KLEIN TOOLS ET140 Pinless Moisture Meter Instruction Manual

June 4, 2024
KLEIN TOOLS

INSTRUCTION MANUAL
Pinless Moisture Meter
• PINLESS, NON-DESTRUCTIVE
MOISTURE METER
MOISTURE METER

ET140

**GENERAL SPECIFICATIONS

**

Klein Tools ET140 is designed to non‑invasively detect moisture based on a material’s capacitance. The mode selection is setup to maximize moisture readings for up to 3/4″ of the depth of the surface of the drywall, cement board, hardwood, and softwood.

  • Moisture Detection Range:
    Hardwood: 0 to 35%
    Softwood: 0 to 55%
    Drywall & Masonry: 0 to 100% Relative

  • Resolution: 0.1%

  • Accuracy: Wood: ± 4%
    Drywall & Masonry: Relative readings

  • Inductive Sensor Size: 1.25″ x 2″ (32 mm x 51 mm)

  • Sensing Depth: Up to 3/4″ (19 mm)

  • Operating Altitude: 6562 ft. (2000 m)

  • Relative Humidity: <90% non‑condensing

  • Operating Temp: 50° to 104°F (10° to 40°C)

  • Storage Temp: ‑4° to 122°F (‑20° to 50°C)
    • Dimensions: 6″ x 2.5″ x 1″ (152 x 64 x 25 mm)

  • Weight: 7 oz. (200 g) including batteries

  • Certified cations: CE, UKCA

  • Batteries: 1 x 9V (included)

  • Drop Protection: 6.6 ft. (2m)

Specifications are subject to change.

WARNINGS
To ensure safe operation and service of the meter, follow these instructions.

  • Read the instructions to ensure safe operation.
  • The meter is NOT insulated. Avoid contact with areas where energized conductive elements may be present.
  • Always wear approved eye protection.

SYMBOLS ON METER

| Warning or Caution
---|---
| Wear approved eye protection
| WEEE – Dispose of properly
| Conformité Européenne: Conforms with European Economic Area directives
| UKCA: UK Conformity Assessment
| Risk of electrical shock
| Read instructions
| Battery Polarity

FEATURE DETAILS

METER FEATURES LCD FEATURES

1. LCD Display
2. Bar Graph
3. MODE Button
4. Power On/Off Button
5. HOLD Button
6. Sensor
7. Battery Cover| A. Moisture Measurement (%)
B. Modes
C. Low Battery Indicator
D. HOLD Indicator
E. Auto‑Power Off Icon
NOTE: There are no user-serviceable parts inside.

FUNCTION BUTTONS

POWER ON/OFF BUTTON 4
Press the Power On/Off Button to turn the meter on or off. Press and hold the Power On/Off Button for 2 seconds to turn the meter off.
NOTE: To conserve battery life, the meter will automatically power off after 10 minutes of inactivity.

MODE BUTTON 3
Repeated presses of the MODE button will toggle through the available material types: Drywall, Masonry, Hardwood, and Softwood, in that order. Select the desired material prior to testing.
HOLD BUTTON 5
Press the HOLD button to lock the current reading. The reading will remain on the display until the HOLD button is pressed again, or until the meter automatically powers off.

OPERATING INSTRUCTIONS

  1. Power on the meter 4 .
  2. Press the MODE button 3 until the desired material type appears on the display.
  3. Scan the material by pressing sensor 6 against the material to be tested. To ensure the best accuracy, press the back of the meter against a flat area of the material. The sensor measures up to 3/4″ (19 mm) thick. The display will show the percentage of moisture numerically A, and segments of the Bar Graph 2 will illuminate as a visual reference of the level of moisture present in the material per the chart below.

NOTE: “OL” will appear on the display if the moisture content detected lies outside the meter’s measurement range.
NOTE: “ER” will appear on the display if the material type is not supported by the meter.

BAR GRAPH MOISTURE INDICATORS

MATERIAL| GREEN LEDs| YELLOW LEDs| RED LEDs
DRYWALL| 0 to 40%| 40 to 70%| 70% +
MASONRY| 0 to 36%| 36 to 63%| 63% +
HARDWOOD| 0 to 14%| 14 to 24.5%| 24.5% +
SOFTWOOD| 0 to 21%| 21 to 39%| 39% +

MAINTENANCE

BATTERY REPLACEMENT
When the Low Battery indicator is displayed on the LCD, the batteries must be replaced to avoid inaccurate measurements. the batteries must be replaced to avoid inaccurate measurements.

  1. Loosen the screw to remove the battery door 7 .
  2. Remove and recycle the exhausted 9V battery.
  3. Install the new battery.
  4. Replace the battery door and tighten the screw.

**** To avoid the risk of electric shock, do not operate the meter while the battery door is removed.

CLEANING

Be sure the meter is turned off and wipe the sensor with a clean, dry lint‑free cloth. Do not use abrasive cleaners or solvents.

STORAGE
Remove the batteries when the meter is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section, allow the meter to return to normal operating conditions before use.

WARRANTY

www.kleintools.com/warranty
DISPOSAL / RECYCLE

Do not place equipment and its accessories in the trash. Items must be properly disposed of in accordance with local regulations. Please see www.epa.gov/recycle for additional information.

CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street Lincolnshire, IL 60069 1‑800‑553‑4676
[email protected] www.kleintools.com