Blue-White APFCL Chlorine Turbidity Analyzer Panel User Guide
- June 13, 2024
- Blue-White
Table of Contents
Blue-White APFCL Chlorine Turbidity Analyzer Panel
Product Information
The APFCL Chlorine Analyzer Panel is a turn-key monitoring solution designed for clean water applications. It offers highly accurate, real-time measurement, display, and data-logging of Free Chlorine or Total Chlorine, pH, and Temperature. The panel utilizes proprietary smart sensor technology and features a color display and data logging terminal. The APFCL is compliant with EPA 334.0 / ISO 7393 standards.
Features
- Real-time measurement of Free Chlorine or Total Chlorine, pH, and Temperature
- Proprietary smart sensor technology
- Color display and data logging terminal
- Compliant with EPA 334.0 / ISO 7393 standards
Specifications
- Power Supply : 100~240VAC 50/60Hz
- Dimensions (in inches) : A – 12.00, B – 24.00, C – 15.67, D – 14.05, E – 1.98
- Weight: Approximately 45 lbs (20kg)
Product Usage Instructions
Pre-Installation
Before you begin
- Ensure the location of the analyzer complies with existing codes and regulations.
- The mounting location must be strong enough to support the weight of the analyzer and water. It is not designed to support any other structures, piping, or equipment.
Wiring
To access the wiring terminal
- Open the enclosure by loosening the four screws with a screwdriver.
- The analyzer must be powered by 100~240VAC 50/60Hz. Use a suitable power supply.
- Communication wiring should be 20-24 AWG shielded cable.
Plumbing
Plan the incoming water supply and drainage piping before installing the
analyzer. Additional components may be required to connect to the provided
tubing.
Installation
Installation Requirements
- Power Supply: 100~240VAC 50/60Hz
- Drainag e: Connect the WR05 outlet tube (3/4 ID Tubing) and the WR05 weir overflow (3/8-inch O.D. Tubing) to a discharge drain via gravity flow.
- Weir Overflow : Connect the -inch weir overflow tubing.
- Outlet Line : Connect 3/4 ID tubing to the outlet drain. This line must be diverted to drain.
- Wall Mount Space : The APFCL analyzer panel size is roughly 12 H x 24 W x 10 D in dimension. Please accommodate sufficient space for mounting. The panel is equipped with pre-installed rear unistrut for simple wall mounting.
- Wall Mount Weight: Approximately 45 lbs (20kg). Please use appropriate mounting hardware.
Note: For more information or assistance, please contact customer service at customerservice@blue-white.com or call 714-893-8529. You can also scan the QR code to download the full Instruction Manual.
Congratulations on your purchase on a APFCL Chlorine
analyzer!
The APFCL Analyzer Panel is specifically designed as a ‘Turn-Key’ monitor-ing
solution for clean water applications including drinking water networks,
secondary water supply and decorative/swimming water applications.
The APFCL series offers highly accurate, real-time measurement, display and
data-logging of Free Chlorine or Total Chlorine, pH and Temperature uti-lizing
proprietary smart sensor technology, coupled with a color display and data
logging terminal. The APFCL is compliant with EPA 334.0 / ISO 7393 standards.
Reference the below dimensional drawings for installation.
APFCL Series Panel Dimensions
IMPORTANT
Safety instructions
- CAUTION : Follow these instructions before installing your analyzer to avoid failure. Read this guide and Instruction Manual before using.
- CAUTION : The APFCL is designed to be installed and operated by qualified personnel only. Please note that warranty coverage does not include damage due to misuse or improper installation.
- CAUTION : Always wear eye protection when using chemicals.
- PRESSURE :Do not exceed maximum pressure. High pressure and temperature will damage the system.
PRE-INSTALLATION
Before you begin
In the Box
- APFCL Analyzer Panel
- Pre-Mounted Struts
- Pre-mounted Flow Meter
- Chlorine Sensor
- 1/4” tubing OD
- 3/8” tubing OD
- 3/4” tubing ID
- Concrete Anchors
- Electrical Schematics
- Quick Start Guide
Tools Needed
- Phillips Screwdriver
- Adjustable wrench
Planning
Ensure location of Analyzer complies with existing codes and regulations.
Location of mounting must be strong enough to support weight of Analyzer and
water. Analyzer and mounts are not designed to support any other structures,
piping, or equipment.
Wiring
To access the wiring terminal, open the enclosure by loosening the four screws
with the screwdriver. The Analyzer must be powered by 100~240VAC 50/60Hz.
Communication wiring should be 20-24 AWG(American Wire Gauge) shielded cable.
Plumbing
Incoming water supply and drainage piping should be planned before installing
the Analyzer. Additional components may be required to connect to the provided
tubing.
Need more information. Please contact customer service with questions.
customerservice@blue-white.com
714-893-8529
Scan QR code to download full Instruction Manual.
Installation Requirements
-
Power Supply: 100~240VAC 50/60Hz
-
Water Supply : Inlet water pressure should be from 7.25 – 30 psi (0.05-0.2MPa) with an inlet feedwater line diameter of ¼-inch O.D. Tubing. The APFCL is pro-vided with an inlet Rotameter and flow regulating valve for sample water inlet flow control and limited pressure regulation. The range of inlet flow for the WR05 should be consistently maintained between 200 and 400 mL per minute.
