Specifications
- Product Name: 200W Single Output Medical Type
- Series: MSP-200
- Input Voltage: Universal AC input / Full range
- Power Factor (PF): >0.95
- Efficiency: Up to 89%
- Surge Input: Withstands 300VAC for 5 seconds
- Protections: Short circuit, Overload, Over voltage, Over temperature
- Cooling: Free air convection
- Constant Current Limiting Circuit: Built-in
- Profile: 1U low profile (38mm)
- Safety Approval: Medical safety approved (MOOP level)
- Remote ON-OFF Control: Built-in
- Standby Power: 5V at 0.3A
- Remote Sense Function: Built-in
- No Load Power Consumption: <0.95/230VAC, <0.99/115VAC at full load
Product Usage Instructions
1. Installation
To install the product, follow the steps below:
- Select a suitable location for the product.
- Ensure proper ventilation around the product.
- Connect the product to the power source using the provided universal AC input.
- If required, use the built-in remote ON-OFF control to turn the product on or off.
- Verify that the product is securely installed and connected.
2. Operating Temperature and Humidity
The product can operate within a temperature range of -40°C to +70°C. It is important to refer to the provided Derating Curve for temperature-dependent performance. The working humidity range is 20% to 90% RH non-condensing.
3. Safety
The product has been approved for medical safety standards and complies with ANSI/AAMI ES60601-1, IEC60601-1, EAC TP TC 004, BS EN/EN60601-1, and BS EN/EN62368-1. It is recommended to follow all safety guidelines provided in the user manual.
4. Protection Features
The product is equipped with several protection features:
- Short Circuit Protection: Automatically activates when a short circuit is detected and recovers automatically after the fault condition is removed.
- Overload Protection: Activates when the output power exceeds 105-135% of the rated output power.
- Over Voltage Protection: Shuts down the output voltage and recovers after re-powering.
- Over Temperature Protection: Activates when the product exceeds the specified temperature range.
FEATURES:
- Universal AC input / Full range
- Built-in active PFC function, PF>0.95
- High efficiency up to 89%
- Withstand 300VAC surge input for 5 seconds
- Protections: Short circuit / Overload / Over voltage / Over temperature
- Cooling by free air convection
- Built-in constant current limiting circuit
- 1U low profile 38mm
- Medical safety approved (MOOP level)
- Built-in remote ON-OFF control
- Standby 5V@0.3A
- Built-in remote sense function
- No load power consumption<0.5W (Note.6)
- 5 years warranty
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
SPECIFICATION
MODEL| MSP-200-3.3| MSP-200-5| MSP-200-7.5| MSP-200-12| MSP-200-15|
MSP-200-24| MSP-200-36| MSP-200-48
---|---|---|---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 7.5V| 12V| 15V| 24V| 36V| 48V
RATED CURRENT| 40A| 35A| 26.7A| 16.7A| 13.4A| 8.4A| 5.7A| 4.3A
CURRENT RANGE| 0 ~ 40A| 0 ~ 35A| 0 ~ 26.7A| 0 ~ 16.7A| 0 ~ 13.4A| 0 ~ 8.4A| 0
~ 5.7A| 0 ~ 4.3A
RATED POWER| 132W| 175W| 200.3W| 200.4W| 201W| 201.6W| 205.2W| 206.4W
RIPPLE & NOISE (max.) Note.2| 80mVp-p| 90mVp-p| 100mVp-p| 120mVp-p| 150mVp-p|
150mVp-p| 250mVp-p| 250mVp-p
VOLTAGE ADJ. RANGE| 2.8 ~ 3.8V| 4.3 ~ 5.8V| 6.8 ~ 9V| 10.2 ~ 13.8V| 13.5 ~
18V| 21.6 ~ 28.8V| 28.8 ~ 39.6V| 40.8 ~ 55.2V
VOLTAGE TOLERANCE Note.3| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%|
±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.3%| ±0.3%| ±0.2%| ±0.2%| ±0.2%
LOAD REGULATION| ±1.5%| ±1.0%| ±1.0%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
SETUP, RISE TIME| 1000ms, 50ms/230VAC 2500ms, 50ms/115VAC at full
load
HOLD UP TIME (Typ.)| 16ms/230VAC 16ms/115VAC at full load
INPUT| VOLTAGE RANGE Note.5| 85 ~ 264VAC 120 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
POWER FACTOR (Typ.)| PF>0.95/230VAC PF>0.99/115VAC at full load
EFFICIENCY (Typ.)| 80%| 84%| 86%| 88%| 88%| 88%| 89%| 89%
AC CURRENT (Typ.)| 2.2A/115VAC 1.1A/230VAC
INRUSH CURRENT (Typ.)| 35A/115VAC 70A/230VAC
LEAKAGE CURRENT Note.7| Earth leakage current < 300μA/264VAC , Touch
leakage current < 100μA/264VAC
PROTECTION| OVERLOAD| 105 ~ 135% rated output power
Protection type : Constant current limiting, recovers automatically after
fault condition is removed
OVER VOLTAGE| 3.96 ~ 4.62V| 6 ~ 7V| 9.4 ~ 10.9V| 14.4 ~ 16.8V| 18.8 ~ 21.8V|
30 ~ 34.8V| 41.4 ~ 48.6V| 57.6 ~ 67.2V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| Shut down o/p voltage, recovers automatically after
temperature goes down
FUNCTION| 5V STANDBY| 5VSB : 5V@0.3A ; tolerance±5%, ripple :
50mVp-p(max.)
