GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter Instruction Manual

June 12, 2024
GOSSEN METRAWATT

GOSSEN METRAWATT logo Short-form Operating Instructions
5/12.21
METRAHIT ISO AERO
TRMS Multimeter with Insulation Measurement

Metrahit ISO Aero TRMS Multimeter

Please make sure to read the detailed operating instructions at www.gossenmetrawatt.com.
The short-form instructions are no substitute for the detailed instructions!
Symbol indicates parameter settings which are only described in the detailed operating instructions.

Standard Equipment

  • 1 Isolation multimeter
  • 1 Protective rubber cover
  • 1 Cable set KS17-2
  • 1 DAkkS calibration certificate
  • 2 Batteries 1.5 V, type AA inserted in unit
  • 1 Power pack NA X-TRA
  • 1 Short-form operating instructions*

Application of measuring cable set KS17-2

Maximum Rated Voltage 600 V 1000V
Measuring Category CAT III CAT I I
Maximum Rated Current 1 A 16 A
with safety cap applied
without safety cap applied

Safety Instructions

In order to maintain the flawless condition of the instrument, and to ensure its safe operation, it is imperative that you read the operating instructions thoroughly and carefully before placing your instrument into service, and that you follow all instructions contained therein.

Observe the following safety precautions:
The multimeter may not be used in potentially explosive atmospheres.
The multimeter may only be operated by persons who are able to recognize contact hazards and take the appropriate safety precautions. Contact hazards exist wherever voltages of more than 33 V (RMS value) and/or 70 V DC occur.
The maximum voltage allowable according to standard between the voltage inputs or all inputs towards earth respectively is equal to 1000 V, category II / 600 V, category III.
Attention: Unexpected voltages may occur at defective devices, capacitors,…!
The insulation of the measurement cables may not be damaged, cables and plugs may not be interrupted!
No measurements may be made in electrical circuits with corona discharge (high-voltage)!
Special care is required when measurements are made in HF electrical circuits where dangerous pulsating voltages may be present.
Measurements under moist ambient conditions are not permissible.
Do not overload the measuring ranges beyond their allowable capacities!
The input of the current measuring ranges is fitted with a fuse.
Use original fuses only, see label on the housing or technical data section!
Only operate the instrument with batteries inserted. Otherwise dangerous currents or voltages will not be indicated and your instrument may be damaged.
The device may not be operated with the fuse or battery compartment cover removed.

Switching on / off / Light on

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Switching on
off Battery TestGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Battery
Test

Battery Replacement

2 Batteries: IEC LR6 / AA – AM3 – Mignon
Disconnect the instrument from the measuring circuit before opening the battery compartment lid! Turn the slotted head screw counter-clockwise for this purpose. Observe the correct polarity of the batteries!
When refitting the battery compartment lid the side with the guide hooks must be inserted first. Then turn the slotted head srew clockwise.

Select ResolutionGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter -
Select Resolution Select Measuring FunctionGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter -
Select Measuring Function

Measuring Range Selection

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Measuring Range
Selection Measuring InputsGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Measuring
Inputs DATA- Hold/-CompareGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter -
Compare MIN/MAX memoryGOSSEN
METRAWATT Metrahit ISO Aero TRMS Multimeter - MAX memory ZERO
Zero BalancingGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter -
Zero Balancing Ω Resistance Measurement
Continuity Testing
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter -
Resistance Measurement

Diode Testing

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Diode
Testing Temp RTD
Temp TC
Temperature MeasurementGOSSEN METRAWATT Metrahit ISO Aero TRMS
Multimeter - Temperature Measurement V **/ V
Direct Voltage / Pulsating Voltage MeasurementGOSSEN METRAWATT Metrahit
ISO Aero TRMS Multimeter - Voltage Measurement V~/ Hz
AC Voltage – Frequency, without/with Low-pass FilterGOSSEN METRAWATT
Metrahit ISO Aero TRMS Multimeter - Low-pass Filter A / A
DC / Pulsating Current MeasurementGOSSEN METRAWATT Metrahit ISO Aero
TRMS Multimeter - Pulsating Current Measurement A~ / Hz
Alternating Current / Frequency MeasurementGOSSEN METRAWATT Metrahit
ISO Aero TRMS Multimeter - Frequency Measurement V / Hz
Measurement with Clip-on Current Sensor**GOSSEN METRAWATT Metrahit ISO
Aero TRMS Multimeter - Measurement with Clip-on Current
Sensor

