W HLER A 550 Portable Flue Gas Emissions Analyzer User Manual
- June 15, 2024
- W HLER
Table of Contents
- W HLER A 550 Portable Flue Gas Emissions Analyzer
- Getting started
- FAQs
- General Information
- Features and functions
- Specifications
- Setting up the instrument
- Getting started
- Using the Analyzer
- Menu
- Customer data
- Maintenance
- Warranty and Service
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
W HLER A 550 Portable Flue Gas Emissions Analyzer
Specifications
Features and functions
The Wohler A 550 INDUSTRIAL is a portable flue gas emissions analyzer designed for professionals. It offers the following features and functions:
- Readings of various parameters
- Calculated values based on the readings
- Technical data display
- Component explanation
- Probes and sensors
- Tightness test
- Gas sampling path
- Battery charging
- Probe and cable-hose assembly connection
- Analyzer power on and off
- Hotspot search
- Measuring functionality
- Data management options
- Air speed and volume flow measurement for flue gas
- Data exchange with PC
- Troubleshooting guide
- Analyzer check
- Warranty and service information
- Declaration of conformity
- Points of sale and service locations
Setting up the instrument
Component Explanation
Before using the Wohler A 550 INDUSTRIAL, it is important to understand its components:
- Analyzer unit: The main device that displays readings and controls the functionality.
- Probes: Attachable components used to measure specific parameters.
- Sensors: Integrated components within the analyzer unit that assist in measuring various parameters.
- Display: The screen that shows readings and menu options.
- Gas sampling path: The pathway through which the flue gas is sampled for analysis.
Probes and sensors
The Wohler A 550 INDUSTRIAL supports various probes and sensors for measuring different parameters. Make sure to connect the appropriate probe to the analyzer unit before use.
Tightness test
Prior to using the analyzer, perform a tightness test to ensure there are no leaks or issues with the gas sampling path. Follow the instructions provided in the user manual for conducting the tightness test.
Display
The analyzer unit features a display screen that shows readings, menu options, and other information. Familiarize yourself with the display layout and menu navigation for efficient usage.
Getting started
Charging the battery
Before using the Wohler A 550 INDUSTRIAL, make sure to fully charge the battery. Connect the charger to the analyzer unit and plug it into a power source. Allow the battery to charge completely before disconnecting.
Connecting the probe and cable-hose assembly
To start measuring, connect the probe and cable-hose assembly to the analyzer unit using the following steps:
- Connect the probe to the appropriate port on the analyzer unit.
- Connect the hose to the other end of the probe securely.
Using the Analyzer
Turning the analyzer on and off
To turn on the Wohler A 550 INDUSTRIAL, press and hold the power button until the device powers on. To turn it off, press and hold the power button again until the device powers off.
Hotspot search
The hotspot search feature allows you to locate areas with high emission levels. Activate the hotspot search mode and follow the on-screen instructions to identify hotspots.
Measuring
To start measuring flue gas emissions, follow these steps:
- Ensure the analyzer is turned on and connected properly.
- Select the desired parameters to measure from the readings screen.
- Choose the fuel type for accurate calculations.
- Set the fuel parameters if necessary.
- Explore additional options available for customization.
Readings Menu
The readings menu provides various options for managing and analyzing the collected data. Use the following functions:
- Delete data: Remove saved data from the analyzer’s memory.
- Print data: Print out the collected data for reference or analysis.
- Save data in the customer menu: Store data under specific customer profiles for easy access.
- Save data with the Quick save option: Quickly save data without assigning it to a customer profile.
Measuring the Air speed and the Volume Flow of the flue gas
Measuring the Air Speed
The Wohler A 550 INDUSTRIAL allows you to measure air speed and volume flow of flue gas. Follow these steps:
- Select the air speed measurement option from the menu.
- Follow the on-screen instructions to perform the measurement.
Troubleshooting
If you encounter any issues with the Wohler A 550 INDUSTRIAL, refer to the troubleshooting section of the user manual for possible solutions.
FAQs
Q: Can the Wohler A 550 INDUSTRIAL be used by non-professionals?
A: No, the analyzer is intended for professional use only.
Q: How long does it take to charge the battery fully?
A: The charging time may vary, but it is recommended to charge the battery completely before usage.
Q: Can I save and print data collected by the analyzer?
A: Yes, the Wohler A 550 INDUSTRIAL provides options to save and print collected data for further analysis or reference.
Q: How can I measure air speed and volume flow of flue gas?
A: Use the dedicated air speed measurement option on the analyzer and follow the on-screen instructions to perform the measurement.
Operation Manual Portable Flue Gas Emissions Analyzer
Wohler A 550 INDUSTRIAL
The Measure of Technology
Best.-Nr.23707 2023-05-17
General Information
1 General Information
1.1 Operation Manual Infor- This operation manual allows you to work safely
mation
with the Wohler A 550 INDUSTRIAL. Please keep this manual for your information.
The Wohler A 550 INDUSTRIAL should be employed by professionals for its intended use only.
Liability is void for any damages caused by not following this manual.
1.2 Notes
WARNING! Not following this warning can cause injury or death.
1.3 Intended Use
ATTENTION! Not following this note can cause permanent damage to the device.
NOTE! Useful information
Use the Wohler A 550 INDUSTRIAL to perform emissions analysis in industrial
applications including boiler and burner testing, heat processes, engine and
turbine testing. The logger function allows user configurable longterm
operation.
5
General Information
1.4 Basic features (other configurations are possible)
Device
Wohler A 550 INDUSTRIAL
Basic equipment
O2, NO, COHigh, NO2, SO2 sensor
Probe and hose assembly with 3,000 mm hose
Modular Flue Gas Probe 1,000 mm with sinter filter
Ambient Temperature Probe (plug)
USB-Charger with micro-USB-cable
1 Water stop filter
1 Coarse Filter
25 Wadding filters
1.5 Transport and stocking ATTENTION!
Improper transport can harm the instrument.
Always transport the analyzer in the provided carrying case in order to
prevent damage.
6
probe handle
General Information
· To open the probe compartment pull down the snap closure under the black
probe handle.
Snap closure
Fig. 1: Wohler A 550 INDUSTRIAL with open probe compartment
Lock the analyzer
ATTENTION! For the transport deactivate the battery with the transport lock.
· Deactivate the battery by pushing the transport lock to the right with a pointed object (wire or pen).
Fig. 2: Transport lock in the probe compartment
When the instrument is delivered, the battery is always deactivated.
1.6 Information on disposal
Electronic equipment does not belong into domestic waste, but must be disposed
in accordance with the applicable statutory provisions.
You may hand in any defective batteries taken out of the unit to our company
as well as to recycling places of public disposal systems or to selling points
of new batteries or storage batteries.
7
Features and functions
1.7 Manufacturer
Wohler Technik GmbH Wöhler-Platz 1 D-33181 Bad Wünnenberg Germany Tel.: +49 2953 73-279 E-Mail: international@woehler.com
1.8 Important information
ATTENTION! In order to ensure the quality of the application and the measurement result, the analyser may only be used with original Wohler accessories and original Wohler spare parts. This applies in particular to legally regulated measurement tasks.
ATTENTION! Do not expose the instrument to moist flue gas, if it has been exposed to temperatures below 0°C for a certain time. This may harm the instrument.
ATTENTION! The instrument is equipped with strong magnets to fix it during the measurement. The magnetic field may harm heart pacemakers, hard disks, data mediums, magnetic stripe cards, controls etc. Keep sufficient distance to magnet-sensitive objects.
