Lutron TU-2016 Turbidity Meter Instruction Manual
- June 15, 2024
- Lutron
Table of Contents
Lutron TU-2016 Turbidity Meter
Specifications
- Circuit : Custom one-chip of microprocessor LSI circuit
- Display Range : 0.00 to 50.00 NTU, 50 to 1,000 NTU
- Accuracy : Greater than or equal to the standard
- Light source: LED, 850 nm
- Detector: Photo diode
- Standard : Meets ISO 7027
- Response time : Less than 10 seconds
- Sample volume needed: 10 mL
- Data hold: Freezes the display reading
- Memory recall: Maximum & Minimum value
- Display sampling time: Approx. 1 second
- Power off: Auto shut off saves battery life or manual off by push button
- Calibration points: 0 NTU, 100 NTU
- Operating temperature and humidity: Less than 85% R.H.
- Power supply: DC 1.5 V battery (UM4, AAA) x 6 PCs, or equivalent
- Power current: Standby Approx. DC 3.5 mA, Testing Approx. DC 36 mA
- Weight : Approx. 320 g/0.70 LB. (Battery is included)
- Dimension : 155 x 76 x 62 mm (6.1 x 3.0 x 2.4 inch)
FEATURES
- Designed to meet ISO 7027.
- NTU ( Nephelometric TURBIDITY Unit ) measuring unit.
- Wide and auto measurement range : 0 to 1,000 NTU.
- High resolution : 0.01 NTU/1 NTU.
- The unique optics structure, enables the instrument to read low value of TURBIDITY to the high level up to 1,000 NTU.
- Four operation buttons and two calibration points, easy operation and guarantee the spec. accuracy.
- Jumbo LCD, easy readout.
- Microprocessor circuit assures maximum possible accuracy, provides special functions and features.
- Battery operated for field and on-site testing convenience.
- Data hold function for freezing the desired value on display.
- Records Maximum and Minimum readings with Recall.
- Heavy duty & compact housing with hard carrying case, designed for easy carry out & operation.
- Auto shut off is available to save battery life.
- Application : Test municipal water, food and beverage water, or other aqueous solution where fluid clarity is important.
SPECIFICATIONS
Circuit | Custom one-chip of microprocessor LSI circuit. |
---|---|
Display | LCD size : 41 mm x 34 mm. |
Range |
- 0.00 to 50.00 NTU, 50 to 1,000 NTU
- NTU : Nephelometric Turbidity Unit
- Auto range
Accuracy | ± 5 % F.S. or ± 0.5 NTU, which ever is greater. |
---|---|
Light source | LED, 850 nm. |
Detector | Photo diode |
Standard | Meet ISO 7027. |
Response time | Less than 10 seconds. |
Sample volume needed | 10 mL. |
Data hold | Freeze the display reading. |
Memory recall | Maximum & Minimum value. |
Display sampling time | Approx. 1 second. |
Power off | Auto shut off saves battery life or manual |
off by push button.
Calibration points| 0 NTU, 100 NTU.
Operating
temperature and humidity
| 0 to 50 ℃.
Less than 85% R.H.
Power supply| DC 1.5 V battery ( UM4, AAA ) x 6 PCs, or equivalent.
Power current| Stand by| Approx. DC 3.5 mA.
Testing| Approx. DC 36 mA.
Weight| 320 g/0.70 LB. @ Battery is included.
Dimension| 155 x 76 x 62 mm ( 6.1 x 3.0 x 2.4 inch)
Accessories Included| Instruction manaual
1 PC.
Testing bottle with 0 NTU standard, TU-0NTU…… 1 PC.
Testing bottle with 100 NTU standard, TU-100NTU… 1 PC.
Empty testing bottle ( Bottle-061 )…………………… 2 PCs
Clean cloth………………………………………………………. 1 PC.
Clean solution ( Distill water )………………………. 1 bottle
Hard carrying case, CA-08………………………………… 1 PC.
