Operating Instructions
MAVOWATT 210 ROGOWSKI
Flexible AC Current Probe
based on the Rogowski Principle
3-447-115-03
1/4.22
MAVOWATT 210 Rogowski
Gossen Metrawatt GmbH
Safety Precautions
General
-
Carefully and completely read and adhere to these operating instructions.
The document can be found at http://www.gossenmetrawatt.com. Retain this document for future reference. -
Carefully and completely read and adhere to the product documentation of the combined products. Retain these documents for future reference.
-
Observe and comply with all safety regulations which are applicable for your work environment.
Handling
-
The probe may only be used as long as it is fully intact.
Inspect all cables and the probe before use. Pay particular attention to damage, broken insulation or kinked cables. -
If the probe doesn’t function flawlessly, remove it from operation and secure it against inadvertent use.
-
If the probe is damaged during use, e.g. through falling, remove it from operation and secure it against inadvertent use.
-
If the equipment is used in a manner not specified in this reference guide, the protection provided by the equipment may be impaired. These safety precautions are repeated where appropriate throughout this manual.
Operating conditions
- Only use the probe in conjunction with the measuring device MAVOWATT 210.
- Only use the probe in compliance with the specified technical data and under the specified conditions (ambient conditions, IP protection class, measuring category, etc.).
- Do not use the probe after long periods of storage under unfavorable conditions (e.g. humidity, dust or extreme temperature).
Electricity
- Installation, operation, and maintenance of this product must be performed by qualified personnel only.
- Qualified personnel who work on or near exposed energized electrical conductors must follow applicablesafety related work practices and procedures including appropriate personal protective equipment.
- Wear proper personal protective equipment, including safety glasses and insulated gloves when making connections to power circuits.
- Hands, shoes, and floor must be dry when making any connection to a power line.
- Never apply the probe around bare conductors with hazardous voltages without having the appropriate permission to perform such work and without wearing protective clothing and gloves as required.
- Connections must be made to the measuring device first, then to the circuit to be monitored.
- The probe may only be used within the limits of the specifications listed in this manual. Loads exceeding the specified values may destroy the probe and the electrical equipment connected to it.
Applications
Please read this important information!
2.1 Intended Use / Use for Intended Purpose
The MAVOWATT 210 Rogowski current probes are based upon the Rogowski
principle. They can be used exclusively to measure AC current by being
connected to MAVOWATT 210 Three-phase Power and Energy Logger.
The flexible probes allow current measurements on conductors that are hard
toreach.
Installation, operation, and maintenance of this product must be performed by
qualified personnel only. Safety of the operator, as well as that of the
product, is only assured when it’s used for its intended purpose.
2.2 Use for Other than Intended Purpose
Using the product for any purposes other than those described in the con-
densed operating instructions or these product operating instructions is
contrary to use for intended purpose.
2.3 Liability and Guarantee
Gossen Metrawatt GmbH assumes no liability for property damage, personal
injury or consequential damage resulting from improper or incorrect use of the
product, in particular due to failure to observe the product documentation.
Furthermore, all guarantee claims are rendered null and void in such cases.
Nor does Gossen Metrawatt GmbH assume any liability for data loss.
2.4 Opening the Instrument / Repairs
Unauthorized modifications to the product are prohibited.
The Product
3.1 Scope of Application
This documentation applies to the following articles:
Product | Product no. | Description |
---|
MAVOWATT 210
Rogowski 500/40| Z840A| Flexible AC current probe based on the Rogowski
principle, suitable for
MAVOWATT 210, 50 − 500 A, measuring head length 40 cm.
MAVOWATT 210
Rogowski 1500/40| Z840B| Flexible AC current probe based on the Rogowski
principle, suitable for
MAVOWATT 210, 150 − 1500 A, measuring head length 40 cm.
MAVOWATT 210
Rogowski 3000/40| Z840C| Flexible AC current probe based on the Rogowski
principle, suitable for
MAVOWATT 210, 300 − 3000 A, measuring head length 40 cm.
3.2 Product Overview
Figure 1: Flexible AC current probes based on the Rogowski principle
3.2.1 Symbols on the product and the included accessories:
| Warning concerning a point of danger (attention, observe
documentation!)
---|---
| Double insulation (protection category II)
| Do not apply around or remove from hazardous live conductors without
additional protective means.
| European conformity marking
| The device may not be disposed of with household trash a “Returns and
Environmentally Sound Disposal” ****9.
3.3 Relevant Standards
The product has been manufactured and tested in accordance with the following
safety regulations:
DIN EN 61326| Electrical Equipment for Measurement, Control and Laboratory Use
− EMC Requirements
---|---
DIN EN 61010-1| Safety Requirements of Electrical Equipment for Measurement,
Control, and Laboratory Use
− Part 1: General Requirements
DIN EN 61010-1-031| Safety Requirements of Electrical Equipment for
Measurement, Control, and Laboratory Use − Part 031: Safety Requirements for
Hand-held Probe Assemblies for Electrical Measurement and Test
DIN EN 61010-2-032| Safety Requirements of Electrical Equipment for
Measurement, Control, and Laboratory Use − Part 2-032: Particular Requirements
for Hand-held and Hand-manipulated Current Sensors for Electrical Test and
Measurement
3.4 Technical Data
Rated Current | Measurement range | MAVOWATT 210 Rogowski 500/40 | 50 A — 500 A |
---|---|---|---|
MAVOWATT 210 Rogowski 1500/40 | 150 A — 1500 A | ||
MAVOWATT 210 Hogowski 3000/40 | 300 A — 3000 A | ||
Accuracy | 0.5% per IEEE C57.13 class 0.6 | ||
Positioning error | ±0.1 % | ||
Ambient Conditions | Operating temperature | —20 … +70 °C | |
Relative humidity | 5% to 95%, |
no condensation allowed
Electrical| Phase orientation| Arrow points towards load
Frequency range| 45 Hz — 65 Hz
Frequency response| 10 Hz — 20 kHz
Safety| Working voltage| 1000 V CAT III, 600 V CAT IV
Dielectric strength| 7400 Vac at 50/60 Hz for 1 minute
Mechanical Design| Protection| IP61 per DIN EN 60529 / IEC 60529
Coil length| 40 cm
Coil diameter| 15.5 mm
Window size| 10.6 cm
Wire length| 2 m
Protection type| IP61
Installation
4.1 Unpacking the Product
- Carefully remove product and accessories from the packaging.