2 Features and functions
Logger-function 4 Pa-Test Wood moisture Spillage test Tuning guide Possible
number of sensors Start/Stop-Button: Remote control via the probe 8
Wohler A 550 INDUSTRIAL
5
1,000 data sets USB and infrared interface Bluetooth interface Fresh air pump
Features and functions
9
Specifications
3 Specifications
3.1 Readings
Oxygen (O2) concentration in flue gas
Display
volume % referenced to dry flue gas
Measurement principle
electrochemical sensor
Range
0.0 to 21.0 %
Accuracy
± 0.3 Vol. -%
Carbon monoxide in flue gas (CO 100,000 ppm)
Display
volume ppm referenced to dry flue gas
Measurement principle
electrochemical sensor
Range
0 to 100,000 vol. ppm, resolution 1 vol.-ppm
Accuracy
± 100 Vol. ppm (< 1,000 ppm), otherwise 10% of reading (when H2 < 5 % of reading)
Carbon monoxide in flue gas (CO 10,000 ppm) H2 compensated
Display
volume ppm referenced to dry flue gas
Measurement principle
electrochemical sensor, H2-compensated
Range
0 to 10,000 Vol. ppm, resolution 1 Vol.-ppm
Accuracy
± 20 Vol. ppm (< 4,000 ppm), otherwise 5% of reading
10
Specifications
Carbon dioxide concentration (CO2) in flue gas, NDIR
Display
Carbon dioxide concentration
Measurement principle
NDIR
Range
0 …40 Vol. %
Accuracy
0…6 Vol. %: ±0,3 Vol. % 6…40 Vol. %: ±5 % of reading
Hydrogen sulfide concentration H2S in flue gas
Display
volume ppm referenced to dry flue gas
Measurement principle
electrochemical sensor
Range
0 …350 ppm
Accuracy
0…40 ppm: ±2 ppm 40…350 ppm: ±5 % of reading
Nitric oxide concentration (NO) in flue gas
Display
volume ppm referenced to dry flue gas
Measurement principle
electrochemical sensor
Range
0 to 3,000 vol. ppm (continuously up to 1,000); resolution 0.1 vol. ppm (<1,000 vol. ppm), otherwise 1 vol. ppm
Accuracy
Resolution 0.1 ppm, ±5 vol. ppm (<100 vol. ppm), otherwise 5 % of reading
11
Specifications
Nitrogen dioxide concentration (NO2) in flue gas
Display
volume ppm referenced to dry flue gas
Measurement principle
electrochemical sensor
Range
0 to 1,000 Vol. ppm (continuously up to 200 vol. ppm); resolution 0.1 vol. ppm
Accuracy
±5 vol. ppm (<100ppm), otherwise 5 % of reading
Sulfur dioxide (SO2) in flue gas (option)
Display
volume ppm referenced to dry flue gas
Measurement principle
electrochemical sensor
Range
0 to 5,000 vol. ppm resolution 0.1 vol. ppm (< 1,000 vol. ppm), otherwise 1 vol. ppm
Accuracy
± 10 vol. ppm (< 200 ppm), otherwise 5 % of reading
Differential pressure (PD)
Display
Pascal
Measurement principle
Semi conductor diaphragm
Range
0.00 to ± 110.00 hPa, resolution 0.1 Pa (<1000,0 Pa), otherwise 1 Pa
Accuracy
0.3 Pa (<10 Pa), otherwise 3% of reading Drift < 0.2 Pa in 5 minutes
12
Specifications
Flue gas temperature (TS)
Display
°C
Measurement principle
Thermocouple (NiCr-Ni)
Range
-20.0 °C to 800.0 °C, resolution 0.1 °C
Accuracy
0 – 133 °C ± 2°C; 133 to 800 °C: 1.5% of reading
Flue gas temperature (TS) High Temperature Probe
Display
°C / °F
Measurement principle
Thermocouple Type K (NiCr-Ni)
Range
-20.0 °C to + 1.200 °C, resolution 0.1 °C
Accuracy
± 1°C
Combustion air temperature (TA)
Display
°C
Measurement principle
Thermocouple (NiCr-Ni)
Range
-20.0 °C to 100 °C, resolution 0.1 °C
Accuracy
± 1°C
13
Specifications
Wood moisture
Reading
water mass related to the absolute dry fuel mass
Measurement resistance measurement principle
Range
10.0 to 40.0 %, resolution 0.1 %
Accuracy
40 % of reading
wood temperature
5 to 25° C
lifetime of the electrodes
Depending on the frequency of utilization The electrodes will work correctly, if there is no surface damage or bending.
14
Specifications
3.2 Calculated Values
Calculated Value
Explanation
Efficiency /SL
Efficiency and losses in accordance to ASME standards
ETA/QS
Efficiency and losses in accordance to European standards (0.0 to 120%)
CO2 in Vol. -% COc, SO2c; NOc, NO2C
Range 0 CO2max, resolution 0.1 %
air free (corrected) Default: 0% (oil and gas).
Dpt
°C
Dew Point in the flue gas
Medium soot level
Soot number ± 0.1
Excess air coefficient Lambda (e.g. 1,25 when the excess of air is 25%)
Con Toxication index GI Air velocity
Condensate quantity in condensing conditions
CO/CO2
0.1 to 130 m/s, resolution < 0.1 m/s, for the measurement of the ventilation
loss (heating check)
Speed Flow
Range 0 130.0 m/s Range 0 9,990.000 m3/h
15
Specifications
3.3 Technical Data
Description
Data
Power supply
Lithium-Ion, rechargeable battery 3.7 V, 6700 mAh, charges via USB
Battery Operation time Approx. 7 h (depends on operation status and display illumination)
Charging cycles of the After 500 charging cycles, at least 70% of the capacity are
batteries:
still available
Storage Temperature -20 °C to 50 °C
Operation temperature +5 °C to 40 °C to maintain stated accuracy
Relative humidity
30% to 70 %
Weight
1,250 g
Dimensions
220 x 160 x 55 mm (without probe)
Length of cable hose 3 m
Degree of protection IP 40
16
Setting up the instrument
4.1 Component Explanation
2
1 4
5
Setting up the instrument
3
7
Explanation
6
Fig. 3: Device front 1 On/off-button 2 Condensate trap 3 Opening handle for
the condensate trap 4 Probe compartment 5 Opening handle for the probe
compartment
NOTE! Open the probe compartment by pulling down the opening handle.
6 Bottom of the analyzer with connectors 7 Loops to fix the probe
17
Setting up the instrument
9
10
11
8
Fig. 4: Probe handle with gas probe 295 mm
Probe probe handle: 8 ON/Off-Key 9 Gas Probe 10 Union nut 11 Plexiglas plug with Coarse Filter
On/Off-key (8)
It is possible to start and stop the measurement with the On/Off-key (8) of the probe handle. This permits an easy handling in difficult measurement conditions.
Fig. 5: Pressing the ON/Off-Key during the measurement
18
12 14 15 16 17
13 18
Fig. 6: Connectors at the bottom
Setting up the instrument
Connectors at the bottom of the analyzer 12 USB port for data transmission and
charger 13 IR Interface for the printer 14 Moisture probe connector 15 Ambient
temperature probe connector 16 Negative pressure connector 17 Positive
pressure connector 18 Speaker for alarm signals
Probe compartment 19 Gas probe connector
19
Fig. 7: Probe compartment
20 21
I o
20 Red point of orientation 21 Transport lock in the probe compartment
· Deactivate the battery by pushing the transport lock to the right (see
figure opposite) with a pointed object (wire or pen).
Fig. 8: Interior of the probe compartment Wohler A 550 INDUSTRIAL
19
Setting up the instrument
4.2 Probes and sensors
NOTE! Only the Flue Gas Probe 1,000 mm and the Ambient Temperature Plug are included in the Basic Set. Other hoses and probes mentioned in this section can be purchased as accessories.
Flue gas probe 1,000 mm to take emission measurements, even with heavy dust loads in difficult to reach locations.
Fig. 9: Flue gas probe 1,000 mm with sinter filter
High temperature Probe
Fig. 10: High Temperature Probe, 1, 000 mm, extendable up to 2,000 mm
Ambient Temperature Plug
Fig. 11: Ambient Temperature Plug
20
Setting up the instrument S-tube probe for measuring the flow speed
Fig. 12: S-tube probe
Flexible capillaries for the 4 PA test
Fig. 13: Flexible capillaries
21
Setting up the instrument
Ambient Temperature Probe 185 mm
Fig. 14: Ambient Temperature Probe
Wood Moisture Probe
Fig. 15: Wood Moisture Probe
Spillage Probe
Fig. 16: Spillage Probe
22
4.3 Tightness test
Fig. 17: Ball pump
Setting up the instrument
Before the tightness test connect the gas probe (see chapter 5.2). Proceed as
follows: 1. Squeeze the ball pump (part 1). 2. Plug the ball pump (part 1) on
the gas probe
(part 2) 3. Release the ball pump. After the ball pump has been released, it
may not fill with air or only fill very slowly. Otherwise the connection is
not tight.
ATTENTION! Never plug the ball pump on the gas probe and squeeze it
afterwards. In this case the draft probe will be damaged.
23
Setting up the instrument
4.4 Display
The Wohler A 550 INDUSTRIAL is operated via Touchscreen. The measurement menu
is similar to the one of a smartphone. A fingertip on the icon starts the
correspondent measurement mode. The active icons are graphically emphasized.
If there is a bar on the right of the display, the display can be scrolled by
pulling it with the finger.
The display is divided into different segments:
The fuel is shown in the topline on the left. The currently selected customer
and measurement mode are shown on the right of the fuel. Date and time, system
diagnostic status and battery level are shown on the right. The readings
segment displays the current readings. Click on PR key for zeroing the
pressure sensor.
