Swing Flex™
Swing Maxi Flex™
Instructions for use
Swing Flex Swing Maxi Double Electric Breast Pump
BPA FREE
Thank you for choosing the Swing Flex / Swing Maxi Flex breast pump. Breast milk is the natural choice for your baby, but there are reasons and situations where breastfeeding is not possible. For over 50 years, it has been Medela’s conviction that these mothers and babies can be supported. In close cooperation with breastfeeding experts, Medela, one of the market leaders, has researched and developed breast pumps that perfectly meet breastfeeding mother’s needs. For mothers who insist on the best for their babies and themselves. Because you care.
Intended use / Population – Contraindication
The intended use and indications of breast pumps
Breast pumps are intended to be used by lactating women in the hospital or
home environment for expressing breast milk. Breast pumps are indicated to
alleviate the symptoms of engorgement by expressing milk. In case of mastitis
breast pumps support the healing process by removing breast milk from the
affected breast. Breast pumps are indicated to relieve sore and cracked
nipples and to bring out flat or inverted nipples. Also breast pumps enable
mothers to provide breast milk for babies who cannot feed directly at the
breast for a variety of reasons such as latch-on problems, cleft palate or
premature babies.
Intended patient population/user
Adolescent paediatric or adult lactating women. Many women find it convenient,
or even necessary, to use a breast pump to express and store their breast milk
once they have returned to work, are travelling or are otherwise separated
from their baby. Women are often delighted to learn that a breast pump can be
used as a supplement to breastfeeding and that some pumps are designed to
mimic the suckling of a nursing baby.
Contraindications
There are no known contraindications for the Swing Flex / Swing Maxi Flex
breast pump.
Meaning of symbols
The warning symbol identifies all instructions that are important for safety.
Failure to observe these instructions can lead to injury or damage to the
breast pump! When used in conjunction with the following words, the warning
symbol stands for:
Warning
Note Caution Info
Can lead to serious injury or death.Can lead to material damage.
Can lead to minor injury.Useful or important information that is not related
to safety.
Symbols on the packaging
| This symbol indicates that the material is part of a recovery/recycling
process.
---|---
| This symbol indicates a carton package.
| This symbol indicates to keep the device away from sunlight.
| This symbol indicates to handle the fragile device with care.
| This symbol indicates the temperature limitation for operation, transport
and storage.
| This symbol indicates the humidity limitation for operation, transport and
storage.
| This symbol indicates the atmospheric pressure limitation for operation,
transport and storage.
| This symbol indicates to keep the device dry.
| This symbol indicates that the packaging contains products intended to come
in contact with food according to regulation 1935/2004.
| This symbol indicates unique GS! Global Trade Item Numbers (GTIN).
****| This symbol indicates do not dispose the device together with unsorted
municipal waste (for EU only).
| This symbol indicates to follow instructions for use.
Symbols on the device
| This symbol indicates to follow instructions for use.
---|---
0123| This symbol indicates the compliance with the essential
requirements of the Council Directive 93/42/EEC of 14 June 1993 concerning
medical devices.
| This symbol indicates the manufacturer.
| This symbol indicates do not dispose the device together with unsorted
municipal waste (for EU only).
| This symbol indicates the compliance with additional USA and Canada safety
requirements for medical electrical equipment.
| This symbol indicates a type BF applied part.
| This symbol indicates manufacturer’s serial number of the device.
| This symbol indicates the positioning of battery cell.
IP22| This symbol indicates the protection against ingress of solid
foreign objects and against harmful effects due to the ingress of water.
This symbol indicates the date of manufacture (four digits for the year and
two digits for the month).
Symbols on the mains adapter
|
This symbol indicates that the mains adapter is a class II device.
---|---
| This symbol indicates the compliance with China Compulsory Certification.
| This symbol indicates that the mains adapter is for indoor use only.
| This symbol indicates the compliance with USA and Canada safety
requirements.
| This symbol indicates the compliance with Federal Communications
Commission requirements.
| This symbol indicates polarity of d.c. power connector.
| This CE-mark indicates the compliance with the low voltage and
electromagnetic compatibility directive.
| This symbol indicates to not dispose the device together with unsorted
municipal waste (for EU only).
