APPA A1 Mini AC/DC Clamp Meter
Safety Information
Use the Meter only as specified in this manual or the protection by the Meter might be impaired.
- Always use proper terminals, switch position and range for measurements.
- Verify the Meter’s operation by measuring a known voltage. If in doubt, have the Meter serviced.
- Do not apply more than the rated voltage, as marked on Meter.
- To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator blinks.
- Do not use Meter around explosive gas or vapor.
- To reduce the risk of fire or electric shock do not expose this product to rain or moisture.
- Individual protective equipment should be used if HAZARDOUS LIVE parts in the installation where measurement is to be carried out could be ACCESSIBLE.
Caution
Do not expose Meter to extremes in temperature of high humidity.
Symbols as marked on the Meter and Instruction manual
| Risk of electric shock
---|---
| See instruction manual
⎓
| DC measurement
∼
| AC measurement
| Equipment protected by double or reinforced insulation
| Battery
| Earth
| Conforms to EU directives
| Application around and removal from hazardous live conductors is permitted
| Do not discard this product or throw away
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts. Repair or servicing should only be
performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents.
Power on/of
Auto Power Of
Auto Power Off Disable
ACA & DCA
Auto ACA&DCA Detection / ACA & DCA
Dut to the high sensitivity of the clamp meter, perform DCA Zero in the same direction as measurement to avoid interference by external magnetic field.
Data Hold
The display will flash continuously if the measured signal is larger 50counts than the display reading. However, it cannot detect across the AC and DC Current.
DCA ZERO
Inrush Current
The meter calculates the RMS value for 100ms as detecting a more than 5A current.
Low Pass Filter
The cut-off frequency of the low pass filter is about 160Hz with attenuation characteristic of approx-24db/octave.
Backlight on/of
Voltsense
The number of dashes displaying on the LCD indicates the electric field
intensity.
If no indication, voltage could still be present.
Battery Replacement
SPECIFICATIONS
General Specifications
Display Count : 6000 count
Overrange Display : ”OL or “-OL”
Conversion Rate : 2 times / second
Dimensions (W x H x D) : 60 x 147 x 31.5 mm
Weight : 140 g
Power requirement : LR44 Size Button Battery 1.5V ⎓ x 2
Battery Life : 20 hours.
Maximum Conductor Size : 22 mm
LVD : EN61010-1, EN61010-2-032
EMC : EN61326-1
Installation category : CAT.III. 600V.
CAT | Application field |
---|---|
I | The circuits not connected to mains. |
II | The circuits directly connected to Low |
voltage installation.
III| The building installation.
IV| The source of the Low-voltage installation.
Environmental Conditions
Indoor Use
Maximum operating altitude : 2000m (6562ft)
Operating temperature :
0°C ~ 30°C, ≤ 80%RH
30°C ~ 40°C, ≤ 75%RH
40°C ~ 50°C, ≤ 45%RH
Storage temperature :
-20 to +60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Pollution Degree : 2
Shock vibration : MIL-PRF-28800F for A class 2 Instrument
Drop Protection :
4 Feet Drop to hardwood on concrete Floor
Electrical Specifications
Accuracy is given as ±(% of reading + counts of least significant digit) at
23°C ± 5°C,with relative humidity Less than 80% R.H. and is specified for one
year after calibration.
ACV and ACA specifications are ac coupled,true R.M.S. The crest factor may be
up to 3.0 as 4000 counts.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
Position Error of Clamp: ±1.5% of LCD reading.
DC Current
Range | Resolution | Accuracy |
---|---|---|
60.00A (1) | 0.01A | ±(1.5%+10D) (2) |
300.0A | 0.1A | ±(1.5%+5D) |
- There are less than 0.3A variation as measuring in different directions.
- Add 10D to accuracy in Auto AC & DC Sense Mode.
AC Current
Range| Resolution| Accuracy (50 ~ 100Hz)| Accuracy (100 ~
400Hz)
---|---|---|---
60.00A| 0.01A| ±(1.5% +25D) for < 3A
±(1.5% +5D) for ≥ 3A| ±(2.5%+25D) for < 3A
±(2.5%+5D) for ≥ 3A
300.0A| 0.1A
Frequency Response: 50 ~ 400Hz
(Sine Wave)
Low-pass Filter
Range | Resolution | Accuracy (50Hz/60Hz) |
---|---|---|
60.00A | 0.01A | ±(3.5%+25D) for < 3A |
±(3.5%+5D) for ≥ 3A
300.0A| 0.1A
Cut-off Frequency (-3dB) : Approx. 160Hz
Attenuation Characteristic : Approx. -24dB
/ Oct
Inrush Current
Range | Resolution |
---|---|
300.0A | 300.0A |
Integration Time : 100ms
Trigger Current : 5.0 A
VoltSense
Voltage Range : 80V ~ 600V (At the tip of clamp)
Limited Warranty
This Meter is warranted to the original purchaser against defects in material
and workmanship for 3 years from the date of purchase. During this warranty
period, manufacturer will, at its option, replace or repair the defective
unit, subject to verification of the defect or malfunction.
This warranty does not cover fuses, disposable batteries, or damage from
abuse, neglect, accident, unauthorized repair, alteration, contamination, or
abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of this product, including but
not limited to implied warranties of merchantability and fitness for a
particular purpose, are limited to the above. The manufacturer shall not be
liable for loss of use of the instrument or other incidental or consequential
damages, expenses, or economic loss, or for any claim or claims for such
damage, expense or economic loss. Some states or countries laws vary, so the
above limitations or exclusions may not apply to you.