NOTE! The user can set up the display according to his needs. He can arrange
the readings and chose the units. Go to the setup-menu and select “change
units” (see chapter 9.10).
Fig. 18: Display details
The icon segment offers different options. These options are explained in chapter 6.3.1
24
Setting up the instrument 4.5 Gas sampling path Wohler A 550 INDUSTRIAL
Fig. 19: Gas sampling path Wohler A 550 INDUSTRIAL 25
Setting up the instrument The Wohler A 550 INDUSTRIAL has two pumps. A sample
pump sucks the flue gas through the probe and the cable hose. A purge pump
pumps fresh air to all sensors during the calibration phase. As the sample
pump is cut-off during the calibration, the hotspot can already be searched
during this time. The sensors are protected against condensate by a four-step-
filter conditioning: The first step is the coarse filter in the probe handle.
Most of the particles will be removed from the sample. After that the gas
passes the condensate trap, where the gas is cooled by a spiral and the
condensate is collected. The second step is an exchangeable cotton filter.
Fine particles and rests of water will be removed from the sample. The last
step is the Water Stop Filter. This filter will not allow any kind of moisture
to pass. The filter will clog automatically in contact with water, so that the
sensors are protected against moisture. The sample passes all sensors. O2 is
always the first sensor and CO (4.000 ppm) is always the last sensor. The
fourth step is the LT-filter. This chemical filter removes all harmful gas
components. It is situated immediately in front of the CO sensor.
NOTE! When using the optional Wöhler USB Peltier cooler the coarse filter must
not be installed!
26
5 Getting started
probe handle
Getting started
· To open the probe compartment pull down the snap closure under the black
probe handle.
Snap closure
Fig. 20: Wohler A 550 INDUSTRIAL with open probe compartment
·
I o
Before the first start up activate the batteries by pushing the transport lock in the probe compartment (see Fig. 20) to the connector, see figure on the left. Use a pointed object to do so.
Fig. 21: Push transport lock to the left: batteries activated. Push transport lock to the right: batteries deactivated
27
Getting started
5.1 Charging the battery
Fig. 22: USB port at the bottom
The battery indication is displayed in the upper right corner of the screen. A
fully charged battery is shown as a solid green battery symbol.
If the battery symbol is red, the display illumination is automatically
reduced to save energy. In this case, charge the battery or use the Wohler A
550 INDUSTRIAL in mains operation.
· Before you plug the USB-recharger into the outlet, connect it with the
analyzer. To do so, plug the USB connector of the recharger to the USB port at
the bottom of the analyzer.
During the charging process a jack-symbol will blink in the upper right part
of the screen and the ON/OFF key will blink red.
NOTE! Recharging empty batteries can take up to 7 hours. You can still go on
measuring while battery charging is in progress. In this case charging time
will be longer.
After the charging time has finished the ON/OFF key will shine red.
WARNING! Risk of electrical shock!
Never touch the power supply with wet hands!
Protect the power supply against water and moisture!
Do not unplug the recharger by pulling the cable!
Do not use the power supply when the voltage requirements of the recharger and
the supply do not match!
28
5.2 Connecting the probe and cable-hose assembly 2
Getting started 4
3
1
Fig. 23: Probe handle with gas probe 295 mm
Probe handle: 1 ON/Off-Key (only Wohler A 550 INDUSTRIAL) 2 Gas probe 3 Union
nut 4 Plexiglas plug with coarse filter
5.2.1 Connecting the probe
Use different probes depending on the measurement to be done. Connect the
probe as follows: Connecting the gas probe 295 mm
· Remove the blue union nut from the probe handle.
· Plug the probe on the probe handle. Take care that the thermocouple
connectors are plugged into the green slots.
· Screw blue union nut on the probe handle again.
Fig. 24: Connecting the probe
29
Getting started
Connecting the gas probe 1000 · mm
·
Remove the blue union nut from the probe handle.
Remove the yellow plug and the grey union nut from the probe.
Fig. 25: Removing the plug from the gas probe 1,000 mm
·
·
Plug the probe on the probe handle. Take care that the thermocouple connectors
are plugged into the green slots.
Screw the grey union nut on the probe handle.
Fig. 26: Probe handle with gas probe 1,000 mm
30
Getting started
The gas probe 1,000 mm has an in-stack sinter-filter for dust protection. It
removes dust (> 20 ym) from the sample gas to protect the analyzer against
heavy dust loads. · Before measuring make sure that the screw
connection on the sinter filter is firmly tightened. · To remove the sinter-
filter carefully unscrew it from the probe.
NOTE! As the sinter-filter covers the temperature probe, a precise temperature
will only be displayed after an adjustment period. To get an exact temperature
value we recommend to repeat the measurement without sinter filter
ATTENTION! To prevent dust from entering into the gas channel, switch of the
gas pump when measuring the temperature without sinter filter. · Before
inserting the gas probe without sinter
filter into the stack, click on “pumps off”.
31
Getting started
5.2.2 Connecting the hose to the analyzer
probe handle
·
· snap shell
Snap closure
·
·
Fig. 27: Wohler A 550 INDUSTRIAL with
open probe compartment
To open the probe compartment pull down the snap closure under the black probe
handle.
Plug the male connector on the female connector, so that the two red points of
the male and the female connector are accurately positioned on top of each
other. The connector must audibly engage. (To remove the connector, pull the
snap shell down.)
Guide the cable through the cable duct.
Close the probe compartment as follows: Pull down the snap closure, lower the
probe handle and engage the snap closure.
Fig. 28: Female connector for the probe Wohler A 550 INDUSTRIAL
32
Using the Analyzer
Using the Analyzer
ATTENTION! Before use, always test if the instrument is in a proper condition
as described in chapter 4 . Perform a tightness test according to chapter 4.3
6.1 Turning the analyzer on and off
· Press the ON/OFF key shortly to turn ON the analyzer. The startup screen
will appear.
· Long press the ON/OFF-key for 3 seconds to turn the analyzer off.
6.2 Hotspot search
After the analyzer has been turned on, it will automatically start to calibrate the sensors with fresh air. During the calibration, the user can already insert the probe into the stack and search for the proper sampling position. In most cases it will be the position with the highest temperature reading (hotspot).
33
Using the Analyzer
· Search the hotspot regarding the flue gas temperature TS in the graph.
The graphical hotspot search is supported by an acoustic signal. The signal
will be more intensive when the temperature is higher.
NOTE! Do not change the position of the probe after having found the hotspot.
Fix the probe with a cone.
Fig. 29: Hotspot search Restart
Diagnosis Customers
Data Exchange Menu
34
Restarts the hotspot search.
Click on the “diagnosis” icon to finish the hotspot search and enter the
analyzers self diagnosis mode (see chapter 13.3).
Click on the “Customers” icon to finish the hotspot search and enter the
Customers menu. The Customers Menu allows to set up a new costumer or
preselect a customer before the measurement. After the measurement the
measured data can be stored to the customer folder, see chapter 7.3. Click on
the “Data Exchange” icon to finish the hotspot search and enter the data
exchange mode. In this mode data can be exchanged between the Wohler A 550
INDUSTRIAL and the PC via USB or Bluetooth. Opens the main menu. Only those
submenus are available which can be started without calibration.
6.3 Measuring 6.3.1 Readings screen
Using the Analyzer
Upon conclusion of the 60 seconds calibration procedure and the hotspot search
the analyzer will display the main combustion emission test screen. At the
same time the purge pump will turn off and the sample pump will turn on.
Al readings (measured and calculated values) will be shown continuously in the
readings segment.
NOTE! The user can change the order of the values. He can also select the unit
of some of the values (see chapter 9.10) Up to 14 values can be shown on
screen. The following screen is only an example.
