hygiena BAX Q7 System Software Upgrade User Guide

June 16, 2024
Hygiena

hygiena BAX Q7 System Software Upgrade

hygiena BAX Q7 System Software Upgrade

Instruction For Use

AX® System Q7 software release v4.22 contains several user interface features, new target options, and error logging improvements.
We strongly recommend that you always install the most recent version as soon as possible for optimal performance. Please read all instructions thoroughly and ensure all required prerequisites are confirmed prior to beginning installation of the software update.
If you currently have a WIN7 computer system, please contact diagnostics.[email protected] or your local Hygiena® representative for upgrade support. This release is NOT applicable for WIN7 systems.

Software Updates and Features for Release v4.22.2.14335 for WIN10 ONLY

  • New Assays with Result Interpretations to assist with visual interpretation of data:
    • foodproof® Salmonella Genus plus Enteritidis & Typhimurium Detection LyoKit
    • foodproof® Salmonella plus Cronobacter Detection LyoKit
    • foodproof® Enterobacteriaceae plus Salmonella Detection LyoKit
    • foodproof® Vibrio Detection LyoKit
    • foodproof® Yeast and Mold Quantification LyoKit
    • foodproof® Aspergillus Detection LyoKit
    • microproof® Legionella Quantification LyoKit
    • SDS Equipment Driver Improvements
    • Messages and Confidence Levels:

Important: All foodproof and microproof algorithms are designed to assist in result interpretation but are experimental and require further visual verification to increase the confidence in positive and negative result calls. Hygiena is working to improve and certify these algorithms as additional data is generated.

Software Upgrade Instructions

  1. Required Prerequisites:

    • Regional Settings are in English
    • Read/Write Permissions for the below file locations for all users
    • C:\Program Files\MATLAB
    • C:\Program Files (x86)\BAX
    • C:\ProgramData\BAX Q7 System
      • All Power and Sleep Settings are set to Never.
    • Obtain HEX Calibration Kit from Hygiena
    • This additional calibration plate is a standalone product and is not included with other calibration kits.
    • Please visit the contact us page for ordering and technical support.
  2. Calibration storage and removal of current software:

    • Back up the system’s calibration file.

    • Select Operations > BAX Maintenance > Export Calibration from the menu bar.

    • Navigate to a new folder in which to store the backup file and click Save.

    • Name the calibration file (example: Q7Cal_date.bqc).

    • Uninstall the previous version of the BAX Q7 application.

    • Navigate to Windows Settings and select Apps.
      Software Upgrade Instructions

    • Click on BAX System Q7, select Uninstall and agree (click Yes) on next window to initiate removal.
      Software Upgrade Instructions

    • If updating from 4.21, removal of MATLAB is not required. If updating from any previous versions, click on the MATLAB application, select Uninstall and click Agree on the next window (this may take several minutes).
      Restart computer BEFORE continuing to step 3.
      Software Upgrade Instructions

Software Installation Instructions for Steps 3 and 4

  1. Download 4.22.2.14335 software by clicking this link or the button from the BAX Q7 Page.

    • Transfer the download to a memory/USB stick if the computer doesn’t have internet access.
    • Click on the Q7 4.22.2.14335 icon and the program will automatically start execution of install.
    • A Windows protect screen may pop up. Select More Info, then select Run anyway.
    • On the next pop-up, select Yes to allow installation to begin.
      Software Installation Instructions for Steps 3 and 4
  2. MATLAB updates will install first. Follow prompts below and select the default responses:
    Software Installation Instructions for Steps 3 and 4
    In case this install error occurs: VCRUNTIME140.dll was not found. (shown below).

  3. Following MATLAB installation, the BAX Q7 application installation will initiate. Read the prompts carefully. Choose the default responses.

    • Welcome Window – click Next
      Software Installation Instructions for Steps 3 and 4

    • Calibration File Warning – ENSURE YOU HAVE BACKED UP THE CALIBRATION FILE.

    • Check box and click Next.
      Software Installation Instructions for Steps 3 and 4

    • License Agreement.

    • Check Accept Terms box and click Next
      Software Installation Instructions for Steps 3 and 4

    • Custom Setup.

    • Click Next.
      Software Installation Instructions for Steps 3 and 4

    • Password Protection.

    • BAX default is disabled password protection. Choose an option for your lab, then click Next.
      Software Installation Instructions for Steps 3 and 4

    • Ready to Install Program.

    • Click Install.
      Software Installation Instructions for Steps 3 and 4

    • nstalling BAX System Q7 Software.

    • During installation, an Error Log Reporting window will pop up. We recommend NOT checking the box so the system can collect software and/or hardware errors in the Log Files to assist Diagnostics Support in troubleshooting. However, please review and decide what is appropriate for your lab.

    • Select Next to complete the installation
      Software Installation Instructions for Steps 3 and 4

    • InstallShield Wizard Complete.

    • Select Finish.
      Software Installation Instructions for Steps 3 and 4

    • Click on the desktop icon to initiate the program. The new open screen will have software version 4.22.2.14335.

    • Enter Instrument Name and Serial Number.

      • This will appear on reports and should be utilized
        Software Installation Instructions for Steps 3 and 4
  4. CHECK system’s calibration (NEW FEATURE)

    • From the menu bar, select Diagnostics > Calibration to open the Q7 Calibration Diagnostics window
    • Click the Check Calibration button to determine the status:
      • If calibration is available, no further actions are needed.
      • If calibration is NOT optimal or NOT found: Import the backed-up calibration file by selecting Operations > BAX Maintenance > Import Calibration from the menu bar and selecting the backed-up calibration file from step 2.
      • To utilize any foodproof assay, additional calibration using the HEX calibration kit must be performed. Please follow the onboard instructions in the BAX Software to perform calibration.
        Software Installation Instructions for Steps 3 and 4
  5. Ready to perform BAX System Analysis

    • Define proper targets that are used in your lab by selecting View > Select Targets > then in window, highlight targets being used and then use arrow buttons to move targets from Define in System to Available for Selection window.
      Software Installation Instructions for Steps 3 and 4

Customer Support

www.hygiena.com