CHAUVIN ARNOUX C.A 863 Digital Thermometer User Manual

June 17, 2024
CHAUVIN ARNOUX

CHAUVIN ARNOUX C.A 863 Digital Thermometer

Product Information

Specifications

  • Model: C.A 861, C.A 863
  • Type: Hand-held thermometer
  • Display: LCD display with backlight
  • Measurement Type: Temperature
  • Sensor Type: K-type thermocouple
  • Weight: Lightweight

Description

The C.A 861 and C.An 863 are high-precision hand-held temperature measurement instruments. They are lightweight and come with a digital display. These instruments are easy to use and support interchangeable K-type thermocouples.

Converters for K-type thermocouple sensors
The instrument supports converters for K-type thermocouple sensors.

LCD, back-lit
The instrument features an LCD with a backlight for easy reading in various lighting conditions.

Symbols

  • K: Indicates a K-type thermocouple sensor is connected
  • Low battery: Indicates low battery level
  • Last measured value: Displays the last measured temperature value

2 function button

The instrument has a button with two functions:

  • Long press (>2 secs): Allows the user to choose the unit used for measurement
  • Short press: Toggles display backlighting on and off

Push button
The push button has two functions:

  • Short press: Stores the last measurement value
  • Long press (>2 secs): Displays the maximum measured value (only available in C.A 863 model)

3 position selector switch

The selector switch has three positions:

  • C.A 861: Command push button for maximum temperature value
  • C.A 863: Scroll-down menu for switching between T1, T2, and T

Shock-proof molded surround
The instrument is designed with a shock-proof molded surround for added durability.

Usage Instructions

Procedure

  1. Connect one or two sensors to the instrument.
  2. Switch on the thermometer.
  3. Select one or two measurement modes based on the number of sensors connected.

Maintenance
Refer to the user’s manual for maintenance instructions.

Warranty
The warranty is applicable for twelve months from the date of purchase. Please refer to our General Conditions of Sale for more details.

FAQ

What type of thermocouple sensors does the instrument support?
The instrument supports K-type thermocouple sensors.

Can I use the instrument with sensors raised to voltages exceeding 24 V AC/DC?
No, to prevent any risk of deterioration to the thermometer, do not use it with sensors raised to voltages exceeding 24 V AC/DC as the sensors are electrically connected to the thermometer input.

What should I do if the thermometer is subjected to great variations in ambient temperature?
If the thermometer is subjected to great variations in ambient temperature, wait for 20 minutes after it becomes stable again before continuing with the measurement.

Meaning of the symbol
Warning! Please refer to the User’s Manual before using the instrument.In this User’s Manual, the instructions preceded by the above symbol, should they not be carried out as shown, can result in a physical accident or damage the instrument and the installations.

Definition of CAT III
This instrument, in overvoltage category III, meets with the strict reliability and environmental requisites that apply to fixed industrial and domestic installations (see IEC 664-1 ed 92).

Thank you for purchasing a C.An 861 or C.An 863 hand-held thermometer.
To get the best service from this instrument:

  • read this user’s manual carefully,
  • respect the safety precautions detailed

SAFETY PRECAUTIONS!

  • To prevent any risk of deterioration to the thermometer, do not use sensors raised to voltages exceeding 24 V AC/DC, since the sensors are electrically connected to the thermometer input.
  • If the thermometer is subjected to great variations in ambient temperature, wait for 20 minutes after it becomes stable again before continuing with the measurement.
  • Respect the climatic environmental conditions (see § 4)

WARRANTY

Our guarantee is applicable for twelve months after the date on which the equipment is made available (extract from our General Conditions of Sale, available on request).

PRESENTATION

The C.A 861 and C.A 863 are high precision hand-held temperature measurement instruments, lightweight, with a digital display, and easy to use with interchangeable K-type thermocouples.