Fig. 30: Readings segment: touch and slide to scroll the display screen 35
Using the Analyzer
O2 CO2 QS CO
COc Con
NO NOc
NO2 NO2c
SO2 SO2c
H2S Dpt GI TA TS
Efficiency /SL
Possible units
Measured and computed values:
Vol.-%
Oxygen
Vol.-%
Carbon dioxide
%
Flue gas loss
ppm; mg/m3
Carbon monoxide, referred to dry flue gas, diluted value
ppm, mg/m3, mg/kWh, mg/MJ Carbon monoxide, corrected (air free)
kg/m2 or kg/kg (depending on Condensate qty. in condensing condi-
the fuel)
tions
ppm, mg/m3
Nitrogen oxide
ppm, mg/m3 ,mg/kWh, mg/MJ, Nitrogen oxide, corrected (air free) lbs/mmbtu, g/GJ
ppm; mg/m3
Nitrogen dioxide
ppm, mg/m3 ,mg/kWh, mg/MJ, Nitrogen dioxyde, corrected (air free) lbs/mmbtu, g/GJ
ppm; mg/m3
Sulfur dioxide
ppm, mg/m3 ,mg/kWh, mg/MJ, Sulfur dioxide, corrected (air free) lbs/mmbtu, g/GJ
ppm
Hydrogen dioxide
°C/°F
Dew Point
—
Toxication index CO/CO2
°C/°F
Ambient Temperature
°C/°F
Flue Gas Temperature
%
Eta, Efficiency and losses calculated in
accordance to EN standards
%
Efficiency and losses in accordance to
ASME standards
36
Using the Analyzer
ETA/QS
Pr EA Spd Flow
Possible units
Measured and computed values:
%
Efficiency and losses in accordance to
European standards (0.0 to 120%)
Pa, hPa, mbar, mmH2O, inwc, psi Differential Pressure
—
Theoretical air/ excess air
%
Excess air
m/s, ft/min
Speed
m3/h, m3/s, cfm
Volume Flow
37
Using the Analyzer 6.3.2 Selecting the fuel
The selected fuel is shown in the topline on the left of the readings screen.
· Click on the fuel key to enter the fuel options
menu. Optional fuels as below: Natural gas, Fuel Oil 2, Fuel Oil 4, Fuel Oil
6, Biofuel, LPG, Propane, Kerosene, Digester Gas, Coal, Wood, Pellets and four
user defined fuels
NOTE! To see all fuels scroll by sliding your finger over the screen.
· Click on a fuel. The fuel will be selected for the next measurement.
· After turning off and on the analyzer, the last selected fuel will be active
again.
Fig. 31: Fuel options
6.3.3
Setting the fuel parameters
The reference oxygen and the Lamda threshold values of all fuels can be
individually set.
All values shown in the tuning guide display can be set for the user defined
fuels.
· Click on the “Fuel data” icon to control or change the fuel parameters. The “Fuel data” icon will turn green.
· Click on the fuel whose parameters are to be changed.
· The analyzer enters the fuel parameter menu. Change the parameters and confirm with “OK”.
6.3.4 Options Graph
In the readings screen a click on one of the icons at the bottom offers the
following options:
· A click on the Graph icon generates a graph of the measured values.
38
Using the Analyzer
· Above the graph the measured value, which is presented in the graph, is
indicated as a number.
· Select the value to be presented with the arrow keys.
The options in the graph mode are as follows: · Click New to start a new
measurement. · Click Stop to stop the measurement and hold
the values. · Click Back to return to the numeric screen.
Fig. 32: Graph of the CO readings Copy values
Clipboard readings Current readings
· Click on the “Copy values” icon to copy the readings into the clipboard. The
“Copy values” icon will appear in the top right screen.
This way the user can save the readings in case of an early switching off of
the burner. When the user continues the measurement after having copied the
readings to the clipboard, he can decide if he wants to save the actual
readings or those of the clipboard.
· To save the readings of the clipboard, stop the measurement. When the
measurement is stopped click on “Clipboard readings”.
In the clipboard screen, the “Current readings” icon will appear.
NOTE! A click on the “current readings” icon enters the current readings
screen. A click on the “Clipboard readings” icon enters the clipboard.
39
Using the Analyzer Stop
Pumps off Pumps on Purging on Accept
Proceed
· Click Stop to stop the measurement and hold the values. The sample pump
continues.
or · Press the Start/Stop button of the probe han-
dle to stop the measurement and hold the values. (The sample pump continues.)
· Click on “Pumps OFF” to turn off the gas pump.
· Click on “Pumps on” to turn on the gas pump.
· If the measurement has stopped, click on “Purging on”
In this case the sensors will be purged with fresh air without the need to
remove the probe from the stack. The “Accept” icon will only appear if the
measurement is stopped. · Click on the “Accept” icon to enter the Read-
ings menu. In the Readings menu you can save the readings in a customer folder
(see chapter Fehler! Verweisquelle konnte nicht gefunden werden.). · Click on
“Proceed” to continue the measure-
ment.
40
Readings Menu
7 Readings Menu
If the user accepts a record, the analyzer will enter the readings menu
automatically. It is also possible to enter the readings menu by clicking on
the readings icon in the main menu. In the readings menu a list with all
possible measurements and tests is shown. Those tests that have already been
done will appear with a checkmark. Those tests that have already been saved
will be marked by a lock. Select a measurement to delete, print or save it.
“Customer Logo” and “Technician” (last item) will always be marked by a
checkmark. · If you do not want the logo or the technician to
appear on the printout, remove the correspondent checkmark. This setting will
be saved if you turn the analyzer off and on again.
Fig. 33: Readings Menu
41
Readings Menu 7.1 Delete data
7.2 Print data
· Click on the “Delete data” icon to delete the selected data.
· Click on the “Print preview” icon to print the selected information on the
Wohler TD 100 Thermoprinter.
A print preview will appear. · Click on the “Cancel” icon, if you do not want
to print the information shown in the preview or · Click on the “Print” icon
to start the printout.
Fig. 34: Print Preview 42
Mesuring the Air speed and the Volume Flow of the flue gas
NOTE! Place the printer with the IR receiver toward the IR interface of the
analyzer.
After printing, the analyzer will enter the measurement mode again.
Fig. 35: Printing the readings of the Wohler A 550 INDUSTRIAL on the
Thermoprinter Wohler TD 100.
7.3 Save data in the cus- · tomer menu.
·
Click on the “Save” icon to save a customer record. The analyzer will enter
the customer menu. Create a new customer folder or save the record in an
existing customer folder.
Proceed as described in chapter 10.
7.4 Save data with the
If the quick save option is activated in the Setup-
“Quick save” option
Menu (see Chapter 9.10), the analyzer will automatically save the measurement records under a
new customer. The customer name will be the
measurement date and the measurement time.
· After the measurement click on “Accept”.
·
Fig. 36 Data record saved with the quick save function
Click on the “Save” icon to save a customer record.
8 Mesuring the Air speed and the Volume Flow of the flue gas
The Wohler A 550 INDUSTRIAL can measure air speed and volume flow by using the
differential pressure sensor. Prandtl and S-Tube probes are available in
different lengths to measure air speed for example inside the stack. In order
to calculate volume flow the user has to enter additional information.
43
Mesuring the Air speed and the Volume Flow of the flue gas · In the Setup-menu
select “Flow parameter” · Enter the cross-section of the pipe.
Fig. 37: Setup > Flow Parameter · Connect an S-tube probe as follows: – plug
the two pressure hoses on the pressure connectors of the analyzer. – Connect
the temperature plug to the temperature sockets next to the pressure
connectors.
Fig. 38: Analyzer with pressure hoses and temperature plug
44
Mesuring the Air speed and the Volume Flow of the flue gas
8.1.1 Measuring the Air Speed
·
Static pressure ·
Immersion depth
Flow
·
Knurled screws
·
Fig. 39: Measuring the air speed with the S-tube
In calm air, switch on the analyzer and wait 60 seconds.
When closed, the S-tube can be inserted through the measurement opening into
the gas flow.
Fix the S-tube with a cone, open it and direct the sensors as shown in the
figure.
If necessary, adapt the S-tube to the cross section of the gas line with the
help of the two knurled screws.
NOTE! The S-tube must be adjusted before the measurement procedure starts, since it is required to check and/or adjust the oppositely running orientation of the measuring tubes after loosening the knurled screws and adjusting the immersion depth.
· The speed and the volume flow values will be shown in the measurement
screen.
·
NOTE! If the air speed probe is equipped with a thermocouple it has to be connected to the stack probe thermocouple connector (see 5.2.1, Fig. 24)
Fig. 40: Values screen with speed and flow
45
Mesuring the Air speed and the Volume Flow of the flue gas 46
9 Menu
Menu
Menu
· Click on the main “Menu” icon to enter the main menu.
In the main menu you can select the different measurement options or the
setup-menu or the calibration menu. The lower icon bar shows the “Customer”
(see chapter 10) and the “Readings” icon (see chapter Fehler! Verweisquelle
konnte nicht gefunden werden.). · Click on the “Back” icon to return to the
normal
measurement mode.
Fig. 41: Main menu
47
Menu 9.1 Device data
In the “Device data” screen the user can enter information about the heating
system. He can save this information together with the readings in the
customer folder.
· Select a parameter.
Fig. 42: “Device data” screen
Fig. 43: Entering the ph value 48
· Enter the correspondent information. · Press OK to confirm.
NOTE! The value to be entered must be within the preset limits (in the example
on the left between 0.0 and 14.0). If the value is not between the preset
limits, it will not be possible to confirm the value.
After confirming the analyzer will automatically return to the device data
screen. · To save the information, proceed as described
see chapter Fehler! Verweisquelle konnte nicht gefunden werden..
NOTE! It is possible to save the information about the heating system together
with the readings.
9.2 Ambient CO
Fig. 44: Ambient CO
Menu
The option ‘Ambient CO’ offers a graphical assisted ambient CO test. The
current ambient CO concentration is shown over the elapsed measurement time.
The figure on the left shows an ‘Ambient CO’ test over a time period of 230
seconds and the current CO level is 2 ppm.
The options during the ambient CO test are as follows: · Cancel (left icon):
click the left key to exit with-
out performing an “Ambient CO” test and return to the main menu. · New: Start
a new “Ambient CO” test · Option “CO=0”: click to zero-out the CO sensor
offset reading. · To evaluate the test click on “Not OK” or “OK”. The analyzer
will enter the data menu where you can save and/or print the information.
NOTE! Due to temperature effects and aging of the CO sensor it is possible
that the 0 ppm offset level changes. Check regularly in clean air that the
ambient CO level is reading zero. If the reading is not zero in clean air,
click the ‘CO=0’ key to zero-out the readings.
49
Menu
9.3 Wood moisture
The Wohler A 550 INDUSTRIAL can perform a wood moisture test based on a
resistance measurement. In order to obtain an exact measurement, the
respective material will be considered. As the measured value depends on the
material temperature, the meter will perform an automatic temperature
compensation on the basis of the temperature entered by the user.
NOTE! It is necessary to connect the Wood moisture probe Wohler A 550
INDUSTRIAL to the analyzer before performing the wood moisture test.
ATTENTION! Before starting the test, check that the needles/electrodes are not
bent. If the electrodes are bent, the distance will not be 25 mm any more, so
that the measurement result may not be correct.
9.3.1 Theoretical background
The resistance measurement uses the needles of the wood moisture probe. The meter will automatically determine the wood moisture U. This is the water mass mw bound in the fuel and related to the absolute dry fuel mass mB according to equation (1):
=
100%
Equation 1
Explanation: U = wood moisture mw = water mass bound in the wood mb = absolut dry fuel mass
NOTE! According to the above equation the wood moisture may be higher than 100 %.
50
Menu
The wood moisture must not be confused with the water content. The water content is
the water mass related to the total mass of fuel and water (wet basis).
=
100 %
Equation 2
Explanation:
m = water content mw = water mass bound to wood
mb = absolute dry fuel mass
The wood moisture can be converted to the water content as follows:
=
1+
100%
Explanation:
w = water content u = wood moisture
9.3.2 Connecting the Probe
Equation 3
· Plug the connectors of the Wood Moisture Probe Wohler A 550 INDUSTRIAL or
the Wood Moisture Probe Wohler HBF 420 to the female connector at the bottom
of the analyzer.
Fig. 45: Connectors for the wood moisture probe at the bottom of the analyzer
NOTE! The result of the resistance measurement will be more exact, if the large wood moisture probe Wohler HBF 420 is used, because the needles of the probe can be inserted more deeply into the wood.
Fig. 46: Wood moisture measurement with Probe Wohler HBF 420.
9.3.3 Measuring
· In the main menu, click on the “Wood moisture” icon.
The analyzer will enter the wood moisture screen.
51
Menu Selecting the type of wood
· Click on the wood in the upper left corner of the screen.
· A list with different woods will appear :
The following woods can be selected.
Maple, birch, beech, oak, ash, spruce, chestnut, pine, larch, linden, poplar,
plane, fir, elm.
· Select the wood of the object to be measured and confirm with “Accept”.
Fig. 47: Different types of wood
Entering the wood temperature
If a temperature probe or plug is connected, the measured wood temperature
will appear under TA.
· Click on TA.
The analyzer will save the TA temperature as the wood temperature. Therefore
the same temperature value will appear under “TA” and “Wood temperature”.
If no temperature sensor is connected, no temperature value will appear under
TA.
· In this case, click on “wood temperature”, enter the wood temperature and
confirm with OK.
Fig. 48: Wood moisture test
52
Menu
Measuring the wood moisture
· Split the wood log in the middle and perform the measurement immediately
after that at the inside part of the log.
· The electrodes should penetrate into the wood approx. 6 mm.
ATTENTION! Insert the electrodes only if possible without applying force.
Average value
NOTE! Insert the electrodes diagonally to the direction of the fibers; in
other words, a line drawn between the measurement needles would dissect the
fiber of the wood.
The wood moisture content in per cent will be displayed.
In general the wood moisture should be measured at three different points (see
figure on the left). After that, the user has to calculate the average. Please
note the following points when performing the measurement:
Fig. 49: Situation of the measurement points at the log
Amongst other factors, the following should be observed:
NOTE! Chose a measuring point without any knobs, resin or fissures.
· Do not measure at the front side or at the exterior side of the split log.
NOTE! The wood to be measured should have a temperature between 0°C to 30 °C.
Accept
· After the measurement, click on the “Accept” icon to save the readings in the readings menu, see chapter Fehler! Verweisquelle konnte nicht gefunden werden..
53
Menu
9.4 Spillage test
The option “Spillage test” offers a search function which will help the user
to control the flue gas spillage of gas devices dependent on room air. Spilled
flue gas is characterized by a higher temperature and moisture content
compared with the ambient air. Flue gas spillage can be extremely dangerous,
when it contains CO which is highly toxic. Two independent measurements
guaranty a reliable result of the measurement. 1 Measurement of the
condensation on the sensor board. 2 Temperature measurement
NOTE! Connect the spillage probe Wohler A 550 INDUSTRIAL to the analyzer.
· Plug the connector of the Spillage Probe Wohler A 550 INDUSTRIAL to the
female connector at the bottom of the analyzer.
Fig. 50: Connector for the spillage probe at the bottom of the analyzer
Spillage
·
In the main menu, click on the “Spillage test ” icon.
54
Menu · Guide the probe along the areas to be in-
spected.
Fig. 51: Spillage test: Searching for condensate
The dew point will be indicated on a scale from 1 to 100. (0% = no
condensation) In the display the user can observe on the graph, when
condensation rises. Additionally an acoustic signal is emitted, that will
become more intensive when the condensate content is getting higher. The
available menu options are: · Cancel: Click on “Cancel” to cancel the meas-
urement. · New: Click on “New” to start a new graph. · To evaluate the test
click on “Not OK” or “OK”.
The analyzer will enter the data menu where you can save and/or print the
information.
Fig. 52: Spillage test screen
55
Menu
9.5 Tuning Guide
In the “Tuning Guide” menu all combustion test related readings are presented in a graph over the excess air. The tuning guide offers an advanced guidance to the technician to properly adjust the burner.
Tuning Guide
· In the main menu, click on the “Tuning guide” icon.
The graph for adjusting the burner will be shown. The current reading is
marked with a red box and any former reading is tracked with a green line.
The ‘Green Zone’ indicates the best combustion performance excess air range of
the burner. The range is user changeable for any selected fuel (see chapter
6.3.3). The figure on the left shows that the CO content in the flue gas
increases significantly when the available excess air decreases.
· Click on “New” to delete the graph and start a new one.
· Click on “Back” to return to the main menu.
Fig. 53: Tuning Guide
56
9.6 Manifold Pressure
Fig. 54: Pressure connector
Menu
· To perform a manifold pressure test click on the “Manifold Pressure” icon in
the main menu screen.
· Connect a pressure hose to one of the pressure connectors at the bottom of
the analyzer .
The options in the “Manifold Pressure” mode screen are: · Option “Cancel”:
click “Cancel” to exit the
“Manifold Pressure” test and return to the main menu. · Option “New”: click
“New” to clear the graph and all captured readings and start a new “Manifold
Pressure” test. · To evaluate the test click on “Not OK” or “OK”. The analyzer
will enter the data menu where you can save and/or print the information.
Figure 1: Manifold Pressure mode screen
57
Menu 9.7 Temperatures
· To measure the temperature click on the “Temperature” icon in the main menu
screen.
NOTE! Make sure you connect the Ambient Air Plug, see Fig. 11.
The options in the “Manifold Pressure” mode screen are: · Option “Cancel”:
click “Cancel” to exit the
“Temperature” mode and return to the main menu. · Option “Accept”: click
“Accept” to save the temperature readings in the readings menu, see chapter
7.3
Figure 2: Manifold Pressure mode screen
58
Menu
9.8 4 Pa – Test
The 4 Pa Pressure Test is a simple control of the depression limit 4 Pa. With
the 4 PaTest the user controls if there is sufficient combustion air in the
room where the fire place is installed. In a closed room or in connected
rooms, simultaneous operation of non-roomsealed heating appliances and
ventilation and air conditioning systems can cause dangerous depression. With
the 4 Pa Pressure Test the depression value can be controlled. The readings
are plotted in a graph during about 4 minutes. They can be saved and print
out.
