GOSSEN METRAWATT PROFITEST MTECH IQ 3 Phase Test Adapter for Testers Including Instruction Manual
- June 12, 2024
- GOSSEN METRAWATT
Table of Contents
Operating Instructions
PROFITEST REMOTE
3-Phase Test Adapter for Testers Including
PROFITEST MTECH+ IQ, MXTRA IQ and PRIME
3-447-051-15 4/6.22
Important
Read carefully before use!
Keep on file for future reference!
Connections Overview
PROFITEST REMOTE
Connections Overview – Symbols – Scope of Delivery
Meanings of Symbols on the Instrument
300 V CAT III Maximum permissible voltage and measuring category between
connections and ground
Warning concerning a point of danger (attention: observe documentation!)
** Indicates European Conformity
This device may not be disposed of with household trash, see “Returns and
Environmentally Sound Disposal” on page 15.
Scope of Delivery**
1 Test adapter (M514R)
1 RS 232 interface cable, type Z3241
1 Set of operating instructions
Safety Precautions
The test adapter has been manufactured and tested in accordance with the
following safety regulations: IEC/EN 61010-1/VDE 0411-1, DIN EN 61326-1/VDE
0843-20-1 Safety of the operator, as well as that of the test adapter, is only
assured when it’s used for its intended purpose. Use for intended purpose
stipulates application for loop and line impedance measurement ZL-PE and ZL-N
as well as insulation resistance measurement RISO exclusively in combination
with the
– PROFITEST MTECH+ IQ, as of FW V03.04.04
– PROFITEST MXTRA IQ, as of FW V03.04.04
– PROFITEST PRIME, as of FW V03.05.00 test instruments.
Any other application (e.g. voltage measurement or use in combination with
other test instruments) are not permissible. Read the operating instructions
carefully and completely before placing your test instrument into service.
Follow all instructions contained therein. Make sure that the operating
instructions are available to all users of the device. Read the operating
instructions for the respective test instrument as well, in particular the
sections concerning the measurement of insulation, loop and internal line
resistance. Tests may only be performed by a qualified electrician, or under
the supervision and direction of a qualified electrician. The user must be
instructed by a qualified electrician concerning performance and evaluation of
the test (see also training seminars listed at
www.gossenmetrawatt.com ).
OBSERVE THE FOLLOWING SAFETY PRECAUTIONS:
- The device may only be connected to elec- trical systems with a maximum of 230/ 400 V which comply with applicable safety regulations (e.g. IEC 60346, VDE 0100) and are protected with a fuse or circuit breaker with a maximum rating of 16 A.
- The test adapter may only be used for tests described in the section entitled “Applications”.
- When using a test probe with coil cord: Grip the tip of the test probe firmly, for example if it has been inserted into a jack socket. Tensioning at the coil cord may otherwise cause the test probe to snap back resulting in possible injury.
- Insulation resistance can only be measured at voltage-free systems.
- Do not touch the test probes during measurement!
FUSE REPLACEMENT
All fuses for the test adapter’s neutral and phase conductors are accessible
from the outside (see section 6.3). The fuses may only be replaced when the
instrument is voltage-free, i.e. the instrument must be disconnected from
mains supply power and may not be
connected to a measuring circuit. The fuse type must comply with the
specifications in the technical data or the labeling on the instrument (see
section 5).
Opening the Device / Repairs
The test adapter may only be opened by authorized, trained personnel in order
to ensure flawless operation and to assure that the guarantee is not rendered
null and void. Even original replacement parts may only be installed by
authorized, trained personnel. If it can be ascertained that the test adapter
has been opened by unauthorized personnel, no guarantee claims can be honored
by the manufacturer with regard to personal safety, measuring accuracy,
compliance with applicable safety measures or any onsequential
damages. The test adapter may not be used:
- If external damage is apparent, for example if parts which conduct dangerous touch voltage are freely accessible, and in the event of defective LEDs
- With damaged connection and/or measurement cables, e.g. interrupted insulation or kinked cable
- If it no longer functions flawlessly
- After extraordinary stressing due to transport In such cases, the test adapter must be removed from operation and secured against unintentional use.
Applications
Use of the test adapter eliminates bother- some replugging, and thus reversing
of the phases as well during measurement of RISO, ZL-PE and ZL-N.