DESCRIPTION

See § 7. Appendix (found at the end of this user’s manual)

  1. Converters for K-type thermocouple sensors

  2. LCD display, back-lit

  3. Main display: digital temperature value (±) 3½ digits or OL error

  4. Symbols :

  5. K K-type thermocouple sensor

  6. Low battery

  7. Last measured value

  8. °C/°F Unit of displayed value

  9. MAX Maximum temperature value

  10. T1-T2 Difference ΔT between T1 and T2 (C.A 863) (see § 3.2)

  11. 2 function button:

  12. Long press > 2 secs: choice of unit used

  13. Short press: switches display backlighting off and on

  14. Push button:

  15. Short press: stores last measurement display Held press > 2 s : Max. measured value (C.A 863)

  16. 3 position selector switch:

  17. Thermometer off

  18. 1°C or °F Resolution

  19. 0.1°C or °F Resolution

  20. C.A 861 :
    Command push button for max. temperature value C.A 863 : Scroll-down menu for going from T1 to T2 and then ΔT.

  21. Shock-proof moulded surround

USE

Procedure

  1. Connect one or two sensors
  2. Switch on the thermometer: select one or two measurement resolutions, using selector switch ➄, the measurement isinstantaneously displayed.
  3. Set-up the thermometer according to the type of measurement to be carried out, using buttons ➂, ➃ and/or ➅ (see § 2 and 3.2).
  4. Read off the measured value.
    The thermometer is switched off by putting switch ➄ back in the OFF position.

Note :
If no sensor is connected, OL Is displayed.
If the temperature to be measured is outside of the instrument’s measurement range, then OL is displayed

Special Functions (see § 2. Description)

  • HOLD : as soon as this button is pressed, the instrument holds the last measurement displayed.
  • MAX : the instrument displays the maximum value of the temperature during measurement

On the C.A 863 :
ΔT : With two sensors the instrument simultaneously measures temperatures T1 and T2 and displays the difference ΔT between these two measurements.

SPECIFICATIONS

  • The temperature range depends upon the K type thermocouple chosen (see section 5):

    • C.A 861 : -40 to +1350°C (-40 to 2000°F)
    • 40 to +200°C (-40 to +392°F) with K type thermocouple supplied as standard.
    • C.A 863 : -50 to +1300°C (-58 to 2000°F)
    • 50 to +200°C (-58 to 392°F) with the K type thermocouple chosen (see section 5).
  • Resolution: 0.1°/1°

  • Polarity indication: automatic

  • Accuracy (ambient temperature 18°C to 28°C)

    • C.A 861:
      ±0.1% of reading +1°C from -40°C to 1350°C
      ±0.1% of reading +2°F from -40°F to 1999°F

    • C.A 863:
      ±0.3% of reading +1°C from -50°C to 1000°C
      ±0.5% of reading +1°C from -1000°C to 1300°C
      ±0.3% of reading +2°F from -58°F to 1999°F

  • Interchangeable K-type sensors (CHROMEL-ALUMEL)

  • Operating environment
    0 to +50°C, < 80% RH

  • Storage environment
    -20°C to +60°C, 0 to 80% RH, without battery

  • Temperature drift
    0.1 x (specified accuracy) by °C from 0 to +18°C and from
    +28 to +50°C (+32 to +64°F and +82 to +122°F)

  • Overvoltage protection: 24 VAC/DC

  • Supply: 9 V battery (type 6LR61 or 6LF22)

  • Dimensions / Weight: 173 x 60.5 x 38 mm / 190 g with batteries

  • Safety: Cat. III 24 V AC/DC according to IEC664-1 ed. 92

  • Electromagnetic compatibility
    Emission according to EN 50081-1 Ed. 1992
    Immunity according to EN 50082-1 Ed. 1992

SENSORS

The following K-type thermocouple sensors can be used with thermometers C.A 861 and C.A 863.

Sensors with grips and extendible spiral leads

Model Reference Measurement range Response time Dimensions (mm)
SK 11 Needle P03. 6529.17 -50…+600°C 12 s Ø 3 – L = 130

For penetration of viscous thick substances.
SK 13| P03. 6529.18| -50…+1100°C| 12 s| Ø 3 – L = 300
General purpose

All uses

SK 14| P03. 6529.19| -50…+450°C| 8 s| Ø 6 – L = 130
Curved surface

For temperature measurement of poorly accessible surfaces.End Ø 15 x 30 mm

SK 15 Surface| P03. 6529.20| -50…+900°C| 2.3 s| Ø 8 – L = 130
8 mm Ø spring-loaded end, ensuring optimum contact even when the sensor is not perpendicular to the surface.
SK 17 Air| P03. 6529.21| -50…+600°C| 2.6 s| Ø 6 – L = 130
For measurement of ambient air temperature. Thermocouple is protected by a 8.5 mm Ø metal shield
PK 9| P03. 6529.15| -20…+250°C| 5 s| Ø 1 – L = 55
Retractable needle

Designed for automobile pneumatics. Penetrates by up to 15 mm.

