sauermann GH 500 Hygrometry Bench Generator User Manual

June 4, 2024
sauermann

sauermann logo.0 GH 500
Hygrometry generatorsauermann GH 500 Hygrometry Bench Generator -
fig

Security

1.1 Precautions for use
Please always use the device in compliance with the intended use and in the limit of described parameters in the technical
features in order not to compromise the protection ensured by the device.

  • The device must be connected on a low voltage installation in compliance with the regulation of the country where the device is installed (in France: NFC 15-100).
  • Every fuse replacement must be realized with the switched-off device by using a recommended fuse in the documentation (see the technical features of the device: please see the technical datasheet).
  • Disconnect the power supply cable to switch off the device.
  • The device is not adapted for ATEX areas according to the current regulations

1.2 Symbols used
For your safety and in order to avoid any damage to your device, please read the described processes in this user manual and read carefully the notes preceded by the following symbol: ****
The following symbol is also used in this user manual:
Please read carefully the information notes indicated after this symbol.

Instrument description

GH500 instrument is the key instrument to adjust and calibrate in ambient air your measuring instruments in hygrometry. Easy-to-use, it will guarantee fast and stable hygrometry generations.
2.1 Schematic drawing

sauermann GH 500 Hygrometry Bench Generator

2.2 Dimensions (in mm)

sauermann GH 500 Hygrometry Bench Generator -
Dimensions

2.3 Description

sauermann GH 500 Hygrometry Bench Generator -
Description

01. K38 controller with adjustable setpoints| 08. Overpressure control exhaust outlet
---|---
02. K38 keypad: selection of the setpoint value| 09. Measuring chamber with front opening
03. CPE 310-S transmitter for measurement of the chamber| 10. Cable gland for probes
04. CPE 310-S keypad: configuration of the measurement transmitter 05. Blower| 11. Moinstening cartridge
05. Blower| 12. Drying cartridge
06. SHSI thermo-hygro measuring probe for CPE 310-S| 13. On/Off key
07. Metallic mounting plate to magnetize the instruments| 14. Power supply

3.1 K38 display description

1 : OUT1: pump of the wet air circuit in operation
2 : OUT2: pump of the dry air circuit in operation
3: -: the value of the measured humidity is lower than the setpoint value
4 =: value of the measured humidity is stabilized at the setpoint value
5 : +: value of the measured humidity is higher than the setpoint value

3.2 Operating principles
The GH 500 is a device that allows generating points of relative humidity in a range from 10 to 85% HR. It works on the principle of a mix of moist air and dry air in different proportions, in order to obtain the required relative humidity value. For the generator works optimally, it must be placed at the most stable room temperature (between 20 and  25 °C). This condition will define the homogeneity in the humidity of the generator. Once the generator is properly installed and turned on, simply select the threshold value of the required relative humidity.

3.3 Enter a setpoint value
To enter a setpoint value :
➢ Press on P The screen starts flashing in displaying SP1 and the actual setpoint value.
➢ Press on to increase the setpoint value or on to decrease the setpoint value.
➢ Press on P to validate.
➢ Press on U  for 3 seconds to return to the normal display.
3.4 Performances
The stability of the GH500 regulation allows having a regulated setpoint value at ± 0.1% RH. It is able to pass from 10 to 85 % RH in 35 minutes and from 85 to 10% RH in 30  minutes. However, the durations can vary according to the consumables wear condition, if the rise and fall times become longer, verify their conditions.
The homogeneity in dry temperature is 0,1°C in a restricted volume, it is therefore advised in order to have a precise measurement to put the probe in the middle of the housing.
3.5 Use advice
The measuring chamber must be always airtight. The ball of the overpressure control exhaust outlet (see 08 instrument description chapter) must turn when the pump is in operation.
If not, check that the front part of the measuring chamber is well closed and that the cable glands are well blocked.

Change the consumables

4.1 Regeneration of the desiccant pearls
Before using the device, put a cotton filter at each end of the desiccant pearls tube.
The desiccant pearl is a material that allows to drain the air; the little balls pick up moisture: their colors pass from orange to transparent. When the color becomes transparent,  the desiccant pearls’ absorption capacity is reduced and it must be regenerated. One hour in an oven at a temperature of 100°C allows a proper recovery of their properties.  However, after several regeneration periods, their capacity can be degraded and it is necessary to change them.

4.2 Maintenance of the bubbler
The bubbler allows the loading of humidity in the air. For an optimal operation, the level of water must be maintained. You must use distilled water to avoid stagnation and clean the bubbler if the water changes color or becomes opaque. It is important not to exceed the maximum level indicated at risk of seeing the water overflow from its tank and infiltrate the pneumatic circuit.

Cable glands

sauermann GH 500 Hygrometry Bench Generator - display1 The M36 thread is adapted to insert a mirror for example.

 Transport case

**** During the transportation, please empty the cartridge of demineralized water. Unclip the drying cartridge and empty it in the flask n°2.

sauermann GH 500 Hygrometry Bench Generator - display2

sauermanngroup.com
Ntang_GH500 – 12/07/16 – RCS (24) Périgueux 349 282 095 Non-contractual document –
We reserve the right to modify the characteristics of our products without prior notice. sauermann logo.0