GOSSEN METRAWATT METRAHIT 2+ Universal TRMS Multimeter Instruction Manual

June 12, 2024
GOSSEN METRAWATT

GOSSEN METRAWATT METRAHIT 2+ Universal TRMS Multimeter Instruction Manual
Multimeter

Please make sure to read the detailed operating instructions in pdf format at www.gossenmetrawatt.com. The short-form instructions are no substitute for the detailed instructions!

Standard Equipment

  • 1 TRMS Multimeter
  • 1 protective rubber holster
  • 1 Set of measuring cables KS17-ONE
  • 2 AA size batteries 1,5 V
  • 1 Short-form Operating Instructions
  • 1 DAkkS calibration certificate

Safety Instructions

In order to maintain the flawless condition of the instrument, and to ensure its safe operation, it is imperative that you read the operating instructions thoroughly and carefully before placing your instrument into service, and that you follow all instructions contained therein. Observe the following safety precautions: The instrument may only be operated by persons who are able to recognize contact hazards and take the appropriate safety precautions. Contact hazards exist wherever voltages of more than 33 V (RMS value) and/or 70 V DC occur. The maximum allowable voltage between the instrument and earth is equal to 600 V, category III.

The nominal voltage of the installation may not exceed the following values:

  • in monophase systems between conductor and neutral: 600 V,
  • in 4-wire 3-phase systems between phase conductors: 690 V,
  • in 3-wire 3-phase systems between phase conductors: 1000 V.

In conformity with standard DIN EN 61010-031, measurements in an environment according to measuring category III may only be performed with the safety cap applied to the test probe of the measurement cable.

Attention: Unexpected voltages may occur at defective devices, capacitors,…!
The insulation of the measurement cables may not be damaged, cables and plugs may not be interrupted!
No measurements may be made in electrical circuits with corona discharge (high-voltage)! Special care is required when measurements are made in HF electrical circuits where dangerous pulsating voltages may be present.
Measurements under moist ambient conditions are not permissible. Do not overload the measuring ranges beyond their allowablecapacities! Use original fuses only

Switching on/off
Instruction

**Select Measuring Function

**

Measuring Inputs
Instruction

Battery Test
Instruction

Battery Replacement

Instruction

 Disconnect the instrument from the measuring circuit before opening the instrument to replace the batteries or the fuse!

Hold / MAX / MIN
Instruction

Memory MAX/MIN
Instruction

Relative Measurement
Instruction

Measuring Range Selection
Instruction

Alternating Voltage Measurement (TRMS)
Instruction

Direct Voltage Measurement
Instruction

Resistance Measurement
Instruction

Continuity Testing
Instruction

Diode Testing
Instruction

Temperature Measurement
Instruction

Measuring Range Selection
Instruction

Current Measurement (TRMS)
Instruction

Current Measurement (TRMS)
Instruction

Technical Data

Measuring range Measuring Error Overload capacity 1)
5)
±( … % v. MW. + … D)
600 mV 0,5 + 5 1 + 5

6     V
60     V
600     V
| | 5)| |
60 mA| 1,0 + 5 (> 10 D)| 1,5 + 5 (> 10 D)| 1,0 A| continuous cont. perma- nente
600 mA| 1,0 + 5| 1,5 + 5 (> 10 D)
6     A| 1,0 + 5 (> 10 D)| 1,5 + 5 (> 10 D)| 10 A 4)
10     A| 1,0 + 5| 1,5 + 5 (> 10 D)
| ±(|… % v. MW.| + … D)|
600  W| 1 + 5 2)|

600 V DC ACrms/ sine

|

max. 10 s

6 kW| 0,7 + 3
60 kW| 0,7 + 3
600 kW| 0,7 + 3
6 MW| 0,7 + 3
40 MW| 2,0 + 3
| 2     V| 1 + 5
600 W| 1 + 5
| ±(|… % v. MW.| + … K)| |
TIPO K| – 50,0 …+400,0 °C| 1,0 + 5 K 3)| 600 V DC/ACrms/ sine| max. 10 s
| | ±(|… % v. MW.| + … D|)| |
Hz| 100 Hz| 0,1 + 2| 600 V 6)| max. 10 s
1000 Hz

  1. at / a / a 0 C … + 40 C 2) with zero balancing;
  2. without zero balancing + 35 digits / con
  3. without sensor
  4. in the range of
  5. 1 … 35 d nello zero con i puntali cortocircuitati, a causa del convertitore TRMS
  6. power limiting : 3 x 106 V x Hz @ U > 100 V Key
    MW = measured value
    d = digit

Electrical Safety

Protection class
Nominal Voltage 5,2 kV~2

IP XY(1st digit X)| Protection against foreign object entry| IP XY(2nd digit Y)| Protection against the penetration of water
---|---|---|---
4| ³ 1,0 mm Æ| 0| not protected

Electrical Safety of cable set KS17-ONE

Maximum Rated Voltage 600 V 600V
Measuring Category CAT III CAT II
Maximum Rated Voltage 1 A 16 A
with safety cap applied ·
without safety cap applied ·

Please observe the maximum values of the electrical safety of the device.

Electromagnetic Compatibility EMC

Interference Emission
61326-1 class B

Ambient Conditions

Accuracy Range 0 C … + 40 C

Operating Temperature –10 C … + 50 C

Storage Temperature – 25 C … + 70 C

Relative Humidity 45 … 75 %,

FusesShock

10 A: FF 10 A/1000V, 10 mm x 38 mm
600 mA: FF1,6 A/1000V, 6,3 mm x 32 mm

f you use other fuses than those indicated above you forfeit your warranty.

Internal Fuse Test
Instruction

Fuse Replacement
Instruction

Please make sure to read the detailed operating instructions in pdf format at www.gossenmetrawatt.com. The short-form instructions are no substitute for the detailed instructions!

Repair and Replacement Parts Service

When you need service, please contact:
GMC-I Service GmbH
Service-Center
Beuthener Straße 41
90471 Nürnberg • Germany
Phone +49 911 817718-0
Fax +49 911 817718-253
E-mail service@gossenmetrawatt.com
www.gmci-service.com

Product Support

When you need support, please contact:
Contacte con:
Rivolgersi a:
Gossen Metrawatt GmbH
Product Support Hotline
Phone +49 911 8602-0
Fax +49 911 8602-669
E-mail support@gossenmetrawatt.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

GOSSEN METRAWATT User Manuals

Related Manuals