The solid bar symbol indicates that mains adapter is put on the market after
13 August 2005.
| This symbol indicates the compliance with energy efficiency requirements.
| This symbol indicates alternating current.
| This symbol indicates direct current.
| This symbol indicates the compliance with energy efficiency requirements.
| This symbol indicates that the mains adapter is safety tested.
| This symbol indicates a short-circuit proof safety isolating transformer.
| Singapore Consumer Protection Registration.
| This symbol indicates that the electronic device incorporating
transformer(s) and electronic circuitry(ies), converts electrical power into
single or multiple power outputs.
| This symbol indicates the compliance with AUS/NZ requirements (regulatory
compliance mark).
| This symbol indicates the compliance with PSE requirements (Japan, Product
Safety Electrical Appliance and Material Safety Law).
| The KC mark indicates that the product meets Korean EMC and electric safety
requirements.
IP20| This symbol indicates the protection against ingress of solid
foreign objects and against harmful effects due to the ingress of water.
Important safety information
Failure to follow these instructions/safety information can lead to danger
from the device.
Subject to technical changes.
Device
Danger of electrocution! Keep the device dry! Never immerse in water or other
liquids!
Do not expose the motor unit to direct sunlight.
The operating life of the Swing Flex breast pump is 275 hours. The service
life is two years.
The operating life of the Swing Maxi Flex breast pump is 250 hours. The
service life is two years.
Electrical use
Separation from the mains is only assured through the disconnection ofthe
mains adapter from the wall socket outlet.
Do not reach for any electrical device if it has fallen into water. Unplug
immediately.
Use
Use the Swing Flex / Swing Maxi Flex breast pump only for its intended use as
described in this manual.
Do not use the Swing Flex /Swing Maxi Flex breast pump while bathing or
showering.
Do not use the Swing Flex / Swing Maxi Flex breast pump while sleeping or
overly drowsy.
This is a single user product.
Use by more than one person may present a health risk.
The Swing Flex / Swing Maxi Flex breast pump is not heat-resistant: Keep away
from radiators and open flames.
Never use a damaged device. Replace damaged or worn parts.
Repairs must be performed only by an authorised service agency. Do not repair
yourself! No modifications to the device are permitted.
Keep the mains adapter away from heated surfaces.
The breast pump should never be left unattended when plugged into a power
source.
Do not drive whilst hands-free pumping.
Supervision is necessary when the Swing Flex / Swing Maxi Flex breast pump is
used in the vicinity of children.
Contact your lactation consultant or breastfeeding specialist if you
experience problems or pain.
Portable and mobile radio frequency communications equipment can affect the
Swing Flex / Swing Maxi Flex breast pump.
Never operate an electrical device if it has a damaged cord or plug, if it is
not working properly, if it has been dropped or damaged, or dropped into
water.
Product description Swing Flex
Swing Flex – single electric 2-Phase breast pump
1x motor unit: article 600.0975
Product description Swing Maxi Flex
Swing Maxi Flex – double electric 2Phase breast pump
1x motor unit: article 200.4932 spare part 099.0270 LED
Cleaning
6.1 Cleaning overview
Washing and sanitising are two different activities. They must be done
separately to protect you, your baby and the performance of your breast pump.
Wash – To clean the surfaces of the parts by physically removing
contamination.
Sanitise – To kill living organisms, such as bacteria or viruses, that may be
present on the surfaces of the parts.
When to clean| Breast shield(s)| Connector and membrane|
Bottles & lids
---|---|---|---
Before 1st use| ✓ 6.2 disassemble
✓ 6.3 wash
✓ 6.4 sanitise| ✓ 6.2 disassemble
✓ 6.3 wash
✓ 6.4 sanitise| ✓ 6.2 disassemble
✓ 6.3 wash
✓ 6.4 sanitise
After each use| ✓ 6.3 wash| ✓ 6.2 disassemble
✓ 6.3 wash| ✓ 6.2 disassemble
✓ 6.3 wash
Once per day| ✓ 6.4 sanitise| ✓ 6.2 disassemble
✓ 6.4 sanitise| ✓ 6.2 disassemble
✓ 6.4 sanitise
When to clean| Tubing| Motor unit
---|---|---
As needed| see chapter 6.5| see chapter 6.6
6.2 disassemble![medela Swing Flex Swing Maxi Double Electric Breast Pump
-
Cleaning 6](https://manuals.plus/wp-content/uploads/2024/01/medela-Swing- Flex-Swing-Maxi-Double-Electric-Breast-Pump-Cleaning-6.jpg)
- Remove and dispose the protective cover* from the breast shield before first use.