Transformation ratio
CLIP| Measuring Ranges| Current sensor
---|---|---
300
mV| 3 V| 30 V
1:1
1mV/1mA| 0,3 A| 3 A| 30 A| WZ12C
1:10
1mV/10mA| 3 A| 30 A| 300 A| WZ12B
Z201A/B
VETRAFLEX
1:100
1mV/100mA| 30 A| 300 A| 3000 A| Z202A/B
VETRAFLEX
1:1000
1 mV/1 A| 300
A| 3000 A| (30kA)| WZ12C
Z202A/B
Z203A/B
VETRAFLEX

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Measurement with Clip-
on Current Sensor 2GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Current
Transformer 1

Transformation ratio
CLIP| Measuring Ranges| Zange Current transformer
---|---|---
30
mA| 300
mA| 3 A
1:1
1mA/1mA| 30
mA| 300
mA| 3 A|
1:10
1mA/10mA| 300
mA| 3 A| 30 A|
1:100
1mA/1 00mA| 3A| 30A| 300A|
1:1000
1 mA/1 A| 30 A| 300 A| 3000 A| WZ12A
WZ12D
WZ11A
73511
73512
73514

CapacitanceGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter -
Capacitance V 1MΩ (Ri = 1 MΩ)
Interference Voltage

MΩ@UISO GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter -
Interference VoltageDevice and Measuring Parameters
SETUP
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Device and
Measuring Parameters 1Main Menus →… SEnd … StorE …GOSSEN METRAWATT Metrahit
ISO Aero TRMS Multimeter - Device and Measuring Parameters
2 Sub-menus / Parameters ↓GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Device
and Measuring Parameters 3

Technical Data

Measuring Range Measuring Uncertainty Overload capacity
****
300 mV ±(0,2% + 3 D)
2) ±1/1,5% + 3 D (> 100 D) 1000 V

DC
AC
sine| continuous
3 V| ±(0,15% + 2 D)| ±1/1,5% + 3 D (> 100 D)
30 V| ±(0,15% + 2 D)
300 V| ±(0,15% + 2 D)
1000 V| ±(0,2% + 2 D)
1 Hz … 300,0 kHz| —| ±(0,1% + 2 D)| 1000 V 3)| max. 10 s
| ****| | |
300 μA| ±(0,5% + 5 D)| ±1,5% + 5 D (> 100 D)| 0,3 A| continuous
3 mA| ±(0,2% + 3 D)
30 mA| ±(0,5% + 3 D)
300 mA| ±(0,2% + 3 D)
3 A| ±(1% + 5 D)| 10 A| 5 min
10 A| ±(1% + 5 D)
1 Hz … 30,00 kHz| —| ±(0,1% + 2 D)| 4)|
| ±(… % v. MW/rdg. + … D)|
300 Ω| 0,5 + 3 2)| 1000 V
DC
AC
sine| max. 10 s
3 kΩ| 0,5 + 2
30 kΩ
300 kΩ
3 MΩ
30 MΩ| 2 + 5
300 Ω| 3 + 5
5,100 V| 2 + 5
| ±(… % v. MW/rdg. + … D)| |
30 nF| ±(1% + 6 D) 2)| 1000 V
DC
AC
sine| max. 10 s
300 nF| ±(1% + 6 D)
3 μF
30 μF
300 μF| ±(5% + 6 D)

  1. bei / at 0 ºC … + 40 ºC
  2. bei / with function ZERO active
  3. Power limiting: 3 · 10⁶ V · Hz @ U > 100 V
  4. max. current values see current measuring range

