APPA A1 Mini AC/DC Clamp Meter User Manual

June 10, 2024
APPA

APPA A1 Mini AC/DC Clamp Meter

Safety Information

Use the Meter only as specified in this manual or the protection by the Meter might be impaired.

  • Always use proper terminals, switch position and range for measurements.
  • Verify the Meter’s operation by measuring a known voltage. If in doubt, have the Meter serviced.
  • Do not apply more than the rated voltage, as marked on Meter.
  • To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator blinks.
  • Do not use Meter around explosive gas or vapor.
  • To reduce the risk of fire or electric shock do not expose this product to rain or moisture.
  • Individual protective equipment should be used if HAZARDOUS LIVE parts in the installation where measurement is to be carried out could be ACCESSIBLE.

Caution

Do not expose Meter to extremes in temperature of high humidity.

Symbols as marked on the Meter and Instruction manual


| Risk of electric shock
---|---


| See instruction manual

| DC measurement

| AC measurement


| Equipment protected by double or reinforced insulation
| Battery
| Earth
| Conforms to EU directives
| Application around and removal from hazardous live conductors is permitted
| Do not discard this product or throw away

Maintenance

Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel.

Cleaning

Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents.

Power on/of

Auto Power Of

Auto Power Off Disable

ACA & DCA

Auto ACA&DCA Detection / ACA & DCA

Dut to the high sensitivity of the clamp meter, perform DCA Zero in the same direction as measurement to avoid interference by external magnetic field.

Data Hold

The display will flash continuously if the measured signal is larger 50counts than the display reading. However, it cannot detect across the AC and DC Current.

DCA ZERO

Inrush Current

The meter calculates the RMS value for 100ms as detecting a more than 5A current.

Low Pass Filter

The cut-off frequency of the low pass filter is about 160Hz with attenuation characteristic of approx-24db/octave.

Backlight on/of

Voltsense

The number of dashes displaying on the LCD indicates the electric field intensity.
If no indication, voltage could still be present.

Battery Replacement

SPECIFICATIONS

General Specifications

Display Count : 6000 count
Overrange Display : ”OL or “-OL”
Conversion Rate : 2 times / second
Dimensions (W x H x D) : 60 x 147 x 31.5 mm
Weight : 140 g
Power requirement : LR44 Size Button Battery 1.5V ⎓ x 2
Battery Life : 20 hours.
Maximum Conductor Size : 22 mm
LVD : EN61010-1, EN61010-2-032
EMC : EN61326-1
Installation category : CAT.III. 600V.

CAT Application field
I The circuits not connected to mains.
II The circuits directly connected to Low

voltage installation.

III| The building installation.
IV| The source of the Low-voltage installation.

Environmental Conditions

Indoor Use

Maximum operating altitude : 2000m (6562ft)
Operating temperature :
0°C ~ 30°C, ≤ 80%RH
30°C ~ 40°C, ≤ 75%RH
40°C ~ 50°C, ≤ 45%RH
Storage temperature :
-20 to +60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Pollution Degree : 2
Shock vibration : MIL-PRF-28800F for A class 2 Instrument
Drop Protection :
4 Feet Drop to hardwood on concrete Floor

Electrical Specifications

Accuracy is given as ±(% of reading + counts of least significant digit) at 23°C ± 5°C,with relative humidity Less than 80% R.H. and is specified for one year after calibration.
ACV and ACA specifications are ac coupled,true R.M.S. The crest factor may be up to 3.0 as 4000 counts.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
Position Error of Clamp: ±1.5% of LCD reading.

DC Current

Range Resolution Accuracy
60.00A (1) 0.01A ±(1.5%+10D) (2)
300.0A 0.1A ±(1.5%+5D)
  1. There are less than 0.3A variation as measuring in different directions.
  2. Add 10D to accuracy in Auto AC & DC Sense Mode.

AC Current

Range| Resolution| Accuracy (50 ~ 100Hz)| Accuracy (100 ~ 400Hz)
---|---|---|---
60.00A| 0.01A| ±(1.5% +25D) for < 3A
±(1.5% +5D) for ≥ 3A| ±(2.5%+25D) for < 3A
±(2.5%+5D) for ≥ 3A
300.0A| 0.1A

Frequency Response: 50 ~ 400Hz
(Sine Wave)

Low-pass Filter

Range Resolution Accuracy (50Hz/60Hz)
60.00A 0.01A ±(3.5%+25D) for < 3A

±(3.5%+5D) for ≥ 3A
300.0A| 0.1A

Cut-off Frequency (-3dB) : Approx. 160Hz
Attenuation Characteristic : Approx. -24dB
/ Oct

Inrush Current

Range Resolution
300.0A 300.0A

Integration Time : 100ms
Trigger Current : 5.0 A

VoltSense

Voltage Range : 80V ~ 600V (At the tip of clamp)

Limited Warranty

This Meter is warranted to the original purchaser against defects in material and workmanship for 3 years from the date of purchase. During this warranty period, manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction.
This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.

Customer Support

www.appatech.com
APAC
MGL APPA Corporation
[email protected]
Flat 4-1, 4/F, No. 35,
Section 3 Minquan East Road,
Taipei, Taiwan
Tel: +886 2-2508-0877

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

APPA User Manuals

Related Manuals