- Disassemble the pump set into its individual parts (breast shield, connector, membrane, milk bottle).
- Open back cap of the connector by squeezing both flaps.
- Check all parts for damage.
- Replace damaged or worn parts.
Caution
- Do not spray or pour liquid directly onto the pump.
- Only use drinking-quality water for cleaning.
- Take apart and wash all parts that come in contact with the breast and breast milk immediately after use to avoid drying up of milk residues and to prevent growth of bacteria.
- Do not clean the tubing with Quick Clean bags in the microwave.
- Unplug the breast pump before
Note
cleaning the motor unit.
- The pump set requires no maintenance.
- Take care not to damage parts of the pump set during cleaning.
- Do not use the breast milk if bottles or components become damaged.
6.3 Wash
Before first use and after each use
Sink
- Rinse all parts with cold, clear drinking-quality water (approx. 20 °C).
- Clean these parts with plenty of warm, soapy water (approx. 30 °C). Use a commercially available washing-up liquid, preferably without artificial fragrances and colouring (pH neutral).
- Rinse the parts with cold, clear drinking-quality water for 10– 15 seconds (approx. 20 °C).
Dry after washing.
OR
Dishwasher
Place all parts, except for the tubing, on the top rack or in the cutlery
section. Use a commercially available dishwashing detergent. Dry after
washing.
6.4 Sanitise
Before first use and once per day
Stovetop
Cover all parts, except the tubing, with water and boil at least for 5
minutes. Dry after sanitising.
OR
Microwave Use Quick Clean* bags in the microwave in accordance with the
instructions. Dry after sanitising.
- Accessories available for order: see chapter 12. For more information, visit www.medela.com.
Drying
Dry with a clean cloth or leave to dry on a clean cloth. Put the clean parts
in a clean storage bag or a clean environment. Do not store parts in an
airtight container/bag if moist. It is important that all residual moisture
dries.
6.5 Cleaning tubing
- No visible residues No cleaning needed.
- With visible residues
- Rinse tubing in cool water
- Wash tubing in warm, soapy water
- Rinse tubing with clear water
- Shake out water droplets
- Hang to air dry
Do not sanitise the tubing.
6.6 Cleaning the motor unit
Note
- If the individual components of the pump set are cleaned in a dishwasher, parts may be discoloured by food pigments. This has no impact on the function.
- The membrane at the connector prevents the milk from flowing into the tubing, however, inspect the tubing after each pumping session for visible residues. If there are visible residues in the tubing, follow the instructions in chapter 6.5.
To clean the motor unit,
wipe the housing with a clean, damp (not wet) cloth. You may use drinking-
quality water and a basic washing-up liquid or a non-abrasive detergent.
Additionally you may use an alcohol-based disinfecting agent.
Preparing for pumping
Warning
-
Mains operation:
Follow the exact order of the steps. -
Only use the mains adapter supplied with the Swing Flex / Swing Maxi Flex.
-
Make sure the voltage of the mains adapter is compatible with the power source.
Info
- You can operate the Swing Flex / Swing Maxi Flex breast pump via battery or mains.
- Battery operation:
7.1 Mains operation
The batteries do not need to be removed when the mains adapter is used for
pumping.
- Regularly check the condition of the batteries in the motor unit.
- If Swing Flex / Swing Maxi Flex will remain unused for a longer period of time, remove the batteries from the battery compartment.
- Batteries last for 4 – 6 pumping sessions (approx. 1 ½ hours pumping time).
Plug the mains adapter cord into the motor unit.
7.2 Battery operation
Open the battery compartment on the back in the direction of the arrow. Plug
the mains adapter into the mains socket. The motor unit is now ready for
operation. Insert four AA Mignon LR6 batteries for Swing Flex .
Insert six AAA Micro 1.5 V batteries via battery. for Swing Maxi Flex. Close
the battery compartment.
The pump is ready for operation
- Hold the tubing by the connecting piece while attaching and detaching it.
Preparing for pumping
Caution
- Wash hands thoroughly (at least one minute) with soap and water before touching the breast and pump parts. Dry your hands with a fresh towel or a singleuse paper towel.
- Also, bottles and component parts may become damaged if mishandled, e.g. dropped, over-tightened, or knocked over.
- Take appropriate care in handling bottles and components.
Note
- Use Medela original accessories only.
- All components must be completely dry before use.