Inlet Water: Connect the ¼-inch inlet water tubing to the quick adapter provided. Consistent flow of 200-400mL is required. -
Drainage: The WR05 outlet tube (3/4” ID Tubing) located on the bottom of the WR05, as well as the WR05 weir overflow (3/8-inch O.D. Tubing) located on the bottom of the WR05, should both be connected to a discharge drain via gravity flow.
-
Weir Overflow : Connect the ⅜-inch weir overflow tubing.
-
Outlet Line : Connect 3/4” ID tubing to the outlet drain. This is the sample water outlet flow. This line must be diverted to drain.
-
Wall Mount Space: The APFCL analyzer panel size is roughly 12” H x 24” W x 10” D in dimension. Please accommodate sufficient space for mounting. The panel is equipped with pre-installed rear unistrut for simple wall mounting.
-
Wall Mount Weight : Approximately 45 lbs (20kg). Please use appropriate mounting hardware.
Wiring
The APFCL analyzer has universal AC power supply equipment allowing users
simply to plug the power supply into a 100~240V AC 50/60Hz power outlet for
normal operation.
The process of electrical connection to contact the 220V single-phase power
supply, should be operated by personnel with an electrician’s license. Failure
to operate according to the electrical code of practice may result in electric
shock injury or even death.
Note: When in doubt regarding your electrical installation, contact a licensed electrician.
INSTALLATION
Step-By-Step
Measure
- Connect Electrical and Tubing
- Use the provided drawing dimensions to determine best mounting location of panel.
- Measure wall and locate mounting holes. Panels should be mounted at an elevation to allow for easy access to control unit.
- Wall Mount Weight : 45 lbs
Install Strut
- Use the provided strut and hardware to mount the panel.
- Strut to be mounted securely to wall using appropriate hardware. Strut spacing is on 18” centerline.
- If using provided anchor bolts, drill holes in concrete and clean thoroughly. Located strut and mount with appropriate tools.
- Check to make sure strut is secure and level.
Mount Panel
- Mounting panel will requires 2 or 3 persons.
- Mount panel to strut using provided hardware.
- Be sure panel is level before securing.
- Check spacing and panel before connecting electrical and piping.
Connect Electrical and Tubing
Use provided electrical diagram to connect power and communication wiring.
Consult with electrical profession before performing any electrical work.
Power Up System
Check all electrical connections before powering on system. Check piping
connections for leaks.
System is activated when power is applied.
Start-Up and Operation
After the TS05 is turned on, the main interface will display the chlorine
residu-al, pH value, and temperature of the solution. The green dot indicates
that the communication between the TS-05 and the probe is normal. To access
other options, press the BACK Key and refer to Instruction Manual for more
information regarding features and user management.
NOTE: The FCL Sensor has a 5 minute “Initialization period” before readings begin.
Real Time Monitoring
Settings
Upon pushing the Back Key, the Settings options will appear. For further
programming instructions and a functional overview of each section, please
refer to complete Instruction Manual.
Engineering Specifications
WARRANTY
LIMITED WARRANTY
Your Blue-White product is a quality product and is warranted for a specific
time from the date
of purchase (proof of purchase is required). The product will be repaired or
replaced at our discretion. Failure must have occurred due to a defect in
material or workmanship and not as a result of the operation of the product
other than in normal operation as defined in the product manual. Warranty
status is determined by the product’s serial label and the sales invoice or
receipt. The serial label must be on the product and legible. The warranty
status of the product will be verified by Blue-White or a factory-authorized
service center.
APFCL is warranted for 13 Months from the date of purchase. Sensors are
warranted for 6 months.
WHAT IS NOT COVERED
- Freight to the factory, or service center.
- Products that have been tampered with, or in pieces.
- Damage resulting from misuse, carelessness such as chemical spills on the enclosure, abuse, lack of maintenance, or alteration which is out of our control.
- Damage by faulty wiring, power surges, or acts of nature.
- Damage that occurs as a result of : meter misalignment, improper installation, over tightening, use of non-recommended chemicals, use of non-recommended adhesives or pipe dopes, excessive heat or pressure, or allowing the unit to support the weight of related piping.
BLUE-WHITE does not assume responsibility for any loss, damage, or
expense directly or indirectly related to or arising out of the use of its
products. Failure must have occurred due to a defect in material or
workmanship and not as a result of the operation of the product other than in
normal operation as defined in the manual.
Warranty status is determined by the product’s serial label and the sales
invoice or receipt. The serial label must be on the product and legible. The
warranty status will be verified by Blue-White or a factory-authorized service
center.
PROCEDURE FOR IN-WARRANTY REPAIR Warranty service must be performed by the
factory or an authorized service center. Contact the factory or local repair
center to obtain a RMA (Return Material Authorization) number. Decontaminate,
dry, and carefully pack the product to be repaired. Please enclose a brief
description of the problem and proof of purchase. Prepay all shipping and
insurance costs. COD shipments will not be accepted. Damage caused by improper
packaging is the responsibility of the sender. When In-Warranty repair is
completed, the factory pays for return shipping to the dealer or customer.