NOTE! For the 4 Pa Test two equal capillary hoses must be connected, see Fig.
13.
The 4 PA-test measures the pressure difference between the room where the
fireplace is installed and another room which is not connected (e.g. outside
air, stairway etc.) .
Before starting the measurement:
· Connect the capillary hose, that leads to the reference point (stairway or outside air), to the negative pressure connector.
·
Fig. 55: Negative Pressure Connector at
the bottom of the analyzer.
·
Connect the capillary hose, which will remain in the room, on the positive
pressure connector.
Depression in the fireplace room in relation to the pressure at the reference
point will lead to a negative differential pressure.
4 PA test
· In the main menu select 4 Pa test.
59
Menu Fig. 56: Plan of the 4 Pa Test
How to perform the 4 Pa Pressure Test:
1. Switch on appliance and all air conditioning (fan, dryer) with maximum
power.
2. Open an outside window or a door to the reference room and test proper
operation of the appliance, ensure that there are no backdraft conditions.
3. Zero the pressure sensor.
NOTE!
During the zeroing proces the at the top of the screen.
icon will appear
4. How to position the reference capillary hose: Lead the hose outside through a window seal or into the stairways through the door rebate or the keyhole.
NOTE! Especially on stormy days, the stairways may be a stable reference room.
If the stairways are used as reference room, all windows, doors, cellar doors
and trap doors have to be closed.
60
Menu
The second capillary hose will stay uncut in the fireplace room. 1. Click on
“Start” to start the 4 PA-test.
The meter will now record the pressure profile for 4 minutes. 2. Open the
window/door for about 30 seconds, so that the zero line can be registered.
(Every 30 seconds an acoustic signal will sound.) 3. Close the window/door for
about 30 seconds, control depression. 4. Open the window/door for about 30
seconds, the zero line should be reached again. 5. Close the window/door for
about 30 seconds, control depression. 6. Open the window/door for about 30
seconds, the zero line should be reached again. 7. Close the window/door for
about 30 seconds, control depression. For a better orientation, every 30
seconds there are auxiliary lines in the diagram. After a maximum of 4
minutes, the measurement will stop automatically.
NOTE! Click on “Stop” to stop the 4 PA Test earlier.
61
Menu Fig. 57: Screen during 4 Pa test
Normally a diagram as shown in figure on the left will appear. The pressure
peaks in the diagram are caused by the rapid movement of the window or the
door and therefore they are not relevant for the interpretation of the
diagram. In the figure on the left the pressure drop is about 2,8 Pa. The
reliability of the system is considered sufficient, when the pressure drop
caused by opening and closing the window is less or equal to 4 Pa (8 Pa in the
case of wood burning depending on room air.)
· After the 4 Pa test is completed, click on “Measurement OK” or “Measurement
not OK” according to the test result.
NOTE! “Measurement OK” or “Measurement not OK” will appear on the printout.
The analyzer will automatically enter the data menu: “4 Pa test” will be
marked by a checkmark. · To save and print the information, proceed as
described in chapter Fehler! Verweisquelle konnte nicht gefunden werden..
NOTE! For the detailed procedure according to the Technical Rules for Gas
Installations (TRGI), working sheet G 625 (2009), it is necessary to replace
the fireplace by the Wohler DP 600 (see manual Wohler DP 600).
62
Menu
9.9 Logger Measurement The Logger option allows to perform long term
measurements. The user can configure the logger measurement to the application
needs ( see Fig. 59: Configuration of the logger measurement). It is possible
to change the start date and time, the interval and the total amount of
measurements. Depending on the measurement interval the analyser can perform
additional purging sequences. Interval < 8 min: The analyser will not turn off
between two measurements. A purging interval can be configured.
Interval > 8 min: The analyser will turn off between two measurements.
Measurement cycle: Automatic start of the analyser 1 min calibration 3 min
stabilization Take measurement readings and save data 3 min purging Automatic
shut down of the analyser When the logger configuration is completed the
Wohler A 550 INDUSTRIAL will calculate the ending date and time of the
measurement for information purposes.
63
Menu
Fig. 58: schematic representation of a logger measurement
· Enter the start date and time
· Enter the measurement interval, the interval duration in the sample in the
figure on the left is 10 seconds
· Enter the number of measurements (up to 9,999 measurements are possible).
NOTE! The analyzer calculates the date and time of completion of the logger
measurement for information purposes.
· Select Purging ‘On’, to activate the fresh air pump in a selected interval.
(only available when interval duration is < 8 min).
In this case the sensors will be purged with fresh air in the selected purging
interval.
NOTE!
During a purging cycle the analyzer will not log
Fig. 59: Configuration of the logger meas- measurements. A purging cycle needs about 3
urement
minutes to complete.
64
Menu
Long time measurements with the USB Peltier cooler
For continuous long term measurements with interval < 8 min it is recommended to use the USB Peltier cooler. The peltier cooler will remove condensate from the sample gas. During operation the condensate trap of the cooler needs to be emptied.
Fig. 60: Probe grip with Peltier Cooler and USB power bank
Before starting
· Connect the Peltier Cooler and ist portable USB power bank as described in
the Manual USB Peltier cooler.
·
NOTE! Take care that the Peltier Cooler has enough power during the long time
measurement. For measurements that take more than 5 hours we recommend to
connect the USB power bank.
Before starting a long term logger measurement:
· Configure the logger measurement.
· Ensure sufficient battery power is available or connect the analyzer to main
power.
· Press Start button to activate the logger measurement
· The ON/OFF button will flash during intervals > 8 min
· When the logger measurement is finished connect the analyzer to a computer
to download the data (please refer to Wöhler A 550 Data Management PC Software
manual).
65
Menu
9.10 SETUP
In the setup-menu the user can adjust the screen or the measurement process. In order to see all parameters, scroll to the top or the bottom of the screen by draging the left column with a finger.
SETUP
· In the main menu, click on the setup icon.
· The analyzer will enter the setup menu. 7 of the 13 parameters are
displayed.
· Sweep the left column with a finger to navigate through the setup menu.
· To change a parameter click on the correspondent key in the right column.
· Click on “OK” to save the new setting before leaving the setup menu.
The following options are available:
Fig. 61: Setup menu, the scroll bar shows that there are more parameters available.
Option “Date”
Change the current date of the analyzer’s internal calendar with day, month and year.
Option “Time”
Change the current time of the analyzer’s internal clock (24 h mode) with hours and minutes.
Option “Date format”
Select a date format. Possible options are dd.mm.yyyy / mm.dd.yyyy
Option “Time format”
Select a time format. Possible options are 24 h / 12 h
66
Option “Brightness”
Menu
Change the brightness of the display in 9 steps. · Click on the upwards arrow
to set the back-
light brighter, click on the downwards arrow to set the display darker.
Option “Buzzer” Option “Quick save”
Option “Customer Search by” Option “Devicen Search by” Option “NO2
calculation”
NOTE! A bright display will increase the power consumption.
Change the sound of the buzzer. The options are Standard, Loud or Off,
If the quick save option is activated, the analyzer will automatically save
the measurement records under a new customer. The customer name will be the
measurement date and the measurement time. (see Chapter 7.4)
Default: “Quick save” off
Change the order of the saved records between customer name or ID (Number)
Change the order of the saved records between name or ID (Number) of the
fireplace
In case the Wohler A 550 INDUSTRIAL is not equipped with a NO2 sensor the user
can enter the NO2 percentage on which the calculation of the NOX value is
based. The default NO2 percentage is 5 %. NOx calculation. The user can select
the formula for calculation NOx: “Sum NO + NO2” NOx = NO x 1,34 + NO2 x 2,05
“NOx as NO2” NOx = (NO + NO2) x 2,05
NOTE! If the Wohler A 550 INDUSTRIAL is equipped with a NO2 sensor, the NO2
factor parameter will not be active, because the NOx calculation will be
directly based on the measurement signals.
67
Menu Option “Flow para”
4 Pa duration
When a Prandtl or S-Tube is connected, the Wohler A 550 INDUSTRIAL measures
the air speed for example inside the stack. In order to calculate volume flow
the user has to enter the cross section.
Here the user can determine how long the analyzer shall log the pressure
during the 4 PA- test. The default 4 Pa-duration is 4 minutes.
Signals
Fig. 62: Configuration of the readings segment.
The user can set up the reading segment according to his needs.
· Click on “change” to enter the signals menu.