The following measurements are supported:
L1-PE, L2-PE, L3-PE,
N-PE, L+N-PE,
L1-N, L2-N, L3-N,
L1-L2, L1-L3, L2-L3
The following test instruments are suitable for use with the test adapter:
– PROFITEST MTECH+ IQ, as of FW V03.04.04
– PROFITEST MXTRA IQ, as of FW V03.04.04
– PROFITEST PRIME, as of FW V03.05.00
Indication of Conductor Connections
The LEDs indicate conductor continuity between the connected system and the
test instrument’s input:
– RED LED: Switch closed, conductor continuity
– GREEN LED: Switch open, no connection
The line circuit-breaker is opened and closed via the TEST ADAPTER CONTROLLER
at the respective test instrument.
Measuring Insulation Resistance RISO in Electrical Systems In order to be able
to detect faults in electrical systems which are caused by defective
insulation, insulation resistance must be measured between each active
conductor and the protective earth conductor. Observe the corresponding
chapters in the operating instructions of your test instrument concerning
correct measurement procedures.
NOTE!
Thanks to 4-wire measurement, there’s no need to determine the test adapter’s
offset value.
Measuring Loop and Line Impedance ZL-PE and ZL-N Loop impedance is the sum of
all resistances within the distribution network and the conductors in the
branch circuit. In order to be able to determine the effectiveness of residual
current devices, loop impedance ZL-PE must be ascertained between the phase
conductor and the protective conductor. The same measuring procedure is used
for line impedance ZL-N as is also the case for loop impedance ZL-PE. However,
the current loop is completed via neutral conductor N, rather than protective
conductor PE as is the case with loop impedance measurement. Observe the
corresponding chapters in the operating instructions of your test instrument
concerning correct measurement procedures.
NOTE!
Thanks to 4-wire measurement, there’s no need to determine the test adapter’s
offset value.
Initial Startup
3.1 Testing the LEDs
Before performing any measurements, the LEDs should be tested for correct
functioning. The test adapter has to be connected to the test instrument via
the serial port to this end. Use the serial interface cable for this purpose
(included). After switching the test
instrument on, the LEDs must light up green in all switch positions except for
RISO, ZL-P and ZL-N.
3.2 Connecting the Test Instrument to the Test Adapter
Connecting the Measurement Inputs The respective test instrument is always
connected to the test adapter via 3 poles using the PRO-SCHUKO plug insert or
via safety sockets L, N and PE.
PROFITEST MTECH+ IQ: test plug
PROFITEST MXTRA IQ: test plug
PROFITEST PRIME: probes 1(L), 2(N), 3(PE)
This also applies in the case of 2-pole measurements, because the
corresponding probes are controlled from the test instrument via the test
adapter controller – the measurement cables or probes no longer have to be
manually replugged.
Data Interface Connection
The test adapter must also be connected to the test instrument via the serial
port in order to control the line circuit-breaker, for the purpose of
indication by means of the conductor LEDs and for power supply. Only use the
serial interface cable included in the scope of delivery to connect the test
adapter with the test instrument.
3.3 Connection to the System or the Mains
The test adapter must be connected to the system for all tests (see
connections overview on page 2). The following adapter cables are available
for this purpose:
Description | Type | Article No. |
---|
Adapter cable with 5-pole 16 A CEE plug and 4 mm safety plugs (L1, L2, L3, N,
PE), CAT III 300 V, cable length: 4.8 m| Connecting-Cable-16| Z570B
Adapter cable with 5-pole 32 A CEE plug and 4 mm safety plugs (L1, L2, L3, N,
PE), CAT III 300 V, cable length: 4.8 m| Connecting-Cable-32| Z570C
Adapter cable with 5-pole 63 A CEE plug and 4 mm safety plugs (L1, L2, L3, N,
PE), CAT III 300 V, cable length: 4.8 m| Connecting-Cable-63| Z570D
➭ Disconnect the system from all sources of voltage and secure against
restart.
➭ Check the measuring points for absence of voltage before connecting them to
the test adapter.
➭ Using the appropriate adapter cable, connect the test adapter to the system
via connector sockets L1, L2, L3, N and PE: Refer to the characteristic values
on page 14 for nominal mains values.
3.4 Key and Significance of the LEDs
Significance of the LEDs
LED lights up red: Switch closed
LED lights up green: Switch open
LED Test
The LEDs must light up green in all switch positions except for RISO, ZL-P and
ZL-N.
Repeat the LED test after each test run.
Measurements
The test adapter can remain connected to the system or the mains during the
course of all measurements, assuming absence of voltage during the insulation
resistance measurement.