Insulation Measurement 1)

Measuring Range| nominal voltage UISO [V]| Measuring Uncertainty
±(… % v. MW/rdg. + … D)
---|---|---
0,3 V … 1000 V  2)| Ri=1MΩ| 3 + 30 > 100 Digit
5 … 310,0 kΩ| 10//50/100/250/500 V| 5 + 30 // 3 + 5
0,280 … 3,100 MΩ| 10//50/100/250/500 V| 5 + 30 // 3 + 5
02,80 … 31,00 MΩ| 10//50/100/250/500 V| 5 + 30 // 5 + 5
028,0 … 310,0 MΩ| 10//50/100/250/500 V| 5 + 30 // 5 + 5
0280 … 3100 MΩ| 500| 5 + 5

  1. During insulation resistance measurement (MΩ@UISO): If „Error“ is displayed >> limits: Uinterference > 10 … 20 V and Uinterference ≠ Uiso,
    Ri < 10 kΩ@ Uiso 10 V, Ri < 50 kΩ@ Uiso 50 V,
    Ri < 100 kΩ @ Uiso 100 V, Ri < 250 kΩ@ Uiso 250 V,
    Ri < 500 kΩ@ Uiso 500 V

  2. Interference voltage measurement TRMS (V AC + DC) with 1 MΩ input resistance, Bandwidth 15 Hz … 500 Hz, measuring error 3% + 30 Digit

Electrical Safety

Protection class II
– nach / per IEC 61010-1:2010/DIN EN 61010-1:2011/VDE 0411-1:2011
Measuring Category CAT II / CAT III
Nominal Voltage 1000 V / 600 V
Pollution degree 2
Test Voltage (nach/per IEC 61010-1) 5,2 kV~
Protection Housing: IP54 (pressure equalization by means of the housing)

Electromagnetic Compatibility EMC

Interference Emission
EN 61326-1:2013 class B
Interference Immunity
EN 61326-1:2013, EN 61326-2-1:2013

Ambient Conditions

Accuracy range 0°C … + 40°C
Operating temperature –10 °C … + 50°C
Storage temperature without battery – 25 °C … + 70°C
relative humidity 40 … 75 %
no condensation allowed
Elevation up to 2000 m maximum

Fuse
FF(ultrarapid) 10 A / 1000 V AC DC
10 mm x 38 mm
breaking capacity: 30 kA (Min 10 kA)

If you use other fuses than the one indicated above you forfeit your product guarantee.

Internal Fuse Test

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Internal Fuse
Test Fuse Replacement

Disconnect the instrument from the measuring circuit before opening the fuse compartment lid! Turn the (captive) slotted head screw counter-clockwise for this purpose. Remove the fuse with the flat end of the fuse compartment lid.
When refitting the fuse compartment lid the side with the guide hooks must be inserted first. Then turn the slotted head srew clockwise.

Repair and Replacement Parts Service
Calibration Lab and Rental Instrument Service
When you need service, please contact:

GMC-I Service GmbH
Service-Center
Beuthener Straße 41
90471 Nürnberg • Germany
Phone +49 911 817718-0
Fax +49 911 817718-253
E-Mail service@gossenmetrawatt.com
www.gmci-service.com

Product Support
When you need support, please contact:

Gossen Metrawatt GmbH
Product Support Hotline
Telefon D 0900 1 8602-00
A/CH +49 911 8602-0
Phone +49 911 8602-0
Fax +49 911 8602-709
E-Mail support@gossenmetrawatt.com

© Gossen Metrawatt GmbH
Edited in Germany • Subject to change without notice / errors excepted • A pdf version is available on the Internet
All trademarks, registered trademarks, logos, product names, and company names are the property of their respective owners.

GOSSEN METRAWATT logoGossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Telefon +49 911 8602-0
Fax +49 911 8602-669
E-Mail info@gossenmetrawatt.com
www.gossenmetrawatt.com