Sensors without grips

SK 1 Needle P03. 6529.01 -50…+800°C 1 s Ø 3 – L = 150

For penetration of viscous, thick substances.
SK 2| P03. 6529.02| -50…+1000°C| 2 s| Ø 2 – L = 1000
Bendable

Bendable as required for application.

SK 3| P03. 6529.03| -50…+1000°C| 6 s| Ø 4 – L = 500
Semi-rigid

Slightly bendable.

SK 4 Surface| P03. 6529.04| 0…+250°C| 1 s| Ø 5 – L = 150
Suited to measurement on small surface areas.
SK 5 Surface| P03. 6529.05| -50…+500°C| 1 s| Ø 5 – L = 150
8 mm Ø spring loaded end, ensuring optimum contact even when the sensor is not perpendicular to the surface.

to be used between 0 and 50°C

__

SK 6

Flexible

| __

P03. 6529.06

| __

-50…+285°C

| 1 s/contact 3 s in| __

Ø 1 – L = 1000

ambient air

---|---|---|---|---
Recommended for difficult-to-get-at measurement points
SK 7 Air| P03. 6529.07| -50…+250°C| 5 s| Ø 5 – L = 150
For ambient air temperature measurement. Thermocouple protected by

8.5 mm Ø shield

SK 8

For pipes Self-gripping

| __

P03. 6529.08

| __

-50…+140°C

| 10 s on

12 mm Ø stainless pipe

| __

10 £ Ø £ 90

The thermocouple is located on a copper sheet at the end of a double- sided velcro tape that wraps around the pipe, ensuring good contact.
SK 19| P03. 6529.22| -50…+200°C| 7 s| Ø 14 – L = 1000
Magnetic

‘Automatic’ mounting on metal surfaces thanks to magnet

Extension leads with grips

Model Reference Diameter Length
CK 1 P03. **6529.09*** 4 mm 1 m

Extension lead

Ended with male/female connectors

CK 2| P03. **6529.10***| 4 mm| 1 m
Extension lead

Ended with male connector / 2 bare wires

CK 3| P03. **6529.13***| 4 mm| 1 m
Extension lead

Ended by DIN 5-pin socket / female connector

CK 4| P03. **6529.14***| 4 mm| 1 m
Extension lead

Ended by 2 banana plugs/ female connector

PP 1| P03. **6527.12***| –| 11 cm
Hand grip

Compatible with extension leads CK 1 to CK 4

Extension leads can withstand temperatures : -40…+100 °C

MAINTENANCE

For maintenance, use only specified spare parts. The manufacturer will not be held responsible for any accident occurring following a repair done other than by its After Sales Service or approved repairers.

Maintenance
Battery replacement

  • Select the OFF position on the switch
  • Take off the lid on the back of the instrument
  • Replace the worn-out battery with a 9 V battery (type 6LR61 or 6LF22)

Cleaning the casing

Clean using a cloth lightly dampened in soapy water. Rinse off with a damp cloth.
Do not use solvent.

Metrological verification

All measuring instruments must be regularly calibrated.
For checking and calibration of your instrument, please contact our accredited laboratories (list on request) or the Chauvin Arnoux subsidiary or Agent in your country.

Maintenance
Repairs under or out of guarantee: please return the product to your distributor.

TO ORDER

  • C.A 861 ……………………………………………………..P01.6501.01.Z
  • C.A 863 ……………………………………………………..P01.6502.01.Z
  • Comes with a shock-proof surround, a 9 V battery, a sensor (C.A 861) or two sensors (C.A 863), and this user’s manual.

Spares:

  • 9 V battery …………………………………………………… P01.1007.32

Accessories :
Refer to §. 5 Sensors

ATTACHMENT

C.A 861

C.A 863

CA UK Ltd – Waldeck House – Waldeck road – Maidenhead SL6 8br – Tel : (01628) 788 888 – Fax : (01628) 628 099
190, rue Championnet – 75876 PARIS Cedex 18 – FRANCE