NOTE!
The fuses have to be checked before each measurement. Blown fuses must be
replaced with new ones (see “Fuse Replacement” on page 15).
4.1 Measuring Insulation Resistance (RISO)
ATTENTION!
Single-Phase Mains Connection
The phase must be connected to L1 for single-phase mains connection.
➭ Make sure that the system remains voltage-free for the entire duration of
the measurement.
ATTENTION!
Insulation resistance can only be measured at voltage-free systems.
ATTENTION!
The N and PE conductors must be disconnected at all times. Consuming devices
should not be connected.
4.1.1 Measuring Insulation Resistance with the PROFITEST MTECH+ IQ or
PROFITEST MXTRA IQ
Preparing for Measurement
➭ Connect the test instrument’s test plug to the PRO-SCHUKO plug insert.
➭ Disconnect the system from all sources of voltage before connecting the
measuring points to the testadapter.
➭ Connect the system to sockets L1, L2, L3, N and PE.
Performing Measurement
ATTENTION!
Insulation test voltage may not exceed 500 V.
➭ Perform the measurement as described in the operating instructions for the
respective test instrument:
PROFITEST MTECH+ IQ: sections 5.8 and 11 PROFITEST MXTRA IQ: sections 5.8 and
11
Overview of Connected Conductors
The red illuminated LEDs in the connections overview indicate the respectively
connected (closed) conductors.
Continuity, PROFITEST MTECH+ IQ/ PROFITEST MXTRA IQ RISO
Test Sequence | RINS | LED Displays (red) |
---|---|---|
01/11 | N-PE | LED4 + LED11 |
02/11 | L1 -PE | LED1 + LED11 |
03/11 | L2-PE | LED2 + LED11 |
04/11 | L3-PE | LED3 + LED11 |
05/11 | L+N -PE | LED1 + LED2 + LED3 + LED4 + LED11 |
06/11 | L1-N | LED1 + LED10 |
07/11 | L2-N | LED2 + LED10 |
08/11 | 13-N | LED3 + LED10 |
09/11 | L1-L2 | LED1 + LED8 |
10/11 | L2 -L3 | LED2 + LED9 |
11/11 | L1-L3 | LED1 + LED9 |
4.1.2 Measuring Insulation Resistance with the PROFITEST PRIME
Preparing for Measurement
➭ Using the corresponding test probes, connect the test instrument to the L, N
and PE sockets at the test adapter.
➭ Disconnect the system from all sources of voltage before connecting the
measuring points to the test adapter.
➭ Connect the system to sockets L1, L2, L3, N and PE.
Performing the Measurement
ATTENTION!
Insulation test voltage may not exceed 500 V.
➭ Perform the measurement as described in the operating instructions for the
test instrument:
PROFITEST PRIME: sections 8.6 and 11
➭ The following measuring functions are available for the measurement of
insulation resistance: insulation measurement with constantvoltage and
insula- tion measurement with rising test voltage .
Overview of Connected Conductors
The red illuminated LEDs in the connections overview indicate the respectively
connected conductors.
Continuity, PRIME RISO
Test Sequence | RINS | LED Displays (red) |
---|---|---|
01/11 | N-PE | LED4 + LED11 |
02/11 | L1-PE | LED1 + LED11 |
03/11 | L2-PE | LED2 + LED11 |
04/11 | L3-PE | LED3 + LED11 |
05/11 | L+N-PE | LED1 + LED2 + LED3 + LED4 + LED11 |
06/11 | L1-N | LED1 + LED10 |
01/11 | L2- N | LE02 + LEDi 0 |
08/11 | L3-N | LED3 + LED10 |
09/11 | L1-L2 | LED1 + LED8 |
10/11 | L2-L3 | LED2 + LED9 |
11/11 | L1-L3 | LED1 + LED9 |
4.2 Measuring Loop Impedance (ZL-PE)
ATTENTION!
Single-Phase Mains Connection The phase must be connected to L1 for single-
phase mains connection.
4.2.1 Measuring Loop Resistance with the PROFITEST MTECH+ IQ or PROFITEST
MXTRA IQ
Preparing for Measurement
➭ Connect the test instrument’s test plug to the PRO-SCHUKO plug insert.
➭ Disconnect the system from all sources of voltage before connecting the
measuring points to the test adapter.
➭ Connect the system to sockets L1, L2, L3, N and PE.
