ROTRONIC HYGROCLIP 2 Advanced Industrial Instruction Manual
- June 7, 2024
- ROTRONIC
Table of Contents
HYGROCLIP2 ADVANCED INDUSTRIAL
Short Instruction Manual
GENERAL DESCRIPTION
Rotronic HygroClip2 Advanced Industrial probes are specially designed for rush environments and temperatures up to 200°C.
CONNECTION
Plug the HygroClip on the mating connector of the transmitter or connection cable. Make sure that the catches are aligned correctly. Tighten the knurled nut by hand.
Only a properly assembled and fixed connection assures the IP protection
rating.
The HygroClip2 Advanced Industrial is compatible with the following Medtronic
products:
- Handhelds of the HP2x and HP3x families.
- The transmitter of the HF5 and HF8 families
- The data logger of the HL-NT family
MOUNTING
The probe should be immersed in the climate that is to be measured
ADJUSTMENT / CALIBRATION
The probes are adjusted in the factory before delivery. We recommend,
depending on the application, a calibration after 1 year of use. We provide
calibration and adjustment
services or you can do this yourselves on-site with the Rotronic SCS humidity
standards or our humidity generators HG2-S. Further information regarding
calibration and
adjustment can be obtained at www.rotronic.com or
by contacting us or one of the representatives.
TECHNICAL DATA
Device type | HygroClip2 Advanced Industrial |
---|---|
Operating humidity | 0…100 %RH |
Operating temperature electronic | -50…100°C |
Operating temperature sensor head | HC2A-ICxxx -100…200°CI |
HC2A-ICxxx-A -100…200°CI
HC2A-IMxxx-M -100…200°C1
HC2A-IExxx 400…200°C1
Continous load: 190 °C max.
I Peak load: 100 h max.
Accuracy @ 23°C| ±0.8 %RH @23°C ±0.1 K @23°C
Supply voltage| 3.3…5 VDC
Current consumption| Approx. 5 mA (Adjusted at 3.3 VDC)
Load| >10 kf2
Cable length| Max 5 m
Protection rating| IP65 (except the sensor area)
Analog outputs| 2x 0…1V
Digital communication| DART
12.1102.0105E