The left two columns of the screen show the current display configuration. The
right two columns show the available signals. To change a displayed signal:
· Click on the signal you want to change in the display columns. The signal
will turn green.
· Now click on the signal in the signals column that shall substitute the
green signal in the display.
· The new signal will appear in the display columns.
· To change a unit, in the left column (display) click several times on the
key whose unit has to be changed. You will find the possible units in the
table of chapter 6.3.1.
The icons in the footer have the following function:
· Cancel: Escape without saving the modifications.
· Restore defaults: Reset to default settings
· Remove signal: Remove a value from the display.
· OK: Confirm the modification and return to the Setup menu.
68
Customer data
Customer Logo
· Click on “Change”, enter your logo and confirm with OK.
Your logo will now appear on all printouts.
9.11 Calibration
In the calibration menu a digitally assisted calibration of all sensors can be
done. To open the calibration menu it is necessary to enter an internal
password. This way unauthorized persons cannot calibrate the analyzer and
incorrect settings are avoided.
The calibration of the analyzer is allowed to trained personnel ONLY. Please
contact Wohler for more information.
10 Customer data
The analyzer saves data records by systems. A customer can have multiple systems like burners and boilers. Each device has a uniqe identification number. The system name and system number can be entered by the user.
10.1 Set up a new cus- · There are two possibilities:
tomer
1. Go to the main menu and click on the Customer icon or
2. Click on the customer icon during the
search for the hotspot.
· Click on the New Customer Icon to create a New Customer folder.
69
· Enter the name and the number of the customer and the device. Confirm with OK.
Fig. 63: Creating a new customer folder
10.2 Save records
· In the Readings menu (see chapter 7) click on the Save-icon to save all
marked measurement records under a customer and a device. The analyzer will
enter the customer menu.
The customer menu shows a list of all customers.
· Click on a customer. The analyzer will enter the system menu.
· Click on a system.
The analyzer will save the marked measurement records under this system.
Fig. 64: Selecting a customer in the customer menu.
70
10.3 Search function : customer, system, customer number, installation number
· In the customer menu click on the “customer search” icon.
· Enter any part of the customer name. In this case it is not necessary to
observe the upper and lower case.
At the bottom the first result will be shown. Click on the arrow key to see
the other results and select the customer.
Fig. 65: Customer research
Research for the customer number and the installation number
When the customer is selected, search the system in the same way. · Click on
the icon ,,Find system”.
Note! In the setup menu you have the possibility to determine that always the
customer number or the installation number has to be searched.
71
10.4 New customer
· Click on the new customer icon in the main menu to create a new customer folder.
10.5 Delete customer 10.6 Delete all
In the customer menu it is possible to cancel a selected customer.
· Click on the “Delete customer” icon. The icon will become green.
· Select the customer to be canceled and confirm.
· To delete all customers from the analyzer click on the “Delete all” icon.
11 Data exchange with the PC
With the Wohler A 550 Sorftware it is possible to exchange the data with the PC via USB or Bluetooth.
NOTE! The data transfer and the functions of the Software are explained in the Manual Wohler A 550 PC Software.
12 Troubleshooting
Problem Empty battery icon No reaction of the analyzer
Possible reason Battery is empty.
Solution
Recharge battery.
Reset: Push the transport lock to the right and then to the left again (see
Fig. 21)
72
Maintenance
13 Maintenance
Proper operation of the Wohler A 550 INDUSTRIAL requires regular maintenance.
13.1 Maintenance work
Interval
Maintenance work
After every
Remove condensate from the
combustion test coarse filter.
Check the condensate trap and remove condensate and water if necessary.
Check the cotton filter. Replace it if polluted or wet.
After the wood moisture measurement
After every working day
Regularly
Once a year
Clean the sinter filter by rinsing it with water or putting it into an
Ultrasonic bath.
Control the distance between the needles: Replace needles if necessary
Dry coarse filter
Replace the cotton filter.
Return analyzer for check and calibration to a Wohler certified service
station
If necessary
Grease the seal rings with silicone.
13.2 Replacing the filters
The sensible sensors are protected against condensate by different filters.
The filters must be replaced or dried regularly.
73
Maintenance
13.2.1 Coarse Filter
Plexiglas plug
·
· ·
· · · Fig. 66: Probe handle with gas probe
A coarse filter located in the probe handle prevents soot from entering the
hose. After every measurement control if it is wet. Proceed as follows:
Pull the plexiglas plug out of the probe handle.
Use tweezers or a similar tool to remove the filter from its chamber.
Press (not rub) the filter to remove humidity.
Replace the plexiglas filter into its old position.
After the working day, remove the coarse filter and clean it by rinsing it in
warm water. Dry it afterwards.
Note! For most accurate SO2 and NO2 readings the coarse filter should not be installed during measurement.
13.2.2 Removing condensate
The condensate will be collected in the condensate trap. To remove the condensate follow the steps below:
· Open the condensate trap compartment as
2
follows: Press the cover on the right side (1)
and then pull it up (2). 1
Fig. 67: Open the condensate trap compartment
74
Maintenance · Pull the condensate trap from the unit.
Condensate outlet. ·
·
Remove the condensate cup from the condensate spiral. The condensate outlet
remains in the cup.
Spill the condensate out of the cup.
Condensate spiral Condensate cup
Fig. 68: Condensate trap
13.2.3 Replace cotton filter
The cotton filter is located in the condensate spiral. · After every
combustion test control the cotton
filter and replace it if wet or dirty.
Fig. 69: Condensate spiral with cotton filter (internally located).
13.2.4 Water Stop Filter
Fig. 70: Water stop filter
Behind the condensate spiral a water stop filter is located. The water stop
filter blocks water from entering the analyzer. When the water stop filter
closes, it will change color.
· In this case replace the water stop filter as follows:
· Remove the Water Stop Filter from the unit.
· Plug the new water stop filter into the opening so that the red letters are
on the unit’s side.
75
Maintenance
13.3 Sensor diagnosis and replacement
Electrochemical gas sensors are subject to wear. The life of the measuring sensors will be determined by numerous parameters such as the removal of condensate, the frequency of use and regular maintenance.
Therefore this information about the average life of measuring sensors is based on specific experience.
O2 – sensor
4 years
CO-sensor
2 years
NO-sensor
2 years
NO2-sensor 2 years
SO2-sensor 2 years
The sensors can be changed by the customer or by an authorized service point or at the factory.
The Wohler A 550 INDUSTRIAL has an advanced electronic self diagnosis program.
You can enter the diagnosis screen directly after having switched on the
analyzer during self check and calibration .
· Click on the Diagnosis icon when the hotspot screen appears.
76
Maintenance
· The diagnosis screen shows the sensor status (OK or NOT OK) and the firmware
version.
· A click on the sensor opens a window with important sensor information.
· Click on “Back” to start the search for the hotspot again.
Fig. 71: Sensor diagnosis Status: OK:
Status: Service Zeroing sensor
The sensor status is continuously displayed in the right top of the screen .
The list below shows the possible diagnosis feedback:
· Status icon: Yellow sign icon indicates a problem.
· “OK” or “Service” will appear for all sensors.
· Zeroing pressure sensor: During zeroing of the pressure sensor the blue icon
appears on the screen. Do not expose any differential pressure to the analyzer
now.
77
Maintenance
NOTE! If the “service” message appears, first turn the analyzer off and on again. After the analyzer has been turned on, it will automatically start to calibrate the sensors with fresh air. Start the diagnosis screen again. Only if the service message still appears on screen, the sensor will have to be changed.
The sensors can be replaced in the factory, in the service points or by the user himself. The order of the sensors is as follows (from the left to the right): Position 1: O2-Sensor, Position 2 4: NO2-Sensor or CO2 NDIR-Sensor or SO2-Sensor or H2S-Sensor Position 5: CO-sensor
NOTE! If the user himself replaces the sensors, the sensor date shown in the diagnosis menu will update automatically.
Replace the sensor as follows:
· Switch off the analyzer.
· Loosen the three bolds of the cover. Disengage the cover hooks as follows:
· First press with both thumbs on the rubber coating in the middle and then on
the two outer rubber coatings.
· Remove the cover.
Fig. 72: Bottom of the unit – Press to open the cover.
78
Maintenance
Battery Flexible conductor
Connecting tube for gas channel
O2-sensor
2 pin connector of the O2 sen-
optional sensor
optional sensor
optional sensor
CO-sensor
Optional sensor: CO2 NDIR, SO2, H2S, NO, NO2
Fig. 73: Bottom of the analyzer without cover
· Slightly lift the left side of the sensor holder and remove the black
connecting tube for the gas channel.
· Take the sensor holder out of the unit (first the left and then the right
side).