➭ Switch the system on.
Performing the Measurement
ATTENTION!
Insulation test voltage may not exceed 500 V.
➭ Perform the measurement as described in the operating instructions for the
respective test instrument: PROFITEST MTECH+ IQ: sections 5.8 and 9 PROFITEST
MXTRA IQ: sections 5.8 and 9
NOTE!
Measurement of loop impedance ZL- PE with the PROFITEST MTECH+ IQ and/ or
PROFITEST MXTRA IQ is conducted with sine (full-wave) and 15 mA sine.
Measurements with DC and halvewave are not possible.
Overview of Connected Conductors
The red illuminated LEDs in the connections overview indicate the respectively
connected conductors.
Continuity, ZL-PE
Test Sequence | ZL-PE | LED Display (red) |
---|---|---|
3-Jan | L1-PE | LED1 + LED11 |
3-Feb | L2-PE | LED2 + LED11 |
3-Mar | L3-PE | LED3 + LED11 |
4.2.2 Measuring Loop Impedance with the PROFITEST PRIME
Preparing for Measurement
➭ Using the corresponding test probes, connect the test instrument to the L, N
and PE sockets at the test adapter.
➭ Disconnect the system from all sources of voltage before connecting the
measuring points to the test adapter.
➭ Connect the system to sockets L1, L2, L3, N and PE.
➭ Switch the system on.
Performing the Measuremen
ATTENTION!
Insulation test voltage may not exceed 500 V.
NOTE!
If many loop measurements are performed with the PROFITEST PRIME in quick
succession, fine-wire fuses in the PROFITEST REMOTE may blow. This is due to
the increased test current in order to achieve a higher measuring accuracy.
➭ Perform the measurement as described in the operating instructions for the
test instrument:
PROFITEST PRIME: sections 8.6 and 13
➭ Loop impedance is measured using the ZLOOP AC/DC measurement variant.
NOTE!
Measurement with the following mea surement variants is not possible with the
PROFITEST PRIME:
– ZLOOP DC+
– ZLOOP
– ZLOOP
Overview of Connected Conductors
The red illuminated LEDs in the connections overview indicate the respectively
connected conductors.
Continuity, ZLOOP AC/DC
Test Sequence | Zloop AC/DC | LED Display (red) |
---|---|---|
9-Jan | L1-PE | LED1 + LED11 |
9-Feb | L2-PE | LED2 + LED11 |
9-Mar | L3-PE | LED3 + LED11 |
9-Apr | L1-N | LED1 + LED10 |
9-May | L2-N | LED2 + LED10 |
9-Jun | L3-N | LED3 + LED10 |
9-Jul | L1-L2 | LED1 + LED8 |
9-Aug | L2-L3 | LED2 + LED9 |
9-Sep | L1-L3 | LED1 + LED9 |
4.3 Measuring Line Impedance (ZL-N)
ATTENTION!
Single-Phase Mains Connection The phase must be connected to L1 for single-
phase mains connection.
4.3.1 Measuring Line Impedance with the PROFITEST MTECH+ IQ and PROFITEST
MXTRA IQ
Preparing for Measurement
➭ Connect the test instrument’s test plug to the PRO-SCHUKO plug insert.
➭ Disconnect the system from all sources of voltage before connecting the
measuring points to the test adapter.
➭ Connect the system to sockets L1, L2, L3, N and PE.
➭ Switch the system on.
Performing the Measurement
ATTENTION!
Insulation test voltage may not exceed 500 V.
➭ Perform the measurement as described in the operating instructions for the
respective test instrument:
PROFITEST MTECH+ IQ: sections 5.8 and 9 PROFITEST MXTRA IQ: sections 5.8 and 9
Overview of Connected Conductors
The red illuminated LEDs in the connections overview indicate the respectively
connected conductors. Continuity, ZL-N
Test Sequence | RINS | LED Displays (red) |
---|---|---|
6-Jan | L1-N | LED1 + LED7 |
6-Feb | L2-N | LED2 + LED7 |
6-Mar | L3-N | LED3 + LED7 |
6-Apr | L1-L2 | LED1 + LED5 |
6-May | L2-L3 | LED2 + LED6 |
6-Jun | L1-L3 | LED1 + LED6 |
4.3.2 Measuring Line Impedance with the PROFITEST PRIME
Preparing for Measurement
➭ Using the corresponding test probes, connect the test instrument to the L, N
and PE sockets at the test adapter.