Info
- Carry out all the steps with care and assemble the pump set correctly. Other wise a good vacuum may not be achieved.
7.3 A ssembling the pump set
1 Check pump set components for wear or damage before use and replace if
necessary.
Make sure parts are cleaned according to chapter 6 before assembling the pump
set.
Carefully place the membrane into the connector with the flap into the opening
of the connector.
Make sure that the membrane forms a seal around the edge of the connector.
Close the lid of the connector audibly clicks. Insert the breast shield onto
the until it connector. Insert the tubing into the correspondingopening in the
lid of the connector. Insert the other end of the tubing into the motor unit.
Expression
Caution
- Step 2 (chapters 8.1, 8.2, 8.3):
- The product comes with two different breast shield sizes 21 mm and 24 mm.
- Answer the following questions to find out whether the available breast shield sizes fit:
- Does your nipple rub the tunnel sides to the point of causing discomfort?
- Do you see excessive areola being pulled into the tunnel?
- Do you see any redness?
- Is your nipple or areola turning white?
- Do you feel unexpressed milk after breast milk pumping?
If you answered “YES” to any of these questions, consider trying a new size by
following the measuring instructions at
www.medela.com.
If you are still unsure that you selected the right breast shield size,
contact your lactation consultant or breastfeeding specialist.
See brochure “Choosing your Medela breast shield size”.
Note
- Step 3 (chapters 8.1, 8.2, 8.3):
Do not hold the pump set by the bottle. This can lead to blockage of the milk ducts and engorgement.
8.1 Swing Flex – Preparing for single pumping The oval breast shield can be rotated (360°) and placed in the desired position to have the most comfortable fit for you. Insert the unused short tubing into the tubing holder.
8.2 Swing Maxi Flex Preparing for single pumping Place the breast shield
on your breast so that your nipple is properly centred in the tunnel.
Formoreinformation onpumping,see chapter8.5.
Hold the breast shield onto your breast with your thumb and index finger.
Support your breast with the palm of your hand. Place the breast shield of one
bottle on the first breast so that the nipple is prop erly centred in the
tunnel. Formoreinformation onpumping,see chapter8.5.
Hold the breast shield onto your breast with your thumb and index finger.
Support your breast with the palm of your hand.
Caution
- Step 2 (chapter 8.3):
- Wash hands thoroughly (at least one minute) with soap and water before touching the breast, the clean pump parts and clean pump set. Dry your hands with a fresh towel or a single-use paper towel.
- Do not hold the pump set by the bottle.
This can lead to blockage of the milk ducts and engorgement.
Info
- Double pumping saves time and increases the energy content of the milk.
8.3 S wing Maxi Flex – Preparing for double pumping
The oval breast shield can be rotated (360°) and placed in the desired
position to have the most comfortable fit for you. Switch on the breast pump
with . Place the breast shield on your breast so that your nipple is properly
centred in the tunnel. Hold the breast shield onto your breast with your thumb
and index finger.
Support your breast with the palm of your hand. Formoreinformation
onpumping,see chapter8.5.
Place the second breast shield on your breast so that your nipple is
properly centred in the tunnel.
Note
- Easy Expression Bustier must be purchased seperately. More information on www.medela.com
8.4 S wing Maxi Flex – Preparing for hands-free expression with the Easy
Expression Bustier
Put on the Easy Exprssion Connect the assembled pump set to the breast
shields. Leave the zip open slightly. Place the breast shield on your breast
under the bustier such that your nipple is centred in the tunnel of the breast
shield. Formoreinformation onpumping,see chapter8.5.
Start pumping.
Caution
- Contact your your lactation consultant or breastfeeding specialist if you can express only minimal or no milk, or if expression is painful.
Note
- Always disconnect the Swing Flex /Swing Maxi Flex from the power source after expression.
- Only fill the bottle to the 150 ml mark.
Info
- Swing Flex / Swing Maxi Flex switches off automatically if it has been running for
- Do not kink the tubing during expression.
The following table shows the different LED status:
Lights up | On |
---|---|
Does not light up | Off |
Even, regular flashing | Stimulation phase |
Light constantly on | Expression phase |
Rapid flashing | Min./Max. vacuum |
8.5 Pumping The stimulation phase lasts a maximum of two minutes. The stimulation phase Switch on the breast pump with will then automatically change to the expression phase. 30 minutes with no interaction. Switch off the breast pump with .
If your milk starts to flow earlier, press the button to change to the
expression phase.