79
Maintenance Replacing the O2 Sensor
NOTE! See Fig. 73 for the position of the sensor.
· Remove the 2-pin connector of the O2 sensor. · Replace the O2 sensor by a
new one. · Plug the new 2-pin connector where the old
one was.
80
Maintenance
Replacing the other sensors (CO2 NDIR, NO2, SO2, H2S, NO, CO)
2
1
NOTE! See Fig. 73 for the position of the sensor.
Pull the small locking mechanism (1) forward. · Take off the white flexible
conductor (2). · Replace the sensor by a new one. · Plug the flexible
conductor in his place and
push the locking mechanism back.
Fig. 74: Sensor with locking mechanism (1) and flexible conductor (2)
·
·
Put the sensor holder back to its place, first the right side and then the
left side.
Connect the gas channel again by inserting the connecting tube of the gas
channel into the sensor holder.
81
Maintenance
ATTENTION! Make sure that the connecting tube is not bent.
· Make sure that the connecting tube is in its original position marked by the
guiding hooks.
ATTENTION! After the NO sensor has been replaced, an activation time of 24
hours must be observed. During the activation time, the position of the
transport lock in the probe compartment must be “ON”, so that the battery is
active (see Fig. 8).
Fig. 75: Connecting tube, correctly positioned, guiding hooks marked by an
arrow.
13.4 Replacing the battery
WARNING! Incorrect use of the battery can cause injury! Never dispose the
battery to fire or high temperatures. Danger of explosion! Liquids can leak
out of the battery due to mistreatment. Never touch the liquid. If you get in
contact to the liquid, remove it with water. If liquid has got into your eyes,
remove it with water and see your doctor as soon as possible.
The Wohler A 550 INDUSTRIAL is equipped with a rechargeable lithium ion
battery 3.7 V. Only after a long working period the battery has to be changed.
In this case send the analyzer to your dealer or change the battery as
follows:
· Switch off the analyzer. Open the analyzer as described before in the
chapter “sensor replacement”.
82
Maintenance · Pull out the battery plug and remove the bat-
tery. · Install a new battery and plug in the battery
plug.
Fig. 76: Pulling out the battery plug · Make sure that the connecting tube is
in its original position marked by the guiding hooks. · Put the cover back to
its original position. · Fix the cover with the three screws.
Fig. 77: Connecting tube, correctly positioned
83
Analyzer check
13.5 Replacing the electrodes of the wood moisture probe
· Control if the electrodes are damaged. Take care that the distance between
the electrodes is 25 mm.
· Broken or damaged electrodes can be exchanged by the user. Loosen the
damaged electrodes with an appropriate tool (gripper, screw wrench or ring
wrench).
· Replace the damaged electrodes by the electrode-kit and fix the new
electrodes with the tool.
14 Analyzer check
We recommend to check the analyzer at least every 12 month at an authorized service center. Local regulations may require more frequent checking of the analyzer.
The following tests have to be performed:
·
With test gas: O2 measurement, CO measurement, NO measurement, , NO2
measurement and SO2 measurement
·
The flue gas temperature TS has to be checked with a reference standard at two
measuring points: one point in the lower part and one point in the upper part of the
measuring range.
·
The ambient air temperature TA will be checked at one measuring point with a ref-
erence standard.
·
The draft has to be checked with an adequate draft regulator.
·
The flow of the flue gas will be controlled by a rotameter at the sucking side of the
gas probe.
·
The gas probe and the analyzer will be optically checked for pollution, condensa-
tion etc.
·
Control of the measuring channel for wood moisture with a reference resistance.
According to VDI 4206, page 4 the displayed wood moisture may not differ more
than ± 5% from the reference value.
84
Warranty and Service
· 15 Warranty and Service
Each Wohler A 550 INDUSTRIAL will be tested in all functions and will leave
our factory only after extensive quality control testing. The final control
will be recorded in detail in a test report and delivered with any unit. If
used properly, the warranty period for the Wohler A 550 INDUSTRIAL will be 12
months from the date of purchase. Consumables (e.g. thermocouple) are not
covered by this warranty. This warranty does not cover the freight and packing
costs when the device is sent to the factory for repair. Service by non
authorized personnel or making modifications to the analyzer voids any
warranty. Wohler has built our reputation on excellence in customer service.
Therefore, of course, we are readily available to assist you after the
warranty period ends. · Send us the device and we will repair it and return it
to you with our package ser-
vice. · Immediate help is provided by our technical staff over the telephone.
16 Declaration of conformity
The manufacturer: Wohler Technik GmbH Wöhler Platz 1, D-33181 Bad Wünnenberg
declares that the product product name: Flue gas analyzer model number: Wohler
A 550 INDUSTRIAL complies with the key safety requirements set down in the
guidelines of the Council for the Harmonization of the Legal Requirements of
the Member States in relation to the electromagnetic compatibility
(2014/30/EU). The analyzer complies with : EN 50379-2 VDI 4206-1 VDI 4206-4
The following standards were availed of to evaluate the product in respect of
the electromagnetic compatibility:
EN 50270:2006 EN 61000-6-3:2007 + A1:2011
85
Accessories for Wöhler A 550 Industrial
17 Accessories for Wöhler A 550 Industrial
USB Peltier Cooler USB Peltier Cooler for Wöhler A 550 Probes Flue gas probe
1,000 mm with filter holder and stainless steel sinter filter Flue gas probe
1,000 mm with filter holder and stainless steel sinter filter Wohler A 550 Gas
Probe 500 mm Wohler A 550 Gas Probe 295 mm Wohler A 550 Flue Gas Probe 1,000
mm Wohler A 550 Air Temperature Probe 280 mm, with 1.7 m cable Wohler Velocity
Probe Type S for Wohler A 550, to measure gas velocity
Pitot Tube, length 100 cm, 7 mm dia. High Temperature Probe 1,000 mm,
extendable Printer Wohler TD 100 Thermal Printer Paper for Wohler TD 100
Thermal Printer Extras Wohler Magnet Holder Wohler Clamping Cone for probes 8
mm Ø Wohler PTFE Cone for probes 8 mm Ø Wohler CHP-filter Spare parts and
consumables Water Stop Filters, 3 pcs. Coarse Filters, 5 pcs. Wadding Filters,
short, pack with 150 pieces Software PC Software Wohler A 550
P/N 4463
P/N 4435
P/N 4187
P/N 9614 P/N 9622 P/N 9695 P/N 5511
P/N 5579 P/N 9489 P/N 2291
P/N 4160 P/N 4145
P/N 6142 P/N 2494 P/N 2463 P/N 11136
P/N 9621 P/N 9632 P/N 5290
P/N 4428
86
Accessories for Wöhler A 550 Industrial
Field Replacable Sensors NO sensor Wöhler A 550 NO2 sensor Wöhler A 550 COHigh sensor 100.000 ppm Wöhler A 550 O2 sensor Wöhler A 550 SO2 sensor Wöhler A 550 CO2 NDIR-Sensor H2S-Sensor Warranty Warranty extension Wöhler A 550 IND by one year
P/N 5597 P/N 5598 P/N 5596 P/N 5594 P/N 5665 P/N 11011 P/N 11014
P/N 599
87
Points of sale and service
Points of sale and service
Germany
Wöhler Technik GmbH Wöhler-Platz 1 33181 Bad Wünnenberg Tel.: +49 2953 73-100
Fax: +49 2953 73-96100 info@woehler.de
www.woehler.de
USA Wohler USA Inc. 208 S Main Street Middleton, MA 01949 Tel.: +1 978 750
9876 www.wohlerusa.com
Italy Wöhler Italia srl Via Coraine 21 37010 Costermano VR Tel. +39 045
6200080 Fax. +39 045 6201508 info@woehler.it
www.woehler.it
Austria Wöhler GmbH Industriezone Burgerfeld 16 3150 Wilhelmsburg Tel.: 0 27
46 / 313 13 10 Mail: info@woehler.at
Wöhler West Steiger-Stein-Str. 5 44805 Bochum Tel.: +49 234 516993-0 Fax: +49
234 516993-99 west@woehler.de
Wöhler Süd Gneisenaustr.12 80992 München Tel.: +49 89 1589223-0 Fax: +49 89
1589223-99 sued@woehler.de Czech Republic Wöhler
Bohemia s.r.o. Za Naspern 1993 393 01 Pelhrimov Tel.: +420 565 323 076 Fax:
+420 565 323 078 info@woehler.cz
France Wöhler France SARL 17 A impasse de Grousset 31590 LAVALETTE Tel.: +33 5
61 52 40 39 Fax: +33 5 62 27 11 31 info@woehler.fr
www.woehler.fr
Your contact:
88
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>