➭ Disconnect the system from all sources of voltage before connecting the
measuring points to the test adapter.
➭ Connect the system to sockets L1, L2, L3, N and PE. Performing the
Measurement
ATTENTION!
Insulation test voltage may not exceed 500 V.
NOTE!
If many loop measurements are performed with the PROFITEST PRIME in quick
succession, fine-wire fuses in the PROFITEST REMOTE may blow. This is due to
the increased test current in order to achieve a higher measuring accuracy.
➭ Perform the measurement as described in the operating instructions for the
test instrument: PROFITEST PRIME: sections 8.6 and 13
➭ Line impedance is measured using the ZLOOP AC/DC measurement variant.
➭ Switch the system on.
Overview of Connected Conductors
The red illuminated LEDs in the connections overview indicate the respectively
connected conductors.
Continuity, ZLOOP AC/DC
Test Sequence | RINS | LED Displays (red) |
---|---|---|
6-Jan | L1-N | LED1 + LED7 |
6-Feb | L2-N | LED2 + LED7 |
6-Mar | L3-N | LED3 + LED7 |
6-Apr | L1-L2 | LED1 + LED5 |
6-May | L2-L3 | LED2 + LED6 |
6-Jun | L1-L3 | LED1 + LED6 |
4.4 Error Messages at the Test Instrument
LCD Pictogram| Test Plug| Measuring Adapter| Position of
the
Function Switch| Function / Meaning
---|---|---|---|---
| X| X| R1so| Check mains connections for absence of voltage. If voltage is
present, it results in short-circuit- ing between L and N and the device’s
fuses are blown. Remedy: Assure absence of volt- age before switching through!
| X| X| All| No interface connection between the test instrument and the test
adapter.
Remedy: Check the interface connection and establish con- nection if
necessary. Use the serial interface cable for this pur- pose (included).
| X| X| R1so| Contact error or defective fuse. Remedy: Check the test plug for
proper fit with the test adapter or replace fuse in the test instrument or
test adapter.
Characteristic Values
Measurement input| PRO-SCHUKO plug insert for PROFITEST MTECH+ IQ and
PROFITEST MXTRA IQ, three 4 mm safety sockets PROFITEST PRIME| Electromagnetic
Compatibility (EMC)Interference emission EN 61326-1:2013,
class B Interference immunity EN 61326-1:2013 EN 61326-2-1:2013
---|---|---
Mains connection| Five 4 mm safety sockets, (adapter cable for CEE 16A, CEE
32Aand CEE 63A available as accessory)| Mechanical Design ProtectionHousing:
IP 40 per DIN VDE 0470, part 1 / EN 60 529
Table Excerpt Regarding Significance of IP Codes
Ambient Conditions
IP XY st (1 digit X)| Protection Against Foreign Object Ingress|
IP XY nd (2 digit Y)| Protection Against Water Ingress
---|---|---|---
4| ³ 1.0 mm dia.| 0| Not protected
Operating temperaturStorage temperature Relative humidity Elevation
Operating temperature
Storage temperature
Relative humidity Elevation| 5 … + 45 °C -20 … + 60 °C Max. 75%, no
condensation allowed
Max. 2000 m| Dimensions Housing (W × H × D): approx. 27.5 × 10.5 ×
12 cm (with PRO-SCHUKO plug insert)
---|---|---
Power Supply
Auxiliary voltage| 5 V via RS 232 port| Weight| Approx. 1.1 kg (test adapter
without interface cable)
---|---|---|---
Current consumption| 60 mA (all relays dropped out) 240 mA (2 relays picked
up) 600 mA (5 relays
picked up)| Data Interface Transmission
parameters| 9600 baud, 1 stop bit, no parity
Electrical Safety
Measuring category | 300 V CAT III |
---|---|
Pollution degree | 2 |
Protection class | II per IEC 61010-1/ EN 61010-1/ VDE 0411-1 |
Fuse links | F1 to F8: 8 × F 3.15 A FF/500 V |
Test voltage | Insulation test voltage may not exceed 500 V. |
Operating conditions | Continuous operation |
Maintenance
6.1 Housing Maintenance
No special maintenance is required. Keep outside surfaces clean and dry. Use a
slightly dampened cloth for cleaning. Avoid the use of solvents, cleansers and
abrasives.
6.2 Technical Safety Inspections,
Testing per DGUV Regulation 3 Subject your test adapter to technical safety
inspections at regular intervals.
The test adapter is designed in accordance with IEC 61010 and VDE 0413-16 as a
protection category I and II test instrument. Testing of protective conductor
and insulation resistance is described in the following subsections.
6.2.1 Testing Protective Conductor Resistance
Testing the 3-Phase Connection
Protected conduct or resistance values of RPE <2 are permissible. This is due
to the design of the test adapter.
Connecting/Contacting the Test Adapter, 3-Phase (400 V) Protective conductor
resistance is tested between the PE safety contact at the mains connection
sockets and the parallel connected 4 mm safety socket at the PE measurement
input.
6.2.2 Testing Insulation Resistance
Testing is conducted respectively between short-circuited contacts L123-N and
PE for the mains connection safety sockets and between L-N and PE for the
measurement inputs. The usual limit values apply.
6.3 Fuse Replacement
All eight fuses for neutral (F7/F8) and phase conductors (F1 to F6) are
accessible from the outside.
The fuses may only be replaced when the device is voltage-free, i.e. the
device must be disconnected from mains supply power and may not be connected
to a measuring circuit. The respective fuse type must comply with the
specifications in the technical data or the labeling on the device.
➭ Unscrew the appropriate fuse holder with the help of a suitable screwdriver.
➭ Remove the blown fuse from the holder.
➭ Insert the new fuse into the holder.
➭ Reinsert the fuse holder and screw it tight.
6.4 Returns and Environmentally Sound Disposal
This instrument is subject to directive 2012/ 19/EC on Waste Electrical and
Electronic Equipment (WEEE) and its German national equivalent implemented as
the Waste Electrical and Electronic Equipment Act (ElektroG) on the marketing,
return and environmentally sound disposal of electrical and electronic
equipment. The device is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German Waste Electrical and
Electronic Equipment Act).
The symbol at the left indicates that this device and its electronic
accessories must be disposed of in accordance with applicable legal
regulations, and not together with household trash. In order to dispose of the
instrument, bring it to a designated collection point or contact our product
support department. Segregated disposal and recycling conserves resources and
protects our health and the environment. Current and further information is
available on our website at http://www.gossenmetrawatt.com under the
search terms “WEEE” and “environmental protection”.
Contact, Support and Service
You can reach Gossen Metrawatt GmbH directly and uncomplicated, we have one
number for everything! Whether it’s a support question, or individual desire,
we answer every request
at:: +49 911 8602-0
Monday –8:00 a.m. – 4:00 p.m.
Thursday:8:00 a.m. – 2:00 p.m.
Friday: You can also e-mail to:
info@gossenmetrawatt.com
Do you prefer support via e-mail? Measuring and test instruments:
support@gossenmetrawatt.com
Industrial equipment:
support.
industrie@gossenmetrawatt.com
Enquiries concerning English seminars can also be submitted by e-mail:
training@gossenmetrawatt.com
For repairs, replacement parts, and calibrations* please contact GMC-I
Service GmbH: +49 911 817718-0
Beuthener Straße 41
90471 Nürnberg
Germany
service@gossenmetrawatt.com
www.gmci-service.com
https://www.gmci-service.com/en/
* DAkkS calibration laboratory per DIN EN ISO/IEC 17025. Accredited at the Deutsche Akkreditierungsstelle GmbH under registration no. D-K-15080-01-01.
© Gossen Metrawatt GmbH
Edited in Germany • Subject to change without notice / Errors excepted • A pdf
version is available on the Internet All trademarks, registered trademarks,
logos, product names, and company names are property of their respective
owners.
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Phone +49 911 8602-0
Fax +49 911 8602-669
E-Mailinfo@gossenmetrawatt.com
www.gossenmetrawatt.com
Documents / Resources
|
GOSSEN METRAWATT PROFITEST MTECH IQ 3 Phase Test Adapter for Testers
Including
[pdf] Instruction Manual
PROFITEST MTECH IQ, MXTRA IQ, PRIME, PROFITEST MTECH IQ 3 Phase Test Adapter
for Testers Including, 3 Phase Test Adapter for Testers Including, Test
Adapter for Testers Including, Adapter for Testers Including, Testers
Including, Including
---|---
References
- trawatt.com
- Servicepartner für Kalibrierungen, Reparaturen, Ersatzteile, Mietgeräte | GOSSEN METRAWATT GMC-I Service GmbH
- GOSSEN METRAWATT, GMC-Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>