Optional Accessories| Testing bottle with 100 NTU standard solution, TU-100NTU Testing bottle with 0 NTU standard solution, TU-0NTU Empty testing bottle, 0601
FRONT PANEL DESCRIPTION
- 3-1 Cover of Testing bottle
- 3-2 Container of Testing bottle
- 3-3 Display
- 3-4 Hold Button ( Esc Button )
- 3-5 TEST/CAL Button
- 3-6 Power Button
- 3-7 ZERO Button
- 3-8 REC Button ( MAX, MIN Button )
- 3-9 Testing bottle with 0 NTU standard solution 3-10 Testing bottle with 100 NTU standard solution 3-11 Empty testing bottle 1
- 3-12 Empty testing bottle 2
- 3-13 Battery Compartment/Cover
- 3-14 Clean Cloth
- 3-15 Clean Solution ( Distill Water )
MEASURING PROCEDURE
Measurement Consideration
-
There is a ” White mark ” on the edge of the ” Container ” ( 3-2, Fig. 1 ) and also on the up position of ” Testing Bottle ” ( 3-9, 3-10, 3-11, 3- 12, Fig. 1 ), refer Fig. 2.
-
When make the measurement ( or calibration ), it should keep the ” Container white mark ” face to face together with the ” Testing bottle white mark ”
-
Insert the ” Testing bottle ” to the bottom of ” Container ” ( 3-2, Fig. 1 ) completely.
-
Before the measurement, it should envelop in the ” Cover ” ( 3-1, Fig. 1 ) completely.
Remark : Before the measurement, it should keep the outside of Testing Bottle under the dry condition and without existing any dust.
Measurement
-
Fill the measurement liquid into the ” Testing bottle ” ( 3-11, 3-12, Fig. 1 ) .
-
Insert the ” Testing bottle ” to the bottom of ” Container ” ( 3-2, Fig. 1 ) completely.
-
Before the measurement, it should envelop in the ” Cover ” ( 3-1, Fig. 1 ) completely.
-
Power ON the meter by pressing the ” Power Button ” ( 3-6, Fig. 1 ) once.
-
Press ” TEST Button ” ( 3-5, Fig. 1 ) once, the display will show the text ” tESt ” ( TEST )”, following the ” tESt ” will flash approx. 10 seconds , then the TURBIDITY value will present on the display along with the ” ntu ” unit.
Remark
- Under power OFF, if press the TEST Button ” ( 3-5, Fig. 1 ) will power ON and going on to make the test automatically.
- After the testing, within ten minutes approx., the meter will auto power OFF automatically.
Wash ( Reinse ) the Testing bottle
- After the testing, it should wash ( reinse ) the Testing Bottle by the Clean Solution ( Distill Water ) ” ( 3-17, Fig. 1 ).
Zero
- When measure the ” 0 NTU ” liquid, if the display not show the zero value, it can deduct ( offset ) those ” not zero value ” with default, the zero procedures are :
- Press the ” ZERO Button ” ( 3-7, Fig, 1 ) continuously until the display show zero value then release the finger from the button.
Remark
- The zero function can be executed within the zero daft value < 2.0 NTU only.
Data Hold
- During the measurement, press the ” Hold Button ” ( 3-4, Fig. 1 ) once will hold the measured value & the LCD will display a ” HOLD ” symbol.
- Press the ” Hold Button ” once again will release the data hold function.
Data Record ( Max., Min. reading )
The data record function records the maximum and minimum readings. Press the ”
REC Button ” ( 3-8, Fig. 1 ) once to start the Data Record function and there
will be a ” REC. ” symbol on the display.
With the ” REC. ” symbol on the display :
- Press the ” REC Button ” ( 3-8, Fig. 1 ) once, the ” REC. MAX. ” symbol along with the maximum value will appear on the display.
- If intend to delete the maximum value, just press the ” Hold Button ” ( 3-4, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.
- Press the ” REC Button ” ( 3-8, Fig. 1 ) again, the REC. MIN. ” symbol along with the minimum value will appear on the display.
- If intend to delete the minimum value, just press the ” Hold Button ” ( 3-4, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.
CALIBRATION PROCEDURE
-
The meter can be calibrated under two kinds standard solution.
- 0 NTU standard solution 100 NTU standard solution
-
The meter ship along with
- Testing bottle with 0 NTU standard solution x 1 PC
- Testing bottle with 100 NTU standard solution x 1 PC as the standard accessories.
-
The complete calibration should be executed by following two solution
- 0 NTU standard solution 100
- NTU standard solution
-
Shaking the calibration bottle lightly
- Before execute the calibration, it should shake lightly ( not strongly ) the Standard solution ( TU-0NTU, TU-100NTU )
- one to two seconds to let the calibration solution under the uniform condition.
- Remark
- Do not shake the Standard solution strongly due to it may generate the ” Air bubble ” and let the solution existing not accurate value.
-
The buttons that will be used during the Calibration procedures Escape button, select button :
- Fig. 1, 3-4 Hold Button ( Esc Button ) Calibration button, Enter button :
- Fig. 1, 3-5 TEST/CAL Button ( Enter Button )
-
0 NTU calibration
-
Insert the ” 0 NTU standard solution into the ” Container ” ( 3-2 , Fig. 1) and envelope in the ” Cover ” ( 3-1, Fig. 1 ) completely, other procedures please refer to above section 4-1, 4-2.
-
Press ” CAL Button ” ( 3-5, Fig. 1 ) continuously (at least 3 seconds ) until the Display show the text ” CAL ” then release the button.
-
After the LCD show above Fig. 5-1, wait a while the display will show
-
Now the meter is ready for the ” 0 NTU ” calibration
-
Press ” CAL Button ” ( 3-5, Fig. 1 ) once, the Display will show following text with flashing ( approx. 10 seconds ).
-
Then the Display will show
-
Now the meter is finished the ” 0 NTU ” calibration procedures and ready for ” 100 NTU ” calibration procedures.
-
-
100 NTU calibration
-
When finish the ” 0 NTU ” calibration procedures and the Display show :
-
The meter is ready for ” 100 NTU ” calibration.
-
Insert the ” 100 NTU standard solution into the ” Container ” ( 3-2 , Fig. 1) and envelope in the ” Cover ”
( 3-1, Fig. 1 ) completely. Press ” CAL Button ” ( 3-5, Fig. 1 ) once, the Display will show following text with flashing ( approx. 10 seconds ). -
Then the LCD display will return to normal measurement screen, now the meter is finished the calibration procedures ( 0 NTU, 100 NTU calibration ) completely and ready for the measurement.
-
-
During the calibration, if exit something wrong ( Error, can not be calibrated.., please check if the ” Standard solution ” value is wrong, if still can not fix the problem , then execute the execute the ” Calibration clear ” procedures
( Section 6 ) may can fix the problem.
CALIBRATION CLEAR
-
Execute the calibration procedures will clear all the calibration value ( Section 5 ) , the system will return to the Default value.
-
Power On ( no matter if the ” Testing bottle ” insert into the ” Container ” or not ) press the ” CAL Button ” ( 3-5, Fig. 1 ) continuously (at least 3 seconds ) until the Display show the text ” CAL ” then release the button, the display will show :
-
After the LCD show above Fig. 6-1, wait a while the display will show :
-
-
Press ” Hold Button ” ( 3-4, Fig. 1 ) once the display will show :
-
Press ” Hold Button ” ( 3-4, Fig. 1 ) once again the display will show :
-
Now the meter is ready for the ” Calibration clear “
-
Press ” CAL Button ” ( 3-5, Fig. 1 ) once, the Display will show following text with flashing ( approx. 10 seconds ).
-
Then the LCD display will return to normal measurement screen, now the meter is finished the calibration clear procedures and clear all the existing calibration data.
BATTERY REPLACEMENT
- When the left corner of LCD display show ” “, it is necessary to replace the battery. However, in-spec. measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate.
- Loss the ” Battery Cover Screws ” and slide the ” Battery Cover ” ( 3-13, Fig. 1 ) away from the instrument and remove the battery.
- Replace with DC 1.5 V battery ( UM4, AAA, Alkaline/heavy duty ) x 6 PCs, and reinstate the cover.
- Make sure the battery cover is secured after changing the battery.