- Check delivery for completeness and possible damage.
- In case of detected damages, hidden defects and short deliveries, document type and scope and contact the manufacturer or supplier immediately.
- Keep packing material for further transport.
4.2 Making Measurement Connections
To perform measurements, the Rogowski current probes are first connected to
MAVOWATT 210. Subsequently, the Rogowski current probes are connected to the
object to be measured.
Attention!
To avoid the risk of electric shock or burns, always connect the earth ground
before making any other connections.
Attention!
Do not exceed the marked maximum ratings.
4.2.1 Connecting MAVOWATT 210 Rogowski Current Probes to MAVOWATT 210
√ MAVOWATT 210 is powered off.
- Plug the output cable of the current probe to the current input of MAVOWATT 210.
4.2.2 Connecting MAVOWATT 210 Rogowski Current Probes to the Circuit to be
Measured
Attention!
DO NOT attempt to measure current in any circuit in which the circuit to
ground voltage exceeds the insulation rating of the current probe.
Attention!
Make sure the current probes are tightly closed. Keep mating surfaces clean
and free from foreign matter.
Note!
Do not locate the probe coupling close to other conductors carrying high
current.
Note!
Be sure to connect the current probes with the arrow pointing towards the load
or an erroneous power reading will result.
√ MAVOWATT 210 is powered off.
- Open the probe coupling and slip the measuring head over the conductor carrying the current to be measured.
- Close the probe coupling such that it visibly and audibly snaps into place.
The arrow marking on the current probe points in the direction of current flow
and towards the load.
The conductor is centered within the measuring head.
The measuring head forms a perfect circle.
Operation
For the different measurement configurations, refer to the Operating Instructions of MAVOWATT 210.
Maintenance
Cleaning
Keep the probe free of surface contamination. If necessary, clean the probe
with isopropyl alcohol.
Contact, Support and Service
You can reach Gossen Metrawatt GmbH directly and uncomplicated, we have one
number for everything!
Whether it’s a support question, or individual desire, we answer every request
at:
+49-911-8602-0
Monday–Thursday:8:00 a.m. – 4:00 p.m.
Friday: 8:00 a.m. – 2:00 p.m.
You can also e-mail to:
info@gossenmetrawatt.com
Do you prefer support via e-mail?
Measuring and test instruments:
support@gossenmetrawatt.com
Industrial equipment:
support.industrie@gossenmetrawatt.com
For repairs, replacement parts, and calibrations 1please contact GMC-I Ser- vice GmbH:
service@gossenmetrawatt.com
www.gmci-service.com
+49-911-817718-0
Beuthener Str. 41
90471 Nürnberg
Germany
- DAkkS calibration laboratory per DIN EN ISO/IEC 17025.
Accredited at the Deutsche Akkreditierungsstelle GmbH under registration no. D-K15080-01-01.
CE Declaration
The product fulfills all requirements of applicable EU directives and national regulations. We confirm this with the CE mark. The CE Declaration of Conformity is available upon request.
Returns and Environmentally Sound Disposal
This product is subject to directive 2012/19/EC on Waste Electrical and
Electronic Equipment (WEEE) and its German national equivalent implemented as
the Waste Electrical and Electronic Equipment Act (ElektroG) on the marketing,
return and environmentally sound disposal of electrical and electronic
equipment. The product is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German Waste Electrical and
Electronic Equipment Act).
The symbol at the left indicates that this device and its electronic
accessories must be disposed of in accordance with applicable legal
regulations, and not together with household waste. In order to dispose of the
device, bring it to a designated collection point or contact our product
support department ( ). Separate disposal and recycling conserves resources
and protects our health and the environment.
Current and further information is available on our website at
http://www.gossenmetrawatt.com under the search terms “WEEE” and
“environmental pro- tection”.
© Gossen Metrawatt GmbH
Prepared in Germany • Subject to change, errors excepted • PDF version
available on the Internet
All trademarks, registered trademarks, logos, product names, and company names
are the property of their respective owners.
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Phone +49 911 8602-0
Fax +49 911 8602-669
E-mailinfo@gossenmetrawatt.com
www.gossenmetrawatt.com
Documents / Resources
|
GOSSEN METRAWATT 210 Rogowski Flexible AC Current
Probe
[pdf] Instruction Manual
210 Rogowski Flexible AC Current Probe, 210 Rogowski, Flexible AC Current
Probe, AC Current Probe, Current Probe, Probe
---|---
References
- Servicepartner für Kalibrierungen, Reparaturen, Ersatzteile, Mietgeräte | GOSSEN METRAWATT GMC-I Service GmbH
- GOSSEN METRAWATT, GMC-Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>