Blue-White APFCL Chlorine Turbidity Analyzer Panel User Guide

June 13, 2024
Blue-White

Blue-White APFCL Chlorine Turbidity Analyzer Panel

Product Information

The APFCL Chlorine Analyzer Panel is a turn-key monitoring solution designed for clean water applications. It offers highly accurate, real-time measurement, display, and data-logging of Free Chlorine or Total Chlorine, pH, and Temperature. The panel utilizes proprietary smart sensor technology and features a color display and data logging terminal. The APFCL is compliant with EPA 334.0 / ISO 7393 standards.

Features

  • Real-time measurement of Free Chlorine or Total Chlorine, pH, and Temperature
  • Proprietary smart sensor technology
  • Color display and data logging terminal
  • Compliant with EPA 334.0 / ISO 7393 standards

Specifications

  • Power Supply : 100~240VAC 50/60Hz
  • Dimensions (in inches) : A – 12.00, B – 24.00, C – 15.67, D – 14.05, E – 1.98
  • Weight: Approximately 45 lbs (20kg)

Product Usage Instructions

Pre-Installation

Before you begin

  • Ensure the location of the analyzer complies with existing codes and regulations.
  • The mounting location must be strong enough to support the weight of the analyzer and water. It is not designed to support any other structures, piping, or equipment.

Wiring
To access the wiring terminal

  1. Open the enclosure by loosening the four screws with a screwdriver.
  2. The analyzer must be powered by 100~240VAC 50/60Hz. Use a suitable power supply.
  3. Communication wiring should be 20-24 AWG shielded cable.

Plumbing
Plan the incoming water supply and drainage piping before installing the analyzer. Additional components may be required to connect to the provided tubing.

Installation

Installation Requirements

  • Power Supply: 100~240VAC 50/60Hz
  • Drainag e: Connect the WR05 outlet tube (3/4 ID Tubing) and the WR05 weir overflow (3/8-inch O.D. Tubing) to a discharge drain via gravity flow.
  • Weir Overflow : Connect the -inch weir overflow tubing.
  • Outlet Line : Connect 3/4 ID tubing to the outlet drain. This line must be diverted to drain.
  • Wall Mount Space : The APFCL analyzer panel size is roughly 12 H x 24 W x 10 D in dimension. Please accommodate sufficient space for mounting. The panel is equipped with pre-installed rear unistrut for simple wall mounting.
  • Wall Mount Weight: Approximately 45 lbs (20kg). Please use appropriate mounting hardware.

Note: For more information or assistance, please contact customer service at customerservice@blue-white.com or call 714-893-8529. You can also scan the QR code to download the full Instruction Manual.

Congratulations on your purchase on a APFCL Chlorine

analyzer!
The APFCL Analyzer Panel is specifically designed as a ‘Turn-Key’ monitor-ing solution for clean water applications including drinking water networks, secondary water supply and decorative/swimming water applications.
The APFCL series offers highly accurate, real-time measurement, display and data-logging of Free Chlorine or Total Chlorine, pH and Temperature uti-lizing proprietary smart sensor technology, coupled with a color display and data logging terminal. The APFCL is compliant with EPA 334.0 / ISO 7393 standards.
Reference the below dimensional drawings for installation.

APFCL Series Panel Dimensions

IMPORTANT

Safety instructions

  • CAUTION : Follow these instructions before installing your analyzer to avoid failure. Read this guide and Instruction Manual before using.
  • CAUTION : The APFCL is designed to be installed and operated by qualified personnel only. Please note that warranty coverage does not include damage due to misuse or improper installation.
  • CAUTION : Always wear eye protection when using chemicals.
  • PRESSURE :Do not exceed maximum pressure. High pressure and temperature will damage the system.

PRE-INSTALLATION
Before you begin

In the Box

  • APFCL Analyzer Panel
  • Pre-Mounted Struts
  • Pre-mounted Flow Meter
  • Chlorine Sensor
  • 1/4” tubing OD
  • 3/8” tubing OD
  • 3/4” tubing ID
  • Concrete Anchors
  • Electrical Schematics
  • Quick Start Guide

Tools Needed

  • Phillips Screwdriver
  • Adjustable wrench

Planning
Ensure location of Analyzer complies with existing codes and regulations. Location of mounting must be strong enough to support weight of Analyzer and water. Analyzer and mounts are not designed to support any other structures, piping, or equipment.

Wiring
To access the wiring terminal, open the enclosure by loosening the four screws with the screwdriver. The Analyzer must be powered by 100~240VAC 50/60Hz. Communication wiring should be 20-24 AWG(American Wire Gauge) shielded cable.

Plumbing
Incoming water supply and drainage piping should be planned before installing the Analyzer. Additional components may be required to connect to the provided tubing.

Need more information. Please contact customer service with questions. customerservice@blue-white.com 714-893-8529
Scan QR code to download full Instruction Manual.

Installation Requirements

  • Power Supply: 100~240VAC 50/60Hz

  • Water Supply : Inlet water pressure should be from 7.25 – 30 psi (0.05-0.2MPa) with an inlet feedwater line diameter of ¼-inch O.D. Tubing. The APFCL is pro-vided with an inlet Rotameter and flow regulating valve for sample water inlet flow control and limited pressure regulation. The range of inlet flow for the WR05 should be consistently maintained between 200 and 400 mL per minute.
    Inlet Water: Connect the ¼-inch inlet water tubing to the quick adapter provided. Consistent flow of 200-400mL is required.