REMOTE CONTROL| RC+/RC- : 4 ~10V or open = power on ; 0 ~ 0.8V or short =
power off
ENVIRONMENT| WORKING TEMP.| -40 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
SAFETY & EMC
(Note 4)| SAFETY STANDARDS| ANSI/AAMI ES60601-1, IEC60601-1, EAC TP TC 004
approved, Design refer to BS EN/EN60601-1, BS EN/EN62368-1
ISOLATION LEVEL| Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃ / 70%
RH
EMC EMISSION| Compliance to BS EN/EN55011 (CISPR11) Class B, BS
EN/EN61000-3-2,-3, EAC TP TC 020
EMC IMMUNITY| Compliance to BS EN/EN61000-4-2,3,4,5,6,8,11, BS EN/EN55035, BS
EN/EN60601-1-2, EAC TP TC 020
OTHERS| MTBF| 1609.5K hrs min. Telcordia SR-332 (Bellcore) ; 186.2K hrs
min. MIL-HDBK-217F (25℃)
DIMENSION| 1999838mm (LWH)
PACKING| 0.77Kg; 18pcs/14.9Kg/0.87CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC input,
rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted
pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load
regulation.
4. The power supply is considered a component which will be installed into a
final equipment. All the EMC tests are been executed by mounting the unit on a
360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-
confirmed that it still meets EMC directives. For guidance on how to perform
these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. Derating may be needed under low input voltages. Please check the derating
curve for more details.
6. No load power consumption<0.5W when RC+ & RC- (CN100 pin1,2) 0 ~ 8V or
short.
7. Touch current was measured from primary input to DC output.
8. The ambient temperature derating of 3.5℃/1000m with fanless models and of
5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
※ Product Liability Disclaimer:For detailed information, please refer to
https://www.meanwell.com/serviceDisclaimer.aspx
Mechanical Specification
Block Diagram
Derating Curve
Output Derating VS Input Voltage
Function Description of CN100
Pin No. | Function | Description |
---|---|---|
1 | RC+ | Turns the output on and off by electrical or dry contact between pin 2 |
(RC-). Short: Power OFF, Open: Power ON.
2| RC-| Remote control ground
3| AUX| Auxiliary voltage output, 4.75~5.25V, reference to pin 4(AUXG). The
maximum load current is 0.3A. This output has the built-in oring diodes and is
not controlled by the “remote ON/OFF control”.
4| AUXG| Auxiliary voltage output ground. The signal return is isolated from
the output terminals (+V & -V).
5| +S| Positive sensing. The +S signal should be connected to the positive
terminal of the load. The +S and -S leads should be twisted in pair to
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
6| -S| Negative sensing. The -S signal should be connected to the negative
terminal of the load. The -S and +S leads should be twisted in pair to
minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
Function Manual
1.Remote Control
The PSU can be turned ON/OFF by using the “Remote
ON/OFF” function
Between RC-(pin2) and RC+(pin1) | Output Status |
---|---|
SW ON (Short) | OFF |
SW OFF (Open) | ON |
2. Remote Sense
The remote sensing compensates voltage drop on the load wiring up to 0.5V.
Downloaded from Arrow.com.
File Name:MSP-200-SPEC 2022-09-20
Frequently Asked Questions (FAQ)
Q: What are the dimensions of the product?
A: The dimensions of the product are 199mm (length) x 98mm (width) x 38mm (height).
Q: What is the weight of the product?
A: The weight of the product is 0.77kg.
Q: What is the packaging information?
A: The product is packed in a box with a weight of 14.9kg and a volume of 0.87 cubic feet. Each box contains 18 units.
Q: What is the MTBF (Mean Time Between Failures) of the product?
A: The MTBF of the product is 1609.5K hours minimum according to Telcordia SR-332 (Bellcore) and 186.2K hours minimum according to MIL-HDBK-217F (25).