Make sure that the milk is flowing into the bottle properly. Use the bottle
stand to prevent the bottle from tipping over. The Max. Comfort vacuum can be
adjusted in every phase. Max. Comfort vacuum: Use to increase the vacuum until
it is slightly uncomfortable. Press to reduce the vacuum, until it is
comfortable again. Cleaninaccordance withchapter6.
Close the bottle with a lid.
Follow the instructions in chapter 9 “Storing and thawing breast milk”.
Storing and thawing breast milk
9.1 Storage
Storage guidelines for freshly expressed breast milk (for healthy term babies)
Room temperature 16 to 25 °C
(60 to 77 °F)| Refrigerator 4 °C (39 °F) or colder| Freezer –18 °C (0 °F ) or
colder| Breast milk thawed in the refrigerator
---|---|---|---
Up to 4 hours is best
Up to 6 hours for milk expressed under very clean conditions| Up to 3 days is best
Up to 5 days for milk expressed under very clean conditions| Up to 6 months is best
*Up to 9 months for milk expressed under very clean conditions| At room temperature: Up to 2 hours
Refrigerator: Up to 24 hours
Do not refreeze!
*Guidelines for expressing milk under very clean conditions:
Before expressing breast milk, mothers should wash their hands with soap and
water or an alcohol-based hand sanitiser. The pump parts, bottles and pumping
area must be clean.
Breasts and nipples do not need to be washed before pumping.
- These guidelines for storage and thawing of breast milk are a recommendation. Contact your lactation consultant or breastfeeding specialist for further information.
- Store breast milk in the coldest spot of the refrigerator (at the back of the glass shelf above the vegetable compartment).
9.2 Freezing
- Freeze expressed breast milk in plastic bottles or milk storage bags. Do not fill the bottles or bags more than 3/4 full to allow space for possible expansion.
- Label the bottles or bags with the date and volume of expression.
- Plastic bottles and component parts become brittle when frozen and may break when dropped.
- Do not use the breast milk if bottles or components become damaged.
9.3 Thawing
Caution
Do not thaw or warm breast milk in a microwave or a pan of boiling water. This
helps preserve important components and prevents burns.
- To preserve breast milk components, thaw the milk overnight in the refrigerator. Alternatively, hold the bottle or bag under warm water (max. 37 °C / 98.6 °F).
- Gently swirl the bottle or bag to mix any fat that has separated. Avoid shaking or stirring the milk.
Troubleshooting
Problem | Solution |
---|---|
If the motor is not running | Check presence of mains current. |
Check the position of the batteries in the battery compartment.
If there is low or no suction| Verify the membrane is clean and not damaged.
Make sure that the membrane forms a seal around the edge of the connector and
the lid is firmly connected with the connector.
Check all connections.
Check if the ends of the tubing fit securely into the back of the connector
and into the port on the breast pump. During single pumping, insert the unused
tubing into the tubing holder.
There should be no water in the tubing.
Do not kink the tubing during expression.
Ensure the breast shield forms a complete seal around the breast.
All components must be completely dry.
If condensation appears in the tubing| Clean the tubing as described in
chapter 6.5. Shake out any water drops.
The pump becomes wet| Unplug the mains adapter from the socket.
Switch off the pump.
Do not place the pump upside down. The control panel must point upwards. Store
the pump in a dry, warm place for 24 hours.
Note:
Do not expose the pump to direct sunlight.
Do not accelerate the drying process by use of heat.
Warranty / disposal
Warranty
The “International Warranty” brochure contains detailed warranty information.
Disposal
This is a single user product. Use by more than one person may present a
health risk.
The unit is made of metals and plastics and should be disposed of in
accordance with the European directives 2011/65/EU and 2012/19/EU. Additional,
local guide-lines must also be observed. Please take care that you dispose of
the unit and its accessories in accordance with your local disposal
guidelines.
Spare parts and Accessories
The following accessories for Swing Flex / Swing Maxi Flex are available from
Medela distribution partners by specifying the product number.