  • Drainage: The WR05 outlet tube (3/4” ID Tubing) located on the bottom of the WR05, as well as the WR05 weir overflow (3/8-inch O.D. Tubing) located on the bottom of the WR05, should both be connected to a discharge drain via gravity flow.

  • Weir Overflow : Connect the ⅜-inch weir overflow tubing.

  • Outlet Line : Connect 3/4” ID tubing to the outlet drain. This is the sample water outlet flow. This line must be diverted to drain.

  • Wall Mount Space: The APFCL analyzer panel size is roughly 12” H x 24” W x 10” D in dimension. Please accommodate sufficient space for mounting. The panel is equipped with pre-installed rear unistrut for simple wall mounting.

  • Wall Mount Weight : Approximately 45 lbs (20kg). Please use appropriate mounting hardware.

Wiring

The APFCL analyzer has universal AC power supply equipment allowing users simply to plug the power supply into a 100~240V AC 50/60Hz power outlet for normal operation.
The process of electrical connection to contact the 220V single-phase power supply, should be operated by personnel with an electrician’s license. Failure to operate according to the electrical code of practice may result in electric shock injury or even death.

Note: When in doubt regarding your electrical installation, contact a licensed electrician.

INSTALLATION
Step-By-Step

Measure

  • Connect Electrical and Tubing
  • Use the provided drawing dimensions to determine best mounting location of panel.
  • Measure wall and locate mounting holes. Panels should be mounted at an elevation to allow for easy access to control unit.
  • Wall Mount Weight : 45 lbs

Install Strut

  • Use the provided strut and hardware to mount the panel.
  • Strut to be mounted securely to wall using appropriate hardware. Strut spacing is on 18” centerline.
  • If using provided anchor bolts, drill holes in concrete and clean thoroughly. Located strut and mount with appropriate tools.
  • Check to make sure strut is secure and level.

Mount Panel

  • Mounting panel will requires 2 or 3 persons.
  • Mount panel to strut using provided hardware.
  • Be sure panel is level before securing.
  • Check spacing and panel before connecting electrical and piping.

Connect Electrical and Tubing
Use provided electrical diagram to connect power and communication wiring.
Consult with electrical profession before performing any electrical work.

Power Up System
Check all electrical connections before powering on system. Check piping connections for leaks.
System is activated when power is applied.

Start-Up and Operation
After the TS05 is turned on, the main interface will display the chlorine residu-al, pH value, and temperature of the solution. The green dot indicates that the communication between the TS-05 and the probe is normal. To access other options, press the BACK Key and refer to Instruction Manual for more information regarding features and user management.

NOTE: The FCL Sensor has a 5 minute “Initialization period” before readings begin.

Real Time Monitoring

Settings
Upon pushing the Back Key, the Settings options will appear. For further programming instructions and a functional overview of each section, please refer to complete Instruction Manual.

Engineering SpecificationsBlue-White-APFCL-Chlorine Turbidity-Analyzer
Panel 6Blue-White-APFCL-Chlorine Turbidity-
Analyzer Panel 7

WARRANTY

LIMITED WARRANTY
Your Blue-White product is a quality product and is warranted for a specific time from the date
of purchase (proof of purchase is required). The product will be repaired or replaced at our discretion. Failure must have occurred due to a defect in material or workmanship and not as a result of the operation of the product other than in normal operation as defined in the product manual. Warranty status is determined by the product’s serial label and the sales invoice or receipt. The serial label must be on the product and legible. The warranty status of the product will be verified by Blue-White or a factory-authorized service center.
APFCL is warranted for 13 Months from the date of purchase. Sensors are warranted for 6 months.

WHAT IS NOT COVERED

  • Freight to the factory, or service center.
  • Products that have been tampered with, or in pieces.
  • Damage resulting from misuse, carelessness such as chemical spills on the enclosure, abuse, lack of maintenance, or alteration which is out of our control.
  • Damage by faulty wiring, power surges, or acts of nature.
  • Damage that occurs as a result of : meter misalignment, improper installation, over tightening, use of non-recommended chemicals, use of non-recommended adhesives or pipe dopes, excessive heat or pressure, or allowing the unit to support the weight of related piping.

BLUE-WHITE does not assume responsibility for any loss, damage, or expense directly or indirectly related to or arising out of the use of its products. Failure must have occurred due to a defect in material or workmanship and not as a result of the operation of the product other than in normal operation as defined in the manual.
Warranty status is determined by the product’s serial label and the sales invoice or receipt. The serial label must be on the product and legible. The warranty status will be verified by Blue-White or a factory-authorized service center.
PROCEDURE FOR IN-WARRANTY REPAIR Warranty service must be performed by the factory or an authorized service center. Contact the factory or local repair center to obtain a RMA (Return Material Authorization) number. Decontaminate, dry, and carefully pack the product to be repaired. Please enclose a brief description of the problem and proof of purchase. Prepay all shipping and insurance costs. COD shipments will not be accepted. Damage caused by improper packaging is the responsibility of the sender. When In-Warranty repair is completed, the factory pays for return shipping to the dealer or customer.

Have Questions?

blue-white.com/contact-us

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals