TOOLOTS ES Series Electronic LCD Balance or Scale Instruction Manual
- October 30, 2023
- TOOLOTS
Table of Contents
- TOOLOTS ES Series Electronic LCD Balance or Scale
- Introduction
- Description
- Features
- Unpacking
- Installation
- Operation
- Using Menus To Configure The Balance
- Setup Menu
- Calibration LockOut Protection
- Care And Maintenance
- Troubleshooting
- Error Codes
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
TOOLOTS ES Series Electronic LCD Balance or Scale
ES SERIES ELECTRONIC LCD BALANCE
INSTRUCTION MANUAL
Introduction
This manual covers the installation, operation, and troubleshooting for the
Portable Advanced Electronic Balances,models YS302, YS602, YS1101, YS2101,
ES-40, ES-50, ES-200(A),ES-300(A), ES-600(A), ES-1100, ES-2100(A), and ES-
3000A, ES-5000A.
To ensure proper operation of the balance, please read this manual completely.
Description
The ES Series Electronic LCD Balance is a portable advanced electronic balance that helps in weighing and counting parts. It has a compact design and a digital display that shows the weight in different units of measurement.
The Portable Advanced balances are precision weighing instruments,designed to provide years of service with virtually no maintenance .The Portable Advanced housing is constructed of a material which is impact resistant . contains a one piece solid-state precision electronics PC board.and a five digit LCD display which is 0.5 inches in height.All Portable Advanced balances are initially factory set to measure in grams and can be set to measure in other units of measure as desired. Each balance operates through a series of menus which enable precise calibration and the selection of various parameters which enhance operation. A built in lock switch prevents calibration settings from being changed. Power is supplied through an AC adapter or batteries.
Portable Advanced balances contain two display menus which enable you to calibrate and configure the balance for specific operating requirements.
- CALIBRATION Menu– Allows the balance to be calibrated by using standard weights.
- SETUP Menu– allows the balance to be set for environmental conditions and customized for specific weighing functions.
When the balance is first turned on, it can be used to weigh tare items without setting the menus. Refer to the individual menus in this manual for detailed descriptions and procedures for programming.
Features
- Compact design
- Digital display
- Weighing and counting parts
- Multiple units of measurement
- Calibration menu protection
- Error codes for troubleshooting
Unpacking
When unpacking the ES Series Electronic LCD Balance, ensure that all items are present and in good condition. The package should contain the following:
All Portable Advanced balances are supplied with a weighing platform, some of which have been installed, AC power adapter,instruction manual.
- ES Series Electronic LCD Balance
- AC adapter
- Batteries (if applicable)
- User manual
Installation
Environment
The ES Series Electronic LCD Balance should be used in a clean and dry
environment to ensure accurate measurements.
Balance performance can be affected by excessive air currents, corrosives, vibration, temperature or humidity extremes.
DO NOT use the balance:
- Next to open windows or doors causing drafts or rapid temperature changes.
- Near air conditioning or heat vents.
- Near vibrating, rotating or reciprocating equipment.
- On an non-level work surface.
- Near corrosive vapors (example : nitric acid ).
Power Options
The ES Series Electronic LCD Balance can be powered using either batteries or
an AC adapter.
The balance can be operated using 4AA alkaline (not included), or from any AC electrical outlet using the AC adapter supplied.
Battery Installation
If using batteries, open the battery compartment and insert the required
batteries with the correct polarity. Ensure that the battery compartment is
securely closed.
- Turn the balance over.
- Press the tabs on the battery cover inward and lift the cover off.
- Remove the battery holder.
- Install the 4AA alkaline battery in the battery holder, orienting the Plus (+) and minus (-) ends as indicated on the holder.
- Replace the battery cover.
AC Adapter
If using an AC adapter, plug one end into the balance and the other end into a
power outlet.
Operation
Turning The Balance On And Off
To turn on the ES Series Electronic LCD Balance, press the power button. To
turn it off, press and hold the power button for 2 seconds.
When no load on the platform, press “ON/TARE” The balance will display the software version, and then all segments will display will turn on,then the software version is displayed.
NOTE: Allow at least 5 minutes for the balance to temperature stabilize before using. If the balance is moved to a different temperature environment, allow additional time to stabilize.
To turn the balance off, press and hold until OFF is displayed, then release it.
Checking Calibration
To check the calibration of the balance, access the calibration menu and
follow the instructions provided in the user manual.
Before using the balance, its calibration should be checked. The balance has been calibrated before shipment, however, calibration is influenced by factors such as:
- Variations in the earth gravitational field at different latitudes of the world.
- Rough handling.
- Changes in work location.
- Height above sea level.
To check the balance’s calibration, place a known mass on the center of the
platform and read the displayed weight.
If the displayed weight differs from the known mass by more than acceptable
limits, refer to the Calibration Menu and the Specifications at the rear of
the manual.
Weighing
To weigh an item, place it on the platform and wait for the weight to
stabilize. The weight will be displayed on the digital display.
- Press “ON/TARE” to set the display to zero.
- Place the object to be weighed on the platform.
- Wait until the unit indicator appears before reading the displayed weight. When the unit indicator appears, the reading is stable.
Units exchange
The ES Series Electronic LCD Balance allows you to weigh items in different
units of measurement. To change the unit of measurement, access the setup menu
and select the desired unit.
Under weighing,press “OFF/MODE” to change the units.
Taring
To tare the balance, place an empty container on the platform and press the
tare button. This will reset the weight to zero and allow you to weigh only
the contents of the container.
When weighing samples that must be held in a container, taring enables the balance to subtract the weight of the container and display only the net weight of the sample.
- Press “ON/TARE” with no load on the platform to set the display to zero.
- Place the empty container on the platform. Its weight will be displayed.
- Press “ON/TARE”. Zero will be displayed and the container’s weight will be stored in the balance’s memory.
- Add the sample to the container and read its net weight on the display.
- When the container and its contents are removed, the container weight will be displayed as a negative number.The tared weight will remain in memory until “ON/TARE” is pressed again or the balance is turned off.
Parts Counting
The ES Series Electronic LCD Balance allows you to count parts by weighing a
sample and entering the piece weight. The balance will then calculate the
number of pieces based on the weight of the sample.
Parts Counting must be enabled in the Setup menu before using
This procedure. The balance will count parts based on the weight of a
reference sample, 5, 10, 20, 30, 40 or 50 parts. For optimum results, the
parts should be uniform in weight.
-
Repeatedly press “OFF/MODE” until pc CON is displayed.
-
Place a container on the platform, then press “ON/TARE” . the balance displays pcADD 5 which is the preset reference quantity. “5” shows the quantity of the reference sample.
-
To change the reference quantity, repeatedly press “ON/TARE” until the desired uantity displayed.
NOTE: If you need to exit or restart parts counting at any time, press and hold “ON/TARE” until pc CON is displayed, then release it. Return to step 2 to restart, or press “OFF/MODE” for other weighing functions. -
Add the displayed number of parts to the container, then press “ON/TARE”
The balance will display the current number of parts.
NOTE: If PCErr is displayed, the sample is too small to provide accurate results within the selected parts counting error level (PCErr of the Setup menu). -
- Add parts to the container as desired and read the quantity on the display. At this time, only “PC” counting signal is displayed.
NOTE: The reference quantity will remain in memory until it is changed or balance is turned off.
- Add parts to the container as desired and read the quantity on the display. At this time, only “PC” counting signal is displayed.
-
To switch between viewing the quantity and weight of the parts, press “OFF/MODE”
Using Menus To Configure The Balance
The following illustration identifies the items in each menu and The factory default settings are shown in bold type.
To access a menu, press and hold until desired menu appears, then release it.
Use these buttons to step through menus and select submenus:
Calibration Menu
The calibration menu allows you to calibrate the balance for accurate
measurements. Follow the instructions provided in the user manual to access
and use this menu.
All the Portable Advanced balances can be calibrated with standard masses.
press and hold with no load on the platform, until “CAL” is displayed, then
release it. Access “ CAL ” menu . Balance will first display “ CAL ” .
In a few seconds, balance will display the calibration weight value, press to
choose the calibration weight value and press to make sure. After that the
balance will display the weight value, according to the shown weight value,
put the corresponding calibrating weight on the center of the platform. In a
few seconds, the display will stop, and meanwhile the calibration weight value
will be shown, indicating that calibration is over. Normal weighting can be
done.
Calibration Menu Protection
The calibration menu can be locked using the lockout switch located on the
right side of the balance. Follow the instructions provided in the user manual
to use this feature.
- The Calibration menu may be locked out to prevent unauthorized access.
- When calibration end, turn the Lockout switch on the right side of the balance forwards, calibration menu may be stored in balance memory forever unless calibration menu is changed.
- When the Lockout switch is turned backwards, calibration menu cannot be stored in memory. It will disappear with no electricity.
Calibration Masses
Calibration masses should be kept in a safe, dry place when not in use.
Before beginning calibration, make sure calibrated masses are on hand. If you
begin calibration and realize masses are not available (masses are not
accurate), either turn the balance off, or go through the procedure without
masses. The balance will use previously stored calibration date.
Calibration should be performed as necessary to ensure accurate weighing.
Setup Menu
The Setup menu is used to customize the operation of the balance for your specific requirements. It contains submenus which enable you to turn features on or off, and program balance parameters. The illustration below indicates in which submenus appear on the Setup menu.
Accessing The Setup Menu
To access the setup menu, follow the instructions provided in the user manual.
To access the Setup menu, press and hold “ON/TARE” until SETUP is Displayed, then release
To access a submenu :
- Repeatedly press “OFF/MODE” until the desired submenu is displayed.
- Press“ON/TARE”to select the displayed submenu.
NOTE: You must use END to store any changes you make to the Setup menu.The following sections describe each item on the Setup menu in detail.
Accessing The Submenus In The Setup
To access the submenus in the setup menu, follow the instructions provided in
the user manual.
Reset To Factory Settings
To reset the balance to factory settings, access the setup menu and select the
reset option.
-
This submenu enables you to reset all Setup menu selections to the original factory setting outlined in the adjacent table.
To reset to factory settings: -
Access the Reset submenu.
-
Press “ON/TARE” to enter RESET.
-
Press “OFF/MODE” repeatedly to change the setting from YES to NO.
Selecting YES resets the Setup menu to the original factory setting. -
Press “ON/TARE” to accept the displayed setting.
Unit Selection
To select a unit of measurement, access the setup menu and choose from the
available units.
The Unit Selection submenu permits you to specify which weighing units and
operating modes will be enabled for use during operation.
The adjacent table lists the units and modes available on Portable Advanced
balances.
- Access the Sel submenu.
- ress “ON/TARE” , the display will show the grams unit indicator (g) along with the current status (ON/OFF).
- Press “OFF/MODE” to change the status.
- Press “ON/TARE” to accept the Displayed status, when released, the display will show the next unit indicator along with the current status.
- Set each unit or mode ON or OFF as in steps 3 and 4.
Taels
Taels are a unit of measurement for weighing precious metals and should only
be used if necessary.
If taels are enabled, you will be required to choose one of three different taels: Hong Kong, Singapore, or Taiwan.
-
When the display shows
TAEL1, press “OFF/MODE” to change to another tael if desired. -
Press “ON/TARE” to accept the displayed tael.
Stability Range
The stability range determines how long the balance takes to stabilize the
weight before displaying it on the digital display.
Follow the instructions provided in the user manual to adjust this setting.
The tability range specifies how much a displayed weight may change while the stability indicator (unit indicator) remains ON. When displayed weight changes beyond the allowable range, the stability indicator turns OFF indicating an unstable condition. Portable advanced balances permit you to select one of four stability ranges (in divisions) as shown in the table.
To view or change the stability range:
- Access the Stb submenu to display the current setting.
- Press “ON/TARE” to display the current setting.
- Press “OFF/MODE” to change the setting.
- Press“ON/TARE” to accept the displayed setting, when released, Stb will be displayed again.
AutoZero
The autozero feature automatically resets the weight to zero when the
platform is empty. Follow the instructions provided in the user manual to
enable or disable this feature.
Auto-Zero minimizes the effects of temperature changes and shift on the zero reading. By defining a threshold level in divisions, the balance maintains the zero display until the threshold is exceeded. This submenu permits you to select one of three threshold levels, or turn the feature OFF. Auto-Zero only functions when the display reads zero.
To view or change the auto-zero setting:
- Access the Auto 0 submenu.
- Press “ON/TARE” to display the current setting.
- Press “OFF/MODE” to change the setting.
- Press “ON/TARE” to accept the displayed setting, when released, auto 0 will be displayed again.
Parts Counting Error Level
The parts counting error level determines the acceptable margin of error
when counting parts. Follow the instructions provided in the user manual to
adjust this setting.
The parts counting error level is based on the resolution (readability) of the particular model balance and sample weight. The adjacent table lists error levels that you can choose from..
To view, change or disable the parts counting error level:
- Access the PC Err submenu.
- Press “ON/TARE” to display the current setting.
- Press “OFF/MODE” to change the setting.
- Press “ON/TARE” to accept the displayed setting, when released,PC Err will be displayed again.
AutoOff
The autooff feature automatically turns off the balance after a period of
inactivity. Follow the instructions provided in the user manual to enable or
disable this feature.
Auto Off permits the selection of either OFF, 60 seconds, 120 seconds or 180 seconds. When this function is selected, the balance will turn off if it is idle for the selected period of time (60, 120 or 180 seconds). This feature is especially useful when the balance is operated from internal batteries.
To view, change or disable the auto off feature:
- Access the A.Off submenu.
- Press “ON/TARE” to display the current setting.
- Press“OFF/MODE”to change the setting.
- Press “ON/TARE” to accept the displayed setting. When is released, A.Off will be displayed again.
End
Always use the END option to exit the setup menu and save your selections.
You must use End to exit the Setup menu. Changes you made in the Setup menu
are only stored in memory if you use End.
To exit the Setup menu and store your settings, press “ON/TARE” when END is
displayed. When is released, the balance will be returned to normal weighing
operations.
Calibration LockOut Protection
The calibration menu can be locked using the lockout switch located on the right side of the balance. Follow the instructions provided in the user manual to use this feature.
Access to the Calibration menu can be disabled using the lock out Switch located on the right side of the balance.
- Turn the balance off and unplug the power cord.
- Locate the access hole on the right side of the balance where the switch is located and remove the plug from the hole.
- Using a small screwdriver, slide the switch to the front of the balance for LOCKED or towards the rear of the balance for UNLOCKED.
- Replace the plug in the access hole.
Plug in the power cord and turn on the balance.
Care And Maintenance
To keep the ES Series Electronic LCD Balance operating properly,ensure that
the housing and platform are kept clean and free from foreign material. Use a
cloth dampened with a mild detergent if necessary. Calibration masses should
be kept in a safe, dry place.
Unplug the AC adapter when not in use and remove batteries for long term
storage.
Troubleshooting
Refer to the troubleshooting section of the user manual for assistance with any issues you may encounter while using the ES Series Electronic LCD Balance.
SYMPTOM | PROBABLE CAUSE | REMEDY |
---|---|---|
No Display | Power Adapter not connected. Batteries are dead. | Connect power |
Adapter. Replace batteries
LO BAT Indicator| Batteries are weak| Replace batteries
Incorrect Weight Reading| Balance was not re-zeroed before weighing Balance
has not been properly calibrated.| Press ON TARE with no weight on the
platform, then weigh item.
Calibration Procedure does not work| Incorrect weights being Used| Use correct
weights.
Unable to display Weight in a particular weighing unit| Weighing unit not
Enabled in Setup menu.| Use Unit Selection submenu to set desired units ON
(see Setup menu).
Balance won’t store Selections made in Setup menu| END selection was not used
to exit Setup menu.| You MUST use END to Exit Setup menu and Save selections.
Error Codes
The ES Series Electronic LCD Balance has error codes that can appear on the digital display. Refer to the user manual for a list of error codes and their suggested remedies.
The following list describes the various errors which can appear on the display and the suggested remedy.
Display Description
Err 1 Sample being weighed exceeds the capacity of the balance.
If error occurs when the sample is within the balance capacity, balance may be
incorrectly calibrated