Sales articles Swing Flex / Swing Maxi Flex
Article number | Product |
---|
101033773
101036226
101033777
101033793
101033775
101036238
101033817
101036232
101033825
101033845
101033819
101033821| Swing Flex breast pump EN NL FR (Euro adapter)
Swing Flex breast pump EN NL FR (Euro adapter) for India
Swing Flex breast pump EN (AUS/NZ adapter)
Swing Flex breast pump EN (UK adapter) for UK
Swing Flex breast pump EN NL FR (UK adapter)
Swing Flex breast pump EN NL FR (US adapter)
Swing Maxi Flex breast pump EN NL FR (Euro adapter)
Swing Maxi Flex breast pump EN NL FR (Euro adapter) for India
Swing Maxi Flex breast pump EN (AUS/NZ adapter)
Swing Maxi Flex breast pump EN (UK adapter) for UK
Swing Maxi Flex breast pump EN NL FR (UK adapter)
Swing Maxi Flex breast pump EN NL FR (US adapter)
Spare parts Swing Flex
Article number | Product |
---|
099.0016
099.0017
099.0018 099.0276
101033542
101038305| Swing Flex adapter (100–240 V; 4.8 V) Euro
Swing Flex adapter (100–240 V; 4.8 V) UK
Swing Flex adapter (100–240 V; 4.8 V) AUS/NZ
Swing Flex adapter (100–240 V; 4.8 V) CN
Swing Flex adapter (100–240 V; 4.8 V) US/JP
Swing Flex adapter (100–240 V; 4.8 V) KR
099.0015| Swing Flex motor unit
101034002| Swing Flex tubing
101034005| PersonalFit Flex connector
Spare parts Swing Maxi Flex
Article number | Product |
---|
099.0271
099.0272
099.0273
099.0274
101033543
101038303| Swing Maxi Flex adapter (100–240 V; 12 V) Euro
Swing Maxi Flex adapter (100–240 V; 12 V) UK
Swing Maxi Flex adapter (100–240 V; 12 V) AUS/NZ
Swing Maxi Flex adapter (100–240 V; 12 V) CN
Swing Maxi Flex adapter (100–240 V; 12 V) US/JP
Swing Maxi Flex adapter (100–240 V; 12 V) KR
099.027| Swing Maxi Flex motor unit
101034003| Swing Maxi Flex tubing
101034005| PersonalFit Flex connector
Accessories for Swing Flex / Swing Maxi Flex
Article number | Product |
---|
101033953
101033964
101033992
101033997| PersonalFit Flex breast shield S (21 mm)
PersonalFit Flex breast shield M (24 mm)
PersonalFit Flex breast shield L (27 mm)
PersonalFit Flex breast shield XL (30 mm)
800.0796
200.1659| 150 ml breast milk bottles (3 pcs)
250 ml breast milk bottles (2 pcs)
008.004| Quick Clean microwave bags
Other Medela products are available at www.medela.com
- If the result of expression is unsatisfactory or expression is painful, contact your lactation consultant or breastfeeding specialist. A different size of breast shield may help make expression more comfortable and successful.
EMC / Technical description Swing Flex / Swing Maxi Flex
Swing Flex and Swing Maxi Flex are EMC-tested in conformity with the
requirements of IEC 60601-1-2:2007 3rd Edition and IEC 60601-1-2:2014 4th
Edition according to clause 7 and 8.9.
Warning – The electric breast pumps Swing Flex and Swing Maxi Flex should
not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the electric breast pumps Swing Flex and Swing Maxi
Flex should be observed to verify normal operation in the configuration in
which it will be used. Wireless communications equipment such as wireless home
network devices, mobile phones, cordless telephones and their base stations,
walkie-talkies can affect the electric breast pumps Swing Flex and Swing Maxi
Flex and should be kept at least a distance 30 cm away from the equipment.
Technical specifications Swing Flex / Swing Maxi Flex
Swing Flex
vacuum (approx.) –35 … –250 mmHg
– 5 … – 33 kPa
45 … 111 cpm
In | Out |
---|
100–240 V~ 50/60 Hz
max 0.5 A| 4.8 V ****
1.2 A
4 x 1.5 V, Alkaline
AA, Mignon, LR6
121 x 129 x 57 mm
300 g
200 g
Operation/Transport/Storage
Operation/Transport/Storage
Operation
Transport/Storage
In | Out |
---|
100-240 V~ 50/60 Hz
max. 0.6 A| V ****
A
Swing Maxi Flex
vacuum (approx.) –35 … –250 mmHg
– 5 … – 33 kPa
45 … 111 cpm
6 x 1.5 V, Alkaline
AA, Mignon, LR03
121 x 129 x 57 mm
290 g
220 g
Operation/Transport/Storage
Transport/Storage/Operation
Operation
Transport/Storage
References