

Dr. Stephan Ester, Geschäftsführer/Managing Director

Short Guide

Mode Press button Function
Every mode 3 s on/off

Measuring mode

| briefly| zero the pressure sensor
double click| print measuring data on thermoprinter


Settings mode

| briefly| select unit / solution / display